Z-SNAP OFF SCREW, Z-DOUBLE THREAD COMPRESSION SCREW, Z-GUIDE WIRE, Z-STAPLE
K121277 · Z-Medical GmbH & Co. KG · HWC · Nov 7, 2012 · Orthopedic
Device Facts
Record ID
K121277
Device Name
Z-SNAP OFF SCREW, Z-DOUBLE THREAD COMPRESSION SCREW, Z-GUIDE WIRE, Z-STAPLE
Applicant
Z-Medical GmbH & Co. KG
Product Code
HWC · Orthopedic
Decision Date
Nov 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Z-Medical SnapOff screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include: - Fixation of small bone fragments - Weil osteotomy - Mono-cortical fixation - Osteotomies and fractures fixation in the foot and hand. The Z-Medical double thread compression screws are indicated for fixation of bone fractures or bone reconstruction. Examples included: - Fixation of bone fragments or small bone fractures - Fracture management in the foot and hand - Arthrodesis in hand, foot or ankle surgery - Mono or Bi-cortical osteotomies in the foot and hand. The Z-Medical's guide wires are intended to perform as fixation and stabilization unit of bone fractures or as guidance at insertion of implants into the skeletal system. The Z-Medical Staples are indicated for fixation of bone fractures, bone reconstruction, ligament, soft tissue and tendon. Examples included: - Fixation of bone fragments or small bones fractures - Fracture management in the foot and hand.
Device Story
Bone fixation system comprising Snap Off Screws, Double Thread Compression Screws, Guide Wires, and Staples. Used in orthopedic surgery (foot/hand/ankle) by physicians. Screws and staples provide mechanical fixation of bone fragments or reconstruction; guide wires assist in implant positioning. Snap Off technology allows distal end of staple to serve as drilling gauge/grip, then snapped off after placement. Manual insertion for screws; machine insertion for guide wires. Devices provide stable fixation to facilitate bone healing.
Clinical Evidence
Bench testing only. Testing included torque tests, break-off tests, sideways stability tests, maximum torque tests, screw-in tests, screw compression tests, and ASTM F564-10 testing for staples.
Technological Characteristics
Materials: ASTM F136 (Ti-6Al-4V ELI) for screws/staples; ASTM F138 (1.4441) for guide wires. Dimensions: Screws (10-36mm length, 2-4mm diameter); Guide wires (80-380mm length, 0.7-6.35mm diameter); Staples (8-25mm size). Features: Cannulated screws, SnapOff technology for staples, trocar end styles for wires. Manual or machine insertion.
Indications for Use
Indicated for patients requiring bone fracture fixation, bone reconstruction, or soft tissue/tendon fixation in the foot, hand, or ankle. Includes use for arthrodesis, osteotomies, and small bone fragment stabilization. No specific age or gender contraindications stated.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo with the letter Z. The word "MEDICAL" is written diagonally across the letter Z. The text "GMBH + CO.KG" is written below the letter Z. The logo appears to be for a medical company.
K121277
NOV 7 2012
# 510(k) Summary
### DATE OF APPLICATION: 2012-04-25
APPLICANT:
Z-Medical GmbH & Co. KG Gänsäcker 38 78532 Tuttlingen Germany Tel.: +49 7462 9455 40 Fax: +49 7462 9455 49
E-Mail: info@z-medical.de www.z-medical.de
CONTACT PERSON:
Alexander Henninger Quality Manager Tel.: +49 7462 9455 40 Fax: +49 7462 9455 49
SECTION_05_510K_Summary_vB.doc
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with the letter Z in a stylized, geometric design. The word "MEDICAL" is written diagonally across the center of the Z, and the text "GMBH & CO. KG" is located at the bottom of the logo. The logo appears to be for a medical company or organization.
#### Device Name 1.
.
#### Snap Off Screw 1.1.
| Trade Names: | Z-Snap Off Screw |
|----------------------|-----------------------|
| Common Name: | Snap Off Screw |
| Classification Name: | Screw, fixation, bone |
#### Double Thread Compression Screw 1.2.
| Trade Names: | Z-Double Thread Compression Screw |
|----------------------|-----------------------------------|
| Common Name: | Double Thread Compression Screw |
| Classification Name: | Screw, fixation, bone |
#### 1.3. Guide Wire
| Trade Names: | Z-Guide Wire |
|----------------------|-----------------------|
| Common Name: | Guide Wire, Kirschner |
| Classification Name: | Pin, fixation, smooth |
#### 1.4. Staple
| Trade Names: | Z-Staple |
|----------------------|------------------------|
| Common Name: | Staple |
| Classification Name: | Staple, fixation, bone |
#### 2. Classification Product Code / Subsequent Code
#### 2.1. Z-Snap Off Screw
| Device | Medical<br>Specialty | Review Panel | Product Code | Device Class | Regulation Number |
|-----------------------|----------------------|--------------|--------------|--------------|-------------------|
| Screw, fixation, bone | Part 888 | Orthopedic | HWC | 2 | 888.3040 |
#### 2.2. Z-Double Thread Compression Screw
| Device | Medical<br>Specialty | Review Panel | Product Code | Device Class | Regulation Number |
|-----------------------|----------------------|--------------|--------------|--------------|-------------------|
| Screw, fixation, bone | Part 888 | Orthopedic | HWC | 2 | 888.3040 |
#### 2.3. Z-Guide Wire
| Device | Medical<br>Specialty | Review Panel | Product Code | Device Class | Regulation Number |
|-----------------------|----------------------|--------------|--------------|--------------|-------------------|
