← Product Code [EGS](/productcode/EGS) · K121901

# TI-MAX Z45 (K121901)

_Nakanishi, Inc. · EGS · Feb 28, 2013 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K121901

## Device Facts

- **Applicant:** Nakanishi, Inc.
- **Product Code:** [EGS](/productcode/EGS.md)
- **Decision Date:** Feb 28, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

TI-MAX Z45 is powered by either an air-motor or electronic micromotor for use in general dentistry. The device is intended for cutting and grinding teeth, cavity preparations, tooth and crown preparations, finishing and trimming teeth and filling materials, and removal of crowns and filling materials.

## Device Story

TI-MAX Z45 is a contra-angle dental handpiece used in general dentistry; powered by air-motor or electronic micromotor. Device transmits rotational force from motor through clutch and gears to chuck, driving dental bur for cutting/grinding teeth, cavity/crown preparations, and material removal. Operated by dental professionals in clinical settings. Features include 4-port water spray for cooling, fiberoptic lighting, and push-button chuck mechanism. Output is mechanical rotation of bur; healthcare provider uses device to perform restorative/surgical dental procedures. Benefits include efficient tooth/material removal and thermal safety via integrated coolant system.

## Clinical Evidence

Bench testing only. Device performance verified against ISO 7785-2 standards, including eccentricity, bur extraction force, and thermal safety (air/water spray). No clinical data presented.

## Technological Characteristics

Materials: Stainless steel, titanium. Power: Air-motor or electronic micromotor. Coupling: ISO 3964 compliant. Bur: ISO 1797-1 compliant friction grip (max 25mm). Cooling: 4-port water spray/jet. Lighting: Fiberoptic. Mechanism: Push-button chuck. Gear ratio: 1:4.2. Sterilization: Autoclave.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- A-DEC/W&H SYNEA AIR-DRIVEN HIGHSPEED HANDPIECE, MODELS TA-98, TA-97 ([K070663](/device/K070663.md))
- SURGICAL CONTRA-ANGLE HANDPIECES TYPES WS-56E, WS-75-E/KM, WS-92E/3 ([K011061](/device/K011061.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the NSK logo in the upper portion of the image. The logo is white text on a black background. The lower portion of the image contains the text "510(k) Summary" in a large font. The text is black on a white background.

Image /page/0/Picture/1 description: The image contains a sequence of alphanumeric characters. The characters appear to be handwritten and include the letters 'KI' followed by the numbers '21901'. The characters are written in black ink on a white background. The writing style is somewhat cursive, with some characters connected.

# FEB 2 8 2013

- Owner/Company name, address a. NAKANISHI, INC. 700 Shimohinata, Kanuma-shi, Tochigi, 322-8666, Japan
### b. Contact

Toshihiko Shinozaki Assistant Manager, R&D Department, Pharmaceutical Affairs Group NAKANISHI, INC. 700 Shimohinata, Kanuma-shi, Tochigi, 322-8666, Japan

| Phone: | 011-81-289-64-3380              |
|--------|---------------------------------|
| Fax:   | 011-81-289-62-6665              |
| Email: | t-shinozaki@nsk-nakanishi.co.jp |

### Application Correspondent C.

Fumiaki Kanai, Ph.D. President and CEO MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan

Phone: 011-81-3-3818-8577 Fax: 011-81-3-3818-8573 Email: kanaif@mici.co.jp

### d. Date prepared

June 28, 2012

## e. Name of device

Trade Name: Classification Name: Regulation Name: Regulatory Class: Product Code:

TI-MAX Z45 Handpiece, Contra-and Right-angle Attachment, Dental Dental handpiece and accessories Class I EGS

510(k) Summary: Page 1 of 6

Image /page/0/Picture/17 description: The image shows a black and white drawing of a bird in flight. The bird has a long beak and a forked tail. Below the bird, the words "Powerful Partners" are written in a small font. The image is simple and appears to be a logo or emblem.

