The ARCHITECT Total Bilirubin assay is used for the quantitation of total bilirubin in human serum or plasma on the ARCHITECT c8000 system. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block. A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Device Story
ARCHITECT Total Bilirubin is an in vitro diagnostic assay for use on the ARCHITECT c8000 system. It measures total bilirubin (conjugated and unconjugated) in human serum or plasma. Principle of operation: colorimetric chemical reaction where bilirubin couples with diazo reagent in the presence of a surfactant to form azobilirubin. The device measures the increase in absorbance at 548 nm, which is directly proportional to total bilirubin concentration. Operated by laboratory personnel in clinical settings. Output is a quantitative concentration value used by physicians to diagnose and monitor liver, hemolytic, and metabolic disorders, and to assess kernicterus risk in neonates.
Clinical Evidence
Bench testing only. Precision studies (N=80, 20-day; N=50, 5-day) showed total %CV between 1.0% and 3.0%. Linearity confirmed across 0.1–25 mg/dL range (r=1.00). Method comparison with predicate (N=138 adults) yielded slope 0.98, intercept 0.07, r=0.9991. Comparison with Roche assay (N=54 neonates) yielded slope 0.99, intercept 0.32, r=0.9967. Interference testing showed no significant interference from hemoglobin (up to 2000 mg/dL); Intralipid (>750 mg/dL) and Indican (>0.175 mmol/L) interfere.
Indicated for quantitative analysis of total bilirubin in human serum or plasma of adults and neonates to aid in diagnosis/treatment of liver, hemolytic hematological, and metabolic disorders (e.g., hepatitis, gall bladder block) and to assess risk of bilirubin encephalopathy (kernicterus) in newborns.
Regulatory Classification
Identification
A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k121985
B. Purpose for Submission:
Modification of a previously cleared total bilirubin assay (k060574) – reduced sample size from 4.0 uL to 2.6 uL.
C. Measurand:
Total Bilirubin
D. Type of Test:
Quantitative, enzymatic, colorimetric assay
E. Applicant:
Abbott Laboratories
F. Proprietary and Established Names:
ARCHITECT Total Bilirubin
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CIG | Class II | 21 CFR §862.1110, Bilirubin (total or direct) test system | Clinical Chemistry (75) |
H. Intended Use:
1. Intended use(s):
See indications for use below.
{1}
2. Indication(s) for use:
The ARCHITECT Total Bilirubin assay is used for the quantitation of total bilirubin in human serum or plasma on the ARCHITECT c8000 system. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological and metabolic disorders, including hepatitis and gall bladder block.
A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Abbott ARCHITECT c8000
I. Device Description:
The total bilirubin reagent kit is packaged as two liquid, ready-to-use reagents, R1 and R2. The following reactive ingredients are presented in the following table:
| R1 | Reactive Ingredients | Concentration |
| --- | --- | --- |
| | Surfactants | 4.51% |
| | HCl | 8.204 g/L |
| R2 | 2,4 - dichloroaniline | 0.81 g/L |
| | HCl | 5.563 g/L |
| | Sodium Nitrite | 0.345 g/L |
| | Surfactant | 1.96% |
This device is the same as the device cleared previously in the Abbott Total Bilirubin assay (k060574). The only claimed change to the device is the sample volume (4.0μL to 2.6 μL).
