As a TENS device: Symptomatic relief of chronic (long term) intractable pain Symptomatic relief of post-traumatic acute pain and post-surgical acute pain As an Interferential and Premodulated device: Symptomatic relief of chronic intractable pain, acute post traumatic pain or acute post surgical pain As a Russian device: Temporary relaxation of muscle spasms Prevention or retardation of disuse atrophy in post-injury type conditions Increase local blood circulation As an Burst modulated alternaing current (Aussie) device: Temporary relaxation of muscle spasms Prevention or retardation of disuse atrophy in post-injury type conditions Increase local blood circulation Muscle re-education Maintaining or increasing range of motion As a Microcurrent device: Symptomatic relief of chronic intractable pain Symptomatic relief of post-traumatic acute pain and post-surgical acute pain
Device Story
Neurodyn and Neurodyn Aussie are programmable powered muscle stimulators used in clinical settings by healthcare providers. Devices deliver various electrical currents—Russian, Aussie (MFAC), Interferential, TENS, Premodulated, and Microcurrent—to patients via electrodes. Input parameters (intensity, frequency, burst duration) are adjusted based on patient needs and prescription settings. The Neurodyn features 5 preset clinical programs and 10 user-adjustable protocols; the Neurodyn Aussie provides four independent output channels. Devices utilize microprocessor control to manage treatment timers, automatic shut-off, and overload protection. Output affects muscle tissue or nerves to provide pain relief, muscle relaxation, or re-education. Benefits include improved circulation, atrophy prevention, and pain management. Healthcare providers monitor patient response and adjust intensity levels accordingly.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety, electromagnetic compatibility (EMC), and performance verification against established standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10).
Technological Characteristics
Powered muscle stimulator; ABS plastic housing; LCD display. Delivers Russian, Aussie (MFAC), Interferential, TENS, Premodulated, and Microcurrent waveforms. 4 output channels with independent intensity control. Microprocessor-controlled. Connectivity: Standalone. Power: 100/240V Bivolt. Safety: Double isolation, Class II, Type BF. Electrodes: Axelgaard (K970426). Standards: NBR IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 13485.
Indications for Use
Indicated for patients requiring symptomatic relief of chronic intractable pain, post-traumatic acute pain, or post-surgical acute pain; temporary relaxation of muscle spasms; prevention or retardation of disuse atrophy; increase of local blood circulation; muscle re-education; and maintenance or increase of range of motion.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for IBRAMED. The logo consists of the word "IBRAMED" in a bold, sans-serif font, with a stylized symbol above it. The symbol is composed of horizontal lines and a vertical element that resembles the letter "I" with extensions on either side. The entire logo is in black against a white background.
### SECTION 9
## 510(k) Summary
Neurodyn Powered Muscle Stimulator Neurodyn Aussie Muscle Stimulator
510 (k) Number: K121369
Date of Submission December 13, 2012
### Submitter:
IBRAMED EQUIPAMENTOS MEDICOS Avenida Dr. Carlos Burgos 2800 Amparo - Sao Paulo - Brasil
TEL - 5519-3817-9633 FAX - 5519-7816-7980
### Official Contact:
Lilian Llull TechLink International Consulting 18851 NE 29th Avenue Suite 720 Aventura, FL 33180
TEL - (305) 206-6777 FAX - (305) 377-0088
This summary is provided in accordance with the Safe Medical Devices Act of 1990 (SMDA). The information provided in the 510 (k) premarket notification is in accordance with 21 CFR 807.87.
| Common (Standard) Name: | Powered Muscle Stimulator |
|-------------------------|-------------------------------------------|
| Trade Name: | Neurodyn Powered Muscle Stimulator |
| | Neurodyn Aussie Powered Muscle Stimulator |
### Regulation Number & Product Codes:
GZJ - 21 CFR 882.5890-Transcutaneous electrical nerve stimulator for pain relief
IPF - 21 CFR 890.5850-Powered muscle stimulator
LIH - Interferential Current Therapy-Pre-amendment
GZI- 21 CFR 882.5810-External functional neuromuscular stimulator
{1}------------------------------------------------
K12/369
Image /page/1/Picture/1 description: The image shows a logo for IBRAMED DEVICE CLASS II. The logo is black and white and features a stylized symbol above the word "IBRAMED". Below the word "IBRAMED" is the text "DEVICE CLASS II". The symbol above the word "IBRAMED" appears to be a stylized letter "I".