{1}------------------------------------------------

**NSK**

### Predicate devices ﮨﺖ

The Ti-MAX Z45 is substantially equivalent to the following legally marketed devices:

| 510(k):              | K070663                                               |
|----------------------|-------------------------------------------------------|
| Trade name:          | A-DEC/W&H SYNEA AIR-DRIVEN HIGHSPEED HANDPIECE,       |
|                      | MODELS TA-98, TA-97                                   |
|                      | A-DEC/W&H SYNEA HANDPIECE ATTACHMENT, MODELS WA-99LT, |
|                      | WA-86LT, WA-66LT, WA-56LT, HA-43LT                    |
| Product code:        | EFB                                                   |
| Classification Name: | Handpiece, Air-powered, Dental                        |
| 510(k):              | K011061                                               |
| Trade name:          | SURGICAL CONTRA-ANGLE HANDPIECES TYPES WS-56E,        |
|                      | WS-75-E/KM, WS-92E/3                                  |
| Product code:        | EGS                                                   |

Classification Name: Handpiece, Contra-and Right-angle Attachment, Dental

Hereinafter, the predicate devices are called "the SYNEA HANDPIECE (K070663)" and "the SURGICAL CONTRA-ANGLE HANDPIECES (K011061)", respectively, for convenience of discussion in this application. Regarding the SYNEA HANDPIECE (K070663), A-DEC/W&H SYNEA HANDPIECE ATTACHMENT is the exact predicate device for the TI-MAX Z45. Since the HA-43LT type of the A-DEC/W&H SYNEA HANDPIECE ATTACHMENT is a straight handpiece, the type is eliminated as the predicate device for the T1-MAX Z45 hereinafter in this submission. The SYNEA HANDPIECE (K070663) indicates A-DEC/W&H SYNEA HANDPIECE ATTACHMENT MODELS WA-99LT, WA-86LT, WA-66LT, WA-56LT in this application.

## g. Description of the device

The Ti-MAX Z45 is a contra-angle dental handpiece which is powered by either an air-motor or electronic-micromotor for use in general dentistry. The TI-MAX Z45 is intended for cutting and grinding teeth, cavity preparations, tooth and crown preparations, finishing and trimming teeth and filling materials, and removal of crowns and filling materials. The Tr-MAX Z45 transmits rotational force from the motor to the gears through a clutch, and then the rotation force reaches the chuck and the dental bur receives the rotation force. The dental bur is able to cut and grind teeth and dentures using the rotation force.

### h. Statement of Intended Use

TI-MAX Z45 is powered by either an air-motor or electronic micromotor for use in general dentistry.

The device is intended for cutting and grinding teeth, cavity preparations, tooth and crown

{2}------------------------------------------------

NET

preparations, finishing and trimming teeth and filling materials, and removal of crowns and filling materials.

## i. Summary of Technological Characteristics

Summary or receded in rotation of the TI-MAX Z45 is 168,000 min1. The rotation speed is increased by 4.2-fold at the chuck using the gears. Since the TI-MAX Z45 uses rotational force for the treatment, temperature increase may occur on the drilling point. Therefore, the TI-MAX Z45 has 4-port water spray outlet on the head in order to cool down the treatment area. It is possible to switch the type of coolant water flow, water jet or water spray. The TI-MAX Z45 has opticfiber for lighting.

The friction grip bur (not exceed 25 mm) which conforms ISO 1797-1 is intended to use for the TI-MAX Z45. The TI-MAX Z45 uses the push-button chuck mechanism.

The TI-MAX Z45 can be connected only E-type motors which conforms ISO 3964.

### Statement of substantial equivalence i.

The Indications for Use of the TI-MAX Z45 is similar to the SYNEA HANDPIECE (K070663). The principle of operation of the TI-MAX Z45 is identical to the predicates. That is, the TI-MAX Z45 and the predicates transmit rotational force from the motor to the dental bur using the gears. The dental bur is able to cut and grind teeth and dentures using the rotation force. In addition, design features (Contra-Angle, Push-Button Chuck, Motor coupling, Thermal safety system) of the Ti-MAX Z45 are identical to those of the predicates.

