ESON NASAL MASK

K121597 · Fisher & Paykel Healthcare Limited · BZD · Oct 2, 2012 · Anesthesiology

Device Facts

Record IDK121597
Device NameESON NASAL MASK
ApplicantFisher & Paykel Healthcare Limited
Product CodeBZD · Anesthesiology
Decision DateOct 2, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Intended Use

The F&P Eson Nasal Mask is intended to be used by individuals who have been diagnosed by a physician as requiring CPAP or Bi-Level Ventilator treatment. The F&P Eson Nasal Mask is intended for single-patient adult use in the home and multiple-patient adult use in the hospital or other clinical setting where proper disinfection of the device can occur between patient uses.

Device Story

Non-invasive nasal mask for CPAP/Bi-Level ventilation; delivers pressurized, potentially humidified/oxygen-enriched gases to patient nose. Features silicone 'RollFit' self-adjusting bridge seal; 'Ergofit' headgear with crown straps for stability; integrated exhalation vents with diffuser system to reduce draft; ball elbow for range of motion; headgear clips. Used in home or clinical settings; operated by patient or clinician. Device connects to breathing tube; provides interface for gas delivery. Benefits include pressure relief on nasal bridge, improved stability, and reduced air draft for patient comfort.

Clinical Evidence

Bench testing only. Performance and biocompatibility testing compared to predicate Zest Nasal Mask demonstrated substantial equivalence.

Technological Characteristics

Silicone nasal seal; headgear straps; swivel adaptor; exhalation vents; diffuser system. Non-invasive interface. No electronic components or software.

Indications for Use

Indicated for adult patients requiring CPAP or Bi-Level ventilator treatment. Suitable for single-patient home use or multi-patient hospital/clinical use with proper disinfection.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### Section 5 - 510(k) Summary K-121597 Image /page/0/Picture/3 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are on the top line in a bold, sans-serif font. The word "HEALTHCARE" is on the second line in a similar font, but smaller and underlined. 15 Maurice Paykel Place, East Tamaki P O Box 14 348, Panmure Auckland, New Zealand Telephone: +64 9 574 0100 Facsimile: +64 9 574 0148 Website: www.fphcare.com . ОСТ | Contact Person | James Thompson | |---------------------|-------------------------------------------------------------------------| | Date Prepared | 29 May 2012 | | Trade Name | F&P Eson™ Nasal Mask | | Common Name | Nasal Mask | | Classification Name | Non continuous ventilator IPPB<br>(21 CFR § 868.5905, product code BZD) | | Predicate Devices | K083832 Fisher & Paykel Healthcare Zest Nasal Mask | 2 2012 {1}------------------------------------------------ ### 5.1 Description The F&P Eson™ is a respiratory mask which is non-invasive. The silicone seal is positioned over the nose and it seals on the area around the nose and the upper lip. The mask is held on the face with headgear straps. It connects to a single breathing tube via a swivel adaptor to received pressurized gases. On the elbow are exhalation vents (bias holes) that allow exhaled gases to be continually flushed and removed to room air. The F&P Eson™ Nasal Mask has a diffuser system that is designed to diffuse air that is expelled from the mask. This reduces the draft. The diffuser does not significantly change the amount of air flowing from the mask, therefore the mask can be used with or without the diffuser system. The air may also be humidified and/or oxygen enriched. An oxygen pressure port accessory is available to be used with the device for oxygen therapy and/or gas monitoring (previously covered by K003894 Fisher & Paykel Healthcare Limited Oracle Oral Mask). ### 5.2 Intended Use The F&P Eson Nasal Mask is intended to be used by individuals who have been diagnosed by a physician as requiring CPAP or Bi-Level Ventilator treatment. The F&P Eson Nasal Mask is intended for single-patient adult use in the home and multiple-patient adult use in the hospital or other clinical setting where proper disinfection of the device can occur between patient uses. #### 5.3 Technological Characteristics Comparison The F&P Eson™ Nasal Mask is very similar to the predicate Fisher & Paykel Healthcare Zest™ Nasal Mask. Both masks are intended for non-continuous ventilation via CPAP and Bi-Level therapy, and deliver CPAP gasses through the nose. The main difference is the F&P Eson™ has a self-adjusting rolling bridge, marketed as the RollFit Seal, for relieving pressure on the bridge of the nose and strategically thickened and thinned sections for overall structure, support and cushioning, whereas the Zest™ relies on the foam for structure, support and cushioning. Other differences with the F&P Eson™ include: - A permanently attached ball elbow rather than a detachable swivel elbow used by the . Zest™ for increase movement range. - A diffuser system that is designed to diffuse air that is expelled from the mask. This . reduces the draft. The diffuser does not significantly change the amount of air flowing from the mask, therefore the mask can be used with or without the diffuser system - Headgear similar to Zest™ but with the addition of the crown straps which is . designed to go over the top of the head for increased stability. This adds two more adjustment points (six in total for F&P Eson™) as opposed to four for Zest™. Marketed as Ergofit Headgear - The use of the headgear as forehead cushions for the F&P Eson™, rather than . silicone forehead pads used by the Zest™ for less parts. - . Headgear clips are incorporated in the lower headgear straps for F&P Eson™ for ease of use. {2}------------------------------------------------ ## 5.4 Non-Clinical Tests Testing of the F&P Eson™ Nasal Mask was compared to the predicate Fisher & Paykel Healthcare Zest™ Nasal Mask for performance and biocompatibility. These tests demonstrate substantial equivalence of the F&P Eson™ Nasal Mask to the predicate mask. Copies of test reports are included in Appendix B. ### Conclusion 5.5 The comparison of features, performance, and intended use demonstrate that the F&P. Eson™ Nasal Mask is substantially equivalent to the predicate Fisher & Paykel Healthcare Zest™ Nasal Mask for performance and biocompatibility. The F&P Eson™ Nasal Mask has been demonstrated to be safe and effective for CPAP and Bi-Level ventilation therapy. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a stylized representation of human figures. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 2 2012 OCT Fisher & Paykel Healthcare, Limited Mr. James Thompson Regulatory Affairs Manager - OSA 15 Maurice Paykel Place, East Tamaki P.O. Box 14 348, Panmure Auckland, New Zealand Re: K121597 Trade/Device Name: F&P Eson™ Nasal Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: September 12, 2012 Received: September 17, 2012 Dear Mr. Thompson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2- Mr. Thompson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, For Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Section 4 - Indications For Use Statement 510(k) Number: Device Name: F&P Eson™ Nasal Mask Indications for Use: The F&P Eson Nasal Mask is intended to be used by individuals who have been diagnosed by a physician as requiring CPAP or Bi-Level Ventilator treatment. The F&P Eson Nasal Mask is intended for single-patient adult use in the home and multiple-patient adult use in the hospital or other clinical setting where proper disinfection of the device can occur between patient uses. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schuler (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: k121597
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...