WRIST-TYPE FULLY AUTOMATIC DIGITAL BLOOD PRESSURE MONITOR
Applicant
Sejoy Electronics & Instruments Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Jul 18, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Wrist-type Fully Automatic Blood Pressure Monitors BP series are intended for used by adults with 12 years and older to measure the systolic blood pressure and pulse rate.
Device Story
Wrist-type fully automatic digital blood pressure monitor; uses inflatable cuff wrapped around patient's wrist. Internal pump automatically inflates cuff; oscillometric method and silicon integrated pressure sensor technology determine systolic/diastolic blood pressure and pulse rate. Deflation rate controlled by preset mechanical valve. Device features maximum pressure safety setting at 300 mmHg. Measurements displayed on LCD for three minutes; device turns off automatically after three minutes of inactivity. Select models (BP-202H, BP-2208) compare results against WHO blood pressure classifications; select models (BP-202H, BP-2206, BP-2208) detect irregular heartbeats. BP-2208 provides average results for all, AM, and PM measurements; BP-202H and BP-2206 calculate average of last three measurements. Used by adults in home or clinical settings. Output allows users to monitor blood pressure trends and identify potential hypertension or irregular heartbeats, facilitating clinical decision-making by healthcare providers.
Clinical Evidence
Clinical tests performed in compliance with ANSI/AAMI SP10:2002/(R) 2008 & A1:2003/(R) 2008 & A2:2006/(R) 2008 standards for manual, electronic, or automated sphygmomanometers. Results demonstrated compliance with accuracy requirements.
Technological Characteristics
Wrist-type oscillometric blood pressure monitor; silicon integrated pressure sensor; internal pump; mechanical deflation valve. Powered by 2 AAA batteries. LCD display. Models vary in memory capacity, averaging functions, and backlight. Complies with ANSI/AAMI SP10:2002, IEC/EN 60601-1, and IEC/EN 60601-1-2. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for adults 12 years and older to measure systolic/diastolic blood pressure and pulse rate.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
K121355
page 1/5
**SEJOY**
丰田子有限公司 Sejoy Electronics & Instruments Co., Ltd.
# 510(k) Summary
JUL 18 2012
The assigned 510(k) number is:
#### 1. Date Prepared:
2011.11.25
#### 2. Submitter's Identification:
Name: Sejoy Electronics & Instruments Co., Ltd.
Add .: Building 2, No.202, Zhenzhong Road, West Lake Economy & Technology Zone, Hangzhou, China 310030 Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: info@sejoy.com
#### 3. Name of the Device:
Trade Name: Wrist-type Fully Automatic Digital Blood Pressure Monitor Including the following models: - BP-201M, BP-202H, BP-202N, BP-2206, BP-2208 Common Name: Blood Pressure Monitor Classification name: Non-invasive blood pressure measurement System 21 CFR 870-1130, Class II, 74-DXN.
#### Classification Information: 4.
Regulation Number: 870.1130 Product Code: DXN Device Class: II Panel: 74 Cardiovascular
#### న్. Predicate Device Information:
The Wrist-type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following device:
7
{1}------------------------------------------------
page 2/5
Image /page/1/Picture/2 description: The image shows the word "SEJOY" in a bold, sans-serif font. The letters are black against a white background. There is a small, triangle-like mark above the letter 'J'.
世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.
AM/PM Memory Wrist Blood Pressure Monitor (Model: UB-512), FDA 510(k) K042967, distributed by A division of A&D Engineering, Inc.
#### Device Description: 6.
The wrist-type fully automatic digital blood pressure monitors uses an inflatable cuff which is wrapped around the patient's wrist. The cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg. For BP-202H, BP-2208, the blood pressure results are compared with WHO (World Health Organization) Blood Pressure classification, which are severe Hypertension, Moderate Hypertension, Mild Hypertension, High-normal, Normal, and Optimal. The corresponding LCD segment will be turned on along with the systolic, diastolic, and pulse rate information. BP-202H, BP-2206, BP-2208 will display an irregular heartbeat symbol " if an irregular heartbeat was detected during the measurement process. BP-2208 can display average results in three ways: the average of all measurements, the average of all AM (5:00 AM-8:59AM) measurements, and the average of all PM (18:00 PM-19:59 PM) measurements. BP-202H, BP-2206 can calculate the average of the last three measurements. In addition, BP-2208 have LCD backlight. After three minutes without operation, the blood pressure monitors turn off automatically.
| Features<br>Models | A | B | C | D | E | F | G | H | I | J | K |
|--------------------|---|--------------------------------|---|---|---|---|---|--------------|--------------|-------|---|
| BP-201M | ● | 120 Memories | ● | ○ | ○ | ○ | ○ | 76×67.5×28.5 | 13.5-21.5 cm | 45×30 | ○ |
| BP-202N | ● | 120 Memories in<br>four groups | ● | ○ | ○ | ○ | ○ | 76×67.5×28.5 | 13.5-21.5 cm | 45×30 | ○ |
| BP-202H | ● | 120 Memories in<br>four groups | ● | ○ | ● | ● | ● | 76×67.5×28.5 | 13.5-21.5 cm | 45×30 | ○ |
| BP-2206 | ● | 120 Memories in<br>two groups | ● | ○ | ● | ● | ● | 77×64×32.5 | 13.5-21.5 cm | 45×30 | ○ |
| BP-2208 | ● | 120 Memories in<br>two groups | ● | ● | ● | ○ | ● | 77×64×32 | 13.5-21.5 cm | 49×38 | ● |
The detail comparisons among devices are listed below:
A = Powered by 2 AAA batteries
B = Memory Size
C = Time & Date
{2}------------------------------------------------
K121355
**SEJOY**
世佳电子有限公司
# Sejoy Electronics & Instruments Co., Ltd.
- D = Results Average in Three way
E = WHO (World Health Organization) Classification Indicator
- F = Last 3 Results Average
- G = Irregular Heartbeat Detection
- H = Outside Demission (L x W x H in mm)
- I = Cuff Size
- J = LCD Size (Viewing Area in mm)
K = LCD Backlight
• = Yes
o = No
The devices are all designed and manufactured according to ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008. manual, electronic or automated sphygmanometers.
