The Asthmapolis System includes the Asthmapolis Sensor which is an electronic accessory device intended for single-patient use to assist physicians and patients in recording and monitoring the actuations of prescribed MDI usage. The Asthmapolis Mobile Application records, stores, and transmits usage events from the Asthmapolis Sensor to a remote storage system. The Asthmapolis Web Application is software intended to allow users to review the collected information and characteristics of MDI use, to add detail associated with a recorded usage event, and to share that information with their physician in order to provide additional information associated with the condition for which their MDI medication(s) are prescribed. The Asthmapolis System may also be used in clinical trials where researchers need to know information about use of a MDI Medication(s) by a participant. The output of the Asthmapolis System is not intended to diagnose or replace a diagnosis provided by a licensed physician. The Asthmapolis system is not intended for use as a MDI medication dose counter, nor is it intended to indicate the quantity of medication remaining in an MDI.
Device Story
Asthmapolis System is an electronic MDI accessory for single-patient use; tracks MDI actuation events. Sensor attaches to MDI; records usage events via internal clock; transmits data to mobile application. Mobile app stores and forwards data to remote storage. Web application allows users/physicians to review usage history, add event details, and share data. Used in home or clinical trial settings; operated by patients or researchers. Provides longitudinal usage data to support clinical decision-making regarding asthma/respiratory conditions. Does not diagnose or count remaining doses.
Clinical Evidence
Bench testing only. Included software verification and validation, performance testing of MDI actuation sensor, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and wireless/Bluetooth technology performance per FDA guidance.
Technological Characteristics
Electronic MDI accessory; internal power source; Bluetooth wireless connectivity. Software-based data collection and review. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for single-patient use to assist physicians and patients in recording and monitoring actuations of prescribed MDI usage. May be used in clinical trials to track MDI medication use by participants. Not intended to diagnose, replace physician diagnosis, act as a dose counter, or indicate remaining medication quantity.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
. JUL = =2 *2012
# 510(k) Premarket Notification Reciprocal Labs Corporation Asthmapolis System
121609 510(k) Summary
| Submission Date: | March 8, 2012 | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Submitter: | Reciprocal Labs Corporation<br>612 W. Main Street, Ste. 201<br>Madison, WI 53703 | |
| Submitter and<br>Official Contact: | Ms. Inger L. Couture<br>Chief Regulatory Officer<br>Reciprocal Labs<br>612 W. Main Street, Ste. 201<br>Madison, WI 53703<br>+1 (608) 251-0470<br>+1 (608) 338-0883 (fax)<br>Inger.Couture@asthmapolis.com | |
| Manufacturing Site: | Reciprocal Labs Corporation<br>612 W. Main Street, Ste. 201<br>Madison, WI 53703 | |
| Trade Name: | Asthmapolis System | |
| Common Name: | Nebulizer | |
| Classification<br>Name: | NEBULIZER (DIRECT PATIENT INTERFACE) | |
| Classification<br>Regulation: | 21 CFR §868.5630 | |
| Product Code: | CAF | |
| Device Description: | Electronic MDI Accessory | |
| Substantially<br>Equivalent Devices: | SmartTrack System | K091803 |
{1}------------------------------------------------
## 510(k) Premarket Notification Reciprocal Labs Corporation Asthmapolis System
Intended Use:
The Asthmapolis System includes the Asthmapolis Sensor which is an electronic accessory device intended for single-patient use to assist physicians and patients in recording and monitoring the actuations of prescribed MDI usage.
The Asthmapolis Mobile Application records, stores, and transmits usage events from the Asthmapolis Sensor to a remote storage system.
The Asthmapolis Web Application is software intended to allow users to review the collected information and characteristics of MDI use, to add detail associated with a recorded usage event, and to share that information with their physician in order to provide additional information associated with the condition for which their MDI medication(s) are prescribed.
The Asthmapolis System may also be used in clinical trials where researchers need to know information about use of a MDI Medication(s) by a participant.
The output of the Asthmapolis System is not intended to diagnose or replace a diagnosis provided by a licensed physician. The Asthmapolis system is not intended for use as a MDI medication dose counter, nor is it intended to indicate the quantity of medication remaining in an MDI.
Technological characteristics of the Asthmapolis System and the SmartTrack System are largely equivalent. Similarities include the indications for use, basic principle of operation, data collection information, time of data recording via internal clock, utilization of software for varving types of data review and modification, dose counting characteristics and internal power source type.
