CONE-TROL Liquid Chemistry Control is a human liquid control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
Device Story
CONE-TROL Liquid Chemistry Control is a human serum-based control containing biochemical materials, chemicals, drugs, preservatives, and stabilizers. It serves as an assayed quality control material to monitor the precision of laboratory testing procedures. The device is used in clinical laboratory settings by laboratory technicians or clinicians. The control is value-assigned using the Abaxis Piccolo Express Analyzer (utilizing reagent discs with built-in calibration) and traditional clinical chemistry analyzers (Roche Cobas Mira Plus, Mindray BS-200). Healthcare providers use the control results to verify the accuracy and precision of their diagnostic testing systems. By ensuring the reliability of laboratory measurements, the device helps clinicians make informed diagnostic and treatment decisions, ultimately benefiting patients through more accurate clinical laboratory results.
Clinical Evidence
Bench testing only. Stability studies confirmed closed vial stability at 2-8°C for 33 days, open vial stability at 2-8°C for 15 days, and shelf-life stability at -20°C for 254 days. Value assignment was performed using Abaxis Piccolo Express, Roche Cobas Mira Plus, and Mindray BS-200 analyzers.
Technological Characteristics
Human serum matrix with added biochemicals, drugs, preservatives, and stabilizers. Liquid form. Stored at -20°C or colder. Value assignment via Abaxis Piccolo Express reagent discs and standard clinical chemistry analyzers (Roche Cobas Mira Plus, Mindray BS-200).
Indications for Use
Indicated for use as a human liquid control serum to monitor the precision of laboratory testing procedures for specific analytes in clinical chemistry settings. No specific patient population, age, or gender restrictions apply as this is an in vitro diagnostic control product.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
{0}------------------------------------------------
# SECTION V
JUL 2 4 2012
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
# 510(k) Number:
## Submitter:
Cone Bioproducts 1012 N. Austin Street Seguin, TX 78155
Telephone: (830)379-0197 Facsimile: (830)379-0471
#### Contact Person:
William K. Cone President Telephone: (830)379-0197 ext. 203 Facsimile: (839)379-0471
#### Preparation Date:
June 1, 2012
#### Device Information:
Proprietary Name: Regulation Section: Product Code: Classification: Panel:
CONE-TROL Liquid Chemistry Control 21 CFR 862.1660 Control JJY - Multi-analyte Control Class I, Reserved Clinical Chemistry (75)
# Device to Which Substantial Equivalence is Claimed:
Liquid Assayed Multiqual Control Levels 1, 2, and 3 Bio-Rad Laboratories Irvine, California 510(k) Number: k011867
{1}------------------------------------------------
#### Device Description:
CONE-TROL Liquid Chemistry Control is prepared from human serum to which purified biochemical material (human and animal origin), chemicals, drugs, preservatives, and stabilizers have been added. The control is provided in liquid form for user convenience.
## Intended Use:
CONE-TROL Liquid Chemistry Control is a human liquid control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
#### Comparison to Predicate Device(s):
The CONE-TROL Liquid Chemistry Control is substantially equivalent to Liquid Assayed Multiqual Control (k011867) for its stated intended use.
| Device<br>Characteristics | Subject Device | Predicate Device(s)<br>k011867 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | CONE-TROL Liquid Chemistry Control is a human liquid control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert. | Liquid Assayed Multiqual Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert. |
| Analyte(s) | Albumin, Alkaline Phosphatase, Alanine Aminotransferase, Amylase, Apolipoprotein A-1, Apolipoprotein B, Aspartate Aminotransferase, Bicarbonate, Bilirubin-Direct, Bilirubin-Total, Blood Urea Nitrogen, Complement 3, Complement 4, C-Reactive Protein, Calcium, Cholesterol, HDL Cholesterol, LDL Cholesterol, Cholinesterase, Creatinine Kinase, Chloride, Creatinine, Gamma Glutamyltransferase, Glucose, Immunoglobulin A, Immunoglobulin G, Immunoglobulin M, Lactate, Lactate Dehyrdogenase, Haptoglobin, Iron, Unsaturated Iron Binding Capacity, Magnesium, Phosphorous, Potassium, Prealbumin, Sodium, Total Protein, Transferrin, Triglycerides, Uric Acid | All analytes in subject device plus the analytes listed below:<br>Acetaminophen, Acid Phosphatase, a-1-Antitrypsin, aHBDH, Amikacin, Carbamazepine, Ceruloplasmin, CK- MB, Copper, Cortisol, Digoxin, Ethanol, Gentamicin, Globulin, Iron Binding Capacity-Total, LAP Arylamidase, Lipase, Lithium, Osmolality, Phenobarbital, Phenytoin, Phospholipids, Prostatic Acid Phosphatase, Protein Electrophoresis, Salicylate, T3-Total, T3-Free, T4-Total, T4- Free, T-Uptake, Theophylline, Thyroid Stimulating Hormone, Tobramycin, Urea, Valproic Acid, Vitamin B12, Zinc |