### 7. Intended Use:
The Wrist-type Fully Automatic Blood Pressure Monitors BP series are intended for used by adults with 12 years and older to measure the systolic blood pressure and pulse rate.
The intended user and the indication for use of the Wrist-type Fully Automatic Blood Pressure Monitors BP series as described in the labeling are nearly the same as their predicated devices, AM/PM Memory Wrist Blood Pressure Monitor (Model: UB-512).
#### 8. Summary comparing technological characteristics with predicate device:
| Technological Characteristics | Comparison result |
|-------------------------------|-------------------|
| Design principle | Identical |
| Appearance | Similar |
| Patients contact Materials | Identical |
| Performance | Similar |
| Biocompatibility | Identical |
| Mechanical Safety | Identical |
| Energy Source | Similar |
| Standards Met | Identical |
| Electrical Safety | Identical |
| EMC | Identical |
| Function | Similar |
The difference of technological characteristic between the predicate device and the submit
{3}------------------------------------------------
page 4/5
Image /page/3/Picture/2 description: The image shows the word "SEJOY" in bold, black letters. There is a small symbol above the "J" that resembles a triangle or arrow pointing upwards. The letters are closely spaced together, and the overall appearance is clean and simple.
世佳电子有限公 Sejoy Electronics & Instruments Co., Ltd.
wrist-type fully automatic blood pressure monitors is the appearance, the performance and the function.
## 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards including ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 as well as IEC/EN 60601-1: 2005+CORR.1 (2006) +CORR.2 (2007) / EN 60601-1:2006 /AC: 2010, IEC/EN 60601-1-2:2007, ISO 10993-5: 2009 Biological evaluation of medical devices -Part 5:Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
Guidance documents include the "FDA Non-invasive Blood Pressure (NIBP) Monitor Guidance" and "FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993."
### Non-clinical Tests:
- . Electromagnetic Compatibility Test according to IEC/EN 60601-1-2:2007
- General Safety Provisions Test according to IEC/EN 60601-1: 2005+CORR.1 (2006) +CORR.2 (2007) / EN 60601-1:2006 /AC: 2010
- . Performance Test according to ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008. The test result all meet or exceed the requirement of the standards.
- Biocompatibility Test according to FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993, ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
### 10. Discussion of Clinical Tests Performed:
Clinical tests were performed and complied the accuracy requirements of ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 "National Standard for Manual, Electronic or Automated Sphygmomanometers".
#### 11. Conclusions:
Our Wrist-type Fully Automatic Blood Pressure Monitor BP-201M, BP-202H, BP-202N, BP-2206, BP-2208 have the same intended use and similar technological characteristics as the
1 ()
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SEJOY" in bold, black letters. The letters are large and spaced closely together. The background is white, providing a strong contrast to the text.
K121355 电子有限公司 Sejoy Electronics & Instruments Co., Ltd.
A & D AM/PM Memory Wrist Blood Pressure Monitor (Model: UB-512) whose 510 (k) number is K042967.
Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared devices.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the emblem.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 18 2012
Sejoy Electronics & Instruments Co., Ltd. c/o Mr. Edmondo Roark 1259 Monten Street #4 Cincinnati, OH 45208
Re: K121355
Trade/Device Name: Wrist-type Fully Automatic Digital Blood Pressure Monitor Wrist-type Funy Automatio Digital (Five Models: BP-201M, BP-202H, BP-202N, BP-2206, BP-2208) Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: II (two) Product Code: DXN Dated: April 6, 2012 Received: May 4, 2012
Dear Mr. Roark:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 310(x) premaited is substantially equivalent (for the indications
referenced above and have determined the device is substanted in interstate referenced above and have ustchillied the arreed predicate devices marketed in interstate for use stated in the enclosure) to regally manced the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 28, 1970, the chacanene with the provisions of the Federal Food. Drug, devices that have been reciassified in accordance that in proval application (PMA).
and Cosmetic Act (Act) that do not require approval or a provisions of the Act . The and Cosmetic Act (Act) that do not require approvate of the general controls provisions of the Act. The Act. The You may, therefore, market the device, subject to incements for annual registration, listing of general controls provisions of the Act mender reading and prohibitions against misbranding and devices, good manufacturing placity, label information related to contract liability.
adulteration. Please note: CDRH does not evaluate in trutiful and not mislead adulteration. Please note: CDRF does not evaluate information of the many of misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it
and he If your device is classified (See above) into enner elass if (Spoth
may be subject to additional controls. Existing major regulations affecting your device can be
{6}------------------------------------------------
Page 2 - Mr. Roark
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K121355
SFÍO
世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.
# Indications for Use
510(k) Number (if known): K 12 / 355
Device Name:
Blood Pressure Monitors BP series
Indications For Use:
Measure blood pressure (systolic and diastolic) and pulse rate.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use V (21 CFR 801 Subpart C)
5
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Ivision Sign-Off) Division of Cardiovascular Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.