DIAVUE PRUDENTIAL BLOOD GLUCOSE MONITORING SYSTEM
K122307 · Biocare Corporation · NBW · Apr 5, 2013 · Clinical Chemistry
Device Facts
| Record ID | K122307 |
| Device Name | DIAVUE PRUDENTIAL BLOOD GLUCOSE MONITORING SYSTEM |
| Applicant | Biocare Corporation |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Apr 5, 2013 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The DIAVUE Prudential Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. It is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples (from finger). This system is intended to be used by a single person and should not be shared. It should not be used for the diagnosis of or screening for diabetes, or testing on neonates.
Device Story
System consists of meter, test strips, and control solutions; uses electrochemical biosensor technology with glucose oxidase enzyme. Capillary action draws blood into test strip chamber; meter measures electrical current generated by glucose reaction; displays result in 7 seconds. Intended for single-patient home use; operated by patient. Provides quantitative glucose measurement (20-600 mg/dL) to assist in diabetes management. Includes memory for 450 tests and multi-day averaging. Meter requires no coding. Performance validated against YSI reference method; includes disinfection validation for multi-use safety.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison of technological characteristics to predicate.
Technological Characteristics
Electrochemical biosensor with carbon electrodes; glucose oxidase enzyme. Dimensions: 90x52x15 mm; weight: 45.6 g. Sample volume: 0.7 μL. Measurement range: 20-600 mg/dL. Connectivity: None. Standards: ISO 15197:2003, ISO 14971:2007, CLSI EP6-A, CLSI EP7-A2, IEC/EN 61010-1/2-101.
Indications for Use
Indicated for people with diabetes for home monitoring of glucose levels in fresh capillary whole blood from the fingertip. Not for diagnosis or screening of diabetes; not for neonatal use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- URIGHT TD-4254 Blood Glucose Monitoring System (K082482)
Related Devices
- K111890 — FORA DIAMOND PRIMA BLOOD GLUCOSE MONITORING SYSTEM, FORA DIAMOND MINI BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Feb 6, 2012
- K111371 — ON CALL CHOSEN BLOOD GLUCOSE MONITORING SYSTEM · ACON Laboratories, Inc. · Oct 6, 2011
- K122110 — ACON ON CALL VIVID PAL BLOOD GLUCOSE MONITORING SYSTEM · ACON Laboratories, Inc. · Nov 21, 2012
- K093941 — CLEVER CHOICE MINI BLOOD GLUCOSE MONITORING SYSTEM MODEL TD-4265 · Taidoc Technology Corporation · Mar 26, 2010
- K113077 — BIOLAND G-423 GLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Aug 3, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
喬聯科技
www.biocare.com.tw
# 510(k) Summar
#### JULY 15 2012
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The Assigned 510(k) number is: K122307
#### 1. Submitter's Name: BioCare Corporation
Address: 4F, No.12, Lane 5, Sec. 2, Nanshan Rd., Lujhu Township,
Taoyuan County 33852, Taiwan.
Contact person: Shuchi Chang / Administrator Phone: +886-3-2616678#6533
Fax: +886-3-2616679
E-mail:shuchichang@biocare.com.tw
#### 2. Device Name and Classification
2.1 DIAVUE Prudential Blood Glucose Monitoring System
Trade name: DIAVUE Prudential Blood Glucose Monitoring System
Regulation section: 21 CFR 862.1345 Glucose Test System
Classification Code: CGA, Glucose Oxidase,
NBW, System, Test, Blood Glucose, Over The Counter
Classification Panel: Chemistry (75)
Device Classification: Class II
#### 2.2 Control Solution
Trade name: DIAVUE Control Solution
Regulation section: 21 CFR 862.1660, Quality Control Material (Assayed and
Unassaved)
Classification Code: JJX-Single (Specified) Analyze Controls (Assayed and Unassayed)
Classification Panel: Clinical Chemistry (75)
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with Chinese characters and the website address "www.biocare.com.tw" underneath. The Chinese characters are written in a traditional style, and the logo has a simple design. The website address is written in a sans-serif font. The logo is likely for a company or organization based in Taiwan.
The Commitment to nignes vnige
BioCARE Corporation
Tot: +886 3 2616678 Fax:+88632616679
4F, No.12, Lane 5, Sec. 2, Nanshan Rd., Lujhu
Township, Taoyuan County 33852, TAIWAN
Device Classification: Class I
#### 3. Predicate Device
### 3.1 Blood Glucose Monitoring System
The predicate device is URIGHT TD-4254 Blood Glucose Monitoring System (K082482) marketed by TaiDoc Technology Corporation.
### 3.2 Control Solution
The predicate device is TaiDoc Control Solution and FORA Control Solution (K093724) marketed by TaiDoc Technology Corporation.
# 4. Device Description
The DIAVUE Prudential Blood Glucose Monitoring System consists of glucose meter, quick start user guide, user manual, warranty card, and carrying bag. Test strips, control solutions or sterile lancets can be purchased separately.