J. Substantial Equivalence Information:
1. Predicate device name(s):
Abbott Total Bilirubin
{2}
2. Predicate 510(k) number(s):
k060574
3. Comparison with predicate:
Similarities and Differences
| Characteristic | Predicate device: Abbott Total Bilirubin (k060574) | Candidate device: ARCHITECT Total Bilirubin |
| --- | --- | --- |
| Intended Use | The Total Bilirubin assay is used for the quantitative analysis of total bilirubin in human serum or plasma of adults and neonates | Same |
| Test Method | Coupling with the diazo reagent in the presence of a surfactant to form azobilirubin and detection at 548 nm. | Same |
| Detection of Analyte | End-point colorimetric | Same |
| Calibration Curve Type | 3 point curve | Same |
| Specimen Type | Serum or plasma | Same |
| Patient population | Adults and Neonates | Same |
| Platform | ARCHITECT c8000 Analyzer | Same |
| Measuring Interval | 0.1 to 25.0 mg/dL | Same |
| Sample size | 4.0 μL | 2.6 μL |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI documents:
Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline (C28-A3)
Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline (EP5-A2)
{3}
Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach (EP6-A)
Interference Testing in Clinical Chemistry; Approved Guideline (EP 7-A2)
Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP9-A2)
Protocols for Determination of Limits of Detection and Limits of Quantitation (EP17-A)
## L. Test Principle:
The total bilirubin assay is a quantitative colorimetric measurement of total bilirubin in human (adult and neonates) serum or plasma. Total (conjugated and unconjugated) bilirubin couples with the diazo reagent in the presence of a surfactant to form azobilirubin. The increase in absorbance at 548 nm due to azobilirubin is directly proportional to the total bilirubin concentration.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Precision was performed for 20 days using one lot of reagent, one lot of calibrators, and one lot of controls following guidance outlined in CLSI EP5-A2. All testing was performed on the ARCHITECT c8000. Four control levels (level 1, level 2, level 3, and level 4) were tested in replicates of two, in two runs per day for 20 days (N=80). Results are summarized in the following table:
20 day precision for serum controls
| Matrix | Level | N | Mean (mg/dl) | Within-run | | Between-run | | Total precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV |
| Serum Controls | 1 | 80 | 0.75 | 0.01 | 1.23 | 0.00 | 0.43 | 0.01 | 1.96 |
| | 2 | 80 | 4.11 | 0.02 | 0.48 | 0.03 | 0.76 | 0.06 | 1.43 |
| | 3 | 80 | 5.86 | 0.03 | 0.43 | 0.02 | 0.41 | 0.07 | 1.20 |
| | 4 | 80 | 15.70 | 0.05 | 0.31 | 0.08 | 0.52 | 0.16 | 1.00 |
An additional 5 day precision study was performed using patient samples, with one lot of reagent, one lot of calibrators, and one lot of controls. All testing was preformed on the Architect c8000. Three levels of serum concentrations were tested in replicates of five for five days (N=50). Results are summarized in the following table:
{4}
| Matrix | Level | N | Mean | Total SD | Total % CV |
| --- | --- | --- | --- | --- | --- |
| Serum | 1 | 50 | 0.89 | 0.027 | 3.0 |
| | 2 | 50 | 9.52 | 0.159 | 1.7 |
| | 3 | 50 | 22.14 | 0.219 | 1.0 |
# b. Linearity/assay reportable range:
A linearity study was performed following guidance from CLSI EP6-A. Nine samples were prepared by diluting a high patient sample with a low patient sample to obtain concentrations that span the entire claimed measuring range. Each serum sample was tested in replicates of four using one lot of reagent on the ARCHITECT c8000 instrument. The concentrations of samples tested spanned the claimed measuring range 0.1 to $25\mathrm{mg / dL}$ . Three linearity studies were conducted with the following components of total bilirubin: $75\%$ Conjugated and $25\%$ Unconjugated, $50\%$ Conjugated and $50\%$ Unconjugated, $25\%$ Conjugated and $75\%$ Unconjugated. The observed values were plotted against the expected values and an appropriate line fitted by standard linear regression. The following is a representative summary of results from these studies:
| | Serum |
| --- | --- |
| Correlation (r) | 1.00 |
| Slope | 0.98 |
| Intercept | 0.06 |
The linearity data support the sponsor's claim that the measuring range for serum is $0.1 - 25\mathrm{mg / dL}$
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrators were previously cleared- See k060574 for traceability, stability, and expected value information.
# d. Detection limit:
The sponsor conducted a Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) study following the CLSI EP-17A guideline. The LoB was determined using three ARCHITECT instruments (N=60 per instrument) using three blank samples. This study was conducted over one day. LoD and LoQ study was performed on two ARCHITECT c8000 Systems in 5 runs over 3 days with several low samples. LoQ is defined as
{5}
the value obtained with $\leq 20\%$ CV. The following is a summary of results:
| Device | Reportable range | LoB | LoD | LoQ |
| --- | --- | --- | --- | --- |
| Total Bilirubin | 0.1 – 25 mg/dL | 0.01 mg/dL | 0.05 mg/dL | 0.07 mg/dL |
These detection limit studies support the claimed measuring range $0.1 - 25$ mg/dL.