Following table is comparison between the TI-MAX Z45 and the predicates.

|                                     | TI-MAX Z45                                                                                                                                                                                                                                                         | SYNEA HANDPIECE<br>(K070663)                                                                                                                                                                                                                                                                   | SURGICAL<br>CONTRA-ANGLE<br>HANDPIECES<br>(K011061)                                                                                                                                                                                                                                                               |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                 | TI-MAX Z45 is powered<br>by either an air-motor or<br>electronic micromotor<br>for use in general<br>dentistry.<br>The device is intended<br>for cutting and grinding<br>teeth, cavity<br>preparations, tooth and<br>crown preparations,<br>finishing and trimming | The A-dec/W&H Synea<br>Air-Driven Highspeed<br>Handpiece is an<br>air-powered dental<br>handpiece for use in<br>general dentistry. This<br>device is designed for<br>removing carious<br>material and excess<br>filling material, cavity<br>and crown preparation,<br>root canal preparations, | Indications are very<br>widespread in the field<br>of implantology and<br>surgery. The<br>mentioned handpieces<br>have been developed<br>especially for the<br>following applications:<br>WS-56 E:for e,g.<br>osteotomla on the<br>upper and lower jaw.<br>germectomls,                                           |
|                                     | teeth and filling<br>materials, and removal<br>of crowns and filling<br>materials.                                                                                                                                                                                 | finishing tooth<br>preparations,<br>restorations and<br>polishing teeth.                                                                                                                                                                                                                       | sequestrotomla<br>WS-75 E/KM: for e.g.<br>implantology<br>(preparation of the<br>bone cavity, placing<br>the implants), surgical<br>treatment of<br>dysgnathia,<br>osteosynthesis,<br>segmentosteotomla<br>and especially mild<br>osteotomla.<br>WS-92 E/3 : for e.g.<br>hemisection, wisdom<br>tooth extraction. |
| Product Code                        | EGS                                                                                                                                                                                                                                                                | EFB                                                                                                                                                                                                                                                                                            | EGS                                                                                                                                                                                                                                                                                                               |
| Classification<br>Name              | Handpiece, Contra-and<br>Right-angle Attachment,<br>Dental                                                                                                                                                                                                         | Handpiece, Air-powered,<br>Dental                                                                                                                                                                                                                                                              | Handpiece, Contra-and<br>Right-angle<br>Attachment, Dental                                                                                                                                                                                                                                                        |
| Design features                     | Contra-Angle,<br>Push-Button Chuck,<br>Motor coupling,<br>Thermal safety system                                                                                                                                                                                    | Contra-Angle,<br>Push-Button Chuck,<br>Motor coupling,<br>Thermal safety system                                                                                                                                                                                                                | Contra-Angle,<br>Push-Button Chuck,<br>Motor coupling,<br>Thermal safety system                                                                                                                                                                                                                                   |
| Gear ratio                          | 1:4.2                                                                                                                                                                                                                                                              | 1:5, 1:1, 2:1, 10:1                                                                                                                                                                                                                                                                            | 1:1, 20:1. 1:2,7                                                                                                                                                                                                                                                                                                  |
| Maximum<br>motor/drive speed        | 40,000 rpm                                                                                                                                                                                                                                                         | 40,000 rpm                                                                                                                                                                                                                                                                                     | 50,000 rpm                                                                                                                                                                                                                                                                                                        |
| Materials                           | Stainless Steel,<br>Titanium                                                                                                                                                                                                                                       | Unknown                                                                                                                                                                                                                                                                                        | Stainless steel                                                                                                                                                                                                                                                                                                   |
| Motor coupling                      | Compliant with ISO<br>3964                                                                                                                                                                                                                                         | Compliant with ISO<br>3964                                                                                                                                                                                                                                                                     | Compliant with ISO<br>3964                                                                                                                                                                                                                                                                                        |
| Source of power                     | Air-motor,<br>Electronic-micromotor                                                                                                                                                                                                                                | Air-motor,<br>Electronic-micromotor                                                                                                                                                                                                                                                            | Air-motor,<br>Electronic-micromotor                                                                                                                                                                                                                                                                               |
| Light                               | Glass Rod (Fiberoptic)                                                                                                                                                                                                                                             | With light type/Without<br>light type                                                                                                                                                                                                                                                          | With light<br>type/Without light type                                                                                                                                                                                                                                                                             |
| Dimension                           | Head Diameter: 9.6mm<br>Head Height: 14.5mm<br>Length: 96.2mm                                                                                                                                                                                                      | Head Height: 12.3 mm<br>or 13.7mm                                                                                                                                                                                                                                                              | Unknown                                                                                                                                                                                                                                                                                                           |