Specifications - Both devices were designed according to Existing Technical Standards for the Development of Medical Devices (NBR NBR IEC 60601-1 IEC 60601-1-2 and IEC 60601-2-10 NBR). Predicate Device Identification:
Chattanooga Vectra Genisys K031077
Predicate devices had been submitted and cleared by 510(k) for the same intended uses and indications.
## Device Description
Neurodyn and Neurodyn Aussie Neuromuscular Stimulators are intended for the treatment of, relief of chronic (long term) intractable pain as adjunctive treatment of post-surgical and post-traumatic acute pain. Both devices have the same intended uses and incorporate the same technologies as the predicate the Vectra Genisys K031077. The Neurodyn Muscle Stimulator is a programmable device. It comes equipped with 5 preset clinical programs along with 10 user protocols. The user programs are adjustable and can be changed according to the patient's needs, doctor's recommendations and prescription settings.
The Neurodyn Aussie Muscle Stimulator has four output channels with independent intensity controls. Thus, four different areas can be stimulated separately or together during a therapy session. It is adjustable and can be changed according to the patient's needs, doctor's recommendations and prescription settings. It generates the medium frequency alternate current (MFAC), burst modulated alternating current (Aussie)- type of sinusoidal current with a frequency carrying 1,000 Hz or 4,000 Hz and a burst duration of 4 ms or 2 ms, modulated in pulse trains (bursts) with a variable frequency from 1 to 120Hz. ·
### Indications for Use
As a TENS device:
Symptomatic relief of chronic (long term) intractable pain Symptomatic relief of post-traumatic acute pain and post-surgical acute pain
As an Interferential and Premodulated device:
Symptomatic relief of chronic intractable pain, acute post traumatic pain or acute post surgical pain
As a Russian device: Temporary relaxation of muscle spasms Prevention or retardation of disuse atrophy in post-injury type conditions Increase local blood circulation
{2}------------------------------------------------
1512/369
Image /page/2/Picture/1 description: The image shows the logo for IBRAMED. Below the logo is the text "Muscle re-education" and "Maintaining or increasing range of motion". The logo is black and white and features a stylized "I" shape.
As an Burst modulated alternaing current (Aussie) device: Temporary relaxation of muscle spasms Prevention or retardation of disuse atrophy in post-injury type conditions Increase local blood circulation Muscle re-education Maintaining or increasing range of motion
As a Microcurrent device: Symptomatic relief of chronic intractable pain Symptomatic relief of post-traumatic acute pain and post-surgical acute pain
## Essential Performance
Neurodyn Muscle Stimulator produces the following currents: Russian / Aussie / Interferential / Tens / Premodulated / Microcurrent
Neurodyn Aussie Muscle Stimulator produces the burst modulated alternating current (Aussie) medium frequency alternate current. The current intensity required for treatment depends on the patient's sensitivity. The treatment should be started with minimum levels of intensity .with gradual increase until the patient achieves the full effect of the treatment.
## Electrodes
Electrodes used are Axelgaard K970426. The minimal size of the electrode that can used with the Neurodyn and Neurodyn Aussie is 25cm².
# Patient Cables
Utilizes shrouded connectors to meet Lead Wire Connectors Safety Requirements IEC-60601-1 Sub clause 66-3 -c -according to 21 CFR 898. They are designed to be 4.5 feet (1.5 meters) away from the patient. The connector cables are designed to comply with subclause 56.3 ( c ) of the following standard: International Electrotechnical Commission (IEC) 60601-1: Medical Electrical Equipment 60601-1 (1988) Part 1: General requirements for safety Amendment No. 1 (1991) Amendment No. 2 (1995)
# Declaration of Conformity
This device has been assessed and found compliant. It upholds the highest safety and effectiveness standards.
