VARIANT II TURBO HBA1C KIT-2.0 VARIANT II TURBO HEMOGLOBIN TESTING SYSTEM
K122472 · Bio-Rad Laboratories, Inc., Clinical Systems Divis · LCP · Oct 26, 2012 · Hematology
Device Facts
Record ID
K122472
Device Name
VARIANT II TURBO HBA1C KIT-2.0 VARIANT II TURBO HEMOGLOBIN TESTING SYSTEM
Applicant
Bio-Rad Laboratories, Inc., Clinical Systems Divis
Product Code
LCP · Hematology
Decision Date
Oct 26, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7470
Device Class
Class 2
Indications for Use
The Bio-Rad VARIANT™ II TURBO HbA1c Kit-2.0 is intended for the quantitative determination of hemoglobin Ate in human whole blood using ionexchange high performance liquid chromatography (HPLC) on the VARIANT™ II TURBO Hemoglobin Testing System. Measurement of hemoglobin A1c is effective in monitoring long term glycemic control in individuals with diabetes mellitus. The Bio-Rad VARIANT™ II TURBO HbA1c Kit-2.0 is intended for Professional Use Only.
Device Story
The VARIANT™ II TURBO HbA1c Kit-2.0 is an in vitro diagnostic assay used with the VARIANT™ II TURBO Hemoglobin Testing System. The system utilizes ion-exchange high-performance liquid chromatography (HPLC) to separate and quantify hemoglobin A1c in EDTA-anticoagulated human whole blood. The system consists of a chromatographic station and a sampling station, controlled by a PC running Clinical Data Management (CDM) software. It is a fully automated, high-throughput analyzer operated by laboratory professionals in clinical settings. The system processes samples, performs chromatographic separation on a cation exchange cartridge, and calculates HbA1c concentrations. Results are reported in %HbA1c (NGSP), mmol/mol (IFCC), or %HbA1c (JDS). These quantitative results assist healthcare providers in assessing long-term glycemic control for patients with diabetes mellitus.
Clinical Evidence
Bench testing only. Method comparison study (CLSI EP9-A2) compared CDM software v5.1 vs v4.03 using 100 EDTA whole blood samples and 16 spiked samples. Results showed high correlation (R²=0.999) and acceptable bias at decision points (6.0% and 9.0% HbA1c). Interference testing (CLSI EP7-A2) confirmed no significant interference from HbS (≤67%), HbC (≤72%), HbD (≤55%), HbE (≤41%), or β-thalassemia trait (HbA2 up to 10%).
Technological Characteristics
Ion-exchange HPLC; cation exchange matrix. Components: analytical cartridge, prefilters, elution buffers, calibrators, whole blood primer. Electrical safety: BS EN 61010; EMC: BS EN 61326. Software: Clinical Data Management (CDM) v5.1.1. Connectivity: PC-controlled system. NGSP and IFCC certified.
Indications for Use
Indicated for the quantitative determination of hemoglobin A1c in human whole blood to monitor long-term glycemic control in individuals with diabetes mellitus. For professional use only.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k122472
B. Purpose for Submission:
Modification to a marketed device to revise hemoglobin variant interference claim and updated software, hardware and firmware.
C. Measurand:
Hemoglobin A₁c (HbA₁c)
D. Type of Test:
Quantitative ion-exchange high-performance liquid chromatography (HPLC)
E. Applicant:
Bio-Rad Laboratories, Inc.
F. Proprietary and Established Names:
VARIANT II TURBO HbA₁c Kit – 2.0 and VARIANT II TURBO Hemoglobin Testing System.
G. Regulatory Information:
| Regulation Description | Product Code | Device Class | Regulation | Panel |
| --- | --- | --- | --- | --- |
| Glycosylated Hemoglobin Assay | LCP | II | 21 CFR § 864.7470 | Hematology, 81 |
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The Bio-Rad VARIANT II TURBO HbA1c Kit – 2.0 is intended for the quantitative determination of hemoglobin A1c in human whole blood using ion-exchange high performance liquid chromatography (HPLC) on the VARIANT II TURBO Hemoglobin Testing System. Measurement of hemoglobin A1c is effective in monitoring long-term glycemic control in individuals with diabetes mellitus. The Bio-Rad VARIANT II TURBO HbA1c Kit – 2.0 is intended for Professional Use Only.
3. Special conditions for use statement(s):
{1}
For prescription use only.
## 4. Special instrument requirements:
For use with the Bio-Rad VARIANT II TURBO Hemoglobin Testing System (cleared under k040872 and k090699; modified in the current submission).
## I. Device Description:
The VARIANT II TURBO HbA₁c Kit – 2.0 contains an analytical cartridge, five prefilters, Elution Buffers A and B, Calibrator Level 1, Calibrator Level 2, Whole Blood Primer, sample vials and a CD-ROM with test parameters.
