The Biodesign ENT Repair Graft is intended to separate tissue or structures compromised by surgical trauma, help control minimal bleeding, and act as an adjunct to aid in the natural healing process. The device is indicated for use where an open wound dressing material is required in the nasal and/or sinus cavities following nasal and/or sinus surgery where separation of tissues or structures is desired. The device is supplied sterile and is intended for one-time use.
Device Story
Biodesign ENT Repair Graft is a bioabsorbable, extracellular collagen membrane matrix derived from Small Intestinal Submucosa (SIS). Device acts as a scaffold for host cell infiltration during natural repair. Physician shapes graft to appropriate size for nasal/sinus cavity application. Used post-surgery to separate compromised tissues/structures, control minimal bleeding, and support healing. Supplied sterile in lyophilized state within double pouch system. Single-use only.
Clinical Evidence
Clinical study using SIS material (Surgisis) for nasal septal perforation repair provided evidence of substantial equivalence. Preclinical mouse subcutaneous study confirmed tissue separation, cell population, and degradation during ingrowth. Biocompatibility testing (genotoxicity, cytotoxicity, sensitization, etc.) per ISO 10993-1 and mechanical testing (tensile/suture strength) confirmed device performance.
Technological Characteristics
Bioabsorbable extracellular collagen membrane matrix (Small Intestinal Submucosa, SIS) containing Types I, III, IV, and VI collagen. Dimensions: 1 cm x 2 cm to 20 cm x 40 cm; thickness: 100 um to 500 um. Supplied sterile, lyophilized. Scaffold-based technology for host cell infiltration.
Indications for Use
Indicated for patients undergoing nasal and/or sinus surgery requiring an open wound dressing material to separate tissues or structures and aid in natural healing.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
Predicate Devices
MeroGel™ Control Gel ENT Surgical Dressing (K002972)
Submission Summary (Full Text)
{0}------------------------------------------------
121360
# 510(k) Summary
#### 21 February 2013
FEB 2 7 2013
## Cook Biotech Incorporated
# Biodesign® ENT Repair Graft
Manufacturer Name:
Cook Biotech Incorporated 1425 Innovation Place West Lafayette, Indiana 47906 Telephone: +1 (765) 497-3355 FAX: +1 (765) 807-7709
Official Contact:
Perry W. Guinn
#### DEVICE NAME AND CLASSIFICATION
Trade/Proprietary Name: Common Name: Classification Regulations:
Biodesign ENT Repair Graft Ear, nose and throat synthetic polymer material Class II, 21 CFR §874.3620 (KHJ)
## INTENDED USE:
The Biodesign ENT Repair Graft is intended to separate tissue or structures compromised by surgical trauma, help control minimal bleeding, and act as an adjunct to aid in the natural healing process. The device is indicated for use where an open wound dressing material is required in the nasal and/or sinus cavities following nasal and/or sinus surgery where separation of tissues or structures is desired. The device is supplied sterile and is intended for one-time use.
#### DEVICE DESCRIPTION:
The Biodesign ENT Repair Graft is composed of a bioabsorbable, extracellular collagen membrane matrix (Small Intestinal Submucosa, SIS). The Biodesign ENT Repair Graft is similar to its predicate MeroGel™ Control Gel ENT Surgical Dressing (K002972) which is a biomaterial composed of HYAFF®, an ester of hyaluronic acid, a natural occurring constituent of extracellular matrix. The device is available in multilayered sheets with sizes from 1 cm by 2 cm to 20 cm x 40 cm. The Biodesign ENT Repair Graft is a scaffold which becomes infiltrated by the host cells during the body's natural repair process. The Biodesign ENT Repair Graft can be shaped by the physician to the appropriate size for the desired indication.
The Biodesign ENT Repair Graft is similar to its MeroGel predicate in its technology in that it has the ability to be incorporated into the body. The device is packaged in a lyophilized (dried) state and supplied sterile in a sealed double pouch system.
{1}------------------------------------------------
# EQUIVALENCE TO MARKETED DEVICES
The Biodesign ENT Repair Graft is similar with respect to intended use, materials and technological characteristics to the predicate device in terms of section 510(k) substantial equivalence, as shown biocompatibility testing (conducted in accordance to ISO 10993-1 standards), mechanical, pre-clinical and clinical testing.
#### Biocompatibility testing
The following biocompatibility tests were performed on sterilized SIS devices (which have already been cleared in multiple applications), which are identical in composition to the Biodesign ENT Repair Graft (according to the ISO 10993-1 standard):
- . Genotoxicity
- Direct contact in vitro hemolysis .
- Cytotoxicity ●
- Muscle implantation .
- Acute intracutaneous reactivity .