DIAVUE Control Solution contains a known amount of glucose that reacts with test strips made by BioCare. Three levels of control solutions are available : level 1.level 2 and level 3.Operation of the control solution test is similar to blood glucose test except that you use DIAVUE control solution instead of a drop of blood. The control solutions should provide results within an expected range printed on the test strip vial label.
#### 5. Intended Use
DIAVUE Prudential Blood Glucose Monitoring System :
The DIAVUE Prudential Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. It is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples (from finger). This system is intended to be used by a single person and should not be shared. It should not be used for the diagnosis of or screening for diabetes, or testing on newborns.
ર-2
{2}------------------------------------------------
高
www.biocore.com.tw
BIOCARE
*The Commitment to Higher Value*
BioCARE Corporation
Tel: +886 3 2616678 Fax: +886 3 2616679
4F, No.12, Lane 5, Sec. 2, Nanshan Rd., Lujhu
Township, Taoyuan County 33852, TAIWAN
### DIAVUE Blood Glucose Test Strips:
The DIAVUE Blood Glucose Test Strips are to be used with the DIA VUE Prudential Blood Glucose Meter, it measures glucose in capillary whole blood taken from a fingertip. It is for use outside of body (in vitro diagnostic use). It is intended for lay use by people with diabetes and should only be used by a single patient. This system should not be shared. They are not indicated for the diagnosis or screening of diabetes or for neonatal use.
#### DIAVUE Control Solutions:
The DIAVUE Control Solutions are used with the DIAVUE Prudential Blood Glucose Monitoring Systems to indicate appropriate user technique and to indicate that the test strip and meter are functioning properly.
#### 6. Comparison with predicate:
The DIAVUE Prudential Blood Glucose Monitoring System has the following similarities to the predicate device:
- Same operating principle, �
- � Same fundamental scientific technology,
- Incorporate the same material, �
- Same shelf life.
- Package using the same materials
The differences between the two devices list as: weight, dimensions, display, parameters and appearance. There are no significant differences that affect the safety and effectiveness. Therefore, BioCare DIAVUE Prudential Blood Glucose Monitoring System is substantially equivalent to the legally marketed device URIGHT TD-4254 Blood Glucose Monitoring System manufactured by TaiDoc Technology Corporation, K082482.
{3}------------------------------------------------
| Image: [Logo of 僑聯聯科技 and www.biocare.com.tw] | | Image: [BioCARE Logo and BioCARE Corporation contact information] |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| | DIAVUE Control Solution | Predicate Device:<br>TaiDoc Control Solution and<br>FORA Control Solution<br>(K093724) |
| Indications for Use | To indicate appropriate user<br>technique and to indicate that<br>the test strip and meter are<br>functioning properly. | To check the performance of<br>blood glucose monitoring<br>systems manufactured by<br>TaiDoc. |
| Analyte | d-glucose | Same |
| Matrix | Aqueous | Same |
| Number of Level(s) | 3 | 3 |
# 7. Test Principle:
The detection and measurement of glucose in blood is by an electrochemical biosensor technology using glucose oxidase.
# 8. Conclusions
Based on the information provided in this submission, the DIAVUE Prudential Blood Glucose Monitoring System is substantially equivalent to the predicate URIGHT TD-4254 Blood Glucose Monitoring System. The DIAVUE Control Solution is substantially equivalent to the predicate TaiDoc Control Solution and FORA Control Solution (K093724).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 5, 2013
BioCare Corporation C/O Shuchi Chang 4F, NO. 12, LANE 5, SEC. 2 NASHAN RD LUJHU TOWNSHIP TAOYUAN COUNTY CHINA, (TAIWAN) 33852
Re: K122307
Trade/Device Name: DIAVUE Prudential Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CGA, NBW, JJX Dated: February 25, 2013 Received: March 28, 2013
Dear Shuchi Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carol G. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K122307
# DIAVUE Prudential Blood Glucose Monitoring System
The DIAVUE Prudential Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. It is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples (from finger). This system is intended to be used by a single person and should not be shared. It should not be used for the diagnosis of or screening for diabetes, or testing on neonates.
#### DIAVUE Blood Glucose Test Strips
The DIAVUE Blood Glucose Test Strips are to be used with the DIAVUE Prudential Blood Glucose Meter; it measures glucose in capillary whole blood taken from a fingertip. It is for use outside of body (in vitro diagnostic use). It is intended for lay use by people with diabetes and should only be used by a single patient. This system should not be shared. They are not indicated for the diagnosis or screening of diabetes or for neonatal use.
#### DIA VUE Control Solutions
The DIAVUE Control Solutions are used with the DIAVUE Prudential Blood Glucose Meter and DIAVUE Blood Glucose Test Sreips to indicate appropriate user technique and to indicate that the test strip and meter are functioning properly.
Prescription Use (Part 21 CFR 801 Subpart D)
AND / OR
Over-The-Counter Use মি (21 CFR 801 Subpart C)
(PLEASE NO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Image /page/6/Picture/13 description: The image shows the name "Katherine Serrano" in a bold, sans-serif font. The name is written in black ink on a white background. The letters are evenly spaced and easy to read. The name appears to be a signature or a title.
Page 1 of _ l