{3}------------------------------------------------
1512/369
Image /page/3/Picture/1 description: The image shows a logo for IBRAMED. The logo is black and white and features the word "IBRAMED" in a bold, sans-serif font. Above the word is a stylized graphic that resembles a capital "I" with horizontal lines extending from the top and bottom. The logo is simple and modern.
# Summary of Safety and Effectiveness
- The Neurodyn and Neurodyn Aussie Muscle Stimulators are substantially equivalent to the Vetra Genisys K031077 manufactured by Chattanooga. All three devices claim similar Indications for Use and Device Characteristics in technological design and materials.
- The Neurodyn and Neurodyn Aussie Muscle Stimulators do not raise any new issues of Safety and Effectiveness based on their similarities.
- The devices have continually proven to be safe and effective and demonstrate intended product performance.
# Non-Clinical Tests Submitted:
The Ibramed Powered Muscle Stimulators have been tested in accordance with the applicable standards for medical device electrical safety, electromagnetic compatibility and the particular requirements for stimulator nerve and muscle safety.
| EMC<br>l | l est |
|----------|-------|
| | |
| Device Name | Neurodyn Aussie | Neurodyn | Vectra Genisys |
|-------------------------------------------------------|----------------------------------|----------------------------------|----------------------------------|
| K number | 121369 | 121369 | 031077 |
| Manufacturer | Ibramed | Ibramed | Ibramed |
| Intended use | Identical | Identical | Identical |
| Indications for use | Identical | Identical | Identical |
| Target population | Identical | Identical | Identical |
| Human factors | Identical | Identical | Identical |
| Contraindication | Identical | Identical | Identical |
| Device Name | Neurodyn Aussie | Neurodyn | Vectra Genisys |
| K number | --- | --- | K031077 |
| Manufacturer | Ibramed | Ibramed | Chattanooga |
| Technological<br>characteristics | Identical | Identical | Identical |
| Medium-<br>frequency<br>alternating current<br>(MFAC) | | | |
| Device Material | ABS plastic panel<br>LCD display | ABS plastic panel<br>LCD display | ABS plastic panel<br>LCD display |
| Width (in) | 10.6 | 14.6 | 9.75 |
| Height (in) | 4.9 | 4.9 | 8.75 |
| Depth (in) | 10.4 | 12.4 | 12.75 |
| Weight (lbs) | 4.08 | 5.5 | 6 |
| Performance | Identical | Identical | Identical |
| Biocompatibility | FDA cleared<br>electrodes | FDA cleared<br>electrodes | FDA cleared<br>electrodes |
| Mechanical safety | Identical | Identical | Identical |
| Anatomical Sites | Identical | Identical | Identical |
| Number of<br>channels | 4 | 4 | 2/4 |
| Russian | No | Yes | Yes |
| Burst Modulated<br>Alternating<br>Current (Aussie) | Yes | Yes | Yes |
| Interferential | No | Yes | Yes |
| Microcurrent | No | Yes | Yes |
| TENS | No | Yes | Yes |
| Premodulated | No | Yes | Yes |
| Voltage Input | 100/240V<br>50/60Hz<br>Bivolt | 100/240V<br>50/60Hz<br>Bivolt | 100/240V<br>50/60Hz<br>1.0A |
| Output | +24V<br>7.3A+24V<br>7.3A | +24V<br>7.3A+24V<br>7.3A | +24V<br>7.3A+24V<br>7.3A |
| Electrical Class | II | II | II |
| Electrical Type | BF Type | BF Type | BF Type |
| Method of line<br>current isolation | Double Isolation | Double Isolation | Fuse-Two 5.6A<br>Time Lag |
| Patient leakage<br>control-normal<br>condition | 0.0508mA | 0.0508mA | 69μA |
| Patient leakage<br>control-single fault<br>condition | 0.0252mA | 0.0252mA | 31μA |
The above comparison chart shows that all four devices are identical in every aspect regarding intended use, indications for use, contraindications, target population and human factors
{4}------------------------------------------------
112/369
:
.
