MYOSURE LITE TISSUE REMOVAL DEVICE (1 PACK), MYOSURE LITE TISSUE REMOVAL DEVICE (3 PACK)

K121868 · Hologic, Inc. · HIH · Jul 19, 2012 · Obstetrics/Gynecology

Device Facts

Record IDK121868
Device NameMYOSURE LITE TISSUE REMOVAL DEVICE (1 PACK), MYOSURE LITE TISSUE REMOVAL DEVICE (3 PACK)
ApplicantHologic, Inc.
Product CodeHIH · Obstetrics/Gynecology
Decision DateJul 19, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 884.1690
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Myosure Hysteroscopic Tissue Removal System is intended for hysteroscopic intrauterine procedures by a trained gynecologist to resect and remove tissue including submucous myomas and endometrial polyps.

Device Story

MyoSure LITE Tissue Removal Device is a component of the MyoSure Hysteroscopic Tissue Removal System; used for hysteroscopic intrauterine tissue resection. System comprises a control unit, tissue removal device (morcellator), and foot pedal. Control unit contains an electric motor and software controller; drives the morcellator via a flexible drive cable. Cutter blade performs simultaneous rotation and reciprocation; vacuum source aspirates resected tissue through a side-facing cutting window. Distension fluid and tissue are transported to a tissue trap and vacuum canister. Operated by a trained gynecologist in a clinical setting. Compatible with standard fluid management systems and hysteroscopes with a 3 mm working channel. Enables mechanical resection and removal of submucous myomas and endometrial polyps.

Clinical Evidence

No clinical data. Verification and validation testing confirmed the modified system meets functional and performance specifications. Biocompatibility testing performed per ISO 10993-1.

Technological Characteristics

Electromechanical tissue morcellator. Components: control unit with electric motor, flexible drive cable, tissue removal device, foot pedal. Materials meet ISO 10993-1 biocompatibility requirements. Compatible with 3 mm working channel hysteroscopes. Software-controlled motor activation.

Indications for Use

Indicated for hysteroscopic intrauterine tissue resection and removal, including submucous myomas and endometrial polyps, in patients requiring such procedures by a trained gynecologist.

Regulatory Classification

Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) SUMMARY 5. JUL 1 9 2012 1. Submitter: > Hologic, Inc. 250 Campus Dr. Marlborough, MA 01752 Telephone: 508.263.8857 Contact: Sarah Fairfield, Sr. Regulatory Affairs Specialist #### 2. Device: Trade Name: MyoSure LITE Tissue Removal Device (part of the MyoSure Hysteroscopic Tissue Removal System) Common Name: Hysteroscope and accessories Classification Name: Hysteroscope and accessories - Class: II #### 3. Predicate Device: MyoSure™ Hysteroscopic Tissue Removal System (K100559) #### Device Description: 4. The modified Myosure Hysteroscopic Tissue Removal System consists of the following procedural components which are identical to those found in the predicate Myosure System: - Myosure Control Unit o - Myosure Tissue Removal Device o - Myosure Foot Pedal O The Myosure Control Unit contains an electric motor and software controller that drives the Myosure Tissue Removal Device. The Control Unit motor is activated and deactivated by the Myosure Foot Pedal. The Myosure Tissue Removal Device is a tissue morcellator that is connected to the Control Unit via a flexible drive cable. The morcellator's cutter blade is controlled by a drive system that enables simultaneous rotation and reciprocation of the cutter. The cutter is also connected to a vacuum source which aspirates resected tissue through a side-facing cutting window in the device's outer tube. Distension fluid and resected tissue are transported from the Myosure Tissue Removal Device to a tissue trap and vacuum canister via a tube protruding from the proximal end of the Tissue Removal Device. The Myosure Hysteroscopic Tissue Removal System is compatible with commercially available fluid management systems and may be used with hysteroscopes that have a straight 3 mm working channel. {1}------------------------------------------------ 5-3 Verification/validation testing of the modified Myosure System was completed and confirmed that the modified Myosure System meets the same functional and performance specifications as the predicate Myosure System. The new patient contact materials in the proposed new sized morcellator for the Myosure System meet the biocompatibility requirements of ISO 10993-1 Biological Evaluation of Medical devices. #### 8. Conclusion: Based on the intended use, descriptive information and performance evaluation provided in this submission, the modified MyoSure Hysteroscopic Tissue Removal System has been shown to be equivalent in technology, method of operation, functional performance and intended use to the predicate MyoSure Hysteroscopic Tissue Removal System. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Sarah Fairfield Senior Regulatory Affairs Specialist Hologic, Inc. 250 Campus Drive MARLBOROUGH MA 01752 JUL 19 2012 K121868 Re: Trade/Device Name: MyoSure LITE Tissue Removal Device (part of the MyoSure Hysteroscopic Tissue Removal System) Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: June 26, 2012 Received: June 26, 2012 ### Dear Ms. Fairfield: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Sincerely yours, Benjamin K. Twitchell Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # 4. INDICATIONS FOR USE STATEMENT 510(k) Number (if known): |(121868 Device Name: MyoSure LITE Tissue Removal Device (part of the MyoSure Hysteroscopic Tissue Removal System) Indications For Use: The Myosure Hysteroscopic Tissue Removal System is intended for hysteroscopic intrauterine procedures by a trained gynecologist to resect and remove tissue including submucous myomas and endometrial polyps. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR : Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | | 18 July 2012 | |---------------------------------------------|--------------| | (Division Sign-Off) | | | Division of Reproductive, Gastro-Renal, and | | Drision of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number K120868
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%