| Matrix | Human Serum | Human Serum |
| Control Form | Liquid | Liquid |
| Levels | 2 | 3 |
| Storage | -20°C or colder | -20°C or colder |
| (Unopened Frozen) | Until expiration date | Until expiration date |
## Table V-1: Device Comparison Table
{2}------------------------------------------------
| Storage<br>(Unopened Thawed) | 2-8°C<br>30 days | 2-8°C (current Package Insert)<br>30 days with the following exceptions:<br>Direct Billirubin - 11 days,<br>Triglycerides, HDL, Cholinesterase,<br>and Phosphorous - 7 days. Total and<br>Direct Bilirubin values may decrease,<br>Alkaline phosphatase activity may rise.<br>The control must be stored frozen when<br>using AST methods without pyridoxal-<br>5-phosphate. |
|------------------------------|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Storage<br>(Open, Thawed) | 2-8°C for 14 days. | 2-8°C (current Package Insert)<br>14 days with the following exceptions:<br>Direct Bilirubin, Triglycerides, HDL,<br>Cholinesterase, and Phosphorous will<br>be stable for 7 days and LAP<br>Arylamidase will be stable for 3 days. |
# Value Assignment
CONE-TROL Liquid Chemistry Control is value assigned on the Abaxis Piccolo Express Analyzer using Piccolo Xpress panels. Each panel is a self-contained plastic disc with reagents. Intelligent Quality Control and bar coded calibration are built in to each reagent disc. The panels used to value assign CONE-TROL Liquid Chemistry Control include but are not limited to: t. Seat 10 vas. Cat. No. 400-0030; Hepatic Panel, Cat. No. 400-0026; General Chemistry 13, Cat. No. 400-0029; Renal Panel Plus, Cat. No. 400-0027; Basic Metabolic Panel Plus, Cat. No. 400-0031; Liver Panel Plus, Cat. No. 400-0003; Metlyte +CRP, Cat. No. 400-0034.
Endogenous analytes were value assigned on the Roche Cobas Mira Plus and Mindray BS-200 using Kamiya and Sekisui reagents.
Values are assigned by assaying 6 randomly selected vials from each manufacturing lot. Vials are measured in duplicate and across 2 or more instruments for each Abaxis Xpress panel listed above. Value assigument is performed within 30 days of manufacturing date. Vials used for value assignment are to be tested within 5 days after product is thawed.
#### Performance Testing:
The closed vial stability of CONE-TROL Liquid Chemistry Control at 2-8°C has been determined from real time stability testing. CONE-TROL Liquid Chemistry Control was stored closed vial at 2-8°C, as described in the package insert, for 33 days. Results indicate acceptable stability through 33 days.
The open vial stability of CONE-TROL Liquid Chemistry Control at 2-8°C has been determined from real time stability testing. After opening, CONE-TROL Liquid Chemistry Control was stored at 2-8°C for 15 days. To simulate expected handling conditions, vials were removed from refrigerator daily (excluding weekends), opened, and returned to refrigerator. Results indicate acceptable stability through 15 days.
{3}------------------------------------------------
The shelf life of CONE-TROL Liquid Chemistry Control at <-20°C has been determined from real time stability testing. CONE-TROL Liquid Chemistry Control was stored closed vial at -20°C, as described in the package insert, for 254 days. Results indicate acceptable stability through 254 days. Real time studies will be ongoing to support the shelf life of this product.
#### Summary:
The information provided in this pre-market notification demonstrates that CONE-TROL Liquid Chemistry Control is substantially equivalent to Liquid Assayed Multiqual Control (k011867). Substantial equivalence was demonstrated through comparison of intended use and physical properties to the commercially available and analytical predicate device. The information supplied in this pre-market notification provides reasonable assurance that the CONE-TROL Liquid Chemistry Control is safe and effective for its stated intended use.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white and appears to be a high-contrast image.
10903 New Hampshire Avenue Silver Spring, MD 20993
JUL 2 4 2012
Cone Bioproducts c/o William K. Cone 1012 N. Austin St. Sequin, TX 78155
k121759 Re:
Trade Name: CONE-TROL Liquid Chemistry Controls Regulation Number: 21 CFR §862.1660 Regulation Name: Quality Control material (assayed and unassayed) Regulatory Class: Class I, reserved Product Codes: JJY Dated: June 1, 2012 Received: June 15, 2012
Dear Mr. Cone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
S. X. Liu, Ph.D.
Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Form
## 510(k) Number (if known): k121759
CONE-TROL Liquid Chemistry Controls Device name:
#### Indications for Use:
CONE-TROL Liquid Chemistry Control is a human liquid control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rte Clu
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
! 21759 510(k)
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.