The sponsor utilized results from a published prospective randomized trial to demonstrate that a 12 cm sling length is as safe and efficacious as the original transobturator technique.
The original versus a modified inside-out transobturator procedure: 1-year results of a prospective randomized trial (de Leval J, et al.); Prospective randomized trial; Follow-up/Duration: 1 year
Females with stress urinary incontinence
Original transobturator technique
Safety, efficacy, and post-operative groin pain
Indications for Use
Desara SL is intended to be used in females to position a mesh for the treatment of Genuine Stress Urlnary Incontinence (SUI), mixed incontinence resulting from urethral hypermobility or Intrinsic sphincter deficiency.
Device Story
Desara SL is a sterile, single-use 1.1 x 12cm monofilament polypropylene mesh sling; used to provide pelvic support for female stress or mixed urinary incontinence. Device features mid-line marker, integral sleeves, tips, and sutures for placement; components removed post-placement. Used with Caldera Medical reusable transobturator helical, hook, and inside-out introducers via Universal Connection System. Surgeon selects preferred surgical approach; device positioned to support urethra. Shorter length (12cm) compared to traditional mesh reduces dissection requirements and avoids exit through upper thigh/groin incisions, potentially reducing post-operative groin pain while maintaining efficacy.
Clinical Evidence
Bench testing only. Evaluated mesh thickness, knit characteristics, pore size, density, tensile strength, stiffness, flexural rigidity, tear resistance, burst strength, and suture pullout. Biocompatibility verified per ISO-10993. Reprocessing validated for reusable instrumentation. Literature cited (de Leval et al., 2011) supports safety and efficacy of 12cm sling length in modified inside-out transobturator procedures, noting reduced post-operative groin pain compared to original techniques.
Technological Characteristics
Monofilament polypropylene mesh; 1.1 x 12cm dimensions. Reusable stainless steel instrumentation (ASTM F899-09). Sterile, single-use implant. Biocompatibility per ISO-10993. Accelerated aging per ASTM F-1980-07. Mechanical testing includes sleeve-to-tip pull-off, suture loop pull-out, and mesh break strength.
Indications for Use
Indicated for females with Genuine Stress Urinary Incontinence (SUI) or mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}------------------------------------------------
Pg. 1 of 2
# 510(k) Summary #K121928
| Date of Summary: | February 18, 2013 | FEB 19 2013 |
|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Applicant: | Caldera Medical, Inc.<br>5171 Clareton Drive<br>Agoura Hills, CA 91301<br>P (818) 879-6555<br>F (818) 879-6556 | |
| Contact: | Vicki Gail<br>QA/RA and Operations Manager<br>P (818) 483-7602<br>F (818) 736-9083<br>E vgail@calderamedical.com | |
| Trade Name:<br>Device Class:<br>Product Code:<br>C.F.R Section:<br>Common name:<br>Classification Panel: | Desara® SL<br>Class II<br>OTN<br>21 CFR 878.3300<br>Surgical Mesh, Gynecologic, For Stress Urinary Incontinence, Female<br>Obstetrics/Gynecology Panel | |
| Predicate Devices: | Desara Mesh, K112609 (Primary)<br>Gynecare TVT Abbrevo Continence System, K100936 (Secondary) | |
### Description of Device:
Desara® SL is a sterile, single-use pubourethral 1.1 x 12cm silng used to provide support in the pelvic region to treat stress urinary Incontinence or mixed Incontinence. The device is manufactured out of a monofilament polypropylene yarn, which is knitted into a mesh. The device has a mid-line marker, integral sleeves, tips and sutures to assist the surgeon in placement of the device. The sleeves, tips and sutures are removed after placement of the device.
Desara® SL Is designed to work with Caldera Medical's reusable transobturator helical, hook, and Inside-out introducers, which utilize a Universal Connection System offering surgeons the flexibility to choose their preferred surgical approach.
## Intended Use of Device:
Desara SL is intended to be used in females to position a mesh for the treatment of Genuine Stress Urlnary Incontinence (SUI), mixed incontinence resulting from urethral hypermobility or Intrinsic sphincter deficiency.
