← Product Code [KDI](/productcode/KDI) · K121341

# FRESENIUS 2008T HEMODIALYSIS MACHINE WITH BIBAG SYSTEM (K121341)

_Fresenius Medical Care North America · KDI · Dec 6, 2012 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K121341

## Device Facts

- **Applicant:** Fresenius Medical Care North America
- **Product Code:** [KDI](/productcode/KDI.md)
- **Decision Date:** Dec 6, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Fresenius bibag system is used with Fresenius three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.

## Device Story

The 2008T Hemodialysis Machine with bibag System is a high-permeability hemodialysis delivery system. It processes dialysis-grade water and dry sodium bicarbonate powder (from a single-use bibag disposable) to generate saturated bicarbonate concentrate online. The system manages fluid and solute removal for patients with renal failure. Operators (clinicians) interface with the machine to set treatment parameters, monitor conductivity/temperature, and manage alarms. Modifications include active pressure regulation for concentrate inlet, active venting to eliminate dialysate flow bypass, and an in-line particulate filter. User interface updates include a 'Mute Once' alarm feature, a one-touch acid/heat disinfection button, and status indicators for heparin/SVS. The device provides automated mixing and delivery of dialysate, supporting hemodialysis, hemofiltration, and hemoconcentration. Clinical benefit is derived from the precise, automated preparation of bicarbonate concentrate and controlled extracorporeal blood treatment.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, software verification/validation, system performance testing, heat disinfection/chemical testing, electromagnetic compatibility, reliability testing (HALT, mechanical life, elevated temperature), biocompatibility, and a summative usability study. All tests met pre-determined acceptance criteria.

## Technological Characteristics

High permeability hemodialysis system; three-stream proportioning. Features: active pressure regulation (up to 10 psi), active venting, in-line particulate filtration. Connectivity: standalone. Software: UI modifications for convenience and status monitoring. Sterilization: heat/chemical disinfection. Complies with ISO 14971 risk management and 21 CFR 820 quality system regulations.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- Fresenius 2008T Hemodialysis Machine with bibag System ([K120017](/device/K120017.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

**FRESENIUS**
**MEDICAL CARE**

Submitter's Information

2008T Hemodialysis Machine with bibag™ System Special 510(k) Notification

DEC

6 2012

# 510(k) Summary

. A.

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.

| Name:                | Fresenius Medical Care North America                                                       |
|----------------------|--------------------------------------------------------------------------------------------|
| Address:             | 920 Winter Street<br>Waltham, MA 02451-1457                                                |
| Phone:               | (781) 699-4479                                                                             |
| Fax:                 | (781) 699-9635                                                                             |
| Contact Person:      | Denise Oppermann, Senior Director<br>Regulatory Affairs - Devices<br>Renal Therapies Group |
| Date of Preparation: | 2 May 2012                                                                                 |

B. Device Name

| Trade Name:                        | Fresenius 2008T Hemodialysis Machine with<br>bibag System |
|------------------------------------|-----------------------------------------------------------|
| Common Name:                       | Hemodialysis Delivery System                              |
| Classification Name:               | High Permeability Hemodialysis System                     |
| Classification Number:             | Class II per § 876.5860                                   |
| Product Code/Classification Panel: | 78KDI/Gastroenterology/Urology Panel                      |

C. Legally Marketed Predicate Device (unmodified device) Fresenius 2008T Hemodialysis Machine with bibag System (K120017).

{1}------------------------------------------------

K121341
page 2 of 5

Image /page/1/Picture/1 description: The image shows the logo for Fresenius Medical Care. On the left is a stylized symbol consisting of three downward-pointing chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in bold, sans-serif font. The word "FRESENIUS" is on the top line, and "MEDICAL CARE" is on the bottom line.

#### D. Device Description

The Fresenius 2008T Hemodialysis Machine with bibag System is indicated for use in bicarbonate hemodialysis for acute and chronic renal failure.

