The MrJ 3300 is indicated for use as a diagnostic imaging device that produces transverse, sagittal, coronal and oblique cross-sectional images that display the internal structure of the following joints: hip, knee, ankle, foot, shoulder, elbow, wrist, hand, calf, thigh, arm, forearm, Temporo Mandibular Joint (TMJ), C-spine, L-Spine with limitation to joint pathologies (no tumors, no angiography). The images produced reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The MR parameters that determine image appearance are proton density, spin-lattice relaxation time (T), spin-spin relaxation time (T2), chemical shift and flow velocity. When interpreted by a trained physician, these images can yield information that can be useful in the determination of a diagnosis.
Device Story
MrJ 3300 is an open-structure MRI system with a 0.3T permanent 'C'-shaped magnet; designed to minimize claustrophobia and installation footprint. Input: magnetic resonance signals from hydrogen nuclei in patient joints. Operation: system generates transverse, sagittal, coronal, and oblique cross-sectional images based on proton density, T1/T2 relaxation times, chemical shift, and flow velocity. Used in hospitals/clinics; operated by trained medical professionals. Output: diagnostic images displayed for physician interpretation to assist in clinical diagnosis of joint pathologies. Benefits: provides non-invasive diagnostic imaging with reduced claustrophobic impact compared to closed systems.
Clinical Evidence
Bench testing only. Performance verified against NEMA standards (MS-1, MS-2, MS-3, MS-5) and IEC safety standards (60601-1, 60601-1-2, 60601-2-33). Clinical acceptability of images was demonstrated through comparison of phantom and healthy volunteer scans against predicate device outputs. No clinical trials on patients were required for this submission.
Technological Characteristics
0.3T permanent 'C'-shaped magnet; 13 MHz working frequency. Includes positioning LED and touch screen. Uses C-UL-US approved fire protection materials for covers. Connectivity: standard MRI imaging chain. Software release 1.A.3. Complies with IEC 60601-1, 60601-1-2, 60601-1-4, and 60601-2-33.
Indications for Use
Indicated for diagnostic imaging of joints (hip, knee, ankle, foot, shoulder, elbow, wrist, hand, calf, thigh, arm, forearm, TMJ, C-spine, L-spine) in patients requiring evaluation of joint pathologies. Contraindicated for tumor imaging and angiography.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
**PARAMed**
MEDICAL SYSTEMS
# 510(K) Summary
Date 05/07/2012
The following Special 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92(c).
# Manufacturer Information
Manufacturer:
Address:
Paramed Srl Corso Perrone 73R 16152 Genova, Italy 3004994584.
Establishment registration number:
# 807.92(a)(1)
Submitter Information Correspondent:
Contact Person
ichard Olson, Correspondent USA Paramed Medical Systems, Inc 6204 W. Oakton Street Morton Grove, IL 60053 Toll Free: 1 866 840-7565 T 1 847 470-0580nF 1 847 470-0612 Luisella De Benedetti Paramed S.r.l. Corso F.M. Perrone 73R 16152 Genova +39 010 6489 358 luisella.debenedetti@paramed.it
## 807.92(a)(2)
Classification Name(s):
Classification Number:
Trade Name:
Common Name:
MrJ 3300
Magnetic resonance diagnostic system System, Nuclear Magnetic Resonance Imaging 21 CFR 892.1000, class II Classification and class of device:
90LNH
## 807.92(a)(3)
Predicate Devices
Section A
| Paramed | MrJ | K033507 |
|---------|---------------------|---------|
| Paramed | MrJ Inspire | K100164 |
| Paramed | MrOpen 05T | K101295 |
| Paramed | Performance Package | K121249 |
10
681245
20
DEC 0 6 2012
{1}------------------------------------------------
## 807.92(a)(4)
ction A
The MrJ 3300 is a magnetic resonance imaging device characterized by an open structure to minize claustrophobic reactions. The magnet is a permanent "C" shaped joke designed to minimize the installation area and the Controlled Access area in order to fit also very small hospital and clinics.
It is indicated for use as a diagnostic imaging device that produces transverse, sagittal. coronal and oblique cross-sectional images that display the internal structure of the following joints: hip, knee, ankle, foot; shoulder, elbow, wrist; hand, calf, thigh, arm, forearm. Temporo Mandibular Joint (TMJ), C-spine, L-Spine with limitation to joint pathologies (no tumors, no angiography). ....
The images produced reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance.
The MR parameters that determine image appearance are proton density, spin-lattice relaxation time (Tl), spin-spin relaxation time (T2), chemical shift and flow velocity. When interpreted by a trained physician, these images can vield information that can be useful in the determination of a diagnosis. ... .
