← Product Code [JXG](/productcode/JXG) · K121248

# LIQUOGUARD CSF LIQUOGUARD CSF TUBING SET (1500MM) LIQUOGUARD CSF TUBING SET (2000MM) SAFETY CUSHION FOR TUBING SET ATTAC (K121248)

_Moller Medical GmbH · JXG · Dec 13, 2013 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K121248

## Device Facts

- **Applicant:** Moller Medical GmbH
- **Product Code:** [JXG](/productcode/JXG.md)
- **Decision Date:** Dec 13, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5550
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K121248 · Dec 13, 2013 | LIQUOGUARD CSF LIQUOGUARD CSF TUBING SET (1500MM) LIQUOGUARD CSF TUBING SET (2000MM) SAFETY CUSHION FOR TUBING SET ATTAC | Moller Medical GmbH | Hospital patient records | A retrospective analysis of 413 patient records from multiple hospitals was conducted to compare the incidence of adverse events between the LiquoGuard system and traditional drip chamber systems to support substantial equivalence. | Retrospective chart review; Adverse event comparison; Hospital records |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective analysis of patient records | Patients undergoing CSF drainage using LiquoGuard or predicate drip chamber systems; Sample Size: 413 patient records; Number of Sites: Several hospitals | Predicate drip chamber systems | Incidence of adverse events |

## Indications for Use

The LiquoGuard CSF system is indicated for the external drainage of cerebrospinal fluid (CSF). It connects to any drainage catheter (not part of the product) which is usually inserted by the doctor into the lateral or third ventricle of the brain or lumbar subarachnoid space in selected patients to reduce intracranial pressure. The LiquoGuard CSF system controls CSF pressure using pressure sensors and a pump, thus taking the role of the conventional CSF dripping chamber and collection bag (predicate devices). The LiguoGuard CSF system also monitors intracranial pressure up to 75 mmHg and cerebrospinal fluid flow during CSF drainage, and provides many alarm functions not offered by the passive dripping chamber systems.

## Device Story

LiquoGuard CSF System is an active, pump-driven external CSF drainage and monitoring device. It accepts input from an external intrathecal drainage catheter via a disposable tube set equipped with dual piezo-resistive pressure sensors. The device uses a peristaltic pump to actively control CSF drainage based on preselected target pressures, replacing passive dripping chambers. It operates in clinical settings (e.g., ICU, OR) and is managed by healthcare professionals. The system provides real-time monitoring of ICP (up to 75 mmHg), CSF flow, and pressure trends, with integrated alarms for pressure deviations, catheter occlusion, sensor failure, and battery status. Data logging and bedside monitor connectivity are supported. By automating drainage and providing active monitoring, the device aims to improve safety, simplify handling, and enhance patient mobility via battery backup.

## Clinical Evidence

Retrospective study of 413 patient records comparing LiquoGuard to traditional drip chambers. Primary endpoint was incidence of adverse events. Results showed similar safety profiles between the two methods, with small variations in favor of the LiquoGuard system. Bench testing included pressure measurement precision, alarm functionality (pressure, pulsation, occlusion), mechanical durability, and battery performance; all tests passed predetermined criteria.

## Technological Characteristics

System consists of a pump unit and disposable tube set. Sensing: dual piezo-resistive pressure transducers. Energy: battery-backed AC power. Dimensions: 315mm x 210mm x 280mm; weight 4.8 kg. Connectivity: external bedside monitor interface. Software: microcontroller-based for pump control, data logging, and alarm management. Sterilization: disposable tube set.

## Regulatory Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

## Predicate Devices

- Integra LifeSciences Corporation ([K030289](/device/K030289.md))
- Integra LifeSciences Corporation ([K032817](/device/K032817.md))
- Codman & Shurtleff, Inc. ([K061568](/device/K061568.md))

## Reference Devices

- Camino NeuroCare, Inc. ([K962928](/device/K962928.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K121248

Image /page/0/Picture/1 description: The image contains the logo for MöllerMedical. The logo consists of a stylized graphic to the left of the company name. The graphic appears to be a cluster of three circular shapes.

