← Product Code [MPB](/productcode/MPB) · K122091

# COOK TURBO-FLO HD ACUTE HEMODIALYSIS CATHETER SET/TRAY COOK SPECTRUM TURBO-FLO HD ACUTE HEMODIALYSIS CATHETER SET/TRAY (K122091)

_Cook, Inc. · MPB · Jul 29, 2013 · Gastroenterology, Urology · SESK_

**Canonical URL:** https://fda.innolitics.com/device/K122091

## Device Facts

- **Applicant:** Cook, Inc.
- **Product Code:** [MPB](/productcode/MPB.md)
- **Decision Date:** Jul 29, 2013
- **Decision:** SESK
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Cook Turbo-Flo® HD Acute Hemodialysis Catheter is intended for acute hemodialysis, apheresis and hemofiltration. It is intended for percutaneous insertion into the subclavian, jugular or femoral veins. The device is a short-term use catheter (less than 30 days).

## Device Story

12 Fr. hydrophilically coated, radiopaque polyurethane central venous catheter; two independent, non-communicating lumens; lengths 15, 20, 25 cm. Used for acute hemodialysis, apheresis, and hemofiltration. Inserted percutaneously into subclavian, jugular, or femoral veins by clinicians. Dual-lumen design with staggered termination points facilitates blood flow and minimizes recirculation. Kit includes accessories (needles, dilators, wire guides, syringes, etc.). Short-term use (<30 days).

## Clinical Evidence

Bench testing only. Testing included tensile strength, flow rate, burst pressures, liquid/air leakage, collapsibility, cyclic bending, leakage following clamp cycling, recirculation rate, biocompatibility, sterilization, and performance following aging.

## Technological Characteristics

12 Fr. radiopaque polyurethane central venous catheter; hydrophilic coating; dual non-communicating lumens. Sterilized. Standalone device.

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Two-Lumen Hemodialysis Catheterization Kit with Blue FlexTip® ARROWg+ard Blue ([K993933](/device/K993933.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif font at the top. Below that, the word "MEDICAL" is in a smaller, white, sans-serif font. The background is black.

COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.5.A.
SONE 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM

JUL 2 9 2013

## 510(k) SUMMARY Cook Turbo-Flo® HD Acute Hemodialysis Catheter 21 CFR §807.92 Date Prepared: July 15, 2013

Submitter Information:

Applicant:

Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402 Phone: (800) 468-1379 Fax: (812) 332-0281

Contact:

Sean Spence, RAC Phone: 812-339-2235 x5127 Fax: 812-332-0281

#### Device Information:

| Trade name:     | Cook Turbo-Flo® HD Acute Hemodialysis Catheter |  |
|-----------------|------------------------------------------------|--|
| Common name:    | Blood access device and accessories            |  |
| Classification: | II                                             |  |
| Regulation:     | 21 CFR §876.5540                               |  |
| Product Code:   | MPB: catheter, hemodialysis, non-implanted     |  |

### Predicate Device:

K993933 Two-Lumen Hemodialysis Catheterization Kit with Blue FlexTip® ARROWg+ard Blue . A two-lumen, short-term access central venous hemodialysis catheter.

#### Device Description:

The Cook Turbo-Flo® HD Hemodialysis Catheters are short-term, 12 Fr., hydrophilically coated, radiopaque polyurethane central venous catheters with two independent, non-communicating lumens. Lengths of 15, 20 and 25 cm are available based on the anatomical needs of the patient.

{1}------------------------------------------------

Cook Incorporated - 510(k) Summary Cook Turbo-Flo® HD Acute Hemodialysis Catheter - K122091 15 July 2013

Various components may be included that are associated with placement of the device.

| Components                 |                   |                  |
|----------------------------|-------------------|------------------|
| Introducer Needles         | Drapes            | Antiseptics      |
| Catheter Introducer Needle | Safety Components | Scalpels         |
| Dilators                   | Suture            | Local Anesthetic |
| Wire Guide                 | Syringes          | Tubing           |

### Table 5-1: Set/Tray Components

Needle Free Connector

## Intended Use:

- a. The Cook Turbo-Flo® HD Acute Hemodialysis Catheter is intended for acute hemodialysis, apheresis and hemofiltration. It is intended for percutaneous insertion into the subclavian, jugular or femoral veins. The device is a short-term use catheter (less than 30 days).
## Comparison to Predicates:

The proposed devices are substantially equivalent to the predicates in terms of intended use, duration of use, principles of operation, technological characteristics, insertion method, anatomical location, and method of sterilization.

## Technological Characteristics:

The Cook Turbo-Flo® HD Acute Hemodialysis Catheters are central venous access catheters intended for patients requiring acute hemodialysis, apheresis, and/or hemofiltration. Design elements such as the dual non-communicating lumens and different points of lumen termination support the flow and minimize recirculation. The following tests were conducted to ensure reliable design and performance:

- . Tensile
- . Flow Rate
- Burst Pressures .
- Liquid/Air Leakage ●
- Collapsibility .
- Cyclic Bending ●
- Leakage following Clamp Cycling .
- . Recirculation Rate
- Biocompatibility .
- Sterilization .
- . Performance following Aging

The results of these tests support a conclusion that the proposed Cook Turbo-Flo® HD Acuto Hemodialysis Catheters are as safe and as effective as the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure with three flowing lines representing the body and arms.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 29, 2013

Cook Incorporated % Sean Spence Regulatory Affairs Team Lead 750 Daniels Way, P.O. Box 489 Bloomington, IN 47402

Re: K122091

Trade/Device Name: Cook Turbo-Flo® HD Acute Hemodialysis Catheter Regulation Number: 21 CFR \$ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: MPB Dated: July 15, 2013 Received: July 17, 2013

Dear Sean Spence,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., stcrilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{3}------------------------------------------------

Page 2 - Sean Spence

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

In addition, we have determined that your device kit contains Lidocaine and ChloraPrep, which are subject to regulation as drugs.

Qur substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:

Director, Division of Drug Labeling Compliance Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

{4}------------------------------------------------

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Cook Incorporated - Traditional 510(k) Premarket Notification Cook Turbo-Flo® HD Acute Hemodialysis Catheter - K122091 15 July 2013

#### INDICATIONS FOR USE STATEMENT

K122091 510(k) Number (if known):

Device Name:

Cook Turbo-Flo® HD Acute Hemodialysis Catheter

#### Indications for Use:

The Cook Turbo-Flo® HD Acute Hemodialysis Catheter is intended for acute hemodialysis, apheresis and hemofiltration. It is intended for percutaneous insertion into the subclavian, jugular or femoral veins. The device is a short-term use catheter (less than 30 days).

> Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

## Herbert P. Lerner -S

K122091

---

**Source:** [https://fda.innolitics.com/device/K122091](https://fda.innolitics.com/device/K122091)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).