ESA615

K121860 · Fluke Biomedical · DRT · Jan 25, 2013 · Cardiovascular

Device Facts

Record IDK121860
Device NameESA615
ApplicantFluke Biomedical
Product CodeDRT · Cardiovascular
Decision DateJan 25, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2

Indications for Use

The Product is an electronic signal source and measurement device for verifying the electrical safety of medical devices. The Product also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. The Product provides the following function categories: ECG Functions . ECG-Performance Testing . The intended user is a trained biomedical equipment technician who performs periodic preventative maintenance checks on patient monitors in service. Users can be associated with hospitals, clinics, original equipment manufacturers and independent service companies that repair and service medical equipment. The end user is an individual, trained in medical instrumentation technology. This Product is intended to be used in the laboratory environment, outside of the patient care area, and is not intended for use on patients, or to test devices while connected to patients. This Product is not intended to be used to calibrate medical equipment. ESA615 is intended for over the counter use.

Device Story

ESA615 Electrical Safety Analyzer serves as electronic signal source and measurement device for verifying electrical safety of medical devices; provides ECG simulation and performance waveforms to verify patient monitor operating specifications. Device used in laboratory environments by trained biomedical equipment technicians; not for use on patients or for calibration. Input includes electrical safety parameters and ECG signals; device processes these via embedded firmware on printed circuit board assemblies. Output displayed on monochrome LCD; device supports USB communication for computer-controlled testing. Benefits include ensuring medical equipment safety and performance during maintenance. Device is AC line-powered.

Clinical Evidence

No clinical data. Bench testing only; laboratory studies verified performance against published specifications using a representative patient monitor. Software performance was successfully validated.

Technological Characteristics

Components: Printed circuit board assemblies with surface mount components, plastic injection molded case, LCD interface, AC power cord (120V/60Hz). Sensing/actuation: Electrical safety measurement and ECG signal simulation. Connectivity: USB port for computer control. Software: Embedded firmware. Dimensions: 6.94" W x 3.3" D x 11.2" H. Weight: 3.5 lbs.