Page: 1 (9)

# 510(k) Summary

# DEC 1 3 2013

Submitter Information: Möller Medical GmbH Wasserkuppenstrasse 29-31 36043 Fulda, Germany

Establishment Registration Number: 3006611221

יי

Contact Person:

Grant Geckeler Free lance consultant 4375 Prado Road, Suite 103 Corona, CA 92880 Phone: +1 714 283 1683 Fax: +1 714 637 2460 E-mail: ggeckeler@geoinc.org

### Name of Device:

| Common Name:      | External CSF Drainage System                                                                        |
|-------------------|-----------------------------------------------------------------------------------------------------|
| Proprietary Name: | LiquoGuard CSF System<br>Consists of: LiquoGuard CSF System pump and LiquoGuard CSF System tube set |
| Classification:   | Shunt, Central Nervous System and Components, Class II, 21 CFR § 882.5550                           |
| Product Code:     | JXG                                                                                                 |

### Device Description:

Cerebrospinal fluid (CSF) drainage and pressure monitoring in lumbar and ventricular applications is the "gold standard" for cases involving trauma, post-operation, shunt infection, subarachnoid hemorrhage (SAH) therapy, neurophysiological monitoring during vascular surgery (e.g. TAA and TAAA) and normal-pressure hydrocephalus (NPH) diagnosis. For CSF drainage, dripping chambers are still being used widely. In contrast, dripping chambers for IV infusion have been mostly replaced by perfusors (syringe pumps).

The LiquoGuard CSF System consists of a pump and a corresponding tube set with pressure sensors. The dynamic range of this system is from -15mmHg to +75 mmHg. The tube set is inserted into the pump, connected to the intrathecal drainage catheter (not part of the product) via Luer-Lock and the LiquoGuard CSF System device via cable connector. The device then continuously measures the pressure inside the tube and operates a tube pump whenever the actual pressure is higher than a preselected target pressure. Thus the LiquoGuard CSF System combines CSF drainage and intracranial pressure (ICP) monitoring with integrated alarm and documentation/data logging functions, improving safety, simplifying the handling and enhancing patient mobility (by its battery back-up).

{1}------------------------------------------------

### Intended Use:

The LiquoGuard CSF system is indicated for the external drainage of cerebrospinal fluid (CSF). It connects to any drainage catheter (not part of the product) which is usually inserted by the doctor into the lateral or third ventricle of the brain or lumbar subarachnoid space in selected patients to reduce intracranial pressure. The LiquoGuard CSF system controls CSF pressure using pressure sensors and a pump, thus taking the role of the conventional CSF dripping chamber and collection bag (predicate devices). The LiguoGuard CSF system also monitors intracranial pressure up to 75 mmHg and cerebrospinal fluid flow during CSF drainage, and provides many alarm functions not offered by the passive dripping chamber systems.

### Substantial Equivalence:

The Moller Medical LiquoGuard CSF System is substantially equivalent to the CSF drainage systems listed below with regard to intended use, design, operating principles, and materials.

Integra NeuroSciences Corporation Codman & Shurtleff, Inc.

K030289. K032817 K061568

The Moller Medical LiquoGuard CSF System shares features with, the intracranial pressure monitor listed below (the shared features include CSF pressure monitoring, semiconductor sensor principle, display, alarm functions and data recording).

Camino NeuroCare. ™ Inc.

### K962928

By using the LiquoGuard CSF System, in the same way as in existing systems, intrathecal pressure can be measured; critical intrathecal pressure in the patient can be recognized, thereby enabling the required diagnostic and treatment steps to be taken. Since the pressure measurement in systems is done via piezo-resistive pressure sensors that are attached at reference height to the patient, there is no difference in intrathecal pressure measurement.