The Calibrators and the Whole Blood Primer contain lyophilized human red blood cell hemolysate with gentamicin, tobramycin, and EDTA as preservatives.
Each unit of whole blood used in the manufacture of the calibrators and whole blood primer was tested by FDA accepted methods and found non-reactive for HIV-1, HIV-2, Hepatitis B (HBV), Hepatitis C (HCV), and syphilis.
The items in the VARIANT II TURBO HbA₁c Kit – 2.0 are described in the table below.
| Component | Description |
| --- | --- |
| Analytical Cartridge | One cation exchange cartridge (2500 tests), 4.6 ID x 27.5 mm. Five prefilter elements (500 tests each) are included with the cartridge. |
| Elution Buffer A | Five bottles containing 2500 mL of a sodium perchlorate buffer. Contains <0.05% sodium azide as preservative. |
| Elution Buffer B | One bottle containing 2500 mL of a sodium perchlorate buffer. Contains <0.05% sodium azide as preservative. |
| Calibrator/Diluent Set | One set consisting of two vials of Calibrator Level 1, two vials of Calibrator Level 2, and one bottle of Calibrator Diluent. The calibrator vials contains lyophilized human red blood cell hemolysate with gentamicin, tobramycin, and EDTA as preservatives. Reconstituted volume is 7 mL per vial. Calibrator Diluent contains 100 mL of deionized water with <0.05% sodium azide as a preservative. |
| Whole Blood Primer | Two vials of lyophilized human red blood cell hemoglysate with gentamicin, tobramycin, and EDTA as preservatives. Reconstituted volume is 1.0 mL per vial. |
| Sample Vials | 100 polypropylene vials with pierceable caps, 1.5 mL. |
| CD-ROM | VARIANT II TURBO HbA1c Kit – 2.0 parameters. |
The VARIANT II TURBO Hemoglobin Testing System provides an integrated method for sample preparation, separation and the percent determination of HbA1c in EDTA human whole blood. The VARIANT II TURBO Hemoglobin Testing System is a fully automated, high-throughput hemoglobin analyzer. It consists of two modules - the VARIANT II Chromatographic Station (VCS) and the VARIANT II Sampling Station (VSS). There have been hardware updates due to obsolescence of parts and firmware updates to support the
{2}
replacement hardware components. A personal computer is used to control the VARIANT II TURBO Hemoglobin Testing System using Clinical Data Management (CDM) software version 5.1.1.
J. Substantial Equivalence Information:
| Predicate device name | Predicate 510(k) number |
| --- | --- |
| Bio-Rad VARIANT II TURBO HbA1c Kit – 2.0 | k090699 |
Comparison with predicate:
| Item | Candidate Device | Predicate Device |
| --- | --- | --- |
| Intended Use | Measurement of hemoglobin A1c is effective in monitoring long-term glycemic control in individuals with diabetes mellitus. | Same |
| Test Principle | Ion-exchange high performance liquid chromatography | Same |
| Sample Type | Anticoagulated whole blood (EDTA) | Same |
| Detection Wavelength | 415 nanometers | Same |
| Reporting Units | % HbA1c (NGSP) mmol/mol HbA1c (IFCC) | Same |
| Expected Range | 3.5 to 19.0% HbA1c (NGSP) | Same |
| Interference from HbF | HbF up to 25% had no significant effect on HbA1c determination. | Same |
| Interference from variants HbC, HbD, HbE, HbS | No significant interference was observed at the following concentrations: • HbC <72% • HbD <55% • HbE <41% • HbS <67% | Hemoglobin variants: Two out of 7 hemoglobin AD-trait, 2 out of 11 hemoglobin AS-trait, 1 out of 12 hemoglobin AE-trait, and 3 out of 9 hemoglobin AC-trait patient samples at the clinically significant levels of 6% and 9% HbA1c exhibited differences of more than +10% from values obtained using boronate affinity reference method. |
| Interference from HbA2 | β-thalassemia trait, as indicated by increased HbA2 concentrations up to 10%, does not interfere with the assay. | No claim previously |
| Instrument | VARIANT II TURBO Hemoglobin Testing System | Same |
| Software | CDM Version 5.1.1 | CDM Version 4.03 |
| Calibrator | 2 levels | Same |
{3}
| Calibration frequency | Once every 500 injections/ 2500 injections total column life | Same |
| --- | --- | --- |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP7-A2: Interference Testing in Clinical Chemistry.
CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples.
