Retrospective postmarket surveillance review of medical records
A retrospective review of 49 eyes was used to compare the clinical outcomes of laser-assisted arcuate incisions (using the VICTUS Platform) against manual arcuate incisions (using a diamond blade) in a routine clinical setting.
Postmarket surveillance; Retrospective review; Arcuate incisions; Cataract surgery
The VICTUS Femtosecond Laser Platform is indicated for use in: - the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea - for anterior capsulotomy during cataract surgery - the creation of penetrating arcuate cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.
Device Story
VICTUS Platform is an ophthalmic surgical femtosecond laser; creates corneal flaps, anterior capsulotomies, and penetrating arcuate incisions. Surgeon programs incision parameters (depth, radius, opening/position angles) into system. Eye fixated via sterile disposable PMMA Patient Interface (contact lens and suction clip) connected to laser via vacuum tube. Laser delivers femtosecond pulses through interface; scanning optic system focuses beam to create continuous incisions via tissue photodisruption. Used in clinical settings by ophthalmologists. Output provides precise, reproducible corneal cuts; assists surgeon in cataract and refractive procedures; potentially improves surgical accuracy and outcomes compared to manual techniques.
Clinical Evidence
Retrospective postmarket surveillance study at a single site (n=49 eyes, 34 subjects; mean age 57.4 years). Compared manual diamond blade arcuate incisions (n=25) to laser-assisted incisions (n=24). Endpoints included BCDVA, refractive change, induced cylinder, and adverse events. Results showed laser-assisted procedure comparable to manual standard. Bench testing (porcine eyes, agarose, PETG, PMMA) confirmed accuracy/reproducibility of rim depth, diameter, and angles via SEM analysis.
{1}------------------------------------------------
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Applicant: | Technolas Perfect Vision GmbH<br>Messerschmittstr. 1.+3<br>80992 Munich, Germany<br>Tel: 49 174 3494062 |
|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Betty M. Johnson、<br>Sr. Manager, Regulatory Affairs<br>Technolas Perfect Vision<br>c/o PO Box 17190<br>Anaheim, CA 92817<br>Ph: (714) 273-8254<br>Fax: (714) 855-2794<br>Email : b.johnson@technolaspv.com |
| Trade Name:<br>Common Name:<br>Classification Name:<br>Device Classification:<br>Product Code | VICTUS Femtosecond Laser Platform<br>Ophthalmic Laser<br>Laser, Ophthalmic<br>Class II<br>OOE (Ophthalmic Femtosecond Laser)<br>HQF (Laser, Ophthalmic) |
## Substantially Equivalent To:
| 510(k) Number | Product Trade Name | Current Manufacturer |
|---------------|-------------------------------------------------|----------------------------------|
| K120426 | VICTUS Femtosecond Laser<br>Platform | Technolas Perfect Vision |
| K110427 | Femtec Laser System for Anterior<br>Capsulotomy | Technolas Perfect Vision<br>GmbH |
| K113151 | iFS Laser System | Abbott Medical Optics |
| K101626 | LenSx Laser | LenSx, an Alcon company |
{2}------------------------------------------------
### Description of the Device Subject to Premarket Notification:
The VICTUS Femtosecond Laser Platform (hercafter referred to as the VICTUS Platform) is a precision ophthalmic surgical laser indicated for use in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.
The VICTUS Platform is an ophthalmic surgical femtosecond laser designed for cutting a penetrating arcuate cut of pre-selected depth, radius, opening angle, and position angle. The system works by first being programmed with the depth, radius, opening angle and position angle at which the penetrating arcuate cut is desired. The surgeon then fixates the eye with a sterile disposable PMMA contact lens (hercafter referred to as the Patient Interface) that is connected to the laser via a vacuum tube. Depths of -120µm, radius of 1500 - 4750 um, opening angles of 1-360°, and position angles of 0 - 359° are available.
Laser pulses are delivered through the sterile Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.
### Indications for Use:
The VICTUS Platform is indicated for use in:
- the creation of a corneal flap in patients undergoing LASIK surgery or other treatment . requiring initial lamellar resection of the cornea
- for anterior capsulotomy during cataract surgery .
- the creation of penetrating arcuate cuts / incisions in the cornea in patients undergoing ● cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.
### Technical Characteristics Comparison:
The design principle of the VICTUS Platform is fundamentally the same as the predicate devices. The VICTUS Platform mode of operation is the same as the previously cleared VICTUS Femtosecond Laser Platform cleared in K120426, iFS Laser System cleared in K113151, and the LenSx Laser System cleared in K101626, all of which deliver femtosecond pulses to produce a pattern of photodisruption to create cuts / separation in ophthalmic tissue. The VICTUS Platform delivers femtosecond pulses to produce a pattern of photodisruption for creation of penetrating arcuate cuts / incisions in the cornea, as do the predicate devices.
The means of fixation of the patient contact portion of the VICTUS Platform is substantially equivalent to that present in the VICTUS Femtosecond Laser Platform cleared in K120426, the Femtec Laser System for Anterior Capsulotomy cleared in K110427, iFS Laser System cleared in K113151, and the LenSx Laser cleared in K101626. All of these predicate systems use suction vacuum to affix a suction ring to the corneal surface prior to use.
