← Product Code [DIF](/productcode/DIF) · K121122

# DETECTABUSE LIQUID CONTROL, DETECTABUSE STAT-SKREEN LIQUID CONTROL (K121122)

_Biochemical Diagnostic, Inc. · DIF · Jul 3, 2012 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K121122

## Device Facts

- **Applicant:** Biochemical Diagnostic, Inc.
- **Product Code:** [DIF](/productcode/DIF.md)
- **Decision Date:** Jul 3, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3280
- **Device Class:** Class 1
- **Review Panel:** Clinical Toxicology

## Indications for Use

The Detectabuse® Liquid control is intended for use as quality control urine to monitor the precision of laboratory urine toxicology testing procedures for the analytes listed in the package insert. For prescription use only

## Device Story

Detectabuse® Liquid Control is a stabilized human-based urine product used as a quality control material in clinical toxicology laboratories. Positive controls are gravimetrically spiked with reference drug standards and metabolites; negative controls are certified via immunoassay, GC/MS, and/or LC/MS. The product contains sodium azide or a proprietary preservative. It is used by laboratory personnel to monitor the precision of urine toxicology testing procedures. The device provides known concentrations of analytes, allowing clinicians to verify the accuracy and reliability of their testing platforms. It does not perform diagnostic tests itself but serves as a reference material to ensure the validity of laboratory results.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Stabilized human-based urine matrix. Preservatives: sodium azide (<1%) or proprietary alternative. Storage: -10 to -20°C or 2-8°C. Analytes include drugs of abuse, metabolites, creatinine, pH, and specific gravity. No electronic or software components.

## Regulatory Identification

A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE

A. 510(k) Number:
k121122

B. Purpose for Submission:
New Device

C. Analyte:
Quality control materials for Delta-9-THC-COOH, Benzoylecgonine, Phencyclidine (PCP), Codeine, Propoxyphene, Methaqualone, Morphine -3-glucuronide, 6-Monoacetylmorphine, Morphine, d-Amphetamine, d-Methamphetamine, Secobarbital, Butalbital, Phenobarbital, Oxazepam, Methadone, MDMA, MDA, MDEA, Oxycodone, Buprenorphine, EDDP, Nordiazepam, Norpropoxyphene, Cotinine, Fentanyl, Ethanol, Nortriptyline, Creatinine, pH, Specific Gravity

D. Type of Test:
Not applicable

E. Applicant:
Biochemical Diagnostics, Inc.

F. Proprietary and Established Names:
Detectabuse® Liquid Control

G. Regulatory Information:
1. Regulation section:
21 CFR 862.3280, Clinical Toxicology Control Material
2. Classification:
Class I, reserved
3. Product Code:
DIF
4. Panel:
Toxicology (91)

H. Intended Use:
1. Intended use(s):
See indications for use below
2. Indication(s) for use:
The Detectabuse® Liquid control is intended for use as quality control urine to monitor the precision of laboratory urine toxicology testing procedures for the analytes listed in the package insert.

{1}

3. Special condition for use statement(s):
For prescription use only

4. Special instrument Requirements:
Not applicable.

I. Device Description:

Detectabuse® Liquid Control

Each bottle contains stabilized human based urine. Positive control urines have been gravimetrically spiked with reference drug standards and/or appropriate metabolites. Negative control urines are certified negative by combination of immunoassay, GC/MS and/or LC/MS for the constituents listed on our target sheets. The products contain less than 1% sodium azide as a preservative. For assays sensitive to sodium azide such as ELISA we substitute a proprietary preservative approved by the manufacturers and DEA.

J. Substantial Equivalence Information:

1. Predicate device name(s):
Detectabuse® Liquid Control Urine

2. Predicate K number(s):
k925586

3. Comparison with predicates:

|  Device | Predicate Device (K925586) | Modified Device (K121122)  |
| --- | --- | --- |
|  Intended Use | Same | Is intended for use as quality control urine to monitor the precision of laboratory urine toxicology testing procedures  |
|  Form | Same | Liquid  |
|  Matrix | Same | Human Urine  |
|  Storage | Closed vial: The controls are stable until the expiration date when stored at -10 to -20°C and protected from light.
The controls are stable until the expiration date when stored at 2-8°C.
Open vial: not available | Closed vial: The controls are stable until (1) 36 months when stored at 2-8°C with the exception of oxazepam; (2) 48 months when stored at -10 to -20°C and protected from light with the exception of oxazepam.
Open vial: The controls are stable 30 days when stored at 2-8°C.  |

{2}

3

|  Analytes | Delta-9-THC-COOH, Benzoylecgonine, Phencyclidine (PCP), Codeine, Propoxyphene
Methaqualone, Morphine -3-glucuronide, 6-Monoacetylmorphine, Morphine, d-Amphetamine, d-Methamphetamine, Secobarbital, Butalbital, Phenobarbital, Oxazepam, Methadone | Same analytes as predicate device A with the additional claims:
MDMA, MDA, MDEA, Oxycodone, Buprenorphine, EDDP, Nordiazepam, Norpropoxyphene, Cotinine, Fentanyl, Ethanol, Nortriptyline, Creatinine, pH, Specific Gravity  |
| --- | --- | --- |

K. Standard/Guidance Document Referenced (if applicable):
None were referenced.

L. Test Principle:
Not applicable.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability (controls, calibrators, or method):
a) Traceability
The controls are gravimetrically spiked with reference standards purchased from commercial vendors. The drug values printed within the package insert are validated by using GC/MS, LC/MS, HPLC, or Immunoassay screening.
b.) Stability
Stability studies have been performed to determine the open vial stability and shelf life stability of the Detectabuse® Liquid control.
Product claims are as follows: open vial stability is 30 days when stored tightly capped at 2-8°C; closed vial stability is 10 months at 2-8°C and 4 years at -10°C to -20°C.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.

{3}

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.

3. Clinical studies:
a. Clinical sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a and b are not applicable):
Not available.

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
Not applicable.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports substantial equivalence decision.

4

---

**Source:** [https://fda.innolitics.com/device/K121122](https://fda.innolitics.com/device/K121122)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