The Asthmapolis System employs these technological characteristics in a similar way as the predicate device. Differences include the method used to detect sensor actuation, and the method used to send the usage data from the sensor to the database. These aspects of the device have been verified and validated in order to establish equivalent performance to the equivalent device. This information indicates that the Asthmapolis System is equivalent to the predicate device in terms of device safety and effectiveness.
Based upon this comparison of the predicate, and the accompanying testing results for the Asthmapolis System, the Asthmapolis System is substantially equivalent to the predicate device.
### Technology Comparison:
{2}------------------------------------------------
# 510(k) Premarket Notification Reciprocal Labs Corporation Asthmapolis System
| Performance<br>Testing Summary: | Non-clinical testing has been carried out to cover functional verification<br>and device performance. This included completion of software<br>verification and validation procedures, with performance testing of the<br>MDI actuation sensor system to ensure data is logged accurately for<br>MDI usage. This established correct functionality of the Asthmapolis<br>System according to the requirements. |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Software: | Software and Firmware for the Asthmapolis System was designed and<br>developed according to a robust software development process<br>aligned with "Design Control Guidance for Medical Device<br>Manufacturers", "The Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical Devices", "Guidance<br>for Off the Shelf Software Use in Medical Devices", and verified and<br>validated using guidance from the "General Principles of Software<br>Validation" as recommended by FDA.<br><br>Test results indicate that the Asthmapolis System complies with its<br>predetermined specifications. |
| Electrical Safety: | The Asthmapolis Sensor has successfully completed patient safety<br>testing according to IEC 60601-1. |
| Electromagnetic<br>Compatibility<br>Testing: | The Asthmapolis Sensor has successfully completed EMC testing<br>according to IEC 60601-1-2. |
| Performance<br>Testing - Bench: | The Asthmapolis System has successfully completed performance<br>testing according to applicable standards and internal testing.<br>Important to highlight in this summary, is the successful performance<br>testing that was completed for wireless/Bluetooth technology in<br>accordance with specifications and also with, "FDA's Guidance on<br>Radio-Frequency Wireless Technology in Medical Devices". In<br>addition, tests required for FCC licensing were successful. |
{3}------------------------------------------------
## 510(k) Premarket Notification Reciprocal Labs Corporation Asthmapolis System
Conclusion:
Hardware testing carried out for the Asthmapolis System indicates it meets design and performance functional requirements. Software verification demonstrates that device features are effective, and that the system configuration functions equivalently to the predicate device. The Asthmapolis System also meets standard requirements for electrical safety, electromagnetic compatibility, biocompatibility, cleaning validation, and wireless technology in medical devices.
This information indicates that the Asthmapolis System is equivalent to the predicate device in terms of device safety and effectiveness.
Based upon this comparison of the predicate, and the accompanying testing results for the Asthmapolis System, the Asthmapolis System is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
2 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Reciprocal Laboratories Corporation c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW Buffalo, Minnesota 55313
Re: K121609
Trade/Device Name: Asthmapolis System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: June 18, 2012 Received: June 19, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
•
નીમ
{5}------------------------------------------------
#### Page 2- Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucml 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Rh fo.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
### Indications for Use
510(k) Number (if known): K121609
Device Name: Asthmapolis System
Indications for use:
The Asthmapolis System includes the Asthmapolis Sensor which is an electronic accessory device intended for single-patient use to assist physicians and patients in recording and monitoring the actuations of prescribed MDI usage.
The Asthmapolis Mobile Application records, stores, and transmits usage events from the Asthmapolis Sensor to a remote storage system.
The Asthmapolis Web Application is software intended to allow users to review the collected information and characteristics of MDI use, to add detail associated with a recorded usage event, and to share that information with their physician in order to provide additional information associated with the condition for which their MDI medication(s) are prescribed.
The Asthmapolis System may also be used in clinical trials where researchers need to know information about use of a MDI Medication(s) by a participant.
The output of the Asthmapolis System is not intended to diagnose or replace a diagnosis provided by a licensed physician. The Asthmapolis system is not intended for use as a MDI medication dose counter, nor is it intended to indicate the quantity of medication remaining in an MDI.
Prescription X AND/OR Use
Over-The-Counter Use
(21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
510/k) Numb
L. Schultheis
sion of Anesthesiology, General Hospital
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.