← Product Code [NVP](/productcode/NVP) · K122252

# SPACE CHAMBER PLUS COMPACT SPACE CHAMBER PLUS (K122252)

_Medical Developments International · NVP · Nov 13, 2013 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K122252

## Device Facts

- **Applicant:** Medical Developments International
- **Product Code:** [NVP](/productcode/NVP.md)
- **Decision Date:** Nov 13, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5630
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Space Chamber Plus and Compact Space Chamber Plus are intended to be used by patients who are under the care or treatment of a licensed health care provider or physician. The devices are intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers, prescribed by a physician or health care professional. The intended environments for use include home, hospitals and clinics.

## Device Story

Space Chamber Plus and Compact Space Chamber Plus are valved holding chambers (VHCs) designed to assist patients in inhaling aerosolized medication from pressurized Metered Dose Inhalers (pMDIs). The device consists of a cylindrical chamber with an inhaler interface at one end and a mouthpiece (optionally used with a mask) at the other. A cross-slit valve system regulates airflow; the valve opens during inhalation and closes during exhalation, allowing the patient to visualize valve movement through the transparent chamber to confirm proper inhalation technique. The device is used by patients in home, hospital, or clinical environments. By providing a holding space for the aerosol, the device improves the respirable fraction of medication delivered to the patient compared to pMDI use alone, facilitating consistent dosing and effective treatment for respiratory conditions.

## Clinical Evidence

Bench testing only. Particle size distribution testing (MMAD and respirable fraction) was conducted using three drug classes (ipratropium bromide, albuterol, beclomethasone dipropionate) to compare the subject devices against the predicate. Results showed equivalent respirable fractions and MMAD values with narrow 95% CIs, confirming consistent medication delivery. Additional testing included materials characterization, particulate analysis, drop testing, cleaning validation, and ISO 10993 biocompatibility testing (cytotoxicity, sensitization, irritation, genotoxicity, implantation).

## Technological Characteristics

Valved holding chamber; cylindrical form factor; cross-slit valve system; compatible with standard pMDIs; mouthpiece or mask interface; transparent body for visual feedback; biocompatible materials (ISO 10993 compliant).

## Regulatory Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

## Predicate Devices

- AeroChamber Plus aVHC with Flow-Vu IFI ([K070674](/device/K070674.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K122252

ﺎ

Image /page/0/Picture/1 description: The image shows the logo for Medical Developments International. The logo consists of a stylized plus sign on the left, followed by the text "Medical Developments International" on the right. The text is stacked in three lines, with "Medical" on the top line, "Developments" on the second line, and "International" on the bottom line.

Space Chamber Plus Traditional 510(k)

## 510(k) Summary

| Date Prepared:<br>510(k) Owner:         | 4 November 2013<br>Medical Developments International Limited<br>7/56 Smith Road<br>Springvale VIC 3171<br>Australia                                                                                                                                                                                                                                                                                                                                                   |              |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Submitter/ Official<br>Contact:         | Maggie Oh - Director, Scientific Affairs<br>7/56 Smith Road<br>Springvale, Victoria 3171<br>Australia                                                                                                                                                                                                                                                                                                                                                                  | NOV 1 3 2013 |
|                                         | Phone: +61 3 9547 1888<br>Fax: +61 3 9547 0262                                                                                                                                                                                                                                                                                                                                                                                                                         |              |
| Device Trade Name:                      | Space Chamber Plus<br>Compact Space Chamber Plus                                                                                                                                                                                                                                                                                                                                                                                                                       |              |
| Device Common /<br>Classification Name: | Spacer / Valved Holding Chamber;<br>Nebulizer (Direct Patient Interface)                                                                                                                                                                                                                                                                                                                                                                                               |              |
| Device Classification:                  | 21 CFR868.5630, Class II                                                                                                                                                                                                                                                                                                                                                                                                                                               |              |
| Product Code:                           | NVP                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |              |
| Predicate:                              | AeroChamber Plus aVHC with Flow-Vu IFI (K070674)                                                                                                                                                                                                                                                                                                                                                                                                                       |              |
| Description:                            | The Space Chamber Plus and Compact Space Chamber Plus are devices<br>designed to assist patients with inhaling aerosolized medication dispensed<br>through a Metered Dose Inhaler.<br>The devices consist of a cylindrical chamber with an opening at each end.<br>One opening is for the aerosol medication to be inserted, the other is<br>tapered into a mouthpiece for inhalation. The mouthpiece can be used with<br>or without a mask.                           |              |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |              |
|                                         | The Compact Space Chamber Plus is a smaller version of the Space<br>Chamber Plus that allows for easier storage and handling.                                                                                                                                                                                                                                                                                                                                          |              |
| Intended Use:                           | The Space Chamber Plus and Compact Space Chamber Plus are intended<br>to be used by patients who are under the care or treatment of a licensed<br>health care provider or physician. The devices are intended to be used by<br>these patients to administer aerosolized medication from most pressurized<br>Metered Dose Inhalers, prescribed by a physician or health care<br>professional. The intended environments for use include home, hospitals<br>and clinics. |              |

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Medical Developments International. The logo consists of a black cross symbol on the left, followed by the company name in a sans-serif font. The text is stacked vertically, with "Medical" on top, "Developments" in the middle, and "International" at the bottom.

