← Product Code [FLL](/productcode/FLL) · K121427

# UNIMED TEMPERATURE PROBE (K121427)

_Unimed Medical Supplies, Inc. · FLL · May 29, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K121427

## Device Facts

- **Applicant:** Unimed Medical Supplies, Inc.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** May 29, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Unimed Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors compatible with YSI 400 series temperature probes. These devices are indicated for used by qualified medical personnel only.

## Device Story

Unimed Temperature Probes (Skin and General Purpose models) are reusable medical devices used to monitor patient temperature. The device consists of a patient-end thermistor and a monitor-end connector, sometimes utilizing an extension cable. The thermistor acts as a temperature-sensitive resistor, converting thermal changes into electrical signals processed by compatible patient monitors (e.g., Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W). Used in clinical settings by qualified medical personnel, the probes provide continuous temperature data to the monitor, enabling healthcare providers to track patient thermal status and inform clinical decision-making. The device benefits patients by facilitating accurate, continuous temperature monitoring.

## Clinical Evidence

Bench testing only. Compliance demonstrated with IEC 60601-1 (general safety), EN 12470-4 (performance of electrical thermometers), ISO 10993-5 (in vitro cytotoxicity), and ISO 10993-10 (irritation and delayed-type hypersensitivity).

## Technological Characteristics

Reusable temperature probes featuring thermistor sensing technology. Accuracy maintained in 25-45°C range. Compatible with YSI 400 series systems. Materials include various cable and plug configurations. Tested per IEC 60601-1, EN 12470-4, ISO 10993-5, and ISO 10993-10.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- YSI 400 Series Autoclavable Temperature Probe ([K962070](/device/K962070.md))
- GE Reusable Temperature Probes ([K050837](/device/K050837.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K.121427

MAY 2 9 2012

Unimed Medical Supplies Inc No. 37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26695165 Fax: 86 755 26697984 Email: info@unimed.cn Website: www.unimed.cn

Image /page/0/Picture/3 description: The image shows a logo with the word "UNIMED" in bold, sans-serif font. Below "UNIMED" is smaller text, possibly in a different language. To the right of the text is a circular graphic with a complex design, possibly representing a medical symbol or emblem. The overall impression is that of a company logo for a medical or healthcare-related organization.

### 510(K) Summary Section 5

### Date: 2012-01-15

## Submitter Information:

Unimed Medical supplies Inc. No.37, Yanshan Road, Shekou, Shenzhen, China 518067

# Contact Person:

Xinmei Tan, QA manager No.37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: (86) 755 26695165 Fax: (86) 755 26697984 Email: tanxinmei@unimed.cn & caijiye@unimed.cn

# Proposed Device:

Trade Name: Unimed Temperature Probe (Unimed Skin Temperature Probe, Unimed General Purpose Temperature Probe) Common Name: Temperature Probe Product Classification: FLL; 880.2910; Class II Classification Panel: General Hospital

| Model    | Name                                     |
|----------|------------------------------------------|
| T2252-AS | Unimed Skin Temperature Probe            |
| THP-AS   | Unimed Skin Temperature Probe            |
| TMQ-AS   | Unimed Skin Temperature Probe            |
| TSW-AS   | Unimed Skin Temperature Probe            |
| TSM-15AS | Unimed Skin Temperature Probe            |
| TSL-AS   | Unimed Skin Temperature Probe            |
| TMR-AS   | Unimed Skin Temperature Probe            |
| T2252-AG | Unimed General Purpose Temperature Probe |
| THP-AG   | Unimed General Purpose Temperature Probe |
| TMQ-AG   | Unimed General Purpose Temperature Probe |
| TSW-AG   | Unimed General Purpose Temperature Probe |

{1}------------------------------------------------

### Unimed Medical Supplies Inc

No. 37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26695165 Fax: 86 755 26697984 a www.unimed.co. Email: info@unimed co

Image /page/1/Picture/2 description: The image shows the logo for Unimed Medical Supplies. The logo features the word "Unimed" in a bold, sans-serif font, with the words "medical supplies" in a smaller font underneath. To the right of the word "Unimed" is a circular graphic with a design inside. The logo is black and white.

| webs e. w |                                          |
|-----------|------------------------------------------|
| TSM-15AG  | Unimed General Purpose Temperature Probe |
| TSL-AG    | Unimed General Purpose Temperature Probe |
| TMR-AG    | Unimed General Purpose Temperature Probe |

### Predicate Device:

YSI 400 Series Autoclavable Temperature Probe (K962070) GE Reusable Temperature Probes (K050837)

### Indications for use

Unimed Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors compatible with YSI 400 series temperature probes.