The MrJ 3300 has a 0,3 T magnet, which has a higher intensity than the predicate one (0,2T). Of course the new 0,3 T magnet with slightly different dimensions needs a new design of covers and some modest upgrades due to design ageing of the MrJ Inspire 01-2000-01 product.
New covers of the system have been designed by the same company which designed MRJ Inspire's covers and they have been manufactured using the same C-UL-US approved fire protection materials.
The MrJ 3300 employs both the positioning LED (K033507) and the positioning touch screen . (K100164) to give to the user the maximum chances to get rapidly to the correct patient of positioning.
The scanning sequences are the same employed in MrJ Inspire model (K100164) unless the FLAIR sequences which are equivalent to the same FLAIR sequences employed on the MrOpen 05 T K101295 device.
The MrJ 3300 is furnished with the Performance Package of new concept receiving coils . (code 06-0096-00) equivalent to the Performance Package K121249.
{2}------------------------------------------------
## 807.92(a)(5)
## Device Intended Use(s)
The intended use of Paramed's MrJ 3300 product is the same than the intended use of the predicate one: it is indicated for use as a diagnostic imaging device that produces transverse, sagittal, coronal and oblique cross-sectional images of hip, knee, ankle, foot, shoulder, elbow, wrist, hand, calf, thigh, arm, forearm, Temporo Mandibular Joint (TMJ), C-spine, L-Spine with limitation to joint pathologies (no tumors, no angiography).
The device produces transverse, sagittal, coronal and oblique cross-sectional images, displaying the internal structure of the limbs and joint being imaged. The images that are produced correspond to the spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon the MR parameters (spin-lattice relaxation i time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and chemical shift), If interpreted by a medical expert, these images can provide diagnostically useful information.
### 807.92(a)(6)
Section A
#### Technological Characteristics
The MrJ 3300 MRI system is substantially equivalent to:
- Paramed Mr Inspire K033507 for intended use. � architecture, software and positioning LED system. . . . . . ::::: .
- Paramed Mr Inspire K100164 for intended use, covers' materials, and Touch panel positioning system.
- Paramed MrOpen 05T K101295 for the FLAIR sequences
- Paramed Performance Package K121249 (C-Spine coil, L-Spine coils, MP_wrap coil) complete equivalence except the working frequency which here is centered at 13 MHz due to the higher magnetic field.
12
10/? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ?
Neither the intended use nor the technological characteristics of this model differ from those of the referenced models. A table is supplied later on regarding design similarities.
{3}------------------------------------------------
## 807.92(b)(1)
The discussion of the nonclinical, tests submitted, referenced, or in the premarket notification submission for a determination of substantial equivalence.
The nonclinical tests are summarized in this submission.
The aim of the nonclinical test is to demonstrate performance towards the relevant standards listed in form 3514. :
| Parameter | MrJ 3300 | MrJ Inspire with<br>Extended Kit<br>K100164 |
|------------------------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| IEC 60601-1 Electrical safety | Conform | Conform |
| IEC 60601-1-2 Electromagnetic<br>compatibility | Conform | Conform |
| IEC 60601-1-4 Programmable devices | Conform<br>Sw release n.1.A.3 | Conform<br>Sw release n. 1.A.3 |
| IEC 60601-2-33 Operation mode | Normal Controlled Operation Mode | Normal Controlled<br>Operation Mode |
| IEC 60601-2-33 SAR<br>Whole Body | 0.66 +/- 0.10 W/Kg | 0,57 +/- 0,15 W/Kg |
| Head | 0.9 +/- 0.6 W/Kg | - |
| Exposed Body | 0.66 +/- 0.10 W/Kg | - |
| IEC 60601-2-33 max dB/dt | Conform<br>19,8 T/s with $\tau$ = 600 $\mu$ s<br>Gradient coil diameter 850 mm | Conform<br>19,8 T/s with $\tau$ = 600 $\mu$ s<br>Gradient coil diameter 850<br>mm |
| IEC 60601-2-33 max Acoustic noise<br>Level A-weighted L eq | Conform (<99dBA)<br>89.8 +/- 0.5 dBA | Conform (<99dBA)<br>79.0+/- 0.1 dBA |
| IEC 60601-2-33 max Acoustic noise<br>Peak | Conform (<140 dB)<br>(102.5 +/-0.5) dB | Conform (<140 dB).<br>82.0+/- 0.1 dB |
| NEMA MS-1 SNR multichannel coils | Conform see Laboratory testing for<br>the results | Conform |
| NEMA MS-3 Image uniformity | Conform see Laboratory testing for<br>the results | Conform |
| NEMA MS-1 SNR for single channel<br>coils | Conform see Laboratory testing for<br>the results | Conform |
| NEMA MS-2 Geometric Distortion | Conform see Appendix L | Conform |
| NEMA MS-5 SLICE THICKNESS | Conform see Appendix L | Conform |
The non clinical test demonstrate that the new tomograph respects the standards as the previous models did.
