DOMED TRI-PEG PATELLA, HIGHLY CROSS LINKED DOMED TRI-PEG PATELLA WITH VE

K121835 · Encore Medical L.P. · JWH · Jul 20, 2012 · Orthopedic

Device Facts

Record IDK121835
Device NameDOMED TRI-PEG PATELLA, HIGHLY CROSS LINKED DOMED TRI-PEG PATELLA WITH VE
ApplicantEncore Medical L.P.
Product CodeJWH · Orthopedic
Decision DateJul 20, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Joint replacement is indicated for patients suffering from disability due to: - degenerative, post-traumatic or rheumatoid arthritis; . - avascular necrosis of the femoral condyle; . - . post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; - . moderate valgus, varus or flexion deformities; - . treatment of fractures that are unmanageable using other techniques. This device may also be indicated in the salvage of previously failed surgical attempts. This system is to be used for cemented applications

Device Story

Domed Tri-Peg Patella is a prosthetic component for knee joint replacement systems (Foundation and 3DKnee). Device replaces damaged patellar surface to restore joint function. Manufactured from standard UHMWPE or highly cross-linked UHMWPE infused with Vitamin E. Used in orthopedic surgery; implanted by surgeons in clinical settings. Device is cemented into place. Provides mechanical articulation surface to address arthritis, necrosis, or post-traumatic deformity. Benefits include pain relief and restoration of joint configuration.

Clinical Evidence

No clinical data provided. Bench testing only. Previous mechanical testing (K113756) included push out, lever out, torsion, Izod impact, small punch, tensile, FTIR, wear, animal implant for toxicological response, and cytotoxicity.

Technological Characteristics

Patellar prosthesis; materials: standard UHMWPE and highly cross-linked UHMWPE infused with Vitamin E; cemented application; semi-constrained design; sterilization and packaging consistent with predicate systems.

Indications for Use

Indicated for patients with joint disability due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (patellofemoral erosion, dysfunction, or prior patellectomy); moderate valgus, varus, or flexion deformities; or fractures unmanageable by other techniques. Also indicated for salvage of failed surgical attempts. For cemented use.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121835 # 510(k) Summary JUL 20 2012 Date: July 12, 2012 Manufacturer: Encore Medical, L.P. 9800 Metric Blvd Austin, TX 78758 Contact Person: Michaela Norris Regulatory Affairs Associate Phone: (512) 834-6291 Fax: (512) 834-6313 Email: michaela.norris@djoglobal.com | A . C . C<br>Click College of Children Comments of Children Comments of Children Comments of Children Comments of Children Comments of Children Comments of Children Comments of Children C<br>roque<br>March Collection of School All Art Art Children Children Children | PATTER PARTY of Property and would be American And Accommend of Act According games or work<br>1<br>assificatio | "roduct c.<br>odes | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|--| | Jomes<br>Patella<br>000<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | The County of Children Company of Children Company of Children Company of Children Company of Children Comments of Children<br>Jass<br>THE STATE THE COLLECT CALL CALL CALL CALL CALL CONTRACT COLLECT COLLECT COLLECT COLLECTION CONTRACT COLLEGION CONTRACT COLLEGION CONTRACT CONTRACT CARTER COLLECTION CONTRACT | MO THIN | | | Product<br>Code | Regulation and Classification Name | |-----------------|---------------------------------------------------------------------------------------------------------------| | JWH | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis per 21 CFR 888.3560 | | OIY | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis per 21 CFR 888.3560 | #### · Regulation Number/Name 21 CFR 888.3560 - Knee joint patellofemoratibial polymer semi-constrained cemented prosthesis. Description: The purpose of this application is to include a new domed design patella for the Foundation and 3DKnee systems manufactured with both the standard UHMWPE and highly crossed linked UHMWPE infused with Vitamin E. #### Indications for Use: Joint replacement is indicated for patients suffering from disability due to: - degenerative, post-traumatic or rheumatoid arthritis; . - avascular necrosis of the femoral condyle; . - . post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; - . moderate valgus, varus or flexion deformities; - . treatment of fractures that are unmanageable using other techniques. This device may also be indicated in the salvage of previously failed surgical attempts. This system is to be used for cemented applications ### Predicate Devices: DJO Surgical Domed Tri-Peg Patella - K905613 DJO Surgical Movation Domed Patella - K100900 DJO Surgical 3DKnee Standard Poly Material - K020114 DJO Surgical e+ Patella HXL VE Poly - K113756 DJO Surgical 3DKnee Sterilization and Packaging of Standard Poly - K020114 DJO Surgical e+ Patella Sterilization and Packaging of HXL VE Poly - K113756 {1}------------------------------------------------ Comparable Features to Predicate Device(s): Features comparable to predicate devices include the same design features, materials, indications sterilization, packaging and intended use. Non-Clinical Testing: Previous mechanical testing outlined in K 1 13756 demonstrated the device's ability to perform under expected conditions. Testing included mechanical characterization testing, push out, lever out, torsion, Izod impact, small punch, tensile, FTIR, wear, animal implant for toxilogical response, and cytotoxicity. Clinical Testing: None provided. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure embracing a bird, which is a symbol of care and protection. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Encore Medical, L.P. % Ms. Michaela Norris Regulatory Affairs Associate 9200 Metric Boulevard Austin, Texas 78758 JUL 20 2012 Re: K121835 Trade/Device Name: Domed Tri-Peg Patella Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis. Regulatory Class: Class II Product Code: JWH, OIY Dated: June 22, 2012 Received: June 22, 2012 Dear Ms. Norris: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ ## Page 2 - Ms. Michaela Norris forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, FOR Ant. Macedo Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Domed Tri-Peg Patella Indications for Use: ## Domed Tri-Peg Patella Indications for Use Joint replacement is indicated for patients suffering from disability due to: - degenerative, post-traumatic or rheumatoid arthritis; . - avascular necrosis of the femoral condyle; . - post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, . dysfunction or prior patellectomy; - moderate valgus, varus or flexion deformities; � - treatment of fractures that are unmanageable using other techniques. . This device may also be indicated in the salvage of previously failed surgical attempts. This system is to be used for cemented applications Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) A (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K121835
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...