CHABNER XRT GARMENTS

K121284 · Chabner Xrt, LLC · IYE · Jun 14, 2013 · Radiology

Device Facts

Record IDK121284
Device NameCHABNER XRT GARMENTS
ApplicantChabner Xrt, LLC
Product CodeIYE · Radiology
Decision DateJun 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

Chabner XRT™ Garments are intended to be used as an aid in the support of the patient during X-ray, computed tomography, magnetic resonance imaging, radiotherapy, and other diagnostic radiological procedures. The Chabner XRT™ Garments are not intended for pediatric use.

Device Story

Chabner XRT™ Garments are non-sterile, wearable compression garments (brassiere, shorts, or girdle) designed to aid patient support and positioning during radiological procedures (X-ray, CT, MRI, radiotherapy). Constructed from radiotranslucent materials (<1 mm thick) without metal, the garments include clear thermoplastic polyurethane (TPU) windows allowing clinicians to visualize index tattoos, midlines, and bony landmarks projected by machine light fields. The compressive nature and adjustment features assist in reproducing body geometry and patient-to-machine alignment. Used in clinical settings by healthcare providers, the garments provide the secondary benefit of protecting patient dignity and comfort during procedures. Radiation passes through the device without causing attenuation or bolus effects.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Non-sterile wearable compression garments (brassiere, shorts, girdle). Materials: Thermoplastic polyurethane (TPU) windows; metal-free construction. Thickness: <1 mm. Radiotranslucent. Surface device, intact skin, limited contact duration (<24 hours). No software or electronic components.

