← Product Code [NDC](/productcode/NDC) · K121542

# PICIS WEIGHT BASED DOSE CONVERTER (K121542)

_Picis, Inc. · NDC · Aug 27, 2012 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K121542

## Device Facts

- **Applicant:** Picis, Inc.
- **Product Code:** [NDC](/productcode/NDC.md)
- **Decision Date:** Aug 27, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1890
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The Picis Weight Based Dose Converter is designed for use by trained clinicians to calculate any individual patient's dose for a given agent based on a weight based determination by the clinician. Druq dosing for a patient must be made only after careful consideration of the full clinical status of the patient by the ordering clinician and the software provides no validation as to the appropriateness of the entered weight based dose. The Picis Weight Based Dose Converter is not a substitute for clinical reasoning. The Picis Weight Based Dose Converter is a convenience feature for trained clinicians based upon accurately entered data. No medical decision should be based solely upon the results provided by this software program.

## Device Story

Software feature integrated into electronic medical record systems; calculates total drug dose based on patient weight and clinician-provided dosage units (e.g., mg/kg). Input: patient weight, numerical dose, units, agent form, route of administration. Operation: simple mathematical multiplication of weight by dose-per-weight factor. Output: total drug dose. Used in hospital settings by trained clinicians. Serves as a convenience tool to reduce manual calculation errors; does not validate clinical appropriateness of dose. Clinical decision-making remains the sole responsibility of the ordering clinician.

## Clinical Evidence

No clinical data. Evidence consists of software verification and validation testing to confirm functionality and reliability of the dose calculation feature.

## Technological Characteristics

Software-based drug dose conversion feature. Operates as an integrated component of an electronic medical record system. Uses simple mathematical equations for unit conversion. No specific hardware, energy source, or materials of construction defined.

## Regulatory Identification

A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.

## Predicate Devices

- TRxF Intelligent Dosing System ([K011571](/device/K011571.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K121542

### 510(k) SUMMARY

#### Picis, Inc.'s Weight Based Dose Converter

AUG 27 2012

## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Picis, Inc. 100 Quannapowitt Parkway Wakefield, MA 01880 Phone: (781) 557-3034 Facsimile: (781) 557-3118

Contact Person: Karen A. Iorio

Date Prepared: May 24, 2012

#### Device Trade Name

Picis Weight Based Dose Converter

### Common or Usual Name / Classification Name

Drug Dose Calculator (Product Code: NDC; 21 C.F.R. 868.1890)

#### Predicate Devices

The Rx Files Corp., TRxF Intelligent Dosing System (K011571)

### Intended Use / Indications for Use

The Picis Weight Based Dose Converter is designed for use by trained clinicians to calculate any individual patient's dose for a given agent based on a weight based determination by the clinician. Druq dosing for a patient must be made only after careful consideration of the full clinical status of the patient by the ordering clinician and the software provides no validation as to the appropriateness of the entered weight based dose.

The Picis Weight Based Dose Converter is not a substitute for clinical reasoning. The Picis Weight Based Dose Converter is a convenience feature for trained clinicians based upon accurately entered data. No medical decision should be based solely upon the results provided by this software program.

#### Technological Characteristics

The Picis Weight Based Dose Converter is a drug dose conversion feature designed for use with an electronic medical record system. The feature allows a user to convert a drug dose in units per body weight to a total dose. For example, the Weight Based Dose Converter provides for a conversion of an agent dose in milligrams per kilogram of body weight to an agent dose in total milligrams according to the patient's weight.

{1}------------------------------------------------

## Performance Data

Software verification and validation testing activities were conducted to establish the performance, functionality, and reliability of the Picis Weight Based Dose Converter software. The results of the testing confirmed that the Picis Weight Based Dose Converter software performs as intended.

### Substantial Equivalence

The Picis Weight Based Dose Converter software is substantially equivalent to the identified predicate device listed above. The Weight Based Dose Converter software has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the subject Weight Based Dose Converter software and its predicate device raise no new issues of safety or effectiveness. Thus, the Picis Weight Based Dose Converter software is substantially equivalent.

