K121280 · Hologic, Inc. · KCT · Feb 8, 2013 · General Hospital
Device Facts
Record ID
K121280
Device Name
MYOSURE INTRUMENT TRAY
Applicant
Hologic, Inc.
Product Code
KCT · General Hospital
Decision Date
Feb 8, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The MyoSure™ Instrument Tray is intended to enclose, protect, and organize the MyoSure Rod Lens Hysteroscope, MyoSure Removable Outflow Channel, and associated components during sterilization and storage. The MyoSure Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycle, and may be stored for up to 30 days in accordance with the wrap manufacturer's instructions.
Device Story
Plastic, perforated sterilization tray designed to house MyoSure Rod Lens Hysteroscope and associated components. Features internal posts and cradles for device positioning; evenly distributed hole pattern facilitates sterilant penetration and air removal. Used in clinical settings by healthcare staff to prepare instruments for sterilization via Prevacuum Steam, Gravity Steam, or STERRAD 100S cycles. Tray must be used with FDA-cleared sterilization wrap. Protects instruments during sterilization and storage (up to 30 days).
Clinical Evidence
Bench testing only. Validation included steam sterilization (Pre-Vacuum 132°C/4 min; Gravity 132°C/15 min) and STERRAD 100S (59% H2O2). Results demonstrated 6-log reduction of spore strips and inoculated devices. Thermocouple and chemical indicator data confirmed sterilant penetration. Biocompatibility testing (intracutaneous irritation and hemolysis) and post-processing materials compatibility testing were successful.
Technological Characteristics
Plastic, perforated base and lid with internal posts/cradles. Dimensions: 13.45" x 8.41" x 2.0". Compatible with Prevacuum Steam, Gravity Steam, and STERRAD 100S sterilization methods.
Indications for Use
Indicated for the organization, protection, and sterilization of MyoSure Rod Lens Hysteroscope, Removable Outflow Channel, and associated components.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
PolyVac Surgical Instrument Delivery System (K012105 and K040223)
Submission Summary (Full Text)
{0}------------------------------------------------
K1 2.12.80
FEB 8 2013
## 510(k) Summary
10.1 Submitter: Hologic, Inc 250 Campus Drive Marlborough, MA 01752 USA
> Contact: Sarah Fairfield GYN-Surgical Division 250 Campus Drive Marlborough, MA 01752 USA 508-263-8857
Date prepared: January 16, 2013
10.2 Device Information: Trade Name: MyoSure Instrument Tray Common Name: Sterilization Tray Classification II (21 CFR 880.6850) Product Code: KCT
- 10.3 Predicate Device:
Trade Name: PolyVac Surgical Instrument Delivery System Submitter/510(k) holder: Symmetry Medical, Inc. 510(k) Numbers: K012105 and K040223
#### 10.4 Device Description
The MyoSure Instrument Tray consists of an approximate 13.45 inch x 8.41 inch x 2.0 inch plastic, perforated base and lid with posts and cradles to position the devices to be sterilized. Each tray and lid has an evenly distributed hole pattern that allows sterilant penetration and air removal.
10.5 Indications for Use
> The MyoSure™ Instrument Tray is intended to enclose, protect, and organize the MyoSure Rod Lens Hysteroscope, MyoSure Removable Outflow Channel, and associated components during sterilization and storage. The MyoSure Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycle, and may be stored for up to 30 days in accordance with the wrap manufacturer's instructions.
