← Product Code [OZT](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/OZT) · K121083

# TRUE DILATATION PERCUTANEOUS TRANSLUMINAL VALVULOPLASTY CATHETER, 20MM X 4.5CM, TRUE DILATATION PERCUTANEOUS TRANSLUMINA (K121083)

_Loma Vista Medical · OZT · Oct 11, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/OZT/K121083

## Device Facts

- **Applicant:** Loma Vista Medical
- **Product Code:** [OZT](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/OZT.md)
- **Decision Date:** Oct 11, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1255
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K121083 · Oct 11, 2012 | TRUE DILATATION PERCUTANEOUS TRANSLUMINAL VALVULOPLASTY CATHETER, 20MM X 4.5CM, TRUE DILATATION PERCUTANEOUS TRANSLUMINA | Loma Vista Medical | European clinical study records; Post-market surveillance data | Clinical data from 69 procedures and post-market experience from over 400 procedures were used to demonstrate safety, effectiveness, and substantial equivalence to the predicate device. | European clinical studies; Post-market surveillance; Clinical outcomes; Procedural data |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| European clinical studies; Clinical study (observational/routine care) | Patients undergoing balloon aortic valvuloplasty (BAV); Sample Size: 69 procedures | Predicate device (NuCLEUS) | Device performance, procedural data, clinical outcomes, adverse events |
| Post-market surveillance; Post-market surveillance; Study Period: Since European commercial launch | Patients undergoing BAV procedures; Sample Size: >400 procedures | Not applicable for this study | Adverse events |

## Indications for Use

The TRUE Dilatation™ Catheter is indicated for balloon aortic valvuloplasty.

## Device Story

TRUE Dilatation™ Balloon Valvuloplasty Catheter is a coaxial catheter with a fixed balloon at the tip; used for balloon aortic valvuloplasty (BAV). Input includes radiopaque contrast media delivered via syringe inflation device to inflate the balloon; guidewire used for positioning. Device operated by physicians in clinical settings. Inflation/deflation of balloon dilates the aortic valve; facilitates pre-dilatation prior to TAVI. Output is mechanical dilation of the valve. Benefits include successful valve dilation and preparation for TAVI procedures. Safety and performance confirmed via bench testing and clinical experience.

## Clinical Evidence

Two European clinical studies involving 69 procedures were conducted. Data included device performance, procedural outcomes, and clinical safety. Results showed 0/69 adverse events. Clinical experience confirmed successful use for pre-dilatation prior to TAVI. Post-market data from over 400 procedures reported no device failures or device-related adverse events, supporting substantial equivalence to the predicate.

## Technological Characteristics

Coaxial catheter design with two lumens (inflation/deflation and guidewire). Materials tested per ISO 10993-1. Mechanical performance verified per ISO 10555-1. Sterilization per ISO 11135-1. Packaging integrity per ASTM D4169. Features include angled luer-lock hub for inflation and straight luer-lock hub for guidewire. Rated for 6 atm maximum inflation pressure; body withstands 30 ATM.

## Regulatory Identification

A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components.
(3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal.
(4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis.
(5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.

In combination with the general controls of the FD&C Act, the NuMED NuCLEUS and NuCLEUS-X BAV Catheters are subject to the following special controls:

## Predicate Devices

- NuCLEUS ([K082776](/device/K082776.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K121083

# OCT 1 1 202

| Revised 510(k) Summary            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                        | Loma Vista Medical<br>863A Mitten Road, Suite 100A<br>Burlingame, CA 94010<br>Phone: (650) 490-4747<br>Fax: (650) 240-0761<br>Email: paul@lomavistamedical.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Contact Person:                   | Tiffini Diage<br>1307 South Mary Avenue, Suite 280<br>Sunnyvale, CA 94087<br>Phone: (707) 799-6732<br>Fax: (408) 462-9132<br>Email: tdiage@namsa.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Date Prepared:                    | 8/27/2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name:                       | TRUE Dilatation™ Balloon Valvuloplasty Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification:                   | Class II<br>Balloon Aortic Valvuloplasty<br>21 CFR 870.1255                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code:                     | OZT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate Device(s):              | The subject device is equivalent to the following devices:<br>NuCLEUS, 510(k) number K082776                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Description:               | The Loma Vista Medical (LVM) TRUE Dilatation™ Catheter is<br>a coaxial catheter with a balloon fixed at the tip used for<br>Balloon Aortic Valvuloplasty (BAV) of the aortic valve. The<br>effective length of the catheter is 110 cm and it has two lumens:<br>one lumen is used to inflate and deflate the balloon and the<br>other permits the use of a guidewire to position the catheter. The<br>balloon inflation luer-lock hub (angled) connects to a syringe<br>inflation device to deliver radiopaque contrast media for<br>inflation. The guidewire luer-lock hub (straight) connects to the<br>guidewire lumen.                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indication for Use:               | The TRUE Dilatation™ Catheter is indicated for balloon aortic<br>valvuloplasty.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Functional and<br>Safety Testing: | To verify that the device design met its functional and<br>performance requirements, representative samples of the device<br>underwent biocompatibility, sterility, packaging integrity, and<br>mechanical testing in accordance with ISO 10993-1 2009, ISO<br>11135-1 2007, ASTM D4169:2009, ISO 10555-1 2009.<br>Additional performance bench testing was performed and is<br>summarized below.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| In Vivo Performance<br>Testing    | Two separate European clinical studies were performed to<br>confirm bench test data, to ensure design features continue to                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                   | meet clinical needs, and to provide in vivo data to demonstrate<br>safe and effective use of the TRUE Dilatation Catheter as to the<br>predicate