## Predicate Device Comparison Table

|                                         | Weight Based Dose Converter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate<br>TRxF Intelligent Dosing System<br>(K011571)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/<br>Indications for<br>Use | The Picis Weight Based Dose<br>Converter is designed for use by trained<br>clinicians to calculate any individual<br>patient's dose for a given agent based<br>on a weight based determination by the<br>clinician. Drug dosing for a patient must<br>be made only after careful consideration<br>of the full clinical status of the patient by<br>the ordering clinician and the software<br>provides no validation as to the<br>appropriateness of the entered weight<br>based dose. The Picis Weight Based<br>Dose Converter is not a substitute for<br>clinical reasoning. The Picis Weight<br>Based Dose Converter is a<br>convenience feature for trained<br>clinicians based upon accurately<br>entered data. No medical decision<br>should be based solely upon the results<br>provided by this software program. | The Intelligent Dosing System ("IDS")<br>is a three-part software suite<br>comprised of DoseRxTM,<br>InterchangeRxTM and<br>PracticePrescribeRxTM. The<br>DoseRxTM is designed for use by<br>trained clinicians to calculate any<br>individual patient's optimal next dose<br>for any given agent. The<br>InterchangeRxTM is designed to<br>switch a patient from one brand of<br>agent to another while maintaining the<br>therapeutic effect of the original<br>agent. The PracticePrescribeRxTM is<br>a dosing simulator that offers graded<br>prescriber training of the next dose<br>calculation scenarios with scalable<br>patient response and surrogate<br>marker inputs that allows the<br>healthcare provider to gain guided<br>and measured experience in<br>calculating the next dose for a new or<br>infrequently used drug. The<br>Intelligent Dosing System is not a<br>substitute for clinical reasoning. The<br>IDS is an aid for trained clinicians<br>based upon significant and properly<br>entered data. Final drug dose<br>recommendations for a patient must<br>be made only after careful<br>consideration of the fully clinical |

{2}------------------------------------------------

|                      | Weight Based Dose Converter                                                                                      | Predicate<br>TRxF Intelligent Dosing System<br>(K011571)                                                                                 |
|----------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Target<br>Population | In-hospital patients                                                                                             | In-hospital patients                                                                                                                     |
| User<br>Population   | Trained healthcare professionals                                                                                 | Trained healthcare professionals                                                                                                         |
| User Input           | Patient weight, numerical dose and unit,<br>agent form and route of administration                               | Patient's last agent dose, surrogate<br>marker response information, the<br>current dose and the desired<br>surrogate marker response    |
| Dose<br>computation  | Software-based application used to<br>compute total drug dose                                                    | Software-based application used to<br>compute next optimal drug dose                                                                     |
| Algorithms           | Uses simple mathematical equation to<br>calculate dosing based on patient<br>weight and numerical dose and units | Uses mathematical clinical algorithm<br>to give proposal for dosing based on<br>scalable patient response and<br>surrogate marker inputs |
| Testing              | Software validation and verification                                                                             | Software validation and verification                                                                                                     |

and the control control control of the control of

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three legs, representing service to the people. The eagle is positioned to the right of a circular inscription that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 27 2012

Picis Incorporated C/O Ms. Janice Hogan Hogan Lovells US, Limited Liability Partnership 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103

Re: K121542

Trade/Device Name: Picis Weight Based Dose Converter Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: II Product Code: NDC Dated: August 8, 2012 Received: August 8, 2012

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

# Page 2- Ms. Hogan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

For Alex

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: Picis Weight Based Dose Converter

Indications for Use:

The Picis Weight Based Dose Converter is designed for use by trained clinicians to calculate any individual patient's dose for a given agent based on a weight based dose determined by the clinician. Drug dosing for a patient must be made only after careful consideration of the full clinical status of the patient by the ordering clinician and the software provides no validation as to the appropriateness of the entered weight based dose.

The Picis Weight Based Dose Converter is not a substitute for clinical reasoning. The Picis Weight Based Dose Converter is a convenience feature for trained clinicians based upon accurately entered data. No medical decision should be based solely upon the results provided by this software program.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rill C. Chyz

(Division Sign-Off) Ivision of Anesthesiology, General Hospital tion Control. Dental I

**510(k) Number:** K121542

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K121542](https://fda.innolitics.com/device/K121542)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