> > 10-1
{1}------------------------------------------------
| Prevacuum Steam | Gravity Steam | STERRAD 100S |
|---------------------------------|---------------------------------|---------------------------------|
| 132°C (270°F) | 132 °C (270 °F) | 59% H2O2 |
| 4 minutes exposure | 15 minutes exposure | Normal Cycle Setting |
| 30 minutes dry time | 30 minutes dry time | 30 minutes dry time |
| Contents - 1 MyoSure Rod Lens | Contents - 1 MyoSure Rod Lens | Contents - 1 MyoSure Rod Lens |
| Hysteroscope, 1 Removable | Hysteroscope, 1 Removable | Hysteroscope, 1 Removable |
| Outflow Channel, 2 Seal Caps, 2 | Outflow Channel, 2 Seal Caps, 2 | Outflow Channel, 2 Seal Caps, 2 |
| Light Guides | Light Guides | Light Guides |
The MyoSure™ Instrument Tray has been validated for the following sterilization cycles:
#### 10.6 Comparison to Predicate
The design, principles of operation, primary functional specifications, and materials of composition of the MyoSure Instrument Tray are identical to those of the predicate PolyVac Surgical Instrument Delivery System. Both travs are made from the same materials, have the same dimensions, and are designed to be compatible with the same sterilization methods and parameters. Verification and Validation testing confirm that the MyoSure Instrument Tray meets the same performance specifications as the predicate PolyVac Surgical Instrument Delivery System.
Summary of Performance Testing: 10.7
### Steam sterilization qualification:
The MyoSure Instrument Tray was successfully validated in the following steam sterilization cycles. Test results demonstrated a 6 log reduction of all spores strips and inoculated devices. Thermocouple data shows excellent sterilant penetration within the wrapped packages.
Validated Steam Sterilization Cycles:
Pre-Vacuum: 4 minutes @ 132C
Gravity Air Displacement: 15 minutes @ 132C
### STERRAD Sterilization qualification
Successful completion of half cycle testing demonstrated that the MyoSure™ Instrument Tray containing a MyoSure Rod Lens Hysteroscope, Removable Outflow Channel, 2 Seal Caps, and 2 Light Guides components can be effectively sterilized using a the Sterrad® 100S System, full hospital cycle. The sterilization tests demonstrated a six log reduction capability of all spore strips and inoculated devices. Chemical indicator data indicated excellent sterilant penetration within the wrapped packages
Additional testing also included post-processing materials compatibility, intracutaneous irritation and hemolysis biocompatibility testing, all of which indicated successful results.
{2}------------------------------------------------
### 10.8 Conclusion
Based on the intended use, descriptive information, and performance evaluation provided in this submission, the MyoSure Instrument Tray has been shown to be substantially equivalent in materials, technology, method of operation, functional performance, and intended use to the predicate PolyVac Surgical Instrument Delivery System.
10-3
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 8, 2013
Mr. Daniel F. Phelan Regulatory Affairs Manager Hologic, Incorporated 250 Campus Drive MARLBOROUGH MA 01752
Re: K121280
Trade/Device Name: MyoSure Instrument Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: January 17, 2013 Received: January 18, 2013
Dear Mr. Phelan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
hor
Theth
Harshita.D.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{5}------------------------------------------------
# Indications for Use Statement
K121280 510(k) Number:
Device Name: MyoSure Instrument Tray
Indications for Use:
The MyoSure™ Instrument Tray is intended to enclose, protect, and organize the MyoSure Rod Lens Hysteroscope, MyoSure Removable Outflow Channel, and associated components during sterilization and storage. The MyoSure Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycle, and may be stored for up to 30 days in accordance with the wrap manufacturer's instructions.
The MyoSure™ Instrument Tray has been validated for the following sterilization cycles:
| Prevacuum Steam | Gravity Steam | STERRAD 100S |
|-------------------------------|-------------------------------|-------------------------------|
| 132°C (270°F) | 132 °C (270 °F) | 59% H2O2 |
| 4 minutes exposure | 15 minutes exposure | Normal Cycle Setting |
| 30 minutes dry time | 30 minutes dry time | |
| Contents - 1 MyoSure Rod Lens | Contents - 1 MyoSure Rod Lens | Contents - 1 MyoSure Rod Lens |
| Hysteroscope and accessories | Hysteroscope and accessories | Hysteroscope and accessories |
Prescription Use (per CFR 801.109) Over-the-counter use X
Concurrence of CDRH
OL
Elizabeth F. Claverie
2013.02.07 18:33:18 -05'00'
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: k121480
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.