Indications for Use

Indicated for adult patients requiring support and positioning during X-ray, CT, MRI, radiotherapy, and other diagnostic radiological procedures. Not indicated for pediatric use.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121284 Page 1 of 4 # 510(k) Summary "510(k) Summary" As required by section 807.92(c) For Chabner XRT™ Garments June 3, 2013 1. 'Company Name and Address a. Sponsor/Manufacturer Bffl Co. Inc. 20 Kensington Road Scarsdale, NY 10583 b. Consultant/Contact Evan P. Phelps OFW Law 600 New Hampshire Ave, Ste. 500 Washington D.C. 20037 ephelps@ofwlaw.com Tel: (202) 789-1212 Fax: (202) 234-3550 2. Establishment Registration Number: Not yet assigned - 3. Device Name: - a. Trade Name: Chabner XRT™ Garments Therapy System Patient Support and Positioning Garments Accessory to Medical Charged-Particle Radiation Common/Usual Name: b. Classification Name: C. ## 4. Device Classification: a. 21 C.F.R. § 892.5050 (Class 11) b. Product Code: IYE c. Classification Panel: Radiology JUN 1 4 2013 {1}------------------------------------------------ #### 5. Legally Marketed Predicate Devices: - a. MR Patient Positioning Devices - i. 510(k) Owner: Med-Tech, Inc. (d.b.a. CIVCO Medical Solutions) - ii. 510(k) Number: K111340 - iii. 21 C.F.R. § 892.5050 (Class II) - iv. Product Code: IYE - b. Moldcare Head & Neck Cushion - i. 510(k) Owner: Med-Tec., Inc. - ii. 510(k) Number: K982624 - iii. 21 C.F.R. § 892.5050 (Class II) - iy. Product Code: IYE # 6. Device Description The device is a non-sterile garment worn by patients as an aid in the support and positioning of patients during X-ray, computed tomography, magnetic resonance imaging, radiotherapy, and other diagnostic radiological procedures. There are three versions of the garment: 1) brassiere; 2) athletic style shorts; and 3} girdle. All versions of the device are devoid of metal and can be worn for use with all treatment or diagnostic machines without causing a bolus (i.e., the materials do not mimic the properties of tissue during a radiological procedure) or attenuation effect. All materials used in the device's construction are < 1 mm in thickness. Clear thermoplastic polyurethane (TPU) windows are incorporated into the device to enable clinicians to view index tattoos, the midline, bony landmarks, and other treatment parameters, as created by the light field that is projected by the treatment or diagnostic machine onto the patient's body. The compressive nature of the materials used in the device's construction, along with various adjustment features incorporated into the device, assist the clinician in positioning the body and to reproduce the desired body geometry, treatment position, and patient-to-machine alignment prior to each radiological procedure. Radiation passes through the TPU material without affecting therapy or diagnosis. Because the device is worn as a garment, they offer the added benefit of protecting the patient's dignity during such procedures that, in turn, increases the patient's overall comfort during the process. - 7. Intended Use Chabner XRT™ Garments are intended to be used as an aid in the support of the patient during X-ray, computed tomography, magnetic resonance imaging, radiotherapy, and other diagnostic radiological procedures. {2}------------------------------------------------ The Chabner XRT™ Garments are not intended for pediatric ,use. # 8. Technological Characteristics and Substantial Equivalence Evaluation The Chabner XRT™ Garments are substantially equivalent in terms of intended use, operating principles, and operational specifications to CIVCO's MR Patient Positioning Devices (K111340) and Med-Tec. Inc.'s Moldcare Head and Neck Cushion (K982624) which are both legally marketable patient support and positioning aids classified by the agency as accessories to medical charged-particle radiation therapy systems. Like the CIVCO MR Patient Positioning Devices and Med-Tec. Inc. Moldcare Head and Neck Cushion, the Chabner XRT™ Garments are intended to be used as an aid in the support and positioning of patients during MR, radiological, and other procedures, are radiotranslucent, conform to the individual patient's body shape, and contact the body for less than a 24 hour period of time. | Feature | Chabner XRT™<br>Garments | MR Patient<br>Positioning Devices | Moldcare Head &<br>Neck Cushion | |-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | K111340 | K982624 | | | | CIVCO | Med-Tec, Inc. | | Indications for Use | Chabner XRT™<br>Garments are<br>intended to be used<br>as an aid in the<br>support of the<br>patient during X-ray,<br>computed<br>tomography,<br>magnetic resonance<br>imaging,<br>radiotherapy, and<br>other diagnostic<br>radiological<br>procedures.<br>The Chabner XRT™<br>Garments are not<br>intended for<br>pediatric use. | CIVCO patient<br>Positioning devices<br>are used to aid in the<br>support and<br>positioning of patients<br>during MR,<br>radiological, and other<br>procedures. | The intended use of<br>this device is to<br>provide an additional<br>aid to the fast and<br>accurate repeat<br>positioning of the<br>patient for radiation or<br>other treatment. The<br>device forms to the<br>back of the patient's<br>head and eliminates<br>the usual range of<br>movement of ordinary<br>head cups. | | Radiotranslucent | Yes | Yes | Yes | | Conforms to patient's<br>body shape | Yes | Yes | Yes | {3}------------------------------------------------ K121284 Page 4 of 4 | Device Body Contact<br>Category (ISO 10993-1) | Surface devices,<br>intact skin; limited<br>contact duration<br>(<24 hours). | Surface devices, intact<br>skin; limited contact<br>duration (<24 hours). | Surface devices, intact<br>skin; limited contact<br>duration (<24 hours). | |-----------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------| |-----------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------| There are no significant differences between the Chabner XRT™ Garments and the predicate devices. However, they do differ from the predicate devices in that they are produced as wearable compression garments that have the added benefit of protecting patient dignity; whereas, the predicate devices are not worn on the body. This difference does not affect the device's safety and efficacy. - 9. Substantial Equivalence Conclusion The Chabner XRT™ Garments are substantially equivalent to the CIVCO MR Patient Positioning Devices and Med-Tec. Inc. Moldcare Head and Neck Cushion with respect to intended use, safety, and effectiveness. {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized representation of a caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it, topped with a pair of wings. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002 June 14, 2013 BFFL Co., Inc. % Mr. Evan P. Phelps Counsel OFW Law 600 New Hampshire Avenue, NW, Suite 500 WASHINGTON DC 20037 Re: K121284 Trade/Device Name: Chabner XRT Garments Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: May 10, 2013 Received: May 15, 2013 Dear Mr. Phelps: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls.provisions.of.the.Act.include.requirements.for.annual.registration,.listing.of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {5}------------------------------------------------ Page 2 - Mr. Phelps forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.lda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Sincerely yours, Michael D. O'Hara for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure- {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): K121284 Device Name: Chabner XRT Garments Indications for Use: Chabner XRT™ Garments are intended to be used as an aid in the support of the patient during X-ray, computed tomography, magnetic resonance imaging, radiotherapy, and other diagnostic radiological procedures. The Chabner XRTTM Garments are not intended for pediatric use. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Porti (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health 510(k) K121284 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...