K121152 · Jiangsu Guangda Medical Material, Co., Ltd. · FYA · Dec 12, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121152
Device Name
SURGICAL GOWN, SMS
Applicant
Jiangsu Guangda Medical Material, Co., Ltd.
Product Code
FYA · General, Plastic Surgery
Decision Date
Dec 12, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Jiangsu Guangda's Reinforced Surgical Gowns, Model Number GD-SG-01, are non-sterile, single use surgical gowns intended to protect surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material. This product may be sterilized using Ethylene Oxide (EO) following the validation and routine control under ANSI/AAMI/ISO 11135.
Device Story
Disposable, reinforced surgical gown (Model GD-SG-01) designed for use in surgical settings to provide a barrier against microorganisms, body fluids, and particulates. Constructed from SMS fabric with a PE+PP two-layer compound reinforcement and 100% white Terylene cuffs. Device is provided non-sterile but is compatible with Ethylene Oxide (EO) sterilization per ANSI/AAMI/ISO 11135. Worn by operating room personnel to protect both the wearer and the patient during surgical procedures. Performance verified through bench testing for liquid barrier properties, tensile/tear/seam strength, and flammability.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-5, -10), liquid barrier performance (AATCC 127, AATCC 42, ASTM F1671, ASTM F1670), physical strength (ASTM D5034, D5733, D1683), and flammability (16 CFR 1610). All tests met acceptance criteria.
Technological Characteristics
Materials: SMS fabric and PE+PP two-layer compound; 100% Terylene cuffs. Form factor: XL size, blue color. Barrier performance: ANSI/AAMI PB70 compliant. Physical specs: ASTM D5034, D5733, D1683. Flammability: Class I (16 CFR 1610). Biocompatibility: ISO 10993-1, -5, -10. Sterilization: Compatible with EO (ANSI/AAMI/ISO 11135).
Indications for Use
Indicated for use as a single-use, non-sterile surgical gown to protect surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material. Suitable for use in surgical settings.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{1}------------------------------------------------
# K121152
# Section 5: 510(k) Summary
#### April 12, 2012 1. Preparation Date:
# DEC 1 2 2012
## 2. Submitted by:
Jiangsu Guangda Medical Material Co., Ltd. 18 BaoChang Road North, LiBao Town, Haian County, Jiangsu Province, China 226631 Owner/Operator #: 10023676 Registration Number: 3006847952
## 3. Contact Person/Prepared by:
Darren Reeves (804) 307-7706 Phone: Fax: (866) 393-4954 Email: dreeves@dpdconline.com
## 4. Device Identification:
Trade Name: Surgical Gown Common Name: Gown, Surgical Classification: Surgical Apparel, FYA, Class 2 (878.4040)
#### GRI MEDICAL Surgical Gowns (K102652) 5. Predicate Device:
#### 6. Device Description:
Jiangsu Guangda's Disposable XLarge Reinforced Surgical Gowns, Model Number GD-SG-01, are made from SMS fabrics and PE + PP two layer compound protective reinforcement with 100% white Terylene cuffs to provide user protection in surgical settings.
### 7. Intended Use:
Jiangsu Guangda's Reinforced Surgical Gowns, Model Number GD-SG-01, are non-sterile, single use surgical gowns intended to protect surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material.
This product may be sterilized using Ethylene Oxide (EO) following the validation and routine control under ANSI/AAMI/ISO 11135.
Page 3 of 7
Image /page/0/Picture/18 description: The image shows the text "P. 1 of 3" written in black ink on a white background. The text appears to be handwritten. The letter "P" is followed by a comma, then the number "1", the word "of", and finally the number "3". The text likely indicates a page number or a sequence within a document.