| Weight                              | 66.5 g                                                                                                                                                                                                                                                             | Unknown                                                                                                                                                                                                                                                                                        | Unknown                                                                                                                                                                                                                                                                                                           |
| Performance                         | Compliant with ISO<br>7785-2                                                                                                                                                                                                                                       | Compliant with ISO<br>7785-2                                                                                                                                                                                                                                                                   | Compliant with ISO<br>7785-2                                                                                                                                                                                                                                                                                      |
| -Eccentricity                       | 0.016 mm                                                                                                                                                                                                                                                           | Unknown                                                                                                                                                                                                                                                                                        | Unknown                                                                                                                                                                                                                                                                                                           |
| -Force for bur<br>extraction        | 25.3N                                                                                                                                                                                                                                                              | Unknown                                                                                                                                                                                                                                                                                        | Unknown                                                                                                                                                                                                                                                                                                           |
| Sterilization                       | Sterilized by user<br>(Autoclave)                                                                                                                                                                                                                                  | Sterilized by user<br>(Autoclave)                                                                                                                                                                                                                                                              | Sterilized by user<br>(Autoclave)                                                                                                                                                                                                                                                                                 |
| Thermal safety                      | Air/Water spray,<br>compliant with<br>ISO7785-2                                                                                                                                                                                                                    | Air/Water spray,<br>compliant with<br>ISO7785-2                                                                                                                                                                                                                                                | Air/Water spray,<br>compliant with<br>ISO7785-2                                                                                                                                                                                                                                                                   |
| -Air/Water Spray<br>port            | 4-port                                                                                                                                                                                                                                                             | Single or 5-port                                                                                                                                                                                                                                                                               | Single or 3-port                                                                                                                                                                                                                                                                                                  |
| -Coolant<br>Water<br>Type           | Water spray/Water jet                                                                                                                                                                                                                                              | Water spray                                                                                                                                                                                                                                                                                    | Water spray                                                                                                                                                                                                                                                                                                       |
| -Flow Rate                          | More than 50 ml/min                                                                                                                                                                                                                                                | More than 50 ml/min                                                                                                                                                                                                                                                                            | More than 50 ml/min                                                                                                                                                                                                                                                                                               |
| -Water<br>pressure<br>setting range | 2 bar                                                                                                                                                                                                                                                              | 0.5-2 bar                                                                                                                                                                                                                                                                                      | Unknown                                                                                                                                                                                                                                                                                                           |
| -Chip air pressure<br>setting range | 2 bar                                                                                                                                                                                                                                                              | 1.5-3bar                                                                                                                                                                                                                                                                                       | Unknown                                                                                                                                                                                                                                                                                                           |
| Lubricant                           | PANA Spray (510(k)<br>number is K052700)                                                                                                                                                                                                                           | W&H Service oil F1                                                                                                                                                                                                                                                                             | W&H Service oil F1                                                                                                                                                                                                                                                                                                |
| Bur                                 | FG burs                                                                                                                                                                                                                                                            | FG burs, contra-angle<br>burs                                                                                                                                                                                                                                                                  | FG burs, contra-angle<br>burs                                                                                                                                                                                                                                                                                     |
| -Maximum<br>bur<br>length           | 25 mm                                                                                                                                                                                                                                                              | 25mm (FG burs), 34mm<br>(contra-angle burs)                                                                                                                                                                                                                                                    | 34mm (WS-56E<br>contra-angle burs), 45<br>mm (WS-75 E/KM,<br>contra-angle bur), 25<br>mm (WS-92 E/3, FG<br>bur)                                                                                                                                                                                                   |