| Pin, fixation, smooth | Part 888 | Orthopedic | HTY | 2 | 888.3040 |
#### Z-Staple 2.4.
| Device | Medical<br>Specialty | Review Panel | Product Code | Device Class | Regulation Number |
|-----------------------|----------------------|--------------|--------------|--------------|-------------------|
| Screw, fixation, bone | Part 888 | Orthopedic | JDR | 2 | 888.3030 |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with the letter Z in a stylized, geometric design. The word "MEDICAL" is written diagonally across the Z. Below the Z, there is text that reads "GMBH + CO. KG" in a smaller font size. The logo appears to be for a medical company.
# 3. Predicate Device
Z-Medical's bone fixation devices are substantially equivalent to the following predicate devices, most recently cleared by the FDA:
| Z-Medical Device | Predicate Device | 510(k) Number | 510(k) Holder |
|--------------------------------------|-----------------------------------------------------------------------------|--------------------|--------------------------------------------------|
| Z-Snap Off Screw | CHARLOTTE Snap-Off Screw | K050819 | Wright Medical Technology |
| Z-Double Thread<br>Compression Screw | Foot and Hand Motion System<br>Synthes 3.0 mm Headless Compression<br>Screw | K091118<br>K050636 | NewClip Technics<br>Synthes (USA) |
| Z-Guide wire | Orthopaedic Fixation Pins and Wires /<br>Kirschner / Guide Wires | K100736 | SMT Schilling Metalltechnik GmbH |
| Z-Staple | Compression Staple and Simple Staple<br>MLI Modular Staples | K043059<br>K962705 | Wright Medical Technology<br>Medicine Lodge Inc. |
#### 4. Description of the Device
#### Z-Snap Off Screw 4.1.
Main application of the Snap Off Screw is the forefoot surgery. The device is intended to connect bones or bone fragments. Especially in the classic "Well"-osteotomy you can fix the two parts of bones with this kind of screw.
Material: ASTM F136 (Ti-6AI-4V ELI) Length: 11 mm to 14 mm Diameter: 2 mm
#### 4.2. Z-Double Thread Compression Screw
The Double thread compression screw is mainly used in the forefoot surgery. Most common use is in the "Scart"-osteotomy. They are cannulated to work with a guide wire and they are drilled in manually.
Material: ASTM F136 (Ti-6Al-4V ELI) Length: 10 mm to 36 mm Diameter: 3 and 4 mm
#### 4.3. Z-Guide Wire
Guide wires are used in all fields of surgery. Main applications are: determine the correct position for e.g. screws or drills, or fixation of bones, fragments. They are usually put in by machine. Guide wires are also used to drill holes, e.g. to apply staples.
| Material: | ASTM F 138 (1.4441) |
|-------------|---------------------|
| Length: | 80 mm to 380 mm |
| Diameter: | 0.7 mm to 6.35 mm |
| End Styles: | Trocar |
| Type: | Smooth |
#### Z-Staple 4.4.
Staples are implants for fixing bones or fragments. Z-Staples use the SnapOff-technology by Z-Medical. The distal end of the staple can be uses as drilling gauge as well as grip to put in the staple. After reaching the final position, the grip can be snapped off by moving sideways.
{3}------------------------------------------------
## Z-Medical GmbH & Co. KG
SnapOff Screw, Double Thread Compression Screw, Guide Wire, Staple 510(k) Premarket Notification
Image /page/3/Picture/2 description: The image shows a logo with a stylized letter Z. The word "MEDICAL" is written diagonally across the Z, enclosed in parentheses. Below the Z, the text "GMBH + CO. KG" is present, indicating a German company structure. The logo appears to be for a medical-related company.
ASTM F136 (Ti-6Al-4V ELI) Material: Size: 8 mm to 25 mm Method of insertion: Staple inserter and SnapOff technique
## 5. Indications for Use
#### 5.1. Z-Snap Off Screw
The Z-Medical SnapOff screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include:
- . Fixation of small bone fragments
- . Weil osteotomy
- Mono-cortical fixation .
- . Osteotomies and fractures fixation in the foot and hand
#### Z-Double Compression Screw 5.2.
The Z-Medical double thread compression screws are indicated for fixation of bone fractures or bone reconstruction. Examples included:
- ◆ Fixation of bone fragments or small bone fractures
- . Fracture management in the foot and hand
- Arthrodesis in hand, foot or ankle surgery .