L. Test Principle:
The VARIANT II TURBO HbA1c Kit- 2.0 is based on chromatographic separation of HbA1c on a cation exchange cartridge. The various forms of hemoglobin exhibit charge differences (positive) at the acidic pH of the mobile phase, and thus can be separated on a support that is negatively charged (cation exchange). The use of ion-exchange chromatography allows molecules to be separated based upon a molecule's charge. Separation is optimized to eliminate interferences from hemoglobin variants (HbS, HbC, HbD and HbE trait), labile A1c, hemoglobin F and carbamylated hemoglobin.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Previously established in submission k090699.
b. Linearity/assay reportable range:
The reportable range of 3.5 to 19.0% HbA₁c was previously established in submission k090699.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrators were previously cleared in k070452. The recommended control materials were cleared in k070546 and k052838.
The modified device is certified by the National Glycohemoglobin Standardization Program (NGSP) as having documented traceability to the Diabetes Control and Complications Trial (DCCT) reference method.
d. Detection limit:
Previously established in submission k090699.
e. Analytical specificity:
The interference claims for icterus (bilirubin), lipemia (triglyceride), EDTA, hemoglobin F, labile A₁c and carbamylated hemoglobin were previously established in submission k090699.
The possible interferant effects from hemoglobin variants C, D, E, F, S and A2 (HbC, HbD, HbE, HbF, HbS and HbA2) were evaluated. Two fresh, EDTA non-variant
{4}
human blood sample pools at 6.5% and 8.0-9.0% HbA₁c were collected. Fresh, homozygous EDTA-whole blood samples containing each of the hemoglobin variants (E, D, S, and C) were obtained. EDTA-whole blood samples were prepared by spiking purified HbF or HbA2. For the interference testing of each variant, a series of test sample pools (0, 25, 50, 75, 100% and additional intermediate pools) were prepared by the dilution of a homozygous variant patient sample (as interferant) in the non-variant patient sample pool. The samples were run in duplicate on two VARIANT II TURBO Hemoglobin Testing Systems. Non-significant interference was defined by the sponsor as less than ±10% bias from control.
Based upon the data in the study, the sponsor claims no interference at the following concentrations: HbC ≤72%, HbD ≤55%, HbE ≤41%, HbF ≤25%, HbS ≤67% and HbA2 ≤10%.
f. Assay cut-off: Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
To demonstrate the correlation between samples analyzed on devices using CDM Software Versions 5.1 (candidate device) and samples run on devices using CDM Software Version 4.03 (predicate device), 40 EDTA whole blood patient samples, 60 commercially purchased EDTA whole blood patient samples and 16 contrived samples prepared by mixing EDTA whole blood samples with either purified HbA0 or purified HbA1c (8 samples cover 3.5% to 4.5% and 8 samples cover 14.2% to 19.5%) were analyzed by the VARIANT II TURBO HbA1c Kit - 2.0 run on the VARIANT II TURBO Hemoglobin Testing System using both software versions. The samples were run in singlet on the subject device and on the predicate device. The results of the linear regression analysis are presented in the table below.
| Reporting Unit | Regression Data | Sample Range |
| --- | --- | --- |
| %HbA1c (NGSP) | y= 0.99x + 0.06, R²= 0.9996 | 3.5% to 19.5% |
| mmol/mol (IFCC) | y= 0.99x + 0.48, R²= 0.9996 | 15 to 190 mmol/mol |
b. Matrix comparison: Not applicable.
3. Clinical studies:
a. Clinical Sensitivity: Not applicable.
b. Clinical specificity: Not applicable.
{5}
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
4. Clinical cut-off: Not applicable.
5. Expected values/Reference range: Hemoglobin A₁c expected value range was cited from American Diabetes Association Standards of Medical Care in Diabetes 2012, 35 (Supplement 1), S11-S63.
| Hemoglobin A₁c (%) | Glycemic Goal |
| --- | --- |
| <8 | Less Stringent |
| <7 | General Goal |
| <6.5 | More Stringent |
| <5.7 | Non-Diabetic Goal |
N. Instrument Name: Bio-Rad VARIANT II TURBO Hemoglobin Testing System
O. System Descriptions:
1. Modes of Operation: A completely closed, fully automated, high-throughput analyzer consisting of a chromatographic and a sampling station.
2. Software: FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification: Samples are identified from the barcode on blood collection test tube. The code is read automatically by the analyzer. The system software (CDM 5.1.1) detects and prevents mismatch of results with sample ID. If a mismatch is detected, the sample result will not be sent to the Laboratory Information System (LIS) and the operator will be notified.
4. Specimen Sampling and Handling: Blood is drawn into normal vacuum test tube. The blood collection tube is placed directly into a sample rack, which is part of the analyzer. A volume of sample is automatically drawn from the tube through its rubber seal.
{6}
5. Calibration:
The calibrators were previously cleared in k070452.
6. Quality Control:
The recommended control materials were cleared in k052838 and k070452.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
The software documentation was reviewed and it supports that the device was developed under good software cycle processes.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.