K121263 · Consensus Orthopedics, Inc. · LPH · May 22, 2012 · Orthopedic
Device Facts
Record ID
K121263
Device Name
TAPER SET HIP SYSTEM RDP STEM
Applicant
Consensus Orthopedics, Inc.
Product Code
LPH · Orthopedic
Decision Date
May 22, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TaperSet™ Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System. A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. C. Proximal femoral fractures. D. Avascular necrosis of the femoral head. E. Non-union of proximal femoral neck fractures. F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities. The TaperSet™ hip stem is indicated for cementless use.
Device Story
TaperSet Hip System RDP Stems are monolithic titanium alloy femoral hip stems for total or partial hip arthroplasty. Device features dual wedge geometry, 135° neck-shaft angle, 12/14 Morse taper trunnion, and proximal plasma-sprayed porous CPTi coating. Designed for cementless fixation. Used by orthopedic surgeons in clinical settings. Compatible with Consensus Hip System components including CoCr, zirconia, and Biolox delta ceramic heads, unipolar/bipolar heads, and UHMWPE inserts. RDP (Reduced Distal Profile) modification allows for anatomical fit in specific patient populations. Surgeon selects appropriate size (10.5mm to 24mm) and offset to restore hip biomechanics. Provides stable femoral component for joint replacement, aiming to alleviate pain and restore function in patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data. Bench testing only; engineering analysis of design modifications confirmed equivalence to predicate devices.
Indicated for patients requiring total or partial hip arthroplasty due to rheumatoid, osteoarthritis, post-traumatic arthritis, failed previous hip procedures, proximal femoral fractures, avascular necrosis, non-union of femoral neck fractures, congenital dysplasia, or other structural/metabolic bone conditions. Indicated for cementless use.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
## 2. 510(k) SUMMARY
| Sponsor Name: | Consensus Orthopedics, Inc.<br>1115 Windfield Way, Suite 100<br>El Dorado Hills, CA 95762 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Matthew M. Hull, RAC<br>Phone: (916) 355-7156/ Fax: (916) 355-7190<br>mhull@consensusortho.com |
| Date Prepared: | 14 May, 2012 |
| Trade Name: | TaperSet™ Hip System RDP Stems |
| Common Name: | Porous-coated hip prosthesis for cementless use |
| Classification Name: | Class II device<br>Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis: 21 CFR 888.3358, Product Code LPH.<br>Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis: 21 CFR 888.3353, Product Code<br>LZO.<br>Hip joint femoral (hemi-hip) metallic cemented or uncemented<br>prosthesis: 21 CFR 888.3360, Product Code KWL.<br>Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented<br>prosthesis: 21 CFR 888.3390, Product Code KWY. |
### Device Description:
The TaperSet Hip System (THS) is a monolithic, titanium alloy tapered hip stem design with a proximal, plasma sprayed, porous CPTi coating. The stem has a dual wedge geometry and is available in both standard and 7mm lateral offsets in sizes designated as 7.5mm to 24mm. The stems feature a neck shaft angle of 135° and a 12/14 Morse taper trunnion. The TaperSet Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System. The stem is compatible with previously cleared CoCr heads, zirconia or Biolox delta ceramic heads, unipolar heads, bipolar heads, UHMWPE inserts and acetabular cups. The modification addressed here is the addition of a line of stems with a reduced distal profile (RDP) in sizes 10.5mm to 24mm.
### Indications for Use:
The TaperSet™ Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System.
The indications for use are:
- A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
- B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
{1}------------------------------------------------
- C. Proximal femoral fractures.
/
- D. Avascular necrosis of the femoral head.
- E. Non-union of proximal femoral neck fractures.
- Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, F. coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
KI21263
The TaperSet™ hip stem is indicated for cementless use.
## Substantial Equivalence:
## Technological Characteristics/Substantial Equivalence:
The new Reduced Distal Profile (RDP) stems for the TaperSet Hip System have the identical neck and taper design, porous coating, and sizing as the predicate TaperSet stems cleared under K 102399 in design and indications. A reduced distal profile femoral hip stem design was previously cleared by FDA under K101086 for the Biomet Taperloc Complete Hip System. The subject Consensus RDP stems are still compatible with previously cleared CoCr heads, zirconia and Biolox delta ceramic heads, unipolar heads, bipolar heads, UHMWPE inserts and acetabular cups. Based on the material, characterization data, and geometry, the TaperSet Hip RDP stem is substantially equivalent to the legally marketed predicates.
## Non-Clinical Performance Data:
No additional non-clinical testing was performed because there was no proximal design change from the predicate TaperSet and distally the new smallest size is still larger than the smallest predicate TaperSet stem which was previously tested.
Based upon engineering analysis of the design modification the new reduced distal profile stems for the TaperSet Hip System by Consensus are substantially equivalent to devices currently marketed. Therefore, the device is as safe, as effective, and performs at least as safely and effectively as legally marketed predicates.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Consensus Orthopedics, Incorporated % Mr. Matthew Hull, RAC OS & RA Director 1115 Windfield Way, Suite 100 El Dorado Hills, California 95762-9623
MAY 2 2 2012
### Re: K121263
Trade/Device Name: TaperSet™ Hip System RDP Stems Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, KWL, KWY Dated: April 25, 2012
Received: April 26, 2012
Dear Mr. Hull:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
# Page 2 - Mr. Matthew Hull, RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
erely yours,
x N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 1. INDICATIONS FOR USE STATEMENT
K121263 510(k) Number (if known): 455
Device Name: TaperSet™ Hip System RDP Stems
### Indications for Use:
The TaperSet™ Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System.
- A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
- B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- C. Proximal femoral fractures.
- D. Avascular necrosis of the femoral head.
- E. Non-union of proximal femoral neck fractures.
- F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
The TaperSet™ hip stem is indicated for cementless use.
Prescription Use X (21 CFR Part 801 Subpart D) AND/OR
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ansto
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121263
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.