K122201 · Med-Tec, Inc. D/B/A Civco Medical Solutions · JAI · Aug 30, 2012 · Radiology
Device Facts
Record ID
K122201
Device Name
PROTURA COUCH SOFTWARE
Applicant
Med-Tec, Inc. D/B/A Civco Medical Solutions
Product Code
JAI · Radiology
Decision Date
Aug 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5770
Device Class
Class 2
Indications for Use
The Protura Couch Software is intended to interface between a record and verify system. linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).
Device Story
Protura Couch Software is a patient positioning interface for radiation therapy. It connects record and verify systems, Linac software, Linac safeguard systems, and image guidance systems to the Protura Couch hardware. The software enables clinicians to control patient positioning with six degrees of freedom from outside the treatment room. By integrating with image guidance systems, it allows for precise patient adjustments based on diagnostic decisions. The system facilitates accurate patient alignment for radiation delivery, potentially improving treatment precision and patient outcomes. It is intended for use in clinical radiation therapy environments.
Clinical Evidence
No clinical data. Bench testing only, covering couch movement, external system interfacing, couch pedestal/isocenter alignment, and patient record handling.
Technological Characteristics
Interface software for radiation therapy couch hardware. Programmed in C#. Runs on Windows XP SP3 (32-bit), Windows 7 (32-bit), and Windows 7 (64-bit). Enables 6-degree-of-freedom positioning control. Connectivity includes record and verify systems, Linac software, and image guidance systems.
Indications for Use
Indicated for use as an interface between radiation therapy systems (record and verify, Linac, safeguard, and image guidance systems) and the Protura 6-degree-of-freedom patient positioning couch to facilitate patient positioning during radiation therapy.
Regulatory Classification
Identification
A powered radiation therapy patient support assembly is an electrically powered adjustable couch intended to support a patient during radiation therapy.
Predicate Devices
PRO SERIES COUCH SOFTWARE, MODEL MT-PRO100 (K034054)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the text "CIVCO" in bold, sans-serif font. Below the company name, the words "MEDICAL SOLUTIONS" are printed in a smaller, sans-serif font, with spacing between the letters. The overall design is clean and professional.
WWW.CIVCO.COM
AUG 30 2012
## Section 5 - 510(k) Summary
| A. Submitter Information | Submitter Name & Address: MED-TEC, Inc. d/b/a CIVCO Medical Solutions<br>1401 8th Street SE<br>Orange City, lowa 51041 |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Amanda Stahle, Regulatory Affairs Specialist<br>Telephone: 319-248-6628, Fax: 877-218-0324<br>amanda.stahle@civco.com |
| Date Summary Prepared: | July 10, 2012 |
| Trade Name:<br>Common Name:<br>Classification Name:<br>Classification Number:<br>Product Code: | Protura Couch Software<br>Treatment Couch Software<br>Powered radiation therapy patient support assembly<br>892.5770<br>JAI |
### B. Predicate Device
MED-TEC, Inc. d/b/a CIVCO Medical Solutions claims the proposed device to be substantially equivalent to the following device:
| 510(k) Number | Device Name | Product<br>Classification/<br>Code | Submitter<br>Name |
|---------------|---------------------------------------------------|------------------------------------|-------------------|
| K034054 | PRO SERIES COUCH<br>SOFTWARE, MODEL MT-<br>PRO100 | 892.5770/JAI | MED-TEC, Inc. |
Both the proposed device and the predicate device are patient positioning software systems that interface with the Couch Hardware (K031866). Six degrees of freedom patient positioning corrections are sent from the software to the Couch Hardware for implementation. The predicate device also interfaces with a treatment planning system whereas the proposed device interfaces with a record and verify system, linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems. Testing has demonstrated that the proposed device is substantially equivalent to the predicate device in regards to safety, effectiveness, and performance, and the additional interfaces provided by the proposed device did not diminish the safety or effectiveness of the device.
| Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | | 0 319.248.6757 | | 3 319.248.6660 |
|------------------------|------------------------|-----------------------|-----------------|--|---------------------|--|---------------------|
| Coralville Office | 2301 Jones Blvd. | Coralville, IA 52241 | USA | | 0 319.248.6757 | | 3 319.248.6660 |
| Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | | 0 +31(0) 182.394495 | | 1 +31(0) 182.395014 |
| Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | | 1 712.737.8688 | | 9 712.737.8654 |
{1}------------------------------------------------
WWW.CIVCO.COM
### C. Device Description
The proposed device, the Protura Couch Software (MT6XSMEU), is patient positioning software used in radiation treatments in conjunction with an Image Guided Radiation Therapy (IGRT) system. The Protura Couch Software interfaces with the Protura Couch Hardware (K031866) and is intended to position the patient after diagnostic decisions have been made based on results from an IGRT system. The Protura Couch Software allows the user to control patient positioning with six degrees of freedom from outside of the treatment room. The Protura Couch Software also includes the ability to interface with a record and verify system, Linac software, Linac safeguard systems, and/or image guidance systems.