Indications for Use

Indicated for use by trained biomedical equipment technicians to perform periodic preventative maintenance checks on patient monitors in a laboratory environment. Not for use on patients or for calibrating medical equipment.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121860. Image /page/0/Picture/1 description: The image shows the logo for Fluke Biomedical. The logo is black and white and features the word "FLUKE" in large, bold letters. Below the word "FLUKE" is the word "Biomedical" in smaller letters. In the upper right corner of the image is the fraction "1/4" written in black ink. #### Section III -510(k) Summary of Safety and Effectiveness #### Submitter: JAN 2 5 2013 Fluke Biomedical 6920 Seaway Blvd Everett WA. 98203 (440) 498-2579 -Phone (440-349-2307) - Fax John Nelson - Contact Person #### Device Name: - Trade Name - ESA615 Electrical Safety Analyzer - Common Name Analyzer . - Classification Name Cardiac monitor (including cardiotachometer and rate alarm), per 21 CFR & ● 870.2300 - Product Codes -DRT . #### Devices for Which Substantial Equivalence is Claimed: - MPS450 #### Device Description: #### Principles of Operation Fluke Biomedical's ESA615 Electrical Safety Analyzer (hereafter referred to as the ESA615) provides a basis for verifying the electrical safety of medical devices. The Product also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. The ESA615 can be controlled by a suitably equipped computer to carry out tests, via either a cable or wirelessly. #### Technological Characteristics The ESA615 consists of the following components: 1) Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors. 2) Plastic injection molded case parts. - 3) Liquid Crystal Display for user interface. - 4) Power cord for powering the unit at 120V and 60Hz. #### Intended Use of the Device: The Product is an electronic signal source and measurement device for verifying the electrical safety of medical devices. The Product also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. {1}------------------------------------------------ The Product provides the following function categories: - ECG Functions . - ECG-Performance Testing . The intended user is a trained biomedical equipment technician who performs periodic preventative maintenance checks on patient monitors in service. Users can be associated with hospitals, clinics, original equipment manufacturers and independent service companies that repair and service medical equipment. The end user is an individual, trained in medical instrumentation technology. This Product is intended to be used in the laboratory environment, outside of the patient care area, and is not intended for use on patients, or to test devices while connected to patients. This Product is not intended to be used to calibrate medical equipment. ESA615 is intended for over the counter use. ### Summary of Technological Characteristics: The ESA615 is substantially equivalent to one other legally marketed and FDA approved device in the United States. The ESA615 functions in a manner similar to and is intended for the similar use as the MPS450 manufactured by Fluke Biomedical. The ESA615 is similar to MPS450 in that it uses LCD display, and allows user to simulate physiological parameter to verify the operation of patient monitors. The ESA615 differs from MPS450 in that ESA615 works only using AC power cord and has additional options of performing electrical safety analysis. | Features | ESA615 | MPS450V | Difference | |-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | Intended Use | The Product is an electronic<br>signal source and measurement<br>device for verifying the<br>electrical safety of medical<br>devices. The Product also<br>provides ECG simulation and<br>performance waveforms to<br>verify patient monitors are<br>performing within their<br>operating specifications.<br>The Product provides the<br>following function categories:<br>• ECG Functions<br>• ECG-Performance<br>Testing<br>The intended user is a trained<br>biomedical equipment | The intended use of MPS450 is<br>to test and verify the basic<br>operation of patient monitoring<br>devices or systems used to<br>monitor various physiological<br>parameters of patient, including<br>ECG, Respiration, Invasive<br>Blood Pressure, and Cardiac<br>Output.<br>The intended user is a trained<br>biomedical equipment technician<br>who is performing periodic<br>preventative maintenance checks<br>on patient monitors in service.<br>Users can be associated with<br>Hospitals, clinics, original<br>equipment manufacturer or<br>equipment | Fewer functions,<br>Respiration,<br>Invasive Blood<br>Pressure, Cardiac<br>Output | | | | | Biomedical | | | technician who performs<br>periodic preventative<br>maintenance checks on patient<br>monitors in service. Users can<br>be associated with hospitals,<br>clinics, original equipment<br>manufacturers and independent<br>service companies that repair<br>and service medical<br>equipment.<br>The end user is an individual,<br>trained in medical<br>instrumentation technology.<br>This Product is intended to be<br>used in the laboratory<br>environment, outside of the<br>patient care area, and is not<br>intended for use on patients, or<br>to test devices while connected<br>to patients. This Product is not<br>intended to be used to calibrate<br>medical equipment. It is<br>intended for over the counter<br>use. | independent service companies<br>that repair and service medical<br>equipment. The end user is<br>technically trained individual,<br>specializing in medical<br>instrumentation technology.