K-ASSAY (R) FERRITIN REAGENT, K-ASSAY (R) FERRITIN CALIBRATOR

K121232 · Kamiya Biomedical Company · DBF · May 24, 2012 · Immunology

Device Facts

Record IDK121232
Device NameK-ASSAY (R) FERRITIN REAGENT, K-ASSAY (R) FERRITIN CALIBRATOR
ApplicantKamiya Biomedical Company
Product CodeDBF · Immunology
Decision DateMay 24, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5340
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K121232 · May 24, 2012K-ASSAY (R) FERRITIN REAGENT, K-ASSAY (R) FERRITIN CALIBRATORKamiya Biomedical CompanyRetrospective clinical patient serum and plasma samplesRetrospective patient samples were used to conduct method comparison studies against a predicate device and to demonstrate matrix equivalence between serum, EDTA plasma, and heparin plasma.Method comparison; Matrix equivalence; Clinical patient samples

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Method comparison studyHuman serum samples; Sample Size: 54Predicate device (k050944)Regression analysis (y = 1.0057x + 0.0968)
Matrix equivalence studyMatched serum and plasma patient samples; Sample Size: 65 (serum vs. EDTA), 59 (serum vs. heparin)Not applicable for this studyRegression analysis for matrix equivalence

Indications for Use

The K-ASSAY® Ferritin (2nd Gen.) assay is an in vitro diagnostic reagent for the quantitative determination of ferritin (an iron-storing protein) in human serum and plasma by immunoturbidimetric assay on the Beckman AU680 analyzer. Measurements of ferritin aid in the diagnosis of diseases affecting iron overload and iron deficiency anemia. For in vitro diagnostic use. The K-ASSAY® Ferritin Calibrator (2nd Gen.) is an in vitro diagnostic reagent for calibration of the K-ASSAY® Ferritin (2nd Gen.) assay. For in vitro diagnostic use.

Device Story

K-ASSAY® Ferritin (2nd Gen.) is a latex-enhanced immuno-turbidimetric assay for quantitative ferritin measurement in human serum/plasma. Input: patient serum/plasma samples mixed with latex particles sensitized with mouse monoclonal anti-human ferritin antibodies. Principle: immune complex formation measured via turbidimetry on Beckman/Olympus AU680 analyzer; signal correlates to ferritin concentration. Output: quantitative ferritin level (ng/mL) calculated via interpolation on a 6-point calibration curve. Used in clinical laboratories by technicians/pathologists to assist in diagnosing iron-related disorders (overload/anemia). Benefits: provides standardized, automated diagnostic data for iron status assessment.

Clinical Evidence

Bench testing only. Performance evaluated per CLSI guidelines (EP05-A2, EP06-A, EP17-A, EP07-A2, EP09-A2-IR). Precision (CV% < 1% between-lot), linearity (5-1,000 ng/mL), LoQ (5.0 ng/mL), and interference studies (no significant interference for bilirubin, hemoglobin, RF, etc.) were conducted. Method comparison with predicate (n=54) showed strong correlation (y = 1.0057x + 0.0968). Matrix comparison confirmed suitability of serum, EDTA plasma, and heparin plasma.

Technological Characteristics

Latex-enhanced immuno-turbidimetric assay. Reagents: HEPES buffer, latex particles sensitized with mouse monoclonal anti-human ferritin antibodies, sodium azide preservative. Form factor: liquid stable reagents and calibrators. Connectivity: designed for use on Beckman/Olympus AU680 clinical chemistry analyzer. Standardization: traceable to WHO 2nd International Ferritin Standard (lot 80/578).

Indications for Use

Indicated for quantitative determination of ferritin in human serum and plasma to aid in diagnosis of iron overload and iron deficiency anemia. Prescription use only.

Regulatory Classification

Identification

A ferritin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ferritin (an iron-storing protein) in serum and other body fluids. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency amemia.

Submission Summary (Full Text)

{0} Statement for the Record, k121232 This 510(k) was reviewed under OIVD’s Pilot Triage Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA. The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant’s 510(k) summary for a summary of the information that supports this SE determination.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...