Comparison of

Technological Characteristics: Space Chamber Plus Traditional 510(k)

| Technology   | Space Chamber<br>Plus and Compact<br>Space Chamber<br>Plus*                                                                                                                                                                                                                       | AeroChamber<br>Plus Flow-Vu<br>Anti-static valved<br>holding chamber                                                                                               | Comparison                                                                                                                                                                                                                                                 |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Valve        | Cross Slit Valve<br>Technology:<br>Confirms a secure<br>seal if the valve<br>opens as the<br>patient inhales<br>and closes as the<br>patient exhales<br>(as observed<br>through the clear<br>transparent body<br>of the spacers)<br>indicating proper<br>inhalation<br>technique. | Flow-Vu indicator:<br>Confirms a secure<br>seal if the<br>indicator moves<br>indicating proper<br>inhalation<br>technique.                                         | Both technologies<br>provide an<br>indication of<br>proper inhalation<br>technique.                                                                                                                                                                        |
| Inhaler base | The orifice in the<br>overmould<br>component is<br>compatible with<br>all standard types<br>of metered dose<br>inhalers.                                                                                                                                                          | Orifice in the<br>backpiece or<br>inhaler base<br>allows insertion<br>of metered dose<br>Inhalers. Flow<br>Signal whistle<br>sound when<br>inhaling too<br>quickly | Patients inhale<br>the amount of<br>drug within the<br>holding chamber.<br>Particle size<br>testing confirms<br>the proposed<br>device delivers an<br>equivalent<br>amount of<br>respirable<br>fraction<br>compared to the<br>predicate (see<br>report in) |

The technological characteristics of the new devices are similar to the legally marketed predicate as summarised below:

The Space Chamber Plus, Compact Space Chamber Plus and the predicate all use a holding chamber and one-way valve system through which metered dose inhaler medication passes to the patient.

Particle size distribution testing confirms the substantial equivalence of respirable dose delivered to the patient between the proposed devices and the predicate.

Clinical Data:

The use of Valved Holding Chambers to assist with the administration of aerosolized medication is proven technology and is well accepted by the medical community. Non-clinical data is sufficient to demonstrate the safety and efficacy of the Space Chamber Plus and Compact Space Chamber Plus devices.

{2}------------------------------------------------

Space Chamber Plus Traditional 510(k)

Image /page/2/Picture/1 description: The image shows a logo for Medical Developments International. The logo consists of a black cross symbol on the left, followed by the text "Medical Developments International" on the right. The text is stacked vertically, with "Medical" on top, "Developments" in the middle, and "International" at the bottom.

## Non-clinical Performance Data:

Evaluation of the new device and the predicate device was performed in accordance with the relevant sections of the CDRH Guidance Document "Reviewers Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH/ODF/DCRD/ADDB -1993).

Particle size distribution testing was performed using 3 drug classes ipratropium bromide (anti-cholinergic bronchodilator), albuterol (beta-agonist bronchodilator) and beclomethasone dipropionate (anti-inflammatory) comparing the Space Chamber Plus and Compact Space Chamber Plus with the predicate. The respirable fraction delivered by the spacers is greater than that delivered by the pMDI alone. The respirable fractions delivered by the Space Chamber Plus and Compact Space Chamber Plus are equivalent to those delivered by the predicate, Aerochamber Plus Flow Vu Anti-static. Likewise, the MMAD values obtained with Space Chamber Plus and Compact Space Chamber Plus are comparable to those obtained with Aerochamber Plus Flow Vu Antistatic. The 95% Cl is narrow for both the respirable fraction and MMAD results, indicating that medication doses can be consistently delivered by the spacers. Particle size distribution testing confirms the proposed devices to be substantially equivalent to the predicate device.