Additional details for comparing these CSF Systems are contained in Table 8 and Table 9.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the logo for "MöllerMedical". The logo consists of a cluster of four irregular circles on the left side of the text. The text "MöllerMedical" is written in a sans-serif font and is positioned to the right of the circles.

# Traditional 510(k) Submission 510(k) Summary

Page: 3 (9)

.

| Product              |                                                     | Dipping chamber systems | ICP Monitor<br>system            | Combination              |
|----------------------|-----------------------------------------------------|-------------------------|----------------------------------|--------------------------|
| Component            | Hermetic Plus<br>External CSF<br>Drainage<br>System | CODMAN EDS<br>3 CSF     | Multi-Parameter<br>Monitor (MPM) | LiquoGuard<br>CSF System |
| Tube set disposable  | ●                                                   | ●                       |                                  | ●                        |
| CSF drainage bag     | ●                                                   | ●                       |                                  | ●                        |
| Pressure transducer  |                                                     |                         | ●                                | ●                        |
| Micro controller     |                                                     |                         | ●                                | ●                        |
| Software operated    |                                                     |                         | ●                                | ●                        |
| Data storage         |                                                     |                         | ●                                | ●                        |
| Button / Foil keypad |                                                     |                         | ●                                | ●                        |
| Screen               |                                                     |                         | ●                                | ●                        |
| Ports                |                                                     |                         | ●                                | ●                        |
| Casing               |                                                     |                         | ●                                | ●                        |
| Power supply         |                                                     |                         | ●                                |                          |

## Table 8: System Comparison

'

·

{3}------------------------------------------------

SemöllerMedical

Page: 4 (9)