73315/295
Differences in the results are due to the new magnet adopted by this model.
Section A
{4}------------------------------------------------
# 807.92(b)(2)
: : : :
A brief discussion of the clinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence. This discussion shall include, where applicable, a description of the subjects upon whom the device was tested, a discussion of the safety on effectiveness data obtained from the testing, with specific reference to adverse effects and complications, and any other information from the clinical testing relevant to a determination of substantial equivalence. .:.
The proposed images (see clinical output par.) are clinically acceptable as those presented for the same districts in previous 510 (k) files to which equivalence is claimed.
| Coil | Limb | Standard/Optional coils available with MrJ 3300 confronted with corresponding coils available on<br>MrJ Inspire K100164 system |
|---------------|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Knee | Knee | Equivalent |
| Hip | Hip | Equivalent |
| Hand | Hand | Equivalent |
| Shoulder | Shoulder | Equivalent |
| TMJ | TMJ | Equivalent |
| Knee large | Knee large | Equivalent |
| VS | Hand | Equivalent |
| VL | Thigh | Equivalent |
| VXL | L-Spine small dimensions' patient | Equivalent |
| C-Spine | C-Spine | Equivalent |
| L-Spine small | L-Spine on small patient | Equivalent |
| L-Spine Large | L-Spine on big patient | Equivalent |
| MP Wrap | Knee | Equivalent |
. : :
.
{5}------------------------------------------------
| Sequence used<br>with MrJ3300 | Sequence used with<br>MrJ Inspire K100164 | District | Decision with reference to MrJ<br>Inspire |
|-------------------------------|-------------------------------------------|----------|-------------------------------------------|
| SCOUT | SCOUT | Knee | Equivalent |
| SE | SE | Knee | Equivalent |
| E-RASE | E-RASE | Hand | Equivalent |
| DE | DE | Knee | Equivalent |
| GFE | GFE | Knee | Equivalent |
| STIR | STIR | Knee | Equivalent |
| STIR GFE | STIR GFE | Knee | Equivalent |
| SE T2 | SE T2 | Hand | Equivalent |
| 3D GFE | 3D GFE | Hand | Equivalent |
| 3D GFE SPOILED | 3D GFE SPOILED | Knee | Equivalent |
| 3D GFE EMIT | 3D GFE EMIT | Knee | Equivalent |
| 3D GBASS | 3D GBASS | L-Spine | Equivalent |
| RISE T2 | RISE T2 | Hand | Equivalent |
| RIDE | RIDE | Shoulder | Equivalent |
| FAST RISE | FAST RISE | C-Spine | Equivalent |
| FIR | FIR | L-Spine | Equivalent |
| FWS GFE | FWS GFE | Knee | Equivalent |
| Sequence used<br>with MrJ3300 | Sequence used with<br>MrOpen 05T<br>K101295 | District | Decision with reference to<br>MrOpen 05T |
|-------------------------------|---------------------------------------------|----------|------------------------------------------|
| FLAIR T1 | FLAIR T1 | C-Spine | Equivalent |
| FLAIR T2 | FLAIR T2 | C-Spine | Equivalent |
## 807.92(b)(3)
The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph (a)(3) of this section. ... . . :: : ・・・・・・
On the basis of the internal final inspection data reports, summarized in the above 807.92(b)(1) and (b)(2) point and of the test images performed both on phantoms and on healthy volunteers, we declare that the MrJ 3300 device is at least as safe and effective as the predicate devices.
:::::::
71243
{6}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure embracing a globe, with flowing lines suggesting movement or connection.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Paramed Srl % Ms. Luisella De Benedetti Quality Manager Paramed Srl Corso F.M. Perrone 73R 16152 Genova, Italy
December 6, 2012
Re: K122034
Trade/Device Name: MrJ 3300 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II · Product Code: LNH Dated: November 14, 2012 Received: November 19, 2012
Dear Ms. De Benedetti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{7}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please . note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
Janine M. Morris, M.S. Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known):
MrJ 3300 code 06-2000-00 Device Name:
Indications for Use:
The MrJ 3300 is indicated for use as a diagnostic imaging device that produces transverse, sagittal, coronal and oblique cross-sectional images that display the internal structure of the following joints: hip, knee, ankle, foot, shoulder, elbow, wrist, hand, calf, thigh, arm, forearm, Temporo Mandibular Joint (TMJ), C-spine, L-Spine with limitation to joint pathologies (no tumors, no angiography).
The images produced reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance.
The MR parameters that determine image appearance are proton density, spin-lattice relaxation time (T), spin-spin relaxation time (T2), chemical shift and flow velocity. When interpreted by a trained physician, these images can yield information that can be useful in the determination of a diagnosis.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Michael D. O'Hara
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K122034
Page I of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.