These devices are indicated for used by qualified medical personnel only.

### Device description

Unimed Skin Temperature Probe (T2252-AS, THP-AS, TMQ-AS, TSW-AS, TSM-15AS, TSL-AS, TMR-AS), and Unimed General Purpose Temperature Probe (T2252-AG, THP-AG, TMQ-AG, TSW-AG, TSM-15AG, TSL-AG, TMR-AG) are used during patient temperature measurement. These probes consist of a connector on the monitor end and a thermistor on the patient end. These probes are to be used with YSI 400 series compatible temperature measurement systems only.

Temperature probes measure temperature by a resistor that is sensitive to temperature changes.

The probe is connected to the monitor either directly by using the connector or by an extension cable. These probes have a skin or core contact with a patient.

# Summary of technological characteristics of device compared to the predicate device

The Unimed temperature probes are substantially equivalent in safety and effectiveness to the predicate YSI temperature probes (K962070) and GE temperature probes (K050837).

They have the same thermistor, accuracy in 25-45℃ range and they all apply to

{2}------------------------------------------------

### Unimed Medical Supplies Inc

No. 37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26695165 Fax: 86 755 26697984 Email: info@unimed.cn Website: www.unimed.cn

Image /page/2/Picture/2 description: The image shows the logo for Unimed USA medical supplies. The logo is in black and white, with the word "Unimed" in bold letters on the left. Below the word "Unimed" is the phrase "medical supplies" in smaller letters. To the right of the word "Unimed" is a circular graphic with the letters "USA" inside.

monitors compatible with YSI temperature probes and GE temperature probes. They are different in labeling, cable material, plug material, cable length.

From the biocompatibility test, safety test and other bench tests, these differences will not bring any safety and effectiveness problem.

In summary, Unimed temperature probes, described in this submission are substantially equivalent to the predicate YSI temperature probes (K962070) and GE temperature probes (K050837).

### Performance data

Bench and laboratory testing to demonstrate safety and effectiveness per: IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1:1991; Amendment 2:1995

EN 12470-4:2000+A1:2009 Clinical thermometers - Part 4: Performance of electrical thermometers for continuous measurement

ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity

ISO 10993-10:2002+A1:2006 Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity

## Conclusion:

Unimed Temperature Probes have the same intended use, the same technology as the predicate devices. They are claimed to be Substantially Equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine and health. The words "DEPARTMENT OF HEALTH & HUMAN" are arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Unimed Medical Supplies Inc. C/O Mr. Jeffrey D. Rongero Responsible Third Party Official Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle, North Carolina 27709

MAY 2 9 2012

Re: K121427

Trade/Device Name: Unimed Temperature Probe Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: May 14, 2012 Received: May 14, 2012

Dear Mr. Rongero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{4}------------------------------------------------

# Page 2 - Mr. Rongero

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for

Anthony D. Watson, B.S., M.S., M.B.A. . Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

# Enclosure

{5}------------------------------------------------

### Unimed Medical Supplies Inc

No. 37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26695165 Fax: 86 755 26697984 Website: www.unimed.cn Email: info@unimed.cn

Image /page/5/Picture/2 description: The image shows the logo for Unimed Medical Supplies. The logo features the word "Unimed" in a bold, sans-serif font, with the words "medical supplies" in a smaller font underneath. To the right of the word "Unimed" is a circular graphic with the letters "LAB" inside. The logo is black and white.

#### Indications for Use Section 4

510(k) Number: k121427

Device Name: Unimed Temperature Probe

Skin Types: T2252-AS, THP-AS, TMQ-AS, TSW-AS, TSM-15AS, TSL-AS, TMR-AS; General Purpose Types: T2252-AG, THP-AG, TMQ-AG, TSW-AG, TSM-15AG, TSL-AG, TMR-AG

# Indications for Use

Unimed Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors compatible with YSI 400 series temperature probes.

These devices are indicated for used by qualified medical personnel only.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Sub part D)

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of _

Aila C. Chay 5/29/12

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K121427

---

**Source:** [https://fda.innolitics.com/device/K121427](https://fda.innolitics.com/device/K121427)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
