K122088 · Merit Medical Systems, Inc. · MMX · Dec 5, 2012 · Cardiovascular
Device Facts
Record ID
K122088
Device Name
MERIT ONE SNARE SYSTEM
Applicant
Merit Medical Systems, Inc.
Product Code
MMX · Cardiovascular
Decision Date
Dec 5, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ONE Snare™ endovascular snare system is intended for use in the coronary and peripheral vascular system or hollow viscous to retrieve and manipulate foreign objects. Retrieval and manipulation procedures include indwelling venous catheter repositioning, indwelling venous catheter fibrin sheath stripping, and central venous access venipuncture procedure assistance.
Device Story
Percutaneous retrieval system comprising snare, snare catheter, introducer, and torque device. Snare features nitinol cable and gold-plated tungsten loop; super-elastic construction allows introduction through catheters without deformation. Catheter includes Pebax body and platinum/iridium radiopaque marker band. Operated by physicians in clinical settings to retrieve/manipulate foreign objects or reposition indwelling venous catheters. Physician manipulates snare via torque device under imaging guidance; radiopaque markers facilitate visualization. Output is mechanical retrieval/repositioning of objects/catheters. Benefits include minimally invasive foreign body removal and catheter management.
Clinical Evidence
Bench testing only. Testing included surface finish, torque strength, torqueability, compatibility, dimensional inspection, radiodetectability, tensile testing, tip flexibility, fracture/flexing tests, corrosion resistance, and leakage testing. Biocompatibility testing included L929 MEM elution, Kligman maximization, intracutaneous injection, systemic injection, rabbit pyrogen test, genotoxicity (Salmonella/E. Coli), and hemocompatibility (in vitro and in vivo dog thrombogenicity study). Packaging and sterilization validation performed per ISO standards.
Technological Characteristics
Materials: Nitinol (snare cable), gold-plated tungsten (snare loop), Pebax (catheter), platinum/iridium (marker band). Principle: Mechanical retrieval via super-elastic snare loop. Dimensions: Not specified. Connectivity: None. Sterilization: Ethylene oxide (ISO 11135:2007). Standards: ISO 11070, ISO 10555-1, ISO 594-1/2, ASTM F2096, ASTM F1929, ASTM F88, ASTM D4169, ASTM F1980, ISO 10993-1.
Indications for Use
Indicated for patients requiring retrieval or manipulation of foreign objects in the coronary and peripheral vascular system or hollow viscous, including indwelling venous catheter repositioning, fibrin sheath stripping, and central venous access assistance.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Merit Medical Systems, Inc.
Merit ONE Snare™ system
Traditional Premarket Notification 510(k)
.
4 Section 10 Executive Summary
| | Section $5$ | | DEC | 5 2012 | |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|--------|--|
| 510(k) Summary | | | | | |
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4349<br>(801) 253-6967<br>Stephanie Erskine<br>1721504 | | | |
| | Correspondent Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Ireland Ltd.<br>Parkmore Business Park,<br>Galway, Ireland<br>(353) 91 703 761<br>(353) 91 771 888<br>Mark Mullaney<br>23-Nov-2012<br>9616662 | | | |
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | Merit ONE Snare™ System<br>Percutaneous retrieval device<br>device, Percutaneous retrieval | | | |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | AMPLATZ GOOSE NECK SNARE KIT/<br>CATHETER<br>catheter, embolectomy<br>K972511<br>ev3 Inc. | | | |
| Classification | Class II<br>21 CFR § 870.5150<br>Division of Cardiovascular Devices | | | | |
| Intended Use | The ONE Snare™ endovascular snare system is intended for use in the<br>coronary and peripheral vascular system or hollow viscous to retrieve and<br>manipulate foreign objects. | | | | |
| | | Retrieval and manipulation procedures include indwelling venous catheter<br>repositioning, indwelling venous catheter fibrin sheath stripping, and central<br>venous access venipuncture procedure assistance. | | | |
.T.
{1}------------------------------------------------
K122088
page 2 of 4
Section 10
Executive Summary
ONE Snare™ System contains: (1) Snare, (1) Snare Catheter, (1) Introducer and (1) Torque Device. The snare is constructed of nitinol cable and a gold plated tungsten loop. Device The pre-formed snare loop can be introduced through catheters without risk Description of snare deformation because of the snare's super-elastic construction. The snare catheter is constructed of polyether block amide (Pebax®) and contains a platinum/iridium radiopaque marker band. Technological characteristics of the subject Merit ONE Snare™ system are Technological substantially equivalent to those of the predicate, the currently marketed AMPLATZ GOOSE NECK SNARE KIT/ CATHETER, manufactured by ev3 Characteristics Inc., 510(k) K972511. No performance standards have been established under Section 514 of the Food. Drug and Cosmetic Act for these devices. However, a battery of tests performed on the Merit ONE Snare™ system were designed to demonstrate that the device meets critical design specifications as well as clinical performance attributes for its intended use. Where appropriate, the tests were based on the requirements of the following documents: FDA guidance Coronary and Cerebrovascular Guide Wire . Guidance January 1995. ISO 11070: 1998, Sterile Single-Use Intravascular Catheter . Introducers. ISO 10555-1 1995 Sterile, single-use intravascular catheters, . Part 1 - General Requirements. . ISO 594-1:1996, Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part Safety & 1: General requirements Performance ISO 594-2:1998, Conical fittings with 6% (Luer) taper for ● Tests syringes, needles and certain other medical equipment - Part 2: Lock fittings ASTM F 2096-11; Standard Test Method for Detecting Gross . Leaks in Medical Packaging by Internal Pressurization (Bubble Test) . ASTM F 1929-98 (Reapproved 2004); Standard Test Method for Detecting Seal Leaks in porous Medical Packaging by Dye Penetration ASTM F 88-09; Standard Test Method for Seal Strength of . Flexible Barrier Materials. ASTM D4169 - 09 Standard Practice for Performance Testing ● of Shipping Containers and Systems ASTM F1980-07 Standard Guide for Accelerated Aging of . Sterile Barrier Systems for Medical Devices
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "K122088 Page 3 of 4 Section 10 Executive Summary". The text appears to be part of a document, possibly a report or summary. The page number is 3 out of 4, and the section number is 10. The document is an executive summary.