{2}------------------------------------------------
#### 8. Comparison to Predicate:
The following table shows similarities and differences between the predicate identified in Section 5 of this summary. There are no significant differences between the Jiangsu Guangda's Gowns, Model Number GD-SG-01, and their respective predicate products.
| | Item | Proposed Device: Jiangsu Guangda's Gowns,<br>Model Number GD-SG-01 | Predicate Device: GRI MEDICAL<br>Surgical Gowns (K102652) |
|-----------------------------------------------|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Product Code | FYA, Class II | FYA, Class II |
| | Intended Use | Jiangsu Guangda's Reinforced Surgical Gowns are<br>non-sterile, single use surgical gowns intended to<br>protect surgical patients and operating room<br>personnel from the transfer of microorganisms,<br>body fluids, and particulate material.<br>This product may be sterilized using Ethylene<br>Oxide (EO) following the validation and routine<br>control under ANSI/AAMI/ISO 11135. | GRI's Non Reinforced, Film Reinforced, and<br>Fabric Reinforced Surgical Gowns are<br>sterile and non-sterile, single use surgical<br>gowns intended to protect surgical patients<br>and operating room personnel from the<br>transfer of microorganisms, body fluids,<br>and particulate material. |
| | Style | Reinforced | Reinforced and non-reinforced |
| | Durability | Disposable | Disposable |
| | Size | XL | Various |
| | Color | Blue | Various |
| | Material | SMS and PE+PP two layer compound | Same |
| | Weight per square | 45 g/m² | 50 g/m² |
| | Label and Labeling | Conforms to FDA Requirements | Unknown |
| Liquid Barrier Performance:<br>ANSI/AAMI PB70 | Hydrostatic Pressure: AATCC<br>Test Method 127 | All were >20 | Has met acceptance criteria |
| | Impact Penetration: AATCC Test<br>Method 42 | All were ≤1 | Has met acceptance criteria |
| | Resistance by Blood-Borne<br>Pathogen: ASTM- F1671 | All were ≤1 | Unknown |
| | Resistance by Synthetic Blood:<br>ASTM- F1670 | Passed | Has met acceptance criteria |
| Physical Specifications | Breaking Strength: ASTM-<br>D5034 | Passed | Has met acceptance criteria |
| | Tearing Strength: ASTM- D5733 | Passed | Has met acceptance criteria |
| | Flammability Test Method (16<br>CFR 1610) | Class I | Has met acceptance criteria |
| | Seam Strength: ASTM- D1683 | Passed | Has met acceptance criteria |
| Biocompatibility:<br>ISO 10993-1 | Cytotoxicity: ISO 10993-5 | Passed | Has met acceptance criteria |
| | Irritation: ISO 10993-10 | Passed | Has met acceptance criteria |
| | Sensitization: ISO 10993-10 | Passed | Has met acceptance criteria |
| Sterilization | Method | Although sold non-sterile, gowns can be EO<br>Sterilized | Unknown |
| | EO and ECH Residual | Conforms to ISO 10993-7 | Unknown |
Jiangsu Guangda Medical Material Co., Ltd. Document Number K121152 S3.1
Page 4 of 7
{3}------------------------------------------------
#### 9. Summary of Testing:
The information in the Premarket Notification on safety and efficiency supports a finding of substantially equivalence to devices already in commercial distribution. Equivalence is demonstrated through intended use, material, design and testing methods.
