The Orthofix Anterior Cervical Plate System is a temporary implant, intended for anterior fixation to the cervical spine from C2 to C7 and indicated for: - Degenerative disc disease (as defined as back pain of discogenic origin with degenerative a) disc confirmed by patient history and radiographic studies); - Spondylolisthesis; b) - Trauma (i.e., fracture or dislocation); c) - d) Spinal stenosis; - Deformities (i.e., scoliosis, kyphosis, and/or lordosis); e) - Tumor; f) - Pseudoarthrosis; g) - Revision of previous surgery h)
Device Story
Orthofix Anterior Cervical Plate System is a temporary spinal fixation implant. System comprises titanium alloy (Ti6Al4V ELI) and Nitinol plates and screws. Surgeon implants system via anterior approach to cervical spine (C2-C7) to provide stabilization. Instrumentation provided for surgical implantation is manual. Device functions as mechanical support for vertebral bodies. Benefits include stabilization of cervical spine segments to facilitate fusion or address structural instability. Used in clinical/surgical settings by orthopedic or neurosurgeons.
Clinical Evidence
Bench testing only. Mechanical testing conducted per ASTM F1717-11, including static torsion, static axial compression bending, and dynamic axial compression bending tests, to demonstrate substantial equivalence to predicate devices.
Technological Characteristics
Materials: Ti6Al4V ELI (ASTM F136) and Nitinol (ASTM F2063). Components: Plates (1-5 levels) and screws. Fixation: Anterior cervical spine (C2-C7). Energy source: None (manual instrumentation). Sterilization: Non-sterile (implants).
Indications for Use
Indicated for patients requiring anterior cervical spine fixation (C2-C7) for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or revision surgery.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Orthofix logo and the number K121658. The Orthofix logo is a stylized image of a circle with a curved line running through it. The word "ORTHOFIX" is written in bold letters below the logo. The number K121658 is written in black ink in the upper left corner of the image.
JUL 200 2012
・・・
Orthofix Anterior Cervical Plate System
# 510(k) SUMMARY
### Orthofix Anterior Cervical Plate System
## Submitter Information
| Name:<br>Address: | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056 |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone Number:<br>Fax Number:<br>Email: | 214-937-2000<br>214-937-3322<br>alisonbaduel@orthofix.com |
| Registration Number: | 3008524126 |
| Contac Person: | Ally Baduel<br>Regulatory Affairs Specialist |
| Date Prepared: | June 4, 2012 |
| Name of Device | |
| Trade Name/Proprietary<br>Name: | Orthofix Anterior Cervical Plate System |
| Common Name: | anterior cervical plate system |
| Product Code: | KWQ - Appliance, Fixation, Spinal Intervertebral Body |
| Regulatory Classification: | Class II - 888.3060 - Spinal intervertebral body fixation orthosis |
| Review Panel: | Orthopedic Device Panel |
| Predicate Devices: | Hallmark Anterior Cervical Plate System (K050892 and<br>K100614), DePuy/Codman & Shurtleff, Inc. SLIM LOC<br>Anterior Cervical Plate System (K013877), Pioneer Aspect<br>Anterior Cervical Plate (K111528) |
Reason for 510(k) Submission: New product offering
2. lofy
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for ORTHOFIX. The logo consists of a stylized symbol above the word "ORTHOFIX". The symbol is a geometric shape with curved lines, and the text is in a bold, sans-serif font. The logo is black and white.
#### Device Description
The Orthofix Anterior Cervical Plate System is comprised of a variety of non-sterile, single use, titanium alloy (Ti6Al4V ELI per ASTM F136) and Nitinol (per ASTM 2063) components that allow a surgeon to build a temporary anterior cervical implant construct. The system is attached to the anterior aspect of the vertebral body by means of screws to the cervical spine. The system consists of an assortment of screws, plates and associated instrumentation, which assists in the surgical implantation of the devices. The instrumentation provided to facilitate implantation are Class I, manual orthopedic standard surgical instruments.
#### Intended Use / Indications for Use
The Orthofix Anterior Cervical Plate System is a temporary implant, intended for anterior fixation to the cervical spine from C2 to C7 and indicated for:
- Degenerative disc disease (as defined as back pain of discogenic origin with degenerative a) disc confirmed by patient history and radiographic studies);
- Spondylolisthesis; b)
- Trauma (i.e., fracture or dislocation); c)
- d) Spinal stenosis;
- Deformities (i.e., scoliosis, kyphosis, and/or lordosis); e)
- Tumor; f)
- Pseudoarthrosis; g)
- Revision of previous surgery h)
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "KI21658" at the top, followed by the Orthofix logo. The Orthofix logo consists of a stylized black bird above the word "ORTHOFIX" in a bold, sans-serif font. The logo is centered in the image.