## Merit Medical Systems, Inc. Merit ONE Snare™ system Traditional Premarket Notification 510(k)
- ISO 11135:2007 Sterilization of health care products --. Ethylene oxide -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 10993-1: 2009, Biological Evaluation of medical Devices Part 1: Evaluation and Testing within a risk management process, and the FDA Modified ISO 10993 Test Profile FDA Memo G95-1.
The Merit ONE Snare™ system was compared to the predicate device for various performance attributes that demonstrate substantial equivalence. The following is a list of all significant testing that was successfully completed:
ONE Snare System (Snare device, Catheter, Introducer and Torque Device)
- Surface Finish, Torque Strength, Torqueability, Compatibility of . snare with snare catheter, Snare Catheter to Guiding Catheter compatibility, Snare introducer fit, Introducer Removal Force from Snare and Dimensional Inspection
#### Snare
- Radiodetectability (Snare wire), Size Designation, Tensile Test, � Tip Flexibility, Fracture Test & Flexing Test and Corrosion resistance
## Catheter & Introducer
- Radiodetectability (Snare catheter / marker bands), Catheter Atraumatic Tip Finish. Marker Band Retention. Force at break (at Catheter Tip and Junction between Catheter Hub and Tube), Catheter Pushability (kink or accordion resistance) & Catheter flexibility (buckling resistance), Hub with Female Luer testing, Dimensions, Freedom from leakage, Surface Finish, Force at break (Junction between Introducer Hub and Tube), Corrosion resistance and Snare Torque Strength.
Biocompatibility testing included
- L929 MEM Elution: MEM extraction .
- Kligman Maximization: Saline and cottonseed oil extractions .
- Irritation Intracutaneous Injection Test Intracutaneous Injection: . Saline and cottonseed oil extractions
- . Systemic Injection: Saline and cottonseed oil extractions
- Material Medicated Rabbit Pyrogen Test: Saline extraction .
- Genotoxicity Salonella typhimurium and Escherichia Coli Reverse t Mutation Assay - ISO
- . Hemocompatibility - Rabbit Blood - Indirect
- Hemocompatibility Rabbit Blood direct .
- Hemocompatibility Complement Activation Assay ISO direct . Contact
- . Hemocompatibility - In Vivo Thrombogenicity study in Dogs
- USP Physicochemical Tests for Plastics USP <661> Non-Volatile . Residue
Safety & Performance Tests (continued)
{3}------------------------------------------------
Merit Medical Systems, Inc.
Merit ONE Snare™ system
Traditional Premarket Notification 510(k)
______________________________________________________________________________________________________________________________________________________________________________
4 Section 10 Executive Summary
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| | Packaging performance before and after exposure to accelerated aging and<br>simulated shipping and handling conditions | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Safety &<br>Performance<br>Tests<br>(continued) | bubble emission dye penetration seal peel tensile strength burst strength visual inspection | |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, design, safety, and performance testing,<br>the subject Merit ONE Snare™ system meets the requirements that are<br>considered essential for its intended use and is substantially equivalent to<br>the predicate device, the currently marketed AMPLATZ GOOSE NECK<br>SNARE KIT/CATHETER, manufactured by ev3 Inc. | |
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three overlapping human profiles, representing the department's focus on people and health.
#### · Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
> DEC 5 2012
Merit Medical Systems, Inc. Mr. Mark Mullaney Regulatory Affairs Manager 1600 West Merit Parkway South Jordan, UT 4095
Re: K122088
Trade/Device Name: Merit ONE Snare System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: MMX Dated: November 22, 2012 Received: November 26, 2012
### Dear Mr. Mullaney:
We have reviewed vour Section 510(k) premarket notification of intent to market the device i referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J.
# Cavanaugh®
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Section 4
## Indications for Use Statement
K122088
510(k) Number (if known):
Device Name: Merit ONE Snare™ System
## Indications for Use:
The ONE Snare™ endovascular snare system is intended for use in the coronary and peripheral vascular system or hollow viscous to retrieve and manipulate foreign objects.
Retrieval and manipulation procedures include indwelling venous catheter repositioning, indwelling venous catheter fibrin sheath stripping, and central venous access venipuncture procedure assistance
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K
vision Sign-Off) vision of Cardiovascular Devices
510(k) Number \$\underline{K*{1} \geq 2*{0} \gamma 8}\$
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.