← Product Code [MQL](/productcode/MQL) · K121128

# MULTIPURPOSE HANDLING MEDIUM (K121128)

_Irvine Scientific Sales Co., Inc. · MQL · Jun 6, 2012 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K121128

## Device Facts

- **Applicant:** Irvine Scientific Sales Co., Inc.
- **Product Code:** [MQL](/productcode/MQL.md)
- **Decision Date:** Jun 6, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Multipurpose Handling Medium™ (MHM™) is intended for use in assisted reproductive procedures which involve the manipulation of gametes or embryos. Specifically, Multipurpose Handling Medium™ (MHM™) is indicated for use as an oocyte retrieval medium during ovarian follicle aspiration procedures (not for flushing ovarian follicles), washing sperm prior to IVF and ICSI fertilization procedures, and for transport of the embryo to the uterus during embryo transfer procedures.

## Device Story

Multipurpose Handling Medium™ (MHM™) is a ready-to-use liquid medium for assisted reproductive technology (ART) procedures. It serves as an oocyte retrieval medium, sperm-washing medium, and embryo transport medium. The medium utilizes a HEPES, MOPS, and Sodium Bicarbonate buffering system, allowing for pH maintenance in ambient atmosphere without a CO2 incubator. It is used by clinicians in a laboratory or clinical setting during IVF/ICSI procedures. The device supports gametes and embryos during manipulation, washing, and transport, facilitating successful fertilization and embryo transfer. It is supplied in 100 mL and 500 mL bottles.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing: 1-cell Mouse Embryo Assay (MEA) and Human Sperm Survival Assay (HSSA). MEA results showed ≥ 80% development. HSSA demonstrated ≥ 70% of original motility at 24 hours. Testing confirmed the device is non-toxic and functional for its intended use, with performance comparable to the predicate mHTF.

## Technological Characteristics

Liquid medium containing HEPES, MOPS, and Sodium Bicarbonate buffering system. Ingredients include WFI water, salts (NaCl, KCl, CaCl2), amino acids, pyruvate, lactate, and gentamicin sulfate. Formulated for ambient atmosphere use. Sterility, endotoxin (≤ 0.25 EU/mL), pH (7.25-7.54), and osmolality (275-295 Osm/KgH2O) specifications are controlled. Non-pyrogenic, sterile-filtered.

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- Modified HTF (mHTF) ([K983586](/device/K983586.md))
- G-MOPS™/G-MOPS™ PLUS ([K081115](/device/K081115.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K121128

### JUN - 6 2012

#### 510(k) SUMMARY AS REQUIRED BY 21 CR § 807.87(h)

Submitter's Name and Address:

Manufacturing Site:

Contact Person:

Irvine Scientific Sales Co., Inc. 2511 Daimler Street Santa Ana, CA 92705-5588

2511 Daimler Street Santa Ana, CA 92705-5588

Jayme Yamaquchi-Owens Requlatory Affairs Manager Irvine Scientific 2511 Daimler Street Santa Ana, CA 92705-5588 Tel: 949-261-7800 ext 254 949-261-6522 Fax: Email: jfy@irvinesci.com

Date of Application:

April 9, 2012

Establishment Registration Number: 2022379

510(k):

K121128

Trade or Proprietary Name:

Device Name:

Device Classification:

Device Regulation:

Device Classification:

Product Code:

Predicate Device:

Performance Standards:

R121120

Multipurpose Handling Medium™ (MHM TM)

Handling Medium

Reproductive Media and Supplements

21 CFR § 884.6180

Class II (Special Controls)

85 MQL

Modified HTF (mHTF), K983586 G-MOPS™/G-MOPS™ PLUS, K081115

None established

{1}------------------------------------------------

#### Purpose:

Purpose of this application is to obtain 510(k) marketing clearance for the Multipurpose Handling Medium™ (MHM™).

#### Indication for Use:

Multipurpose Handling Medium™ (MHM™) is intended for use in assisted reproductive procedures which involve the manipulation of gametes or embryos. Specifically, Multipurpose Handling Medium™) is indicated for use as an oocyte retrieval medium during ovarian follicle aspiration procedures (not for flushing ovarian follicles), washing sperm prior to IVF and ICSI fertilization procedures, and for transport of the embryo to the uterus during embryo transfer procedures.