Comnarison table

510(k) Summary: Page 3 of 6

Image /page/2/Picture/13 description: The image shows a black and white drawing of a bird perched on a branch. The bird is facing to the left and has a long tail. The branch is thin and has a few leaves on it. The drawing is simple and has a vintage look.

{3}------------------------------------------------

1577

510(k) Summary: Page 4 of 6

Image /page/3/Picture/4 description: The image shows the words "Powerful Partners" in a stylized font. The text is black and the background is white. There is a small graphic above the text.

{4}------------------------------------------------

NSK

Note*: The models for "the A-DEC/W&H SYNEA HANDPIECE ATTACHMENT" are used as the predicate device for the Ti-Max Z45 in the submission.

As shown in the comparison table, the indications for use of the TI-MAX Z45 is similar to the SYNEA HANDPIECE (K070663). As described above, the principle of operation and design features of the TI-MAX Z45 are identical to those of the predicates. However, the dimensions and materials of the TI-MAX Z45 are different from the predicates. The gear ratio of the TI-MAX Z45 is also different from that of the predicates. Therefore, the maximum rotation speed at the chuck is different between the Ti-MAX Z45 and the predicates. In addition, the number of the air/water spray port is different between the TI-MAX Z45 and the predicates. In order to evaluate that those differences do not affect safety and effectiveness of the TI-MAX Z45, bench testing was performed. Bench testing demonstrated that the TI-MAX Z45 met the requirement of ISO7785-2 which was used for evaluation of the predicate devices. Based on above result, the TI-MAX Z45 is substantially equivalent to the predicates and does not raise any new safety and effectiveness concern.

510(k) Summary: Page 5 of 6

Powerts: Partrers

{5}------------------------------------------------

### k. Bench Testing

The bench tests were performed in order to ensure the safety and effectiveness of the TI-MAX Z45, verify conformity to ISO 7785-2 and demonstrate substantial equivalence to the predicates.

All TI-MAX Z45 samples were compliant with ISO 7785-2 and demonstrated substantial equivalence to the predicates.

### ﻨﺴﺘ Conclusion

The indications for use of the TI-MAX Z45 is similar to the SYNEA HANDPIECE (K070663). In addition, the principle of operation and design features of the TI-MAX Z45 are identical to those of the predicates. However, the dimensions and materials of the TI-MAX Z45 are different from the predicates. The gear ratio of the TI-MAX Z45 is also different from that of the predicates. Therefore, the maximum rotation speed at the chuck is different between the TI-MAX Z45 and the predicates. In addition, the number of the air/water spray port is different between the TI-MAX Z45 and the predicates. Although there are those differences in characteristics between the TI-MAX Z45 and the predicates, a number of performance testing indicated that the Ti-MAX Z45 met the requirement of the recognized consensus or voluntary standard. Based on above result, we conclude the TI-MAX Z45 is substantially equivalent to the predicates and does not raise any new safety and effectiveness concern.

Image /page/5/Picture/6 description: The image shows a black and white drawing of a bird perched on a branch. The bird has a long beak and a long tail. The text "Powerful Partners" is written below the bird. The drawing is simple and stylized.

510(k) Summary: Page 6 of 6

{6}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized design resembling an eagle or bird-like figure with three curved lines representing wings or feathers.

### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

### February 28, 2013

Nakanishi, Incorporated C/O Fumiaki Kanai, Ph.D. President and Chief Executive Officer MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, Japan 113-0033

Re: K121901

Trade/Device Name: TI-MAX Z45 Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EGS Dated: February 5, 2013 Received: February 11: 2013

### Dear Dr. Kanai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Orug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractlie ality warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Image /page/7/Picture/7 description: The image contains the words "Kwame Ulmer" in bold, black font on the left side of the image. To the right of the name is the word "HEDA" in a stylized font. The letters are outlined with multiple lines, giving them a three-dimensional appearance. The letter "O" is placed above the "D".

for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

## Indications for Use

k121901 510(k) Number (If known):

Device Name: TI-MAX Z45

### Indications for Use

TI-MAX Z45 is powered by either an air-motor or electronic micromotor for use in general dentistry.

The device is intended for cutting and grinding teeth, cavity preparations, tooth and crown preparations, finishing and trimming teeth and filling materials, and removal of crowns and filling materials.

**Prescription Use**
(Per 21 CFR 801 Subpart D) 

![x mark](data:image/png;base64,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

AND/OR

Over-the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mary S. Runner -S 013:02.28 07:38:56

(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dental Devices

510(k) Number;

---

**Source:** [https://fda.innolitics.com/device/K121901](https://fda.innolitics.com/device/K121901)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