- . Mono or Bi-cortical osteotomies in the foot and hand
#### 5.3. Z-Guide Wire
The Z-Medical's guide wires are intended to perform as fixation and stabilization unit of bone fractures or as guidance at insertion of implants into the skeletal system.
#### 5.4. Z-Staple
The Z-Medical Staples are indicated for fixation of bone fractures, bone reconstruction, ligament, soft tissue and tendon. Examples included:
- Fixation of bone fragments or small bones fractures .
- . Fracture management in the foot and hand
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with the word "MEDICAL" written diagonally across a stylized letter "Z". Below the "Z", the text "ONAH CO.KR" is visible. The logo appears to be for a medical company or organization, possibly based in Korea, given the ".KR" domain extension.
## 6. Testing
Testing in order to help establish safety and effectiveness of Z-Medical's bone fixation devices has been performed accordingly and results are conforming to the respective requirements.
- . Z-SnapOff: Torque test, break-off test, sideways stability test, maximum torque test
- Z-Double Compression Screw: Screw-in test, screw compression test, maximum torque test .
- Z-Staple: ASTM F564-10 test, break-off test .
## 7. Substantial Equivalence Summary / Conclusion
Based on available 510(k) information provided herein, Z-Medical's bone fixation devices are considered substantial equivalent to the predicate devices in terms of indications for use, material, technology, design and performance specifications.
There are no differences between the devices which would raise new issues of safety or effectiveness.
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Letter Dated: November 7, 2012
Z-Medical GmbH & Co. KG % Mr. Alexander Henninger Quality Manager Gänsäcker 38 Tuttlingen, Baden-Wuerttemberg Germany 78532
Re: K121277
Trade/Device Name: Z-Snap Off Screw, Z-Double Thread Compression Screw, Z-Guide Wire. Z-Staple
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC, JDR, HTY Dated: October 25, 2012 Received: October 25, 2012
Dear Mr. Henninger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
### Page 2 - Mr. Alexander Henninger
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo with the word "MEDICAL" in bold, uppercase letters diagonally across a stylized shape. Below the shape, there is text that appears to read "GMBH + CO.KG" in a smaller, sans-serif font. The logo has a simple, geometric design and the text is arranged to fit neatly beneath the graphic element.
## K121277
# Indications for Use Statement
Device Name: Z-Snap Off Screw
Indications for Use:
The Z-Medical SnapOff screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include:
- Fixation of small bone fragments ●
- . Weil osteotomy
- . Mono-cortical fixation
- Osteotomies and fractures fixation in the foot and hand t
Prescription Use _ YES (Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use (21 SFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k121277
SECTION*04*x01*IFU*Statement*SnapOffScrew*v02.doc Page 1 of 1
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a logo for a company called "Z Medical GmbH & Co. KG". The logo features the letter "Z" in a bold, stylized font, with the word "MEDICAL" written across the middle of the "Z". The text "GmbH & Co. KG" is written in a smaller font below the "Z".
### K121277
# Indications for Use Statement
### Device Name: Z-Double Thread Compression Screw
Indications for Use:
The Z-Medical double thread compression screws are indicated for fixation of bone fractures or bone reconstruction. Examples included:
- Fixation of bone fragments or small bone fractures .
- Fracture management in the foot and hand .
- Arthrodesis in hand, foot or ankle surgery ●
- Mono or Bi-cortical osteotomies in the foot and hand .
AND/OR Prescription Use _ yEs (Part 21 CFR 801 Subpart D)
Over-The-Counter Use NO_ (21 SFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121277
SECTION_04_x02_IFU_Statement_DoubleThreadCompressionScrew_v2.doc
Page 1 of 1
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a logo with the word "MEDICAL" written diagonally across a stylized letter "Z". The text "GMBH + CO. KG" is located at the bottom of the logo. The logo has a simple, geometric design with a bold, sans-serif typeface.
## K121277
# Indications for Use Statement
Device Name: Z-Guide Wire
Indications for Use:
The Z-Medical's guide wires are intended to perform as fixation and stabilization unit of bone fractures or as guidance at insertion of implants into the skeletal system.
Prescription Use __ YES (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ NO (21 SFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121277
SECTION_04_x03_1FU_Statement_GuideWire_v02.doc
Page 1 of 1
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows a logo with the word "MEDICAL" written diagonally across a stylized letter "Z". Below the "Z", the text "DMSK + OD.KO" is written in a smaller font. The logo appears to be for a medical-related company or organization. The image is in black and white.
# K121277
# Indications for Use Statement
Device Name: Z-Guide Staple
Indications for Use:
The Z-Medical Staples are indicated for fixation of bone fractures, bone reconstruction, ligament, soft tissue and tendon. Examples included:
- Fixation of bone fragments or small bones fractures .
- Fracture management in the foot and hand .
Prescription Use YES (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ NO (21 SFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
SECTION_04_x04_IFU_Statement_Staple_v2.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.