The 2008T Hemodialysis Machine with bibag System allows operators the option to prepare a saturated sodium bicarbonate solution online through automated mixing of dialysis grade water and dry sodium bicarbonate powder within the bibag source disposable. The bibag System comprises: (1) the sodium bicarbonate concentrate generator (known as the bibag module); (2) the bag of dry sodium bicarbonate concentrate. A specialized bibag connector with a door is used to connect the single-use bibag disposable (650g/900g) filled with USP grade dry sodium bicarbonate powder to the bibag connector. The 2008T Hemodialysis Machine draws dialysis grade water into the bibag to produce a saturated solution of sodium bicarbonate online. This online generation of sodium bicarbonate can only be performed using a specially modified Fresenius 2008T Hemodialysis Machine with bibag System and can only be used with 45x (1:44) dilution. The bibag cannot be used with non-Fresenius hemodialysis machines capable of using cartridge type dry sodium bicarbonate because of the unique connection between the bibag disposable, the bibag connector, and the hemodialysis machine.

Modifications to the previously cleared 2008T Hemodialysis Machine with bibag System include:

- Active Pressure Regulation Feature: Increases the maximum acid/bicarbonate . concentrate inlet pressure specification from 2 psi to 10 psi. Provides an integrated process for regulating concentrate supply (inlet) pressures from central delivery systems.
- Active Venting Feature: Replaces the existing venting process requiring dialysate . flow bypass.
- . In-line Particulate Filter: Addition of an inline particulate filter between the bibag concentrate outlet and bicarbonate pump to eliminate the possibility of particulates from the disposable bag from entering the hydraulics.
- . Mute Once Feature: Users may elect to mute all conductivity and temperature audible alarms for up to 6 minutes (maximum) to allow a newly installed bibag disposable to reach operating temperature and conductivity.
- Acid Clean/Heat disinfection button: Allows users to initiate the acid/heat disinfect process with a single screen selection.

{2}------------------------------------------------

K121341
page 3 of 5

Image /page/2/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a stylized graphic on the left and the text "FRESENIUS MEDICAL CARE" on the right. The graphic is composed of three stacked, downward-pointing chevron shapes. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

This submission also includes a description of software modifications to implement user interface changes. Modifications include: Heparin/SVS status. Dialysate On/Off button. Configurator, SVS Option and are intended to address user preferences and to provide additional user convenience. These changes were also described in K120505 (submitted 17 February 2012; cleared 6 March 2012).

Additionally, this submission includes minor software maintenance changes made to the 2008T Hemodialysis Machine with bibag System since the last clearance (K120017).

Treatment modalities for the modified Fresenius 2008T Hemodialysis Machine with bibag System remain identical to those for the unmodified device (K120017):

The 2008T Hemodialysis Machine with bibag System is a high permeability hemodialysis system used for the treatment of patients with acute or chronic kidney failure, fluid overload or toxemic conditions. Therapies include hemodialysis. hemofiltration and hemoconcentration. The 2008T will accommodate the use of both low flux and high flux dialyzers.

#### Indications for Use E.

The modified Fresenius 2008T Hemodialysis Machine with bibag System has the same indications for use as the unmodified device. The Fresenius Medical Care bibag system is used with Fresenius Medical Care three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag System is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.

#### F. Technological Characteristics

There are no changes in the technological characteristics of the previously cleared Fresenius 2008T Hemodialysis Machine with bibag System (K120017). The Fresenius 2008T Hemodialysis Machine with bibag System incorporates changes (hardware, software, labeling) related to performance (active pressure regulation, active venting, particulate filter) and features to facilitate user convenience (Mute Once, Acid/Heat Disinfection button, Heparin/SVS status, Dialysate On/Off button, Configurator, SVS Option).

The modified and the unmodified devices have identical indications for use and are intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The modified device has the same operating principle, fundamental scientific technology, and is comparable in key safety and effectiveness and quality assurance features.