### Summary of Technological Characteristics of the Device Compared to the Selected Predicate Devices
| Characteristic | Subject<br>Device | Predicates | | | |
|-----------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Device Name | Orthofix<br>Anterior<br>Cervical Plate<br>System | Hallmark Anterior<br>Cervical Plate<br>System<br>(K050892) | Hallmark<br>Anterior<br>Cervical Plate<br>System<br>(K100614) | DePuy/Codman<br>& Shurtleff, Inc.<br>(K013877) | Pioneer Aspect<br>Anterior<br>Cervical Plate<br>(K111528) |
| Method of<br>Fixation | ACP system<br>intended for<br>anterior fixation<br>to the cervical<br>spine from C2<br>to C7. | ACP system<br>intended for<br>anterior fixation<br>to the cervical<br>spine from C2 to<br>C7. | ACP system<br>intended for<br>anterior fixation<br>to the cervical<br>spine from C2 to<br>C7. | ACP system<br>intended for<br>anterior fixation<br>to the cervical<br>spine from C2 to<br>C7. | ACP system<br>intended for<br>anterior cervical<br>fixation. |
| Implantation | Anterior<br>approach | Anterior approach | Anterior<br>approach | Anterior approach | Anterior<br>approach |
| Design | Plates (1-level<br>through 5-<br>level) | Plates (1-level<br>through 4-level) | Plates (5-level<br>plate only) | Plates (1-level<br>through 5-level) | Plates (1-level<br>through 4-level) |
| Material | Ti6Al4V ELI<br>per ASTM<br>F136 and<br>Nitinol per<br>ASTM F2063. | Ti6Al4V ELI per<br>ASTM F136. | Ti6Al4V ELI per<br>ASTM F136. | Titanium | Ti6Al4V ELI<br>per ASTM<br>F136 and<br>Nitinol per<br>ASTM F2063. |
PERFORMANCE DATA-Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence
| Characteristic | Standard / Test/ FDA Guidance |
|----------------------------------------|-------------------------------|
| Static Torsion Test | ASTM F1717-11 |
| Static Axial Compression Bending Test | ASTM F1717-11 |
| Dynamic Axial Compression Bending Test | ASTM F1717-11 |
### Performance Data Summary
Mechanical testing for the subject Orthofix Anterior Cervical Plate System was conducted in accordance to ASTM F1717-11- Standard Test Methods for Static and Fatigue for Spinal Implant Constructs in a Vertebrectomy Model. Test results demonstrated that the new, proposed device is substantially equivalent to the predicate device the Hallmark Anterior Cervical Plate System
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the number "K121658" written in black ink in the upper left corner. To the right and below the number is the Orthofix logo. The logo consists of a stylized black circle with a white crescent shape inside, and the word "ORTHOFIX" in bold black letters below the circle.
(K050892 and K100614) that have the same intended use, similar indications, technological characteristics, and principles of operation.
#### Basis of Substantial Equivalence
The subject Orthofix Anterior Cervical Plate System is substantially equivalent in design, configuration, function, indications for use to the Hallmark Anterior Cervical Plate System (K050892 & K100614), the DePuy/Codman & Shurtleff, Inc. Anterior Cervical Plate System (K013877), and the Pioneer Aspect Anterior Cervical Plate (K111528).
4.fy
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized graphic of three abstract shapes that resemble human figures or waves, stacked on top of each other.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Orthofix, Incorporated % Ms. Ally Baduel Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056
JUL 20 2012
Re: K121658
Trade/Device Name: Orthofix Anterior Cervical Plate Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: June 4, 2012 Received: June 5, 2012
Dear Ms. Baduel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
## Page 2-Ms. Ally Baduel
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a stylized black symbol above the word "ORTHOFIX" in bold, black letters. The symbol is an abstract shape, possibly representing a bone or joint, and the word "ORTHOFIX" is written in a sans-serif font.
Indications for Use
510(k) Number (if known):
K12-1658
Device Name: Orthofix Anterior Cervical Plate System
Indications for Use:
The Orthofix Anterior Cervical Plate System is a temporary implant intended for anterior fixation to the cervical spine from C2 to C7 and indicated for:
- a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies);
- Spondylolisthesis; b)
- c) Trauma (i.e., fracture or dislocation);
- d) Spinal stenosis;
- e) Deformities (i.e., scoliosis, kyphosis, and/or lordosis);
- Tumor; f)
- g) Pseudoarthrosis;
- h) Revision of previous surgery
Prescription Use: X (Part 21 CFR 801 Subpart D) And/Or
Over-The-Counter (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 1
K121658 510(k) Number_
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.