#### Description of the Device:

The Multipurpose Handling Medium™ (MHM™) is based upon the mHTF (K983586) formulation. The formula for the Multipurpose Handling Medium™ (MHM™) is based upon the composition of modified HTF, K983586. The medium uses a buffering system composed of HEPES (N-2-Hydroxyethylpiperazine-N1 - - 2 - - ethanesulfonic acid). MOPS . (3 Morpholinopropane - 1- sulfonic acid) and Sodium Bicarbonate. This buffering system provides pH maintenance over the physiologic range and does not require the use of a CO2 Incubator. - The product also contains 10 ug/mL gentamicin.

Multipurpose Handling Medium™ (MHM™) has utility as an oocyte retrieval medium, in procedures that aspirate oocytes from the patient's ovarian follicles. Multipurpose Handling Medium™ (MHM™) is used only in the oocyte collection vessel and is not for use in flushing oocytes from ovarian follicles. Once the oocyte has been retrieved, it is placed into a culture dish with an appropriate amount of a culture medium (does not include MHM™), as specified by the user's internal procedures, and fertilization is allowed to occur. Multipurpose Handling Medium™ (MHM™) can also be used to wash sperm in preparation for

{2}------------------------------------------------

fertilization by conventional IVF or ICSI procedures. The fertilized oocyte (zygote) is allowed to grow in the culture dish, supported by a culture medium (does not include MHM™), as specified by the user's internal procedures, and an appropriate protein supplement, in a carbon dioxide incubator at 37°C until the desired stage of development is achieved. The embryo is removed from the culture dish, placed into an appropriate amount of Multipurpose Handling Medium™ (MHM™) for transfer into the uterus.

Multipurpose Handling Medium™ (MHM™) is supplied as a ready to use liquid, in 100 and 500 mL bottles.

### Technological Characteristics:

Multipurpose Handling Medium™ (MHM™) can be used as an oocyte collection medium during in procedures that aspirate oocytes from the patient's ovarian follicles. Multipurpose Handling Medium™ (MHM™) is used only in the oocyte collection vessel and is not for use in flushing oocytes from ovarian follicles. Once the oocyte has been retrieved, it is placed into a culture dish with an appropriate amount of a culture medium, as specified by the user's internal procedures, and fertilized. After fertilization, the embryo is allowed to develop in a culture medium, as specified by the user's internal procedures, until an appropriate developmental stage is reached. After incubation of the embryo, to the appropriate developmental stage, the embryo is removed from the incubation dish and placed into a suitable amount of Multipurpose Handling Medium™ (MHM™) for transport and transfer into the patient. Multipurpose Handling Medium™ (MHM™) can be used for washing sperm from its surrounding seminal fluid. Once semen has liquefied it is placed into a conical tube with a volume of room temperature Handling Medium™ (MHM™) and centrifuged to separate the sperm from the seminal fluid prior to IVF and ICSI fertilization procedures. Multipurpose Handling Medium™ (MHM™) is therefore intended for use as an oocyte retrieval medium, as a transport and storage medium and as a spermwashing medium.

{3}------------------------------------------------

Multipurpose Handling Medium™ (MHM™) is similar to mHTF (K983586) it is intended for use in assisted reproductive technology procedures that involve the manipulation of gametes or embryos. mHTF is also intended for use as a spermprocessing medium in washing procedures, as an oocyte retrieval medium, for transport of the embryo, and as a support medium for transfer of the embryo. mHTF is intended to simulate the substances found in the human, female reproductive system.

Multipurpose Handling Medium™ (MHM™) is similar to G-MOPS and G-MOPS" PLUS (K081115) in that is intended for oocyte collection and for handling and manipulating oocytes and embryos in ambient atmosphere.

#### Performance Data:

Multipurpose Handling Medium™ is assayed by one (1) -cell mouse embryo assay (MEA) prior to release to the market. This assay assures that the product is both functional for its intended use and no embryo-toxic components are present in the formulation.