# e. Analytical specificity:
A study of common interferents was conducted. Solutions of human serum albumin were spiked with two levels of bilirubin (a high normal total bilirubin and an abnormal high total bilirubin). These two solutions were spiked with varying levels of interferent. The spiked samples were tested in seven replicates and these results were compared to seven replicates of a reference sample (without spiked interferent). Interference was evaluated at two medical decision points for total bilirubin. The percent recovery was determined by dividing the mean result of replicates for each interferent level by the mean result of the replicates for the reference sample. The sponsor's acceptance criteria was set at $\pm 10\%$ or $\pm 0.3\mathrm{mg / dL}$ difference between the interferent results and the reference results, whichever is greater (The acceptance criteria of $\leq 10\%$ or $\pm 0.3\mathrm{mg / dL}$ is applied using the $\leq 10\%$ criteria first. If the percentage is $>10\%$ , the observed vs. target values must be within $0.3\mathrm{mg / dL}$ ). The following is a summary of results:
Intralipid at concentrations $>750\mathrm{mg / dL}$ have been shown to interfere with the measurement of Total Bilirubin at the target concentration of 1.2744 $\mathrm{mg / dL}$ . Indican concentrations $>0.175\mathrm{mmol / L}$ have been shown to interfere with Total Bilirubin target concentrations at $1.3241\mathrm{mg / dL}$ . Hemoglobin tested up to $2000\mathrm{mg / dL}$ do not interfere with the total bilirubin assay.
Other interferences from medications or endogenous substances may affect results and the sponsor refers user to the literature for more information in the labeling.
Literature cited: Young, D.S. Effects of Preanalytical Variables on Clinical Laboratory Tests, $2^{\mathrm{nd}}$ ed. Washington DC: AACC Press; 1997:3-85
# f. Assay cut-off:
Not Applicable
{6}
# 2. Comparison studies:
# a. Method comparison with predicate device:
A method comparison study was conducted using 138 adult serum samples ranging from 0.11 to $23.96\mathrm{mg / dL}$ . All samples were tested on the Architect c8000 System and the candidate device was compared against the predicate device. The following table provides a summary of the regression analysis:
Adult Sample comparison:
| Sample type | N | Range (mg/dL) | Slope | Intercept | Correlation Coefficient |
| --- | --- | --- | --- | --- | --- |
| Serum | 138 | 0.11 – 23.96 | 0.98 | 0.07 | 0.9991 |
In addition, 54 neonate serum samples ranging from 0.94 to $19.05\mathrm{mg / dL}$ were compared using the candidate device evaluated on the Architect c8000 System against another FDA cleared device (Roche total bilirubin on Hitachi 717). The following table provides a summary of the regression analysis:
Neonatal Sample comparison:
| Sample type | N | Range (mg/dL) | Slope | Intercept | Correlation Coefficient |
| --- | --- | --- | --- | --- | --- |
| Serum | 54 | 0.94 – 19.05 | 0.99 | 0.32 | 0.9967 |
# b. Matrix comparison:
Thirty subjects' total bilirubin results using glass serum tube (control) were compared to plastic K2 EDTA, Lithium heparin with and without a gel barrer, and sodium heparin tubes. All the results generated from the plastic tubes (SST tube, lithium heparin, sodium heparin, and plasma lithium heparin with gel separator) had less than $10\%$ differences when compared to the glass tube (control). Total bilirubin concentrations tested range from 0.1 to $25\mathrm{mg / dL}$ . Passing-Bablok regression analysis results are summarized below:
{7}
| | | | Passing-Bablok | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Slope | | Intercept | |
| Tube Type | N | Corr Coef | Slope | 95% CI | Intercept | 95% CI |
| Serum Separator Tube (SST) | 30 | 0.9999 | 1.00 | (0.99, 1.01) | 0.01 | (0.00, 0.02) |
| K2 EDTA Plasma Tube (non gel) | 30 | 0.9995 | 1.05 | (1.02, 1.07) | -0.01 | (-0.10, 0.00) |
| Lithium Heparin Plasma Tube (non gel) | 30 | 0.9998 | 1.00 | (0.99, 1.01) | 0.01 | (-0.01, 0.01) |
| Lithium Heparin Plasma Tube (Plasma Separator Tube PST) with gel | 30 | 0.9999 | 1.00 | (1.00, 1.00) | 0.00 | (-0.00, 0.01) |
| Sodium Heparin Plasma Tube (non gel) | 30 | 0.9999 | 1.00 | (0.99, 1.00) | 0.00 | (0.00, 0.01) |
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
{8}
Adult Range: A reference range confirmation study was performed following guidance from CLSI C28-A3 using 26 samples. The study confirms with a 95% confidence interval an adult reference range of 0.2 to 1.2 mg/dL.
The expected values for the neonates are based on the literature*
| Reference Range | Premature Newborn: < 24 Hours < 8.0 < 48 Hours < 12.0 3 to 5 Days < 15.0 7 Days < 15.0 Full Term Newborn: < 24 Hours < 6.0 < 48 Hours < 10.0 3 to 5 Days < 12.0 7 Days < 10.0 Adult: 0.2 to 1.2 mg/dL |
| --- | --- |
The following reference was cited: *American Academy on Pediatrics Subcommittee on Hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics 2004; 114(1):297-316.
## N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.