← Product Code [HQC](/productcode/HQC) · K121721

# LAMINAR FLOW PHACO TIPS AND IRRIGATION SLEEVE (21G, REUSABLE) (K121721)

_Abbott Medical Optics, Inc. · HQC · Oct 15, 2012 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K121721

## Device Facts

- **Applicant:** Abbott Medical Optics, Inc.
- **Product Code:** [HQC](/productcode/HQC.md)
- **Decision Date:** Oct 15, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4670
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

The Laminar Flow Phacoemulsification Tips are used to break up (emulsify) the nucleus of a cataractous lens and remove the remaining nuclear fragments. The Irrigation Sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery. The Phacoemulsification tip is intended for use in conjunction with the Irrigation Sleeve during ocular surgery.

## Device Story

Device consists of reusable 21G titanium phacoemulsification tips and medical-grade silicone irrigation sleeves; used during cataract surgery. Tip attaches to ultrasonic handpiece to vibrate at ultrasonic frequencies, emulsifying cataractous lens material; irrigation sleeve directs balanced salt solution (BSS) across tip shaft into eye. Fragmented tissue is aspirated through tip lumen. Operated by surgeons in clinical settings. Reusability supported by cleaning and autoclave validation testing. Output is physical disruption and removal of cataract material. Benefits include effective lens extraction via smaller incision sizes compared to previous iterations.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (functional validation, flow rate), biocompatibility studies, and non-clinical animal studies demonstrating safety and performance equivalent to predicates.

## Technological Characteristics

Materials: Titanium Alloy (Ti 6Al-4V) for tips, medical-grade silicone for sleeves. Energy: Ultrasonic vibration. Dimensions: 21G gauge. Sterilization: Gamma radiation (initial) and autoclave (reprocessing). Standards: ISO 11137-1/2/3, ANSI/AAMI ST79, AAMI TIR12, TIR30, ISO 17664, ANSI/AAMI/IEC 80601-2-58.

## Regulatory Identification

A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.

## Predicate Devices

- AMO Irrigation Sleeve ([K103023](/device/K103023.md))
- Laminar Flow Phaco Tip ([K111882](/device/K111882.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K12172/
Section 3
510(k) Summary

## 3. 510(K) Summary

This 510(k) summary is being submitted in accordance with the Medical Device Amendments of 1976, the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21CFR 807.92.

Applicant:

Contact Person:

Abbott Medical Optics Inc. 1700 E. St. Andrew Place P.O. Box 25162 Santa Ana, CA 92799-5162, USA

Abbas M. Bohra, RAC Regulatory Affairs Specialist 1700 E. St. Andrew Place Santa Ana, CA 92705 Tel: (714) 247-8488 Fax: (714) 247-8784 Email: Abbas Bohra@amo.abbott.com

KIM ROGERS FOR ABBAS BONRA

510(k) Summary Preparation Date:

June 11, 2012

Device that is the subject of this notification:

Trade/Proprietary Name: Laminar Flow Phacoemulsification Tip and Irrigation Sleeve (21G. Reusable)

Classification Name:

Phacoemulsification Needle & Irrigation Sleeve (21 CFR 886.4670, Product code HQC)

The devices to which substantial equivalence are claimed are listed below in Table 3-1:

Table 3-1: Predicate Devices to which Substantial Equivalence is claimed

| Predicate Device<br>Name  | Predicate Trade<br>Name | 510(k)<br>Holder | 510(k)<br>Number | Clearance<br>Date |
|---------------------------|-------------------------|------------------|------------------|-------------------|
| AMO Irrigation<br>Sleeve  | Phaco Sleeve            | AMO              | K103023          | 05/24/2011        |
| Laminar Flow<br>Phaco Tip | Phaco Tip               | AMO              | K111882          | 10/24/2011        |

Table 3-2 compares features of the proposed Phaco Tips to those of the predicate device, The Laminar Flow Phaco Tip (cleared by the FDA under K111882). The Phaco Tip, subject of this premarket notification, is exactly the same in comparison to the predicate phaco tip except for the reusability. The Phaco Tips are compatible with all currently marketed AMO ultrasonic handpieces due to their hub size and the internal right-handed threading that

Abbott Medical Optics Inc.

{1}------------------------------------------------

allows them to be screwed onto the distal end of any AMO handpiece. The only difference pertaining to reusability will be supported by additional testing highlighted in latter sections.