Multipurpose Handling Medium™ (MHM™) is also assayed human sperm survival assay was also performed prior to release to the market. This assay also assures that the product functions in accordance with its intended use with regards to sperm wash procedure and that no sperm-toxic components are present in the formulation.

#### Nonclinical Tests:

One (1) -cell MEA was performed on the Multipurpose Handling Medium™ (MHM™) as part of the design validation. Testing was performed at three (3) different test facilities after 2 hours exposure to the Multipurpose Handling Medium™ (MHM™).

In addition, human sperm survival assay was also performed on donor specimens that were initially processed by gradient separation and resulting motile specimens were equally divided and the average % motility after 24 hours 

{4}------------------------------------------------

at 37°C in ambient air was determined. The field evaluations demonstrate that the Multipurpose Handling Medium™ (MHM™) was "equal" to the proven control medium of mHTF (K983586).

#### Additional Information:

Endotoxin, pH, osmolality, mouse embryo assay, human sperm survival assay and sterility tests will be performed as a condition of release for Multipurpose Handling Medium™ (MHM™). Results of all release assays performed will be reported on a lot-specific certificate of analysis, and will be indicated on the labeling.

#### Conclusion:

The conclusion from the performance testing as well as the nonclinical and clinical data demonstrates that the Multipurpose Handling Medium™ (MHM™) is suitable for its intended use, and meets the criteria in the comparison to predicate devices in which substantial equivalence has been demonstrated and meets the criteria outlined in the Notice of Final Rule, 63 FR 48428, Docket Number 97N-0335.

A comparison of the Predicate Device Intended Use, Product Formula and Product Specifications are summarized below that demonstrate substantial equivalence.

| Product Name                             | K#      | Retrieval /<br>Collection | Culture | Transfer | Ambient |
|------------------------------------------|---------|---------------------------|---------|----------|---------|
| Modified HTF<br>(mHTF) Medium –<br>HEPES | K983586 | +                         | +       | +        | +       |
| G-MOPS™/G-<br>MOPS™ PLUS                 | K081115 | +                         | +       | +        | +       |

#### Table 1: Predicate Intended Use Comparison

{5}------------------------------------------------

| Raw Materials                                                                                  | Multipurpose<br>Handling<br>Medium™<br>Catalog #90163   | Modified HTF<br>(K983586) | G-MOPS™<br>(K081115)  | G-MOPS™<br>PLUS<br>(K081115)  |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------|---------------------------|-----------------------|-------------------------------|
| WFI Quality Water                                                                              | +                                                       | +                         | +                     | +                             |
| Phenol Red, Na Salt                                                                            | +                                                       | +                         | -                     | -                             |
| Sodium Bicarbonate                                                                             | +                                                       | +                         | +                     | +                             |
| Sodium Chloride                                                                                | +                                                       | +                         | +                     | +                             |
| Potassium Chloride                                                                             | +                                                       | +                         | +                     | +                             |
| Dextrose                                                                                       | +                                                       | +                         | +                     | +                             |
| Magnesium Sulfate                                                                              | +                                                       | +                         | +                     | +                             |
| Potassium Phosphate                                                                            | +                                                       | +                         | -                     | -                             |
| Pyruvic Acid, sodium<br>salt                                                                   | +                                                       | +                         | -                     | -                             |
| DL-Lactic Acid, sodium                                                                         | +                                                       | +                         | +                     | +                             |
| Glycine                                                                                        | +                                                       | -                         | +                     | +                             |
| Taurine                                                                                        | +                                                       | -                         | +                     | +                             |
| Calcium Chloride,<br>anhydrous                                                                 | +                                                       | +                         | +                     | +                             |
| HEPES ½ sodium salt<br>[4-(2-hydroxyethyl)-1-<br>piperazineethanesulfonic acid, ½ sodium salt] | +                                                       | +                         | -                     | -                             |
| HEPES Sodium salt<br>[4-(2-hydroxyethyl)-1-<br>piperazineethanesulfonic acid, sodium salt]     | -                                                       | +                         | -                     | -                             |
| MOPS, sodium salt                                                                              | +                                                       | -                         | +                     | +                             |
| Gentamicin Sulfate                                                                             | +                                                       | +                         | +                     | +                             |
| Alanine                                                                                        | +                                                       | +                         | +                     | +                             |
| Alanyl-glutamine                                                                               | +                                                       | +                         | +                     | +                             |
| Asparagine                                                                                     | +                                                       | +                         | +                     | +                             |
| Aspartate                                                                                      | +                                                       | +                         | +                     | +                             |
| Raw Materials                                                                                  | Multipurpose<br>Handling<br>MediumTM,<br>Catalog #90163 | Modified HTF<br>(K983586) | G-MOPSTM<br>(K081115) | G-MOPSTM<br>PLUS<br>(K081115) |
| Glutamate                                                                                      | +                                                       | +                         | +                     | +                             |
| Glycine                                                                                        | +                                                       | +                         | +                     | +                             |
| Lipoic Acid                                                                                    | +                                                       | +                         | +                     | +                             |
| Methionine                                                                                     | +                                                       | +                         | +                     | +                             |
| Proline                                                                                        | +                                                       | +                         | +                     | +                             |
| Serine                                                                                         | +                                                       | +                         | +                     | +                             |
| Sodium citrate                                                                                 | +                                                       | +                         | +                     | +                             |
| Sodium Hydrogen<br>Phosphate                                                                   | +                                                       | +                         | +                     | +                             |
| Sodium Pyruvate                                                                                | +                                                       | +                         | +                     | +                             |
| Human serum albumin                                                                            | -1                                                      | +                         | +                     | +                             |