<br>The MPS450 is intended to be<br>used in the laboratory<br>environment and is not intended<br>for use on patients, or to test<br>devices while connected to<br>patients. The MPS450 is not<br>intended to be used to calibrate<br>medical equipment and not<br>intended for over the counter<br>use. | 3/4 | | Construction | Plastic Case | Plastic Case | None | | Size | 6.94" wide x 3.3" deep x 11.2"<br>high | 5.5" wide x 7.5" deep x 1.8"<br>high | Bigger | | Weight | 3.5 lbs | 2lbs | Heavier | | Display | 3" diagonal screen<br>Monochrome STN<br>transflective | 4 line x 20-character super twist<br>LCD display | Larger screen | | Function Key | Soft | Soft | None | | ECG Leads | 5 applied parts jacks and easy<br>ECG snap connection with<br>optional expander box | 10 binding posts; compatible<br>with disposable snaps, 3.2 mm<br>or 4 mm electrodes, and banana<br>plugs | Fewer applied<br>parts on ESA615 | | Communication<br>s Port | USB | RS232 | Change from<br>RS232 to USB<br>data port with<br>advancement in<br>technology | | Power | No Battery - AC line powered<br>only | 9V alkaline battery with low<br>battery indicator; or battery<br>eliminator; transformer certified<br>to CSA | No battery because<br>of higher power<br>requirements | 000144 Blomedica {2}------------------------------------------------ : - 0001455 . ・・ {3}------------------------------------------------ | FLuke. | |------------| | Biomedical | 000146 | Lead<br>Configuration | 5 leads | 12 leads | None | |------------------------|-------------------------------|-----------------------------------------------------------------------------------------|---------------------------| | Amplitude<br>Accuracy | ± 5% of 1mV setting | ±2% of setting | Less accurate | | Rate Accuracy | ± 2% of setting | ±1% setting | Less accurate | | Normal Sinus<br>Rhythm | 30, 60, 120, 180, 240 bpm | 30, 40, 45, 60, 80, 90, 100, 120,<br>140, 160, 180, 200, 220, 240,<br>260, 280, 300 bpm | Fewer NSR waves | | Sine wave | 10, 40, 50, 60, 100 Hz | 0.5, 5, 10, 40, 50, 60, 100 Hz | Fewer frequencies | | Square wave | 0.125, 2.0 Hz | 0.125, 2.0 Hz | None | | Triangle wave | 2.0 Hz | 2.0, 2.5 Hz | Fewer frequencies | | Pulse wave | 30, 60 bpm, 63 ms pulse width | 30, 60 bpm, 60 ms pulse width | Wider pulse width | | Cable<br>Connector | ECG leads, 5 binding postings | ECG leads, 10 binding postings | Fewer binding<br>postings | #### Substantial Equivalence: The ESA615 is substantially equivalent to one other legally marketed device in the United States. The ESA615 functions in a manner similar to and is intended for the same use as the MPS450 manufactured by Fluke Biomedical. The ESA615 is similar to the MPS450 in that it uses LCD display, and allows user to simulate ECG parameters to verify the operation of patient monitors. The ESA615 differs from the MPS450 in that the ESA615 is not battery operated; it performs Electrical Safety Analysis and does not perform respiration, blood pressure or cardiac output. #### Non-Clinical Test Data: Laboratory studies have been conducted with a representative patient monitor to verify and validate the ESA615 will perform within its' published specifications -- Document: NPI-05012012-00002 The ESA615 software has been successfully validated to confirm the performance of the device. Document: NPI-03012012-00001 and NPI-05042012-00001 #### Clinical Test Data: Clinical testing has not been conducted on this product. #### Conclusion: Based upon the laboratory studies, similar technological/performance characteristics as compared to the predicate device, and successful validation of the ESA615 software, the performance of the ESA615 is deemed to be substantially equivalent to the MPS450. {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUM AN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 JAN 2 5 2013 Fluke Biomedical c/o Mr. John Nelson Director of Regulatory/Quality Affairs 6045 Cochran Rd. Solon, OH 44139 Re: K121860 > Trade/Device Names: ESA615 Regulatory Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (Two) Product Code: DRT Dated: January 7, 2013 Received: January 8, 2013 Dear Mr. Nelson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. John Nelson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Owen P. Faris -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ K12if60 Image /page/6/Picture/1 description: The image shows the Fluke Biomedical logo. The logo consists of the word "FLUKE" in large, bold, sans-serif font, with a stylized design above the letters. Below the word "FLUKE" is the word "Biomedical" in a smaller, bold, sans-serif font. The image is black and white. # Indications for Use # 510(k) Number (if known): 长しさした 6 0 Device Name: ESA 615 Electrical Safety Analyzer ## Indications for Use: The ESA615 Electrical Safety Analyzer is an electronic signal source and measurement device for verifying the electrical safety of medical devices. The ESA615 also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. The ESA615 provides following function categories: - ECG Functions . - ECG-Performance Testing Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 807 Subpart C) 000004 # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) . Concurryal; goff CDRJL Office_of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices Kr21860 510(k) Number_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...