# Table 9: Device Comparison

|                                                                          |                                                                              | Primary Predicate Devices                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                | Reference Device                                                                                                        |                                                | Primary Predicate Devices                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                    | Reference Device                                                       |                                                                              | Primary Predicate Devices  |                                                                                | Reference Device                                          |                                       |                                                                                                                                                 |
|--------------------------------------------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------------|----------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                          |                                                                              | Integra LifeSciences<br>Corporation<br>(K030289, K032817)                                                                                                                                                                                                                      | Codman & Shrutleff, Inc.<br>(K061568)                                                                                                                                                                                          | Camino NeuroCare,™ Inc.<br>(K962928)                                                                                    | Device                                         | Möller Medical<br>LiquoGuard CSF System                                                                                                                                                                                                                                                                                                                                                                 | Integra LifeSciences<br>Corporation<br>(K030289, K032817)                                          | Codman & Shrutleff, Inc.<br>(K061568)                                  | Camino NeuroCare,™ Inc.<br>(K962928)                                         |                            | Möller Medical<br>LiquoGuard CSF System                                        | Integra LifeSciences<br>Corporation<br>(K030289, K032817) | Codman & Shrutleff, Inc.<br>(K061568) | Camino NeuroCare, ™ Inc.<br>(K962928)                                                                                                           |
| Indication                                                               | Möller Medical<br>LiquoGuard CSF System                                      | External draining of<br>cerebrospinal fluid (CSF)<br>from the lateral ventricles<br>of the brain or the lumbar<br>subarachnoid space to<br>reduce intracranial<br>pressure and measurement<br>of CSF pressure.                                                                 | External draining of<br>cerebrospinal fluid (CSF)<br>and other fluids of similar<br>physical characteristics<br>from a patient at a<br>controlled rate based on<br>differential pressure<br>between the device and<br>patient. | Measurement of<br>intracranial pressure in the<br>ventricles and during the<br>cerebrospinal fluid (CSF)<br>drainage.   | Device adjusting<br>fluid flow                 | sensor needs to be aligned.<br>LiquoGuard CSF System<br>peristaltic pump controlled<br>by pressure at sensor<br>location.                                                                                                                                                                                                                                                                               | target pressure.<br>None - fluid flow is not<br>controlled or limited.                             | target pressure.<br>None - fluid flow is not<br>controlled or limited. | incorporated.<br>No drainage.                                                | Measured<br>pressure range | -15mmHg to +75mmHg.                                                            | No measurement.                                           | No measurement.                       | -10mmHg to +250mmHg.                                                                                                                            |
| CSF sterile<br>tubing set with<br>pressure sensor<br>and drainage<br>bag |                                                                              | Tubing material:<br>Polyvinyl chloride (PVC).<br>Pressure sensor:<br>Dual transducers.<br>CSF Collection Bag:<br>500ml.                                                                                                                                                        | Tubing material:<br>Polyvinyl chloride (PVC).<br>Pressure sensor:<br>None.<br>CSF Collection Bag:<br>700ml.                                                                                                                    | Tubing material:<br>Not applicable.<br>Pressure sensor:<br>Single transducer.<br>CSF Collection Bag:<br>Not applicable. | Monitoring<br>functions                        | Drained volume and<br>pressure curve inside<br>tubing (equal to intracranial<br>pressure unless blockage)<br>are monitored and recorded<br>by LiquoGuard CSF<br>System.                                                                                                                                                                                                                                 | Pressure is sensed and<br>may be recorded by<br>external device. Drained<br>volume is unmonitored. | No monitoring functions.                                               | Pressure, pressure curve<br>and temperature are<br>monitored and recorded.   | Physical<br>dimensions     | 315mm x 210 mm x 280 mm (H x W x D).<br>4.8 kg.                                |                                                           |                                       | 274 mm x 216 mm x 89 mm (H x W x D).<br>4.3 kg.                                                                                                 |
| Pressure<br>controlling<br>mechanism'                                    |                                                                              | Based on pressure<br>equivalence (Pascal's law<br>of fluid statics) between<br>patient and pressure<br>sensors inside the tubing.<br>From the sensor location,<br>drainage is performed<br>using a peristaltic pump<br>controlled by the sensed<br>pressure inside the tubing. | Based on differential<br>pressure (Pascal's law of<br>fluid statics) caused by<br>height difference between<br>patient and dripping<br>chamber.                                                                                | Based on pressure<br>equivalence (Pascal's law<br>of fluid statics) inside the<br>head.                                 | Alarm functions<br>enhancing<br>patient safety | Many alarm functions<br>integrated in the system:<br>* Pressure too high.<br>* Pressure too low.<br>* Physiological pressure<br>pulsation lost.<br>* Pressure remains<br>constant for too long.<br>* Dual sensors deviate in<br>readings (defective sensor).<br>* Watchdog or main<br>controller does not respond<br>to each other's requests<br>(device or software<br>malfunction).<br>* Battery low. | No integrated alarm<br>functions.                                                                  | No alarm functions.                                                    | Few integrated alarm<br>functions:<br>* Pressure too high.<br>* Battery low. | Displayed<br>parameters    | Actual ICP value, ICP<br>waveform, ICP trend.<br>Display resolution of 1 mmHg. | No display.                                               | No display.                           | Actual ICP value, ICP<br>waveform, ICP trend,<br>intracranial temperature,<br>Cerebral Perfusion,<br>Pressure.<br>Display resolution of 1 mmHg. |
| Height alignment<br>of patient and                                       |                                                                              | Device can be positioned<br>arbitrarily, only the pressure                                                                                                                                                                                                                     | Alignment is critical<br>because it controls the                                                                                                                                                                               | Device can be positioned<br>arbitrarily; the pressure<br>sensors need to be                                             | Sensor type                                    | Dual transducer.                                                                                                                                                                                                                                                                                                                                                                                        | Single transducer.                                                                                 | No sensor.                                                             | Fiber-optic single pressure<br>transducer.                                   |                            |                                                                                |                                                           |                                       |                                                                                                                                                 |
| Scheduled<br>preventive<br>maintenance                                   | Every 12 months                                                              | No                                                                                                                                                                                                                                                                             | No                                                                                                                                                                                                                             | Every 18 months                                                                                                         |                                                |                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                    |                                                                        |                                                                              |                            |                                                                                |                                                           |                                       |                                                                                                                                                 |
| Battery                                                                  | Rechargeable sealed lead<br>acid.<br>Charge time: 8h.<br>Operation time: 3h. | No battery.                                                                                                                                                                                                                                                                    | No battery.                                                                                                                                                                                                                    | Rechargeable sealed lead<br>acid.<br>Charge time: 8-10h.<br>Operation time: 1-2h.                                       |                                                |                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                    |                                                                        |                                                                              |                            |                                                                                |                                                           |                                       |                                                                                                                                                 |