{3}------------------------------------------------

K121341
page 4 of 5

Image /page/3/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a stylized graphic on the left and the text "FRESENIUS MEDICAL CARE" on the right. The graphic is composed of three downward-pointing chevrons stacked vertically. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

All water requirements, module options, functional options, performance limits, control parameters, compatible bloodlines, and language options remain unchanged from the predicate device.

The following technical specifications of the modified device remain the same as the unmodified device:

- Safety system
- System performance
- Environmental Requirements ●
- User Interface (except for described modifications) .
- Hardware and therapy settings ●
- Accessories
- . Environmental Design
- Alarms (except for additional fault condition alarms) .
- Accuracy and Controls .
- Protection against Mechanical Hazard
- Protection against Electrical Hazard
- Protection against excessive temperature or other hazards ●
- Transportation and Storage conditions ●
- Manufacturing Location and manufacturing processes (assembly, fabrication, testing, shipping, installation and service).

A risk analysis (per ISO 14971) has been completed and potential hazards associated with the modifications are identified and mitigated. Mitigations are verified wherever applicable. All potential risks were deemed acceptable after mitigation. Performance and safety tests were conducted to ensure the safety and effectiveness of the device after the proposed modifications.

#### G. Performance Data

The performance of the modified device described in this submission was evaluated according to existing FMCNA procedures, protocols, declared performance standards and guidelines of the quality system regulation (21 CFR 820). Design verification and validation tests were conducted to ensure that the modifications did not affect the essential performance of the device and the device functions as intended.

{4}------------------------------------------------

K121341
page 5 of 5

Image /page/4/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a stylized graphic on the left and the text "FRESENIUS MEDICAL CARE" on the right. The graphic is composed of three stacked chevron-like shapes pointing downwards. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

The following tests were conducted for the modified device:

#### 6. System and Software Verification and Validation Testing

- Functional Verification and Software Validation
	- > Software Verification (Functional Tests)
	- > Regression
	- > Safety Systems Verification
	- > Simulated Dialysis Treatment
	- > Production Test Procedure
	- > Unstructured and Static Code Verification
- System Performance �
- . Heat Disinfection Testing
- . Chemical Testing

#### 7. System Safety

- . Equipment Safety
- . Electromagnetic Compatibility

#### 8. Reliability

- . High Accelerated Life Testing
- . Mechanical Life Testing
- Elevated Temperature Testing .
- 9. Biocompatibility
- 10. Summative Usability

#### H. Conclusion

Test results demonstrated that the modified 2008T Hemodialysis Machine with bibag System functions as intended and met pre-determined acceptance criteria. Results of system/software verification/validation testing, safety testing, reliability testing, biocompatibility tests, summative usability study and risk analysis indicate that the modified Fresenius 2008T Hemodialysis Machine with bibag System is substantially equivalent to the named predicate device and remains safe and effective for its intended use.

{5}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

December 6, 2012

Fresenius Medical Care, North America % Ms. Denise Oppermann Senior Director 920 Winter Street WALTHAM MA 01854

Re: K121341

Trade/Device Name: 2008T Hemodialysis Machine with bibag™ System Regulation Number: 21 CFR§ 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI, KPO Dated: November 21, 2012 Received: November 23, 2012

Dear Ms. Oppermann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{6}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Herbert R. Lerner

Acting Director for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, , and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a stylized symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three chevrons stacked on top of each other, with the top two chevrons pointing upwards and the bottom chevron pointing downwards.

## Indications for Use Statement

K121341 510(k) Number (if known):

## Device Name:

2008T Hemodialysis Machine with bibag™ System

## Indications for Use:

The Fresenius bibag system is used with Fresenius three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.

XPrescription Use (Per 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

Herbert P. Lerner

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K121341

---

**Source:** [https://fda.innolitics.com/device/K121341](https://fda.innolitics.com/device/K121341)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