### Summary of Technological Characteristics:
The Desara SL and Desara Mesh do not have the same technological characteristics. The new technological characteristics of the Desara SL include its comparatively shorter length (12 cm versus 45 cm), and therefore, different surgical procedure. Specifically, Desara SL mesh does not exit or reach the upper thigh/groin incisions like the Desara Mesh. The upper thigh/groin incisions required are only used to adjust the mesh position during the surgical procedure. The new technological characteristics of the Desara SL could affect safety and effectiveness because the Desara SL does not extend beyond the obturator foramen like the Desara Mesh. Therefore, no tissue ingrowth or granulation will occur between the obturator foramen and upper thigh/groin Incisions to provide additional support to the implant. This could affect the effectiveness of the Desara SL compared to the Desara Mesh. In regards to safety, the Desara SL requires less dissection than Desara Mesh. The new technological characteristics of Desara SL do not ralse new Issues of safety and effectiveness because the FDA has previously cleared a
{2}------------------------------------------------
K121928
Pg. 2 of 2
hybrid sling, the TVT Abbrevo. TVT Abbrevo has the same length and nearly the identical implantation procedure as the Desara SL.
## Performance Summary
Desara® SL Is comprised of the same mesh as the predicate device, Desara® Mesh, (K112609), also a product of Caldera Medical. In accordance with the FDA's Guidance for the Preparation of a Premarket Notification Application for Surgical Mesh, the following mesh characteristics were assessed: mesh thickness, mesh knit characteristics, pore size, mesh density, tensile strength, mesh stiffness, flexural rigidity, tear resistance, burst strength, suture pullout and pyrogen levels.
The Desara® device was used as a reference for Desara® SL biocompatibility since both are comprised of the same materials. Desara® has passed all biocompatibility testing as indicated per the FDA quidance documents, FDA Device Advice, Guidance Documents (Medical Device and Radiation Emitting Product), 3. Biocompatibility and FDA Blue Book Memorandum #G95-1 Entitled "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". In addition, ASTM F899-09 Standard Specification for Wrought Stainless Steels for Surgical Instruments, was referenced to support the biocompatibility of the Desara SL instrumentation.
Desara® SL instrumentation is reusable and physician labeling for the Desara SL instrumentation includes reprocessing instructions validated per the Guidance Document "Draft Guidance for Industry and FDA Staff - Processing Medical Devices in Health Care Settings: Validation Methods and Labeling" issued on May 2, 2011. The reprocessing instructions for the Desara SL Instrumentation are similar to those of the Desara Mini instrumentation (K103418)
Results of both mechanical bench and validation testing demonstrate equivalent device function based upon its intended use to the primary predicate device, Desara® Mesh. The following failure modes were assessed during mechanical and validation testing: sleeve to tip pull-off force, suture loop pull-out force and device mesh break strength and all testing met predefined acceptance criteria.
Testing documentation from Desara was used to support the aging, shelf life, transportation and sterlization Desara® SL In accordance with the FDA Guidance, Updated 510(k) Sterlilty Review Guidance K90-1; Final Guidance for Industry, FDA, FDA Device Advice, Guidance Documents (Medical Device and Radiation Emitting Product), 4. Labeling and FDA Consensus standard, ASTM F-1980-07, Standard Guide for Accelerated Aging of Sterlie Barrier Systems for Medical Devices as It is comprised of the same materials, packaging, manufacturing and sterlization processes as that of Desara®.
One year results of the prospective randomized trial by de Leval J, et al' found that at one year follow-up a modified inside-out transobturator tape procedure utilizing a 12 cm sling or 'tape' length like Desara SL was as safe and efficacious as the original transobturator technique but associated with less immediate post-operative groin pain.
## Summary of Substantial Equivalence
The conclusions drawn from the non-clinical and clinical tests demonstrate that the Desara SL is substantially equivalent to its proposed predicate devices.
¹ de Leval, J, Waltregny TD. The original versus a modified inside-out transobturator procedure: 1-year results of a prospective randomized trial. Int Urogynecol J (2011) 22:145-156. DOI 10.1007/s00192-010-1264-4.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be a caduceus, a symbol often associated with healthcare, featuring a staff with intertwined snakes.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# February 19, 2013
Caldera Medical, Inc. % Ms. Vicki Gail QA/RA Manager 5171 Clareton Drive AGOURA HILLS CA 91301
Re: K121928
Trade/Device Name: Desara SL Regulation Number: 21 CFR& 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTN Dated: January 29, 2013 Received: January 31, 2013
Dear Ms. Gail:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Ms. Vicki Gail
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R.A.Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K121928
Device Name: Desara SL
# Indications for Use:
Desara SL is intended to be used in females to position a mesh for the treatment of Genuine Stress Urinary Incontinence (SUI), mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.
| Prescription Use |
|------------------|
|------------------|
AND/OR
Over-The-Counter Use
| (21 CFR 801 Subpart D) |
|------------------------|
|------------------------|
x
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Benjamin R. Fisher -S 2013.02.19 18:10:06 -05'00'
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K121928 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.