Aside from the difference highlighted in bold in the table below, the Phaco Tips have the same intended use, material, and functionality as its predicate and are therefore deemed to be substantially equivalent.

| Features and<br>Characteristics | Proposed Device<br>The Laminar Flow Phaco Tips                                                                                                                                               | Predicate Device<br>The Laminar Flow Phaco<br>Tips (K111882) |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Intended Use                    | With an Ultrasonic Handpiece during<br>phacoemulsification surgery                                                                                                                           | Same                                                         |
| Description                     | The needle supplies ultrasonic<br>movement at the tip to break up<br>cataract material. The needle also<br>supports an aspiration fluid path for<br>removing cataract particles from the eye | Same                                                         |
| Handpiece<br>Compatibility      | All currently marketed AMO Ultrasonic<br>Handpieces                                                                                                                                          | Same                                                         |
| Patient Contact<br>Material     | Titanium Alloy (Ti 6Al-4V)                                                                                                                                                                   | Same                                                         |
| Sterilization<br>Method         | Gamma Radiation                                                                                                                                                                              | Same                                                         |
| Use                             | Reusable                                                                                                                                                                                     | Single-use                                                   |
| Thread                          | Right-Handed                                                                                                                                                                                 | Same                                                         |
| Tip Bevel Angle                 | Straight: 0-60°<br>Curved: 30-45°                                                                                                                                                            | Same                                                         |
| Hub Design                      | Tapered with 5 lands                                                                                                                                                                         | Same                                                         |
| Tip Length                      | Straight: 0.834" - 0.854"<br>Curved: 0.840" - 0.860"                                                                                                                                         | Same                                                         |
| Outer Diameter                  | 21G 0.029" - 0.031"                                                                                                                                                                          | Same                                                         |
| Inner Diameter                  | 21G 0.020" - 0.022"                                                                                                                                                                          | Same                                                         |
| Curve Range                     | 0-24°                                                                                                                                                                                        | Same                                                         |
| Packaging                       | With Irrigation Sleeve only                                                                                                                                                                  | With Irrigation Sleeve<br>only                               |

## Table 3-2: Comparison Overview of the Phaco Tips to the Predicate Devices

As the subject 21G Phaco Tips are intended to be packaged with the 21G Irrigation Sleeve, which is also the subject of this premarket notification; Table 3-3 referenced below compares the features of the proposed Irrigation sleeve with its respective predicate device, AMO Irrigation Sleeve, cleared by the FDA under K103023.

The Irrigation sleeve, subject of this premarket notification, is similar in comparison to the cleared Irrigation sleeve except for the reusability. The different gauge size (21G) compared to the predicate sleeve is mainly for achieving a smaller incision size during surgery. The smaller incision size thus enables the surgeons to use this device for Micro-implantation cataract surgeries. Additionally, the difference in flow rate specifications is to demonstrate a higher flow rate during surgery for the proposed Irrigation sleeve. Moreover, the difference in sieeve color (Light Blue and Orange vs. Yellow) is to differentiate between various sleeve gauge size offerings within the market (20G vs. 21G). Moreover, the difference in the sleeve. color within the 21G size offering (Light Blue vs Orange) is only to offer the surgeon a preferred choice for use. Lastly, the proposed Irrigation sleeve will be provided as reusable

{2}------------------------------------------------

compared to AMO's disposable sleeve product offering. This additional reusability claim has been supported by testing highlighted in latter sections.

Aside from these differences highlighted in bold, in the table below, the Irrigation sleeves have the same intended use, material, and functionality as its predicate and are therefore. deemed to be substantially equivalent. Additionally, Functional validation testing has been performed to demonstrate higher flow rate for the proposed irrigation Sleeve. This testing can be found below under Section 16, Performance Testing - Bench.

| Features and<br>Characteristics  | Proposed Device<br>The Laminar Flow Irrigation Sleeve                                                                                                                                        | Predicate Device<br>AMO Irrigation Sleeve<br>(K103023)     |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Intended Use                     | To provide Irrigation during<br>phacoemulsification surgery                                                                                                                                  | Same                                                       |
| Description                      | The needle supplies ultrasonic<br>movement at the tip to break up cataract<br>material. The needle also supports an<br>aspiration fluid path for removing cataract<br>particles from the eye | Same                                                       |
| Contact Material                 | Silicone, Medical Grade                                                                                                                                                                      | Same                                                       |
| Colorant                         | Light Blue (Gray 61-801220, Blue 121<br>and White 177 Gum, odorless, stable,<br>insoluble)<br>Orange (Red 042 and Yellow 185<br>Gum, odorless, stable, insoluble)                            | Yellow 185 (Yellow<br>Gum, odorless, stable,<br>insoluble) |
| Use                              | Reusable                                                                                                                                                                                     | Single-use                                                 |
| Sterilization<br>Method          | Gamma Radiation                                                                                                                                                                              | Same                                                       |
| Flow Rate (cc/min)               | $\ge 40$                                                                                                                                                                                     | $\ge 18$                                                   |
| Gauge Size                       | 21G                                                                                                                                                                                          | 19G - 23G                                                  |
| Length Range<br>(inches)         | 1.0                                                                                                                                                                                          | 0.75 - 1.50                                                |
| Inner Diameter<br>Range (inches) | 0.050                                                                                                                                                                                        | 0.030 - 0.065                                              |
| Outer Diameter<br>Range (inches) | 0.063                                                                                                                                                                                        | 0.050 - 0.080                                              |
| Weight<br>(grams/sleeve)         | 0.4                                                                                                                                                                                          | Same                                                       |
| Durometer (units)                | $80 \pm 5$                                                                                                                                                                                   | Same                                                       |