2013-12

{4}------------------------------------------------

SMöllerMedical

raditional 510(k) Submission

.
Page: 5 (9)

. 2013-12

{5}------------------------------------------------

SemöllerMedical

raditional 510(k) Submission

Page: 6 (9)

2013-12

-

. ·

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MöllerMedical. The logo consists of a cluster of four circles on the left side of the text. The text "MöllerMedical" is written in a simple, sans-serif font.

### Summary of substantial equivalence discussion

Many if not all risks leading to accidental, cerebral herniation in both systems (traditional drainage and ICP monitors) can be reduced or even eliminated by the LiquoGuard CSF Svstem.

The LiquoGuard CSF System is working equivalently to a drip chamber, with the additional safety that, using LiguoGuard CSF System, a partial or intermittent catheter occlusion may be detected in many cases: In difference to the dripping chamber which must be an open system during drainage the LiquoGuard CSF System can measure the hydrostatic pressure inside the tube (drainage system) during drainage because the drainage system is closed, combined with a CSF pressure monitor and its features like displaying pressure values, curves and trends and enabling alarm functionality. The principles of measurement are comparable to the ICP monitor and its pressure transducer as both systems measure the pressure of the surrounding liquid. The only difference is the position of the transducers, and the fact that LiquoGuard uses dual transducers for redundancy.

The LiguoGuard CSF System behaves comparable to a traditional drainage system even in the case of a clotted catheter or collapsed ventricles. It is able to sound and display alarms to operators, is portable, can be used while not connected to the mains, records and stores data and can be connected to bedside monitor systems, same as the Multi Parameter Monitor.

By combining the functionality of dripping chambers with the features of a computer operated device (ICP monitor) and adding a peristaltic pump the LiguoGuard CSF System benefit is potentiated as some major problems are solved (elimination of sensor drift problematic, protection of accidental over and / or under drainage, pressure measurement simultaneously with CSF drainage, etc.).

### Summary and Conclusion of the Performance Testing:

The LiguoGuard CSF System has the same indication for use as the predicate devices (measurement and drainage of CSF). In addition, the LiquoGuard CSF System tubing set is made from the same materials as the predicate devices, and because the risks associated with the surgical insertion of a CSF drain are not dependent on the CSF drainage system chosen, no new safety or effectiveness concerns are introduced over the predicate devices. Please refer to Table 10 for the list of bench tests - there you find a list of all bench tests that have been performed.

The test series exhibits a very wide range of performance tests and all necessary and required tests for the LiquoGuard were appropriate performed and all tests passed according to the predetermined passifiall criteria. Particularly highlighting are tests which serve to check the pressure and alarm systems. For this purpose, a plurality of measurements has been carried out to verify the critical components.

2013-12

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for MöllerMedical. The logo consists of the text "MöllerMedical" with a stylized graphic to the left. The graphic appears to be a cluster of oval shapes, possibly representing cells or other biological elements.

. '

. .