- Skin irritation ●
- ISO Sensitization ●
- . Acute systemic toxicity
- Pyrogenicity
- LAL endotoxins ●
- Subchronic systemic toxicity .
The results of these tests provided evidence that the Biodesign ENT Repair Graft meets biocompatibility requirements of the ISO standard.
#### Mechanical Testing
The Biodesign ENT Repair Graft material was tested for the following:
- . Suture retention strength
- Ultimate tensile strength .
The results of the mechanical testing provided evidence that the Biodesign ENT Repair Graft provided adequate mechanical strength in its application.
## Preclinical Testing
A mouse subcutaneous study confirmed that the Biodesign ENT Repair Graft was able to help maintain tissue separation, is quickly populated with cells, and degrades during cellular ingrowth. This animal study provides evidence that the Biodesign ENT Repair Graft is biocompatible and safe in its application.
{2}------------------------------------------------
# Clinical Testing
A clinical study was performed using the SIS material (Surgisis) for nasal septal perforation repair. The clinical study provides evidence that the Biodesign ENT Repair Graft is substantially equivalent to its predicate in nasal and/or sinus procedures.
# Substantial Equivalence
See Table 1 for a comparison of the subject device and its predicate.
| Device | Biodesign ENT Repair | MeroGel™ Control Gel |
|---------------|-----------------------------|---------------------------------------------------------|
| | Graft | ENT Surgical Dressing |
| Manufacturer | Cook Biotech | Medtronic Xomed |
| | Incorporated | |
| 510(k) Number | K121360 | K002972 |
| Intended Use | The Biodesign® ENT | MeroGel Control Gel |
| | Repair Graft is intended to | ENT Surgical Dressing is |
| | separate tissue or | a dressing and/or stent |
| | structures compromised | intended to separate tissue |
| | by surgical trauma, help | or structures compromised |
| | control minimal bleeding, | by surgical trauma, help |
| | and act as an adjunct to | control minimal bleeding, |
| | aid in the natural healing | and act as an adjunct to |
| | process. The device is | aid in the natural healing |
| | indicated for use where an | process. The device is |
| | open wound dressing | indicated for use in ear, |
| | material is required in the | nose, and throat, head and |
| | nasal and/or sinus cavities | neck surgical procedures |
| | following nasal and/or | where an open wound |
| | sinus surgery where | dressing material is |
| | separation of tissues or | required including the |
| | structures is desired. | middle ear and external |
| | The device is supplied | ear canal following |
| | sterile and is intended for | myringoplasty, |
| | one-time use. | tympanoplasty, |
| | | canalplasty, stapes and |
| | | mastoid surgery, also for |
| | | use in the nasal and/or |
| | | sinus cavities following<br>nasal, sinus, and/or throat |
| | | surgery where separation |
| | | of tissues or structures is |
| | | desired. |
| Material | Small intestinal | HY AFF® (ester of |
| | submucosa | hyaluronic acid, a |
| | Primarily Types I, III, IV | naturally occurring |
| | and VI collagen | constituent of the |
| | (constituents of the | extracellular matrix) |
| | extracellular matrix) | |
| Dimensions | 1 cm x 2 cm to | N/A |
| | 20 x 40 cm | |
| Thickness | 100 um to 500 um | N/A |
## Table 1 - Substantial Equivalence Comparison
{3}------------------------------------------------
CONCLUSION: The biocompatibility, mechanical, pre-clinical and clinical tests performed on the Biodesign ENT Repair Graft show that the device is substantially equivalent to its predicate.
:
and the country of the county of
.
、
. .
.
:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
February 27, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Cook Biotech Incorporated % Mr. Perry W. Guinn VP of OA and Regulatory Affairs 1425 Innovation Place West Lafayette, IN 47906-1000
Re: K121360
Trade/Device Name: Biodesign® ENT Repair Graft Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, nose, and throat synthetic polymer material Regulatory Class: Class II Product Code: KHJ Dated: February 4. 2013 Received: February 7, 2013
Dear Mr. Guinn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. Perry W. Guinn
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia YAlexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Biodesign® ENT Repair Graft Device Name:
Indications For Use:
The Biodesign® ENT Repair Graft is intended to separate tissue or structures compromised by surgical trauma, help control minimal bleeding, and act as an adjunct to aid in the natural healing process. The device is indicated for use where an open wound dressing material is required in the nasal and/or sinus cavities following nasal and/or sinus surgery where separation of tissues or structures is desired.
The device is supplied sterile and is intended for one-time use.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of _
Vasant G.
Malshet
Digitally signed by Vasant G.
Malshet
DN: c=US/o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
cn=Vasant G. Malshet,
0.9.2342:19200300.100.1.1=130
0087851
Date: 2013.02.26 13:13:30 -05'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.