← Product Code [DXQ](/productcode/DXQ) · K122259

# BLOOD PRESSURE MONITOR WITH STETHOSCOPE (K122259)

_Hangzhou Reli-On Co., Ltd. · DXQ · Sep 24, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K122259

## Device Facts

- **Applicant:** Hangzhou Reli-On Co., Ltd.
- **Product Code:** [DXQ](/productcode/DXQ.md)
- **Decision Date:** Sep 24, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1120
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Blood pressure monitor with stethoscope, Model HBPK-A is a non-automated, aneroid blood pressure monitor that is used for the indirect measurement (noninvasive) and display of arterial blood pressure. It can be used by professionals as well as trained individual users at hospital or at home to monitor both systolic and diastolic pressure. This device is sold with an adult D-ring cuff and suitable for use on adult.

## Device Story

Non-automated, manual aneroid sphygmomanometer; used for indirect blood pressure measurement. Components: adjustable adult D-ring cuff, manual inflation bulb, 300 mmHg aneroid gauge, and stethoscope. Operation: user inflates cuff around patient's arm; uses stethoscope to auscultate Korotkoff sounds over brachial artery; identifies systolic and diastolic pressures via gauge observation. Used in clinical or home settings by professionals or trained individuals. Provides manual blood pressure monitoring; aids in cardiovascular health assessment.

## Clinical Evidence

Bench testing only. Compliance with ANSI/AAMI SP10:2002/(R)2008 and ANSI/AAMI SP10:2002/A1:2003 for manual sphygmomanometers. Biological compatibility testing performed per ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

Manual aneroid sphygmomanometer. Components: D-ring cuff, stethoscope, 300 mmHg rotary pin gauge, inflation bulb. Standards: ANSI/AAMI SP10:2002/(R)2008, ANSI/AAMI SP10:2002/A1:2003, ISO 10993-5, ISO 10993-10. Non-electronic, no software, no connectivity.

## Regulatory Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

## Predicate Devices

- Wenzhou Kindcare Import & Export Co., Ltd Aneroid Sphygmomanometer, Model KT-A01 ([K081951](/device/K081951.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

PY%

SEP 24 2012

### 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

#### Date: July 14, 2012

1. Company and Correspondent making the submission:

|          | Company                                                                                | Correspondent                         |
|----------|----------------------------------------------------------------------------------------|---------------------------------------|
| Name     | Hangzhou Reli-On Co., Ltd                                                              | IRC/ Charles Mack                     |
| Address  | #204, 3Xin Chang Road,<br>Gongshu District, Hangzhou<br>City, Zhejiang Province, CHINA | 77325 Joyce Way<br>Echo, Oregon 97826 |
| Phone    | +86-571-28036398                                                                       | 931-625-4938                          |
| FAX      | +86-571-28036400                                                                       | 541-376-5063                          |
| Contact  | Ms. Yu Manhua / General<br>Manager                                                     | Charles Mack/Principal<br>Engineer    |
| Internet | charliemack@irc-us.com                                                                 | charliemack@irc-us.com                |

#### 2. Device :

Trade/proprietary name: Blood pressure monitor with stethoscope, Model HBPK-A

| Common Name         | : Blood Pressure Cuff |
|---------------------|-----------------------|
| Classification Name | : Blood Pressure Cuff |

- 3. Predicate Device :
The predicate device is:

Wenzhou Kindcare Import & Export Co., Ltd K081951 Aneroid Sphygmomanometer, Model KT-A01

Recognized Consensus Standards:

The following standard was used to demonstrate compliance to FDA recognized consensus standards:

- SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/( Manual, electronic or . automated sphygmomanometers

{2}------------------------------------------------

- 4. Classifications Names & Citations :
21CFR 870.1120, DXQ, Blood Pressure Cuff, Class 2

#### 5. Description :

Blood pressure monitor with stethoscope, Model HBPK-A is a non-invasive blood pressure measurement system for monitoring blood pressure levels. This Non-automated Sphygmomanometer uses an occluding cuff, an aneroid manometer to measure pressure, and a stethoscope for detecting Korotkoff so unds.

The Aneroid Sphygmomanometer with Stethoscope contains:

- 1. Adjustable D-ring Cuff (Adult Size)
- 2. Stethoscope (Attaches to the cuff)
- 3. Non-stop rotary pin. 300 mmHq gauge
- 4. Instruction booklet and record
- 4. Carrying case

-The Blood pressure monitor with stethoscope enables the user to monitor the pressure of flowing blood that is exerted against the arteries at highest (systolic or contraction) and lowest (diastolic or relaxation) pressure. To operate, the user places the attached stethoscope on the inner arm above the bend in the elbow, to detect the pulse of the brachial artery. After inflation of the cuff, the user does auditory monitoring with the stethoscope to evaluate systolic and diastolic pressure. The two values are usually recorded as a ratio of the two measurements: systolic over diastolic.

{3}------------------------------------------------

6. Indications for use :

The Blood pressure monitor with stethoscope, Model HBPK-A is a non-automated, aneroid blood pressure monitor that is used for the indirect measurement (noninvasive) and display of arterial blood pressure. It can be used by professionals as well as trained individual users at hospital or at home to monitor both systolic and diastolic pressure. This device is sold with an adult D-ring cuff and suitable for use on adult.

## 7. Technical Characteristics:

The submitted Hangzhou Rei-On device is exactly like the predicate. This is a simple manually operated aneroid blood pressure monitor with a bulb for inflation of a bladder which fits around the user or patient's arm. A stethoscope is used to hear the Korotkoff sounds.

## 8. Safety and Performance Data Summary:

Testing was performed in accordance with the Recognized Consensus standards, noted under Item 3, and also testing to Biological Compatibility Standards (ISO 10993-5, ISO 10993-10). The device successfully passed all the requirements within these standards.

#### 9. Conclusions:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Hangzhou Reli-On Co., Ltd, concludes that Blood pressure monitor with stethoscope, Model HBPK-A is safe and effective as proven by compliance to consensus standards.

END

{4}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The eagle is facing to the right.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP
2 4
2012

Hangzhou Reli-On Co., Ltd. c/o Mr. Charlie Mack Principal Engineer IRC 77325 Joyce Way Echo, OR 97826

Re: K122259

> Trade/Device Name: Blood Pressure Monitor with Stethoscope, Model HBPK-A Regulatory Number: 21 CFR 870.1120 Regulation Name: Blood-Pressure Cuff Regulatory Class: II (two) Product Code: 74 DXO Dated: September 13, 2012 Received: September 18, 2012

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{5}------------------------------------------------

Page 2 - Mr. Charlie Mack

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

6

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: Blood pressure monitor with stethoscope, Model HBPK-A

Indications for Use:

The Blood pressure monitor with stethoscope, Model HBPK-A is a non-automated, aneroid blood pressure monitor that is used for the indirect measurement (noninvasive) and display of arterial blood pressure. It can be used by professionals as well as trained individual users at hospital or at home to monitor both systolic and diastolic pressure. This device is sold with an adult D-ring cuff and suitable for use on adult.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

Page **_ of _**

510(k) Number K122259

14

---

**Source:** [https://fda.innolitics.com/device/K122259](https://fda.innolitics.com/device/K122259)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