### D. Indications for Use/Intended Use
The Protura Couch Software is intended to interface between a record and verify system. linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).
### E. Technological Characteristics
Both the Protura Couch Software and the predicate device are interface software systems that are intended to operate the Couch Hardware (K031866). The predicate device interfaces with a treatment planning system whereas the Protura Couch Software interfaces with a record and verify system, Linac software, Linac safeguard systems, and/or image guidance systems. The Protura Couch Software is programmed in C#; the predicate device is programmed in C, C#, C++, and Java. The Protura Couch Software runs on Windows XP SP3 (32 bit), Windows 7 (32 bit), and Windows 7 (64 bit); the predicate runs on Windows 2000.
### F. Non-Clinical Performance Data
Non-clinical performance testing was conducted for the following characteristics:
- Movement of the Protura Couch .
- Interfacing with External Systems .
- . Couch Pedestal and Isocenter Alignment
- . Patient Record Handling
All testing confirmed that the Protura Couch Software is safe and effective for its intended use.
### G. Clinical Performance Data
No clinical testing was performed in the evaluation of this medical device.
### H. Non-Clinical and Clinical Performance Data Conclusions
The conclusions drawn from the tests are that the Protura Couch Software is substantially equivalent to the predicate device in regards to safety, effectiveness, and performance, and the additional interfaces have not diminished the safety or effectiveness of the device.
| Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | | <span style="font-family: sans-serif;">ⓟ</span> 319.248.6757 | | <span style="font-family: sans-serif;">Ⓕ</span> 319.248.6660 |
|------------------------|------------------------|-----------------------|-----------------|--|-------------------------------------------------------------------|--|-------------------------------------------------------------------|
| Coralville Office | 2301 Jones Blvd. | Coralville, IA 52241 | USA | | <span style="font-family: sans-serif;">ⓟ</span> 319.248.6757 | | <span style="font-family: sans-serif;">Ⓕ</span> 319.248.6660 |
| Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | | <span style="font-family: sans-serif;">ⓟ</span> +31(0) 182.394495 | | <span style="font-family: sans-serif;">Ⓕ</span> +31(0) 182.395014 |
| Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | | <span style="font-family: sans-serif;">ⓟ</span> 712.737.8688 | | <span style="font-family: sans-serif;">Ⓕ</span> 712.737.8654 |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular fashion around the eagle. The eagle is depicted in a simple, black and white design.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms.` Amanda Stahle Regulatory Affairs Specialist MED-TEC, Inc. d/b/a CIVCO Medical Solutions 1401 8th Street SE ORANGE CITY IA 51041
AUG 30 2012
Re: K122201
Trade/Device Name: Protura Couch Software Regulation Number: 21 CFR 892.5770 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: JAI Dated: July 17, 2012 Received: July 25, 2012
Dear Ms. Stahle:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We nave reviewed your Section 910(t) premains is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the Electosule for tegarry manation that of the Medical Device Ameral Food Truc commerce prior to May 26, 1977, the excordance with the provisions of the Federal Food. Drug, devices that have been reciassified in accordines while premarket approval application (PMA). and Cosmetic Act (Act) that to not require approval controls provisions of the Act. The Act. The Act. The Act. The Act. The Act. The Act. The Act. The You may, therefore, market the device, saojeet to decidents for annual registration, listing of
general controls provisions of the Act include requirements for annual regist general controls provisions of the receives and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such If your device is classified (Scc above) into eason affecting your device can be found in Title 21, 2011 additional controls. Existing major regulations a 195. In addition, FDA may publish further
Code of Federal Regulations (CFR), Parts 800 to 1995. In addition, FDA may publ Code of Poderal regarning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuality of a succession with other requirements of the Act
that FDA has made a determination that your devices associes. You must that FDA has made a decemmation that your corred by other Federal agencies. You must or any Federal statutes and regulations and including, but not to: registing (21)
comply with all the Act's requirements, including, but not includes areasting (reporting of comply with all the Act s requirements, therauling, our not many of the one of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice mediem as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as bet form inrketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nonicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you desire possible ad res to: Jon Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Four may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Device Name: Protura Couch Software
Indications for Use: The Protura Couch Software is interface between a record and verify system, linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Arland D. Johns
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) KLaaaøl
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.