| Standard or Guidance | Data Generated | Relevant Section of<br>Submission |
|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|-----------------------------------|
| ASTM-F2407-06 Standard Specification for Surgical Gowns intended for Use in Healthcare Facilities | Biocompatibility Testing Evaluation | 15 |
| AAMI/ANSI/ISO 10993-1:2003(E), Biological evaluation of<br>medical devices -- Part 1: Evaluation and testing | Biocompatibility Testing<br>Evaluation | 15 |
| AAMI/ANSI/ISO 10993-5:2009, Biological evaluation of medical<br>devices -- Part 5: Tests for In Vitro cytotoxicity | Cytotoxicity | 15 |
| AAMI/ANSI/ISO 10993-10:2010, Biological evaluation of<br>medical devices -- Part 10: Tests for irritation and sensitization | Skin Irritation, intra-<br>cutaneous reactivity &<br>sensitization | 15 |
| ANSI/AAMI PB70:2003, Liquid Barrier Performance and<br>Classification of protective apparel and drapes intended for use<br>in health care facilities | Barrier Performance | 18 |
| AATCC Test Method 127-2007&2008 Hydrostatic Pressure:<br>Option 2 | Water Resistance | 18 |
| AATCC Test Method 42-2007 Impact Penetration: Type II | Water Resistance | 18 |
| ASTM-F1671:2007 Standard Test Method for Resistance of<br>Materials Used in Protective Clothing to Penetration by Blood-<br>Borne Pathogens | Resistance to Penetration | 18 |
| ASTM-F1670:2008 Standard Test Method for Resistance of<br>Materials Used in Protective Clothing to Penetration by<br>Synthetic Blood | Resistance to Penetration | 18 |
| ASTM-D5034:1995 Standard Test Method for Breaking<br>Strength and Elongation of Textile Fabrics (Grab Test) | Tensile Strength | 18 |
| ASTM-D5733:1999 Standard Test Method for Tearing Strength<br>of Nonwoven Fabrics by Trapezoid Procedure | Tearing Strength | 18 |
| ASTM-D1683:2007 Standard Test Method for Failure in Sewn<br>Seams of Woven Apparel Fabrics | Seam Strength Test | 18 |
| CPSC CS-191-53 Flammability Test Method (16 CFR 1610)<br>Standard for Flammability of Clothing Textiles | Flammability | 18 |
#### Test Data Provided in this Submission
The predicate, GRI MEDICAL Surgical Gowns (K102652) met acceptance criteria for bench testing including biocompatibility, hydrostatic pressure, impact penetration, Resistance to Penetration by Synthetic Blood, flammability, and strength via tensile, tear and seam strength.
#### 10. Conclusion:
The information in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to devices already in commercial distribution. Equivalence is is demonstrated through intended use, materials, design and testing methods.
11. Similarities/Differences of the proposed device when compared to the predicate: The data within this submission demonstrates that there are no significant differences between the application device and the predicate, indicating that the application device is safe, effective and substantially equivalent for marketing in the U.S.
Jiangsu Guangda Medical Material Co., Ltd. Document Number K121152 S3.1
Page 5 of 7
P. 3 of 3
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
#### December 12, 2012
Mr. Darren Reeves Jiangsu Guangda Medical Material Company, Limited 7305 Hancock Village Drive, Suite 109 CHESTERFIELD VA 23832
Re: K121152
Trade/Device Name: Jiangsu Guangda Surgical Gown, Model Number GD-SG-01 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA Dated: November 30, 2012 Received: December 4, 2012
Dear Mr. Reeves:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave a reward your and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to conninered pror to May 20, 1977, in accordance with the provisions of the Federal Food, Drug, de necs that have been require approval of a premarket approval application (PMA). and Cosmetic Act (110.) that to neviee, subject to the general controls provisions of the Act. The 1 ou may, therefore, market the act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability addition. I lease noter a, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it IT your device is blassified (been to be subjecting major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Reeves
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthong Dr. ant
Anthony D. Watson, B.A., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the FDA logo. The logo is white and is set against a black background. The letters FDA are stylized with a double-line effect.
Image /page/5/Picture/2 description: The image shows the logo for the Department of Health and Human Services. The logo includes an emblem on the left and the text "Department of Health and Human Services" on the right. The emblem appears to be a stylized representation of an eagle or other bird.
**CENTER FOR DEVICES AND RADIOLOGICAL HEALTH**
K121157
# Indications for Use
510(k) Number (if known): K121152
Device Name: Jiangsu Guangda Surgical Gown, Model Number GD-SG-01
Indications for Use:
Jiangsu Guangda's Reinforced Surgical Gowns are non-sterile, single use surgical gowns intended to protect surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material.
This product may be sterilized using Ethylene Oxide (EO) following the validation and routine control under ANSI/AAMI/ISO 11135.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth F. Claverie
2012.12.11 18:08:11 -05'00'
General Hospital Anesthesfolog Infection Control, Denta
**510(k) Number:** K121152
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.