:
{5}------------------------------------------------
| | 1 | 1<br>And And States of the States of the States of the States of the States of the States of the States of the States of the States of the States of the States of the States of th<br>1<br>l | | r<br>0<br>ﺳﮯ<br>)<br>2<br>ﻟﻤﻌﺘﻤﺪ | C |
|--|---|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----------------------------------|---|
|--|---|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----------------------------------|---|
| Software<br>Microprocessor | Yes | Yes | Yes |
|---------------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------------------|
| Automatic<br>overload trip | Yes | Yes | Yes |
| Automatic shut off | Yes | Yes | Yes |
| Temperature<br>range during<br>transport and<br>storage | 41°F-122°F | 41°F-122°F | 59°F-104°F |
| Environment<br>operating<br>temperature<br>range | 41°F-113°F | 41°F-113°F | 59°F-104°F |
| Locking feature | Keyboard lock<br>safety feature | Keyboard lock<br>safety feature | Keyboard lock<br>safety feature |
| Treatment timer | Treatment timer<br>with auto shut off | Treatment timer<br>with auto shut off | Treatment timer<br>with auto shut off |
| Auto test and<br>repeat | Automatic setting<br>and repeat<br>treatment | Automatic setting<br>and repeat<br>treatment | Automatic setting<br>and repeat<br>treatment |
| Safety standards<br>requirements | ISO 13485<br>IEC 60601-1<br>IEC 60601-2<br>IEC 60602-10<br>CE | ISO 13485<br>IEC 60601-1<br>IEC 60601-2<br>IEC 60602-10<br>CE | ISO 13485<br>IEC 60601-1<br>IEC 60601-2<br>IEC 60602-10<br>UL 60602 |
| Electromagnetic<br>compatibility | IEC 60601-1-<br>2001/A:2004 | IEC 60601-1-<br>2001/A:2004 | IEC 60601-1-<br>2001/A:2004 |
The preceding comparison chart shows that all three devices are similar in every aspect. We assessed the indications for use, intended use, technological characteristics, voltages, inputs, outputs, sizes, weight, types of materials, patient cable construction, software construction, user interface, control parameters, level of concern (software), waveforms, treatment times and the results of each treatment.
We have concluded that the Neurodyn and the Neurodyn Aussie are as safe and effective as the Vectra Genisys. The Ibramed devices deliver the same currents for the same intended uses as the Vectra Genisys. Thus, we found them to be substantially equivalent.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular fashion around the symbol. The text is in all caps and is oriented to follow the curve of the circle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
### December 19, 2012
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Ibramed Equipamentos Medicos % Ms. Lilian Llull Ibramed U.S. Agent TechLink International 18851 NE 29th Avenue 720 Aventura. FL 33180
K121369 Re:
> Trade/Device Name: Neurodyn/Neurodyn Aussie Powered Muscle Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ, IPF, LIH, GZI Dated: November 29, 2012 Received: December 5, 2012
Dear Ms. Llull:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties: We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
## Page 2 - Ms. Lillian Llull
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### Indications for Use Statement
Neurodyn Series: Neurodyn/ Neurodyn Aussie
### 510(k) Number: K121369
#### As a TENS device:
- Symptomatic relief of chronic (long term) intractable pain l
- Symptomatic relief of post-traumatic acute pain and post-surgical acute pain -
### As an Interferential and Premodulated device:
- Symptomatic relief of chronic intractable pain, acute post traumatic pain or acute post traumatic surqical pain
#### As a Russian device:
- Temporary relaxation of muscle spasms .
- Prevention or retardation of disuse atrophy in post-injury type conditions -
- Increase local blood circulation -
- Muscle re-education
- Maintaining or increasing range of motion -
### As a Burst Modulated Alternating Current (Aussie) device:
- Temporary relaxation of muscle spasms ー
- Prevention or retardation of disuse atrophy in post-injury type conditions ।
- Increase local blood circulation .
- Muscle re-education l
- Maintaining or increasing range of motion
#### As a Microcurrent device:
- Symptomatic relief of chronic intractable pain
- Symptomatic relief of post-traumatic acute pain and post-surgical acute pain
Prescription Use ____________ (Part 21 CFR 801 Subpart D) AND/OR
#### Over-The-Counter Use __________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEED
Concurrence of CDRH, Office of Device Evaluation
(ODE) Page 1 of 1
Division of Neurological and I Medicine Devices 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.