#### Table 3-3: Comparison Overview of the Irrigation Sleeve to the Predicate Device

{3}------------------------------------------------

#### 3.1 Device Description

Phacoemulsification (Phaco) is a technique used during cataract surgery to emulsify and extract a cataractous lens from the eye with an ultrasonic handpiece. The ultrasonic handpiece is connected to, and powered by a phacoemulsification system, which is identified in 21 CFR 886.4670 as "an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract."

The fragmentation needle, also referred to as a phaco tip, vibrates at an ultrasonic frequency and emulsifies the cataract when connected to an ultrasonic phacoemulsification handpiece. The fragmented tissue can then be aspirated and removed through the lumen of the tip. During a surgical procedure, irrigation solution, typically Balanced Salt Solution (BSS), is used to irrigate the eye and this is passed with the help of an Irrigation Sleeve. The Irrigation Sleeve directs the irrigation solution across the shaft of the phacoemulsification tip, allowing it to enter the eye during ocular surgery. This small molded component is placed over the Phacoemulsification tip. The proposed tips and irrigation sleeves are made of fitanium and medical grade silicone respectively.

The subject of this 510(k) premarket notification is the Reusable Phacoemulsification Tips (Phaco Tips) & Irrigation Sleeves (Phaco Sleeves), which will be additions to the family of Disposable Laminar Flow Phacoemulsification Tips & Sleeves. The Phaco tips that are the subject of this 510(k) premarket notification have two different designs, straight and curved tip. The same two designs were the subject of the predicate device and have been cleared by the FDA under K111882 (AMO Laminar Flow Phaco Tips). These tips were cleared by the FDA for single use only and now AMO intends to introduce the reusable version of the same models into the market to offer a broad range of products to our customers. These tips will be compatible with all our currently marketed AMO ultrasonic handpieces, and will be made of the same surgical grade titanium as the predicate device. Moreover, these 21G tips will also be compatible with our 21G Irrigation Sleeves (Phaco Sleeve), which is the other subject device of this 510(k) premarket notification. The 21G Irrigation Sleeve will have a slightly different design as compared to the predicate device, which has been cleared under K103023 (AMO Irrigation Sleeve). The reduced gauge size (21G) is designed for use with the 21G Phaco tips. The proposed Light Blue and Orange sleeve will be made of the same medical grade silicone material as the predicate device (Yellow Sleeve), with a difference in the colorant to distinguish between the various AMO sleeve offerings in the market.

#### 3.2 Indications for Use

The Laminar Flow Phacoemulsification Tips are used to break up (emulsify) the nucleus of a cataractous lens and remove the remaining nuclear fragments.

The Irrigation Sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

The Phacoemulsification tip is interided for use in conjunction with the Irrigation Sleeve during ocular surgery.

#### 3.3 Technological Characteristics

The technological characteristics of the Phaco Tips & Irrigation Sleeves are identical to those of the legally marketed predicate device also manufactured by AMO. With the exception of the reusability claim, the mechanism of action and technological features of the Phaco Tips & Irrigation Sleeves are the same as the predicate device with respect to materials, intended use, claims (except for reusability claim), clinical applications, patient population, method of operation, and performance specifications. Additional testing including Cleaning and

{4}------------------------------------------------

AMO Phacoemulsification Tip & Irrigation Sleeve 510{k} Premarket Notification

Autoclave validation has been performed to demonstrate reusability for both the proposed subject devices.

## 3.4 Summary of Performance Data

The proposed Phaco Tip has undergone various testing and is in compliance with applicable safety standards. Functional testing, sterilization, shelf life, packaging integrity, cleaning and autoclave validation has been performed and based on the results obtained, the proposed AMO Laminar Flow Phaco Tip is deemed to be substantially equivalent in its function and performance to the legally marketed predicate device, the AMO Laminar Flow Phaco Tips (K111882).