# Table 2: Product Formulation Comparison

.

·

:

·

{6}------------------------------------------------

Table 3: Predicate Device Product Specification Comparison

| Specification | Multipurpose<br>Handling Medium™          | Modified HTF<br>(K983586) | G-MOPS™/<br>G-MOPS™ PLUS<br>(K081115) |
|---------------|-------------------------------------------|---------------------------|---------------------------------------|
| pH            | 7.25 - 7.54                               | 7.25 - 7.54               | Unknown                               |
| Osmolality    | 275 - 295 Osm/KgH2O                       | 272 - 288 Osm/KgH2O       | Unknown                               |
| Sterility     | Pass                                      | Pass                      | Pass                                  |
| Endotoxin     | ≤ 0.25 EU/mL                              | ≤ 0.25 EU/mL              | ≤ 0.25 EU/mL                          |
| MEA           | ≥ 80%                                     | ≥ 70%                     | ≥ 80%                                 |
| HSSA          | ≥ 70% of original<br>motility at 24 hours | NT2                       | NT                                    |

.

1 Protein supplementation is to be performed prior to use

.

² NT — Not Tested

{7}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Jayme Yamaguchi-Owens Regulatory Affairs Manager Irvine Scientific Sales Co., Inc. 2511 Daimler Street SANTA ANA CA 92705

JUN - 6 2012

Re: K121128

> Trade/Device Name: Multipurpose Handling Medium™ (MHMTM) Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MQL Dated: April 9, 2012 Received: April 13, 2012

Dear Ms. Yamaguchi-Owens:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 abo career in to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that 1 17 has interes and regulations administered by other Federal agencies. You must of any Feather statest is requirements, including, but not limited to: registration and listing eoulpry .mart 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{8}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin K. Evans

Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Irvine Scientific

.

#### INDICATIONS FOR USE STATEMENT (page 1 of 1) র্ব

# 510(K) Number: K121128

Multipurpose Handling Medium™ (MHM™) Device Name:

#### Indications for Use:

Multipurpose Handling Medium™ (MHM™) is intended for use in assisted reproductive procedures which involve the manipulation of gametes or embryos. Specifically, Multipurpose Handling Medium™ (MHM™) is indicated for use as an oocyte retrieval medium during ovarian follicle aspiration procedures (not for flushing ovarian follicles), washing sperm prior to IVF and ICSI fertilization procedures, and for transport of the embryo to the uterus during embryo transfer procedures.

Prescription Use (Part 21 CFR § 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR § 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of _

Benjamin R. Folds 06 June 2012
(Division Sign-Off)

astro-Renal, and

---

**Source:** [https://fda.innolitics.com/device/K121128](https://fda.innolitics.com/device/K121128)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