{3}------------------------------------------------
The VICTUS Platform mode of operation and the technology used to create the cutting action are similar or identical to these previously mentioned devices, and therefore substantially equivalent to these legally marketed predicate devices.
#### Performance Data:
The VICTUS Femtosecond Laser Platform has undergone testing and is in compliance with applicable safety standards as listed in the following table.
| Standard | Title |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN ISO 60601-1 | Medical electrical equipment - Part I: General requirements<br>for safety |
| EN ISO 60601-1-2 | Medical electrical equipment - Part 1: General requirements<br>for safety; 2. Collateral standard: electromagnetic<br>compatibility; requirements and tests |
| EN ISO 60601-1-4 | Medical electrical equipment - Part 1-4: General requirements<br>for safety; Collateral standard: Programmable electrical<br>medical systems |
| EN ISO 60601-2-22 | Medical electrical equipment – Part 2: Particular requirements<br>for the safety of diagnostic and therapeutic laser equipment |
The VICTUS Femtosecond Laser Platform has been found to perform equivalently to the predicate devices in patients undergoing ophthalmic surgery or other treatment requiring penetrating arcuate cuts / incisions in the cornea. The VICTUS Femtosecond Laser Platform and the predicate devices therefore have similar performance profiles.
#### Non-Clinical Performance Data
A variety of test procedures were conducted to demonstrate the performance of the proposed VICTUS Platform in support of this premarket submission. The collected data were evaluated by comparing the mean values to the specified acceptance criteria and their 95% confidence intervals. Four different materials were used for the bench performance testing; porcine eyes, agarose gel, polyethylene terephthalate (PETG), and polymethyl methacrylate (PMMA). Scanning electronic microscopy has also been utilized to visually assess the results of the arcuate cuts / incisions procedures with the VICTUS Platform. The resulting micrographs are compared qualitatively with images obtained from the predicate devices.
The testing showed that laser-assisted arcuate cuts / incisions performed with the VICTUS Femtosecond Laser Platform resulted in highly reproducible and accurate rim depth, diameter, open angle, position angle, and centration.
{4}------------------------------------------------
K122386
#### Postmarket Surveillance Clinical Review
A retrospective postmarket surveillance review was performed at a single site in Europe to compare manual arcuate incisions using a diamond blade to laser-assisted arcuate incisions using the VICTUS™ Platform. This review was conducted at a single clinical location, and all surgeries were performed by the same surgeon.
A total of 49 eyes in 34 subjects were included in the analysis. The mean age of the study population was 57.4 ± 11.7 years (range: 18 to 84 years). Twenty four (24) eyes received laserassisted arcuate cuts, of which 11 received arcuate cuts alone and 13 received both arcuate cuts and cataract surgery. Twenty five (25) eyes received manual arcuate cuts, of which four received arcuate cuts alone and 21 received both arcuate cuts and cataract surgery. One adverse event was reported in the laser-assisted group, which was considered by the surgeon to not be related to the laser-assisted procedure.
The laser-assisted arcuate cuts procedure was shown to be comparable to the standard manual procedure when assessed for common clinical endpoints such as post-surgical loss of lines of best corrected distance visual acuity (BCDVA), post-surgical overall change in refraction, postoperative induced cylinder, adverse events, and serious adverse events.
The postmarket surveillance review data confirm that laser-assisted arcuate cuts are comparable to the standard manual procedure.
#### Basis for Determination of Substantial Equivalence:
The technological characteristics of the VICTUS Femtosecond Laser Platform are substantially equivalent to the technological characteristics of the VICTUS Femtosecond Laser Platform (K120426), the Femtec Laser System for Anterior Capsulotomy (K110427), iFS Laser System (K113151), the LenSx Laser (K101626).
The proposed additional indication for use of arcuate cuts / incisions for the VICTUS Femtosecond Laser Platform is very similar to the indications for use cleared for the iFS Femtosecond Laser in K113151 and for the LenSx Laser in K101426; both include arcuate corneal incisions.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
February 8, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Technolas Perfect Vision GmbH % Ms. Betty M. Johnson Senior Manager, Regulatory Affairs, Americas PO Box 17190 Anaheim Hills, CA 92817
Re: K122386
Trade/Device Name: VICTUS Femtosecond Laser Platform Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Femtosecond Laser Regulatory Class: Class II Product Code: OOE, HQF Dated: February 1, 2013 Received: February 4, 2013
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Ms. Betty M. Johnson
CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia YAlexander -S.
Malvina B. Eydelman, M.D. Director Division of Opthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K122386
Device Name(s): VICTUS Femtosecond Laser Platform
Indications for Use:
The VICTUS Femtosecond Laser Platform is indicated for use in:
- the creation of a corneal flap in patients undergoing LASIK surgery or other treatment ● requiring initial lamellar resection of the cornea
- . for anterior capsulotomy during cataract surgery
- . the creation of penetrating arcuate cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODF)
(Division Sign-Off) (Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number
Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.