| Device                                  | Particle size,<br>Mean MMAD,<br>μm<br>(95% CI) | Mean total<br>respirable dose,<br>μg<br>(95% CI) | Mean respirable<br>fraction<br>PF% 0.5-4.7<br>(95% CI) |
|-----------------------------------------|------------------------------------------------|--------------------------------------------------|--------------------------------------------------------|
| Albuterol                               |                                                |                                                  |                                                        |
| Albuterol pMDI                          | 2.6 (0.11)                                     | 347.1 (43.3)                                     | 79.7 (3.3)                                             |
| Space Chamber Plus                      | 2.5 (0.04)                                     | 286.9 (36.3)                                     | 86.1 (0.9)                                             |
| Compact Space<br>Chamber Plus           | 2.3 (0.05)                                     | 253.4 (38.7)                                     | 88.0 (0.6)                                             |
| Aerochamber Plus<br>Flow Vu Anti-static | 2.4 (0.03)                                     | 292.2 (19.2)                                     | 87.6 (0.7)                                             |
| Ipratropium Bromide                     |                                                |                                                  |                                                        |
| Ipratropium<br>Bromide pMDI             | 0.72 (0.06)                                    | 51.14 (2.8)                                      | 54.7 (1.7)                                             |
| Space Chamber Plus                      | 0.93 (0.05)                                    | 85.26 (3.3)                                      | 65.3 (0.9)                                             |
| Compact Space<br>Chamber Plus           | 0.80 (0.02)                                    | 65.19 (2.1)                                      | 61.9 (0.7)                                             |
| Aerochamber Plus<br>Flow Vu Anti-static | 0.76 (0.05)                                    | 55.28 (3.0)                                      | 60.0 (1.8)                                             |
| Beclomethasone Dipropionate             |                                                |                                                  |                                                        |
| Beclomethasone<br>Dipropionate pMDI     | 1.4 (0.07)                                     | 218.12 (16.3)                                    | 77.3 (2.4)                                             |

{3}------------------------------------------------

| Space Chamber Plus                      | 1.3 (0.02) | 328.05 (9.9)  | 82.7 (0.8) |
|-----------------------------------------|------------|---------------|------------|
| Compact Space<br>Chamber Plus           | 1.3 (NA^)  | 308.29 (12.7) | 83.7 (1.0) |
| Aerochamber Plus<br>Flow Vu Anti-static | 1.3 (0.03) | 316.25 (17.6) | 83.9 (1.1) |

^ Not applicable (NA) when results for each individual run is the same

Particle size distribution testing was repeated post-cleaning. Performance testing demonstrates that the post-washed Space Chamber Plus and Compact Space Chamber Plus remain equivalent to the pre-washed devices and to the predicate device.

Other performance testing includes materials characterization, particulate analysis, drop test and cleaning validation.

ISO 10993 testing has been performed on the Space Chamber Plus components, including Cytotoxicity, Sensitization, Irritation, Genotoxicity and Implantation tests. The Space Chamber Plus and Compact Space Chamber Plus devices are made of identical materials; the Compact Space Chamber Plus is a smaller version of the Space Chamber Plus that allows for easier storage and handling.

Non-clinical and performance testing demonstrates that the Space Chamber Plus and Compact Space Chamber Plus devices raise no new issues of safety or effectiveness compared to the legally marketed predicate device.

Substantial Equivalence Conclusion:

The new devices, the Space Chamber Plus and Compact Space Chamber Plus, are considered to be substantially equivalent to the predicate device based on the following:

- . they have the same intended use and are indicated for the same user population:
- they have equivalent technological characteristics to the predicate: ●
- they do not raise any new questions of safety or effectiveness; .
- they are at least as safe and effective as the predicate. .

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight. The symbol is composed of three curved lines that form the body and wings of the bird, with three smaller curved lines below representing the tail feathers.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 13, 2013

Medical Developments International Limited Maggie Oh Director, Scientific Affairs 7/56 Smith Road Springvale Victoria 3171 Australia

Re: K122252

Trade/Device Name: Space Chamber Plus Compact Space Chamber Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: NVP Dated: November 8, 2013 Received: November 12, 2013

Dear Ms. Oh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

MarySTBmer -S

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Evaluation Center for Device and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known): KIZZZZZ

Device Name: Space Chamber Plus

Indications for Use:

The Space Chamber Plus is intended to be used by patients who are under the care or treatment of a licensed health care provider or physician. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers, prescribed by a physician or health care professional. The intended environments for use include home, hospitals and clinics.

Device Name: Compact Space Chamber Plus

Indications for Use:

The Compact Space Chamber Plus is intended to be used by patients who are under the care or treatment of a licensed health care provider or physician. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers, prescribed by a physician or health care professional. The intended environments for use include home, hospitals and clinics.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/6/Picture/13 description: The image shows the text "Nayan J. Patel - S" in the top row. Below that, the text "2013.11.13 12:03:43 -05'00'" is displayed. The text appears to be a timestamp or a date and time record, possibly extracted from a document or a system log.

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K122252](https://fda.innolitics.com/device/K122252)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
