K121929 · Diacor, Inc. · FRZ · Jan 17, 2013 · General Hospital
Device Facts
Record ID
K121929
Device Name
ZEPHYR 'X-SERIES PATIENT TRANSFER SLED
Applicant
Diacor, Inc.
Product Code
FRZ · General Hospital
Decision Date
Jan 17, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6775
Device Class
Class 2
Indications for Use
The Zephyr "x-series" Patient Positioning and Transfer System is indicated to aid in the support, positioning, and transfer of a patient for procedures involving imaging, including MRI, and external beam radiation therapy treatment with electrons, photons or protons, as well as other procedures requiring transfer of a patient.
Device Story
System functions as accessory for patient support, positioning, and transfer during imaging and radiation therapy. Utilizes forced air as low-friction bearing to transfer patients between flat surfaces, eliminating manual lifting. Comprises Kevlar base board and low-pressure air pad system. Used in clinical settings (imaging suites, radiation therapy departments); operated by healthcare personnel. Provides stable platform for therapy/imaging; reduces physical strain on staff during patient movement. Benefits patient by facilitating safe, efficient positioning and transfer.
Clinical Evidence
Bench testing only. Attenuation measurements (6MV/10MV photons) showed average factor of 3.33. Build-up measurements equivalent to 6mm tissue. MR safety testing confirmed no detectable magnetic pull on board; blower hose showed slight attraction but remained manageable.
Technological Characteristics
Kevlar base board; low-pressure air pad system; forced air bearing. MR safe (excluding air blower). Dimensions/materials produce dose buildup equivalent to 6mm tissue. Standalone system.
Indications for Use
Indicated for patients requiring support, positioning, or transfer during imaging (including MRI) or external beam radiation therapy (electrons, photons, or protons). No specific age or gender contraindications listed.
Regulatory Classification
Identification
A powered patient transfer device is a device consisting of a wheeled stretcher and a powered mechanism that has a broad, flexible band stretched over long rollers that can advance itself under a patient and transfer the patient with minimal disturbance in a horizontal position to the stretcher.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
Predicate Devices
Patient Transfer Systems, Inc., AirPal Platform Powered Patient Transfer Device (510(k) Exempt)
{0}------------------------------------------------
K121929
Image /page/0/Picture/1 description: The image shows the word "DIACOR" in white letters against a black background. The letters are bold and sans-serif. There is some noise and distortion in the image, particularly around the edges of the letters and the black background. The word appears to be a logo or brand name.
Precision comes standard." Since 1983.
JAN 1 7 2013
2550 Decker Lake Blvd. Suite 26 Salt Lake City, Utah 84119
> Section 5 510(k) Premarket Notification Summary (As required by 21 CFR 807.92)
## A. General Information
Date Prepared: December 19, 2012 Submitter's Name: Diacor Inc. 2550 Decker Lake Blvd. Suite 26, West Valley City, UT 84119 Address: Telephone No: 801-467-0050 Fax: 801-487-3258 Tamara Tialavea Contact Person:
Establishment Registration Number: 1721113
| Device Trade name: | Diacor Zephyr “x-series” Patient Positioning and Transfer System |
|-----------------------------|------------------------------------------------------------------------------|
| Device Common name: | Patient Positioning and Transfer System |
| Device Classification name: | 1. Accessory to Medical Charged-Particle Radiation Therapy System - 892.5050 |
| | 2. Accessory Magnetic Resonance Diagnostic Device - 21 CFR 892.1000 |
| | 3. Powered patient transfer device - 880.6775 |
#### B. Legally Marketed Predicate Devices
- 1. Patient Transfer Systems, Inc., AirPal Platform Powered Patient Transfer Device, Product Code FRZ, Regulation Number: 880.6775 (510(k) Exempt)
- 2. CIVCO MR & Radiological Patient Positioning Devices K111340
#### ﺯ Device Description
The Zephyr "x-series" Patient Positioning and Transfer System functions as an accessory to support a patient and positioning devices during imaging and radiation therapy procedures, and other procedures requiring the transfer of a patient. The Zephyr "x-series" Patient Positioning and Transfer System utilizes forced air as a low-friction bearing to transfer patients from one flat surface to another, eliminating the need to manually lift the patient.
{1}------------------------------------------------
# Indications for Use/Intended Use
The Zephyr "x-series" Patient Positioning and Transfer System is indicated to aid in the support, positioning, and transfer of a patient for procedures involving imaging, including MRI, and external beam radiation therapy treatment with electrons, photons or protons, as well as other procedures requiring transfer of a patient.
### Technological Characteristics E.
The Zephyr "x-series" Patient Positioning and Transfer System uses the same technology as the two predicate devices combined: 1) a patient positioning Kevlar base board (predicate CIVCO Kevlar base plate) and, 2) a low pressure air pad patient transfer system (predicate AirPal Patient Transfer); therefore, Diacor claims substantial equivalence.
### F. Testing
### Clinical Testing: N/A
Animal Testing: N/A
Bench Testing: Bench testing for attenuation, build-up, and MR safety was conducted. Attenuation measurements were taken using a Unidose-E electrometer with a Farmer Chamber FC65-G placed at a depth of 10cm in solid water. Measurements were made with and without the base board using both 6MV and 10MV clinical photon energies. The average attenuation factor measured 3.33, with a median of 2.65.
Build-up measurements were also taken using a Unidose-E electrometer with a PTW Advanced Markus Plane Parallel Chamber at various depths in solid water. Measurements were made with and without the base board. The results show that the materials and dimensions used in the board produce a dose buildup effect with an approximate equivalence to 6mm of tissue.
The Zephyr "x-series" Patient Positioning and Transfer System was tested for MR safety with a magnet and hand-held metal detector. The board was then positioned on the MR patient table and slid into the MR without any detectable pull by the magnet. The blower hose did show a slight attraction when directly next to the magnet. It was placed into the bore near the isocenter and was easily removed.
Conclusion: Based on our industry knowledge and experience, both attenuation and buildup test results were found to be in line with patient positioning devices that are readily found on the market. All components, excluding the air blower, tested MR Safe and are labeled accordingly.
### i - D.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 17, 2013
Ms. Tami Tialayea Chief Operations Officer Diacor, Inc. 2550 Decker Lake Blvd., Suite 26 WEST VALLEY CITY UT 84119
Re: K121929
Trade/Device Name: Zephyr "X-series" Patient Positioning and Transfer System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LNH, LHN, IYE, AND FRZ Dated: November 14, 2012 Received: November 16, 2012
# Dear Ms. Tialavea:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the idence for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drig and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, or annual registations against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleding.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Fedral Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.
{3}------------------------------------------------
Page 2-Ms. Tialavea
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm/15809.html
the Center for Devices and Podical Houles/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH)CDRHOMICE (CDRH) for
note the regulation entitled "Michaeling (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the manner (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K121929
Device Name: Zephyr "x-series" Patient Positioning and Transfer System
Indications for Use:
The Zephyr "x-series" Patient Positioning and Transfer System is indicated to aid in the support, positioning, and transfer of a patient for procedures involving imaging, including MRI, and external beam radiation therapy treatment with electrons, photons or protons, as well as other procedures requiring transfer of a patient.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
. AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
| | Michael D. O'Hara |
|--------|-------------------------------------------------------|
| | (Division Sign Off) |
| | Division of Radiological Health |
| | Office of In Vitro Diagnostic and Radiological Health |
| 510(k) | K121929 |
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.