# Traditional 510(k) Submission

510(k) Summary

| Table 10: List of bench test                    |                                                                                                                                                                                                                                                                                                                                                   |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pressure measurement                            | ✓ Absence of sensor drift during packaging / sterilization<br>✓ Absence of sensor drift during ageing<br>✓ Precision of displayed pressure at various pump rates and pressure loads                                                                                                                                                               |
| Pressure regulation                             | ✓ Endurance test with in-vitro simulated patient<br>✓ Performance comparison with drip chamber<br>✓ Reliability of pump shut-off with negative pressures<br>✓ Device functions independent of positioning                                                                                                                                         |
| Volume and flow<br>measurement                  | ✓ Precision of total volume and flow (pump rate) measurement                                                                                                                                                                                                                                                                                      |
| Alarm functions related to<br>patient condition | ✓ Pressure alarms<br>✓ Pulsation alarm (indication for events like: catheter occlusion collapsed ventricles and disconnected catheter)<br>✓ Pressure too constant (indication for events like: catheter occlusion collapsed ventricles and disconnected catheter)<br>✓ Influence of pressure on catheter walls<br>✓ Detection of clogged catheter |
| Alarm functions related to<br>device functions  | ✓ Defective/disconnected sensors<br>✓ Disconnected/leaking tube set<br>✓ Battery low<br>✓ Main controller/watchdog cross-check<br>✓ Double safety<br>✓ Tube set too old<br>✓ Battery defective                                                                                                                                                    |
| Mechanical handling and<br>operation            | ✓ Disposable check for single-use<br>✓ Wrong tube set insertion<br>✓ Opening of front panel<br>✓ Disposable check for age                                                                                                                                                                                                                         |
| Endurance                                       | ✓ Automatic switch to battery upon power failure<br>✓ Tube set durability                                                                                                                                                                                                                                                                         |
| Other                                           | ✓ External connectivity                                                                                                                                                                                                                                                                                                                           |

### Table 10: List of bench test

{8}------------------------------------------------

### Summary and Conclusion of the Clinical Study:

Drawing on the user base and combining similar, independent studies from several hospitals into one data basis, we collected 413 patient records on treatments using the LiquoGuard (LG) as well as the predicate devices (drip chambers). We identified a number of "adverse events" and compared their rate of incidence for both cases (LG vs. predicate) for the total patient population as well as certain patient categories, e.g. medical indication.

Since the two CSF drainage methods (LiquoGuard and drip chamber) apply the same therapy (drainage) to the patients, we expected to obtain similar results. Indeed, the data show a great similarity. E.g., total rates of incidence of adverse events are for each type of event very close, with most small variations in favor of the LiquoGuard.

The findings support the equivalence of LiquoGuard and the listed predicate devices.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a logo for the Department of Health & Human Services. The logo features the department's emblem, which includes a stylized caduceus-like symbol with three wavy lines and a circle containing the words "DEPARTMENT OF HEALTH & HUMAN SERVICES". Above the emblem, there are the letters "JXG".

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

### December 13, 2013

Möller Medical GmbH c/o Mr. Grant Geckeler 1612 Jenks Drive Corona, CA 92880

Re: K121248

Trade/Device Name: LiquoGuard CSF System Regulation Number: 21 CFR 882.5550 Regulation Name: Shunt, Central Nervous System and Components Regulatory Class: Class II Product Code: JXG Dated: May 13, 2013 Received: September 16, 2013

Dear Mr. Geckeler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA.'s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{10}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Joyce M. Whang -S

for .

for Victor Krauthamer, Ph.D. Acting Director

Division of Neurological

and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

# Indications for Use

# 510(k) Number: K121248

# Device Name: LiquoGuard CSF System

Indications for Use:

The LiquoGuard CSF system is indicated for the external drainage of cerebrospinal fluid (CSF). It connects to any drainage catheter (not part of the product) which is usually inserted by the doctor into the lateral or third ventricle of the brain or lumbar subarachnoid space in selected patients to reduce intracranial pressure. The LiquoGuard CSF system controls CSF pressure using pressure sensors and a pump, thus taking the role of the conventional CSF dripping chamber and collection bag (predicate devices). The LiquoGuard CSF system also monitors intracranial pressure up to 75 mmHg and cerebrospinal fluid flow during CSF drainage, and provides many alarm functions not offered by the passive dripping chamber systems.

Prescription Use A (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Page 1 of 1

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

---

**Source:** [https://fda.innolitics.com/device/K121248](https://fda.innolitics.com/device/K121248)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
