These devices are implantable bone plates and bone screws for facial procedures including: 1. Fractures 2. Osteotomies 3. Reconstructive procedures 4. Revision procedures where other treatments or devices have failed
Device Story
Biomet Microfixation Facial Plating System consists of titanium or titanium alloy bone plates, meshes, and screws for internal fixation of facial fractures and reconstruction. Plates available in various geometries (straight, curved, L, T, Z, X, Y, H, etc.) and thicknesses; offered flat or pre-bent. Surgeons contour flat plates intraoperatively using manual instruments. Screws range 1.0mm-2.3mm diameter and 2.0mm-29.0mm length. System used in clinical settings by surgeons to stabilize bone segments. Output is physical stabilization of bone; aids healing by maintaining alignment. Benefits include structural support for facial fractures and reconstructive sites.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including failure force in bending, rate of deflection, push-through strength, and insertion/fracture torque.
Technological Characteristics
Materials: Titanium or titanium alloy. Form factor: Plates (various shapes/thicknesses), meshes, and screws (1.0-2.3mm diameter). Principle: Mechanical internal fixation. Sterilization: Non-sterile, single-use. Connectivity: None.
Indications for Use
Indicated for patients requiring internal fixation for facial fractures, osteotomies, reconstructive procedures, or revision procedures following failed treatments. Contraindicated in patients with active infection, foreign body sensitivity, limited blood supply, insufficient bone quantity, latent infection, or mental/neurologic conditions preventing adherence to postoperative care.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
Lorenz 1.0mm, 1.5mm, 2.0mm Plating System (K953385)
Paulus Titanium Mini Bone Plates and Bone Screws (K862534)
{0}------------------------------------------------
Attachment 5 - Revised 510(k) Summary
## 510(k) Summary
**510(k) Number:**
**Date Prepared:**
**Submitter:**
K121589 July 23, 2012 Biomet Microfixation 1520 Tradeport Drive Jacksonville, FL 32218
Contact:
Lauren Jasper, Regulatory Affairs Specialist Lauren.Jasper@biomet.com Telephone: 904-741-9259 Fax: 904-741-9425
Common of Usual Name: Classification Name: Device Classification: Device Product Code:
Bone Plate or Plate, Fixation, Bone Class II 76 JEY (21 CFR 872.4760) Bone Screw Screw, Fixation, Bone Class II 87 HWC (21 CFR 888.3040)
SEP 2 1 2012
Device Name:
Biomet Microfixation Facial Plating System
### Intended Use:
These devices are implantable bone plates and bone screws for facial procedures including:
- 1. Fractures
- 2. Osteotomies
- 3. Reconstructive procedures
- 4. Revision procedures where other treatments or devices have failed
#### Contraindications:
- 1. Active infection
- Where material sensitivity is suspected, testing is to be 2. Foreign body sensitivity. completed prior to implantation.
- 3. Patients with limited blood supply, insufficient quantity of bone, or latent infection.
- 4. Patients with mental or neurologic conditions who are unwilling or incapable of following postoperative care instructions.
#### Possible Risks:
- Osteomyelitis, inhibited Osteoporosis, Osteolysis, formation, 1. Poor bone revascularization, or infection can cause loosening, bending, cracking or fracture of the device.
- 2. Nonunion or delayed union, which may lead to breakage of the implant.
- 3. Migration, bending, fracture or loosening of the implant.
- 4. Metal sensitivity or allergic reaction to a foreign body.
{1}------------------------------------------------
- 5. Decrease in bone density due to stress shielding.
- 6. Pain, discomfort, abnormal sensation, or palpability due to the presence of the device.
- 7. Increased fibrous tissue response around the fracture site and/or the implant.
- 8. Necrosis of bone.
- 9. Inadequate healing.
## Description:
The Biomet Microfixation Facial Plating System is comprised of a variety of titanium plates, meshes, and screws with shapes and sizes designed for internal fixation of facial fractures and reconstruction procedures. System implants are manufactured from either titanium or titanium alloy and are intended for single use only.
The Facial Plating System plates that are the subject of this 510(k) submission include variations of straight, angle, curved, L-shape, double T-shape, Z-shape, X-shape, Y-shape, double Y-shape, H-shape, triangle, square, rectangle, matrix, mesh, orbital floor, LeFort, and chin options with various lengths and thickness. Plates are offered flat or pre-bent. Surgeons use cutting and bending instruments intraoperatively to contour flat plates to patient anatomy; prebent plates are contoured by Biomet Microfixation per surgeon specifications or patient anatomy as a convenience. The Facial Plating System screws range in diameters of 1.0mm to 2.3mm and lengths from 2:0mm to 29.0mm.
### Sterility Information:
The plates and screws will be marketed as non-sterile, single-use devices.
## Technological Characteristics:
The subject Facial Plating System devices are similar to the predicate devices in terms of indications, use, and design. The non-clinical testing data shown below demonstrate that the subject devices have equivalent or better mechanical performance when compared to the predicate devices. Minor differences in device geometry do not raise new issues of safety and efficacy.
### Clinical Testing:
Clinical testing was not performed to support this submission.
## Non-Clinical Testing:
Mechanical testing was performed to compare the subject devices to the predicate devices to measure:
- Failure Force (1bf) in Bending of Plates ●
- Rate of Deflection (in/lb) of Plates .
- Push-Through Strength (Ibs) of Plates .
- Insertion Torque (in-oz) of Screws ●
- Fracture Torque (in-oz) of Screws .
{2}------------------------------------------------
The non-clinical test results demonstrate that the mechanical performance of the subject Facial Plating System plates and screws are equivalent or better than the predicate devices and support the substantial equivalence to the predicate devices.
KIZIS89
### Substantial Equivalence:
Biomet Microfixation considers the Facial Plating System modifications equivalent to the Biomet Microfixation's (formerly Walter Lorenz Surgical's) Lorenz 1.0mm, 1.5mm, 2.0mm Plating System cleared under K953385, Paulus Titanium Mini Bone Plates and Bone Screws cleared under K862534. Wuertzburg Plates and Screws cleared under K854886, Lorenz Self-Drilling Screw cleared under K013954, and Lorenz 1.0mm System cleared under K922741.
In conclusion, the subject Biomet Microfixation Facial Plating System devices have the same intended use and similar technological characteristics to the legally marketed predicate devices. Non-clinical testing demonstrates that slight geometrical differences in the devices do not affect safety or effectiveness. The information presented supports substantial equivalence of the subject devices to the predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure. The figure is composed of three curved lines that suggest a person in motion or reaching out. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the emblem.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP
21 2012
Biomet Microfixation Ms. Lauren Jasper Regulatory Affairs Specialist 1520 Tradeport Drive Jacksonville. Florida 32218
Re: K121589
Trade/Device Name: Biomet Microfixation Facial Plating System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: August 23, 2012 Received: August 24, 2012
Dear Ms. Jasper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2- Ms. Jasper
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K121589
# Device Name: Biomet Microfixation Facial Plating System
### Indications For Use:
These devices are implantable bone plates and bone screws for facial procedures including:
- 1. Fractures
- 2. Osteotomies
- 3. Reconstructive procedures
- 4. Revision procedures where other treatments or devices have failed
AND/OR Prescription Use XX (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Susa Puno
(Division Sign-Off) Division of Anestheslology, General Hospital Infection Control, Dental Devices
510(k) Number;
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.