The probosed Irrigation sleeve has also undergone various testing and is in compliance with applicable safety standards. Functional testing, sterilization, shelf life, packaging integrity, cleaning and autoclave validation has been performed and based on the results obtained, the proposed AMO Laminar Flow Irrigation Sleeve is deemed to be substantially equivalent in its function and performance to the legally marketed predicate device, AMO Irrigation sleeve (K103023).

Based on the results obtained, the performance testing was considered successful and demonstrates that Phaco Tips & Irrigation Sleeve perform equivalently to the predicate devices with respect to safety and effectiveness.

The various performance testing performed for our Phaco Tips and Irrigation Sleeves are in compliance with the applicable standards listed below in Table 3-4.

| Standard Number                   | Document Title                                                                                                                                                               |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 11137-1:2006/(R)2010          | Sterilization of Health Care Products - Radiation - Part 1:<br>Requirements for Development, Validation, and Routine<br>Control of Sterilization Process for Medical Devices |
| ISO 11137-2:2006                  | Sterilization of Health Care Products - Radiation - Part 2<br>Establishing the Sterilization Dose                                                                            |
| ISO 11137-3:2006/ (R) 2010        | Sterilization of Health Care Products - Radiation - Part 3:<br>Guidance on Dosimetric Aspects                                                                                |
| ANSI/AAMI ST79: 2010 &<br>A1:2010 | Comprehensive Guide to Steam Sterilization and Sterility<br>Assurance in Health Care Facilities                                                                              |
| AAMI TIR12:2010                   | Designing, testing, and labeling reusable medical devices<br>for reprocessing in health care facilities: A guide for<br>medical device manufacturers                         |
| TIR30:2011                        | A compendium of processes, materials, test methods, and<br>acceptance criteria for cleaning reusable medical devices                                                         |
| ISO 17664:2004                    | Sterilization of medical devices - Information to be<br>provided by the manufacturer for the processing of<br>resterilizable medical devices.                                |
| ANSI/AAMI/IEC 80601-2-<br>58:2008 | Particular requirements for the basic safety and essential<br>performance of lens removal devices and vitrectomy<br>devices for ophthalmic surgery                           |

#### Table 3-4: Standards Used in Testing of Phaco Tip, Irrigation Sleeve & Test Chamber

{5}------------------------------------------------

## 3.5 Summary of Non-Clinical Data

Biocompatibility studies have been performed for the Phaco Tips and Irrigation Sleeves and results have shown the material used in these devices to be biocompatible and safe for use during the cataract surgery. Additionally Non-clinical animal studies were performed for the Straight and Curved Laminar Flow Phaco Tips along with the Irrigation Sleeves. Both studies demonstrate that Phaco tips and sleeves are safe and effective for human use and that they perform well with our various AMO phacoemulsification systems and handpieces and pose no additional risks.

Both these Non-Clinical testing (Animal and Biocompatibility testing, irrespective of their colorants) demonstrate similar effectiveness and performance profiles with respect to their predicates and hence they are deemed to be substantially equivalent to their respective predicate devices.

## 3.6 Overall Summary

The information presented in this 510(k) premarket notification demonstrates that the Laminar Flow Phaco Tips & Irrigation Sleeve are substantially equivalent to the legally marketed predicate device, also manufactured by AMO. The technological characteristics, material, intended use, and manufacturing processes remain unchanged for the Phaco Tips & Irrigation Sleeve when compared to their predicates. AMO concludes that these devices are safe, effective, and substantially equivalent to their predicate devices as described herein.

{6}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

OCT 15 2012

Abbott Medical Optics Inc. % Mr. Abbas Bohra Regulatory Affairs Specialist 1700 E. St. Andrew Place Santa Ana, CA 92705

Re: K121721

Trade/Device Name: Laminar Flow Phaco Tips and Irrigation Sleeve (21G, Reusable) Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HQC Dated: September 6, 2012 Received: September 6, 2012

Dear Mr. Bohra:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{7}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for

N
L
D

O
O

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

#### 2. INDICATIONS FOR USE STATEMENT

#### 510(k) Number (if known): K121721

Device Name: Laminar Flow Phacoemulsification Tip and Irrigation Sleeve (21G, Reusable)

#### Indications For Use:

The Laminar Flow Phacoemulsification Tips are used to break up (emulsify) the nucleus of a cataractous lens and remove the remaining nuclear fragments.

The Irrigation Sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

The Phacoemulsification tip is intended for use in conjunction with the Irrigation Sleeve during ocular surgery.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K/2/21

Abbott Medical Optics Inc.

---

**Source:** [https://fda.innolitics.com/device/K121721](https://fda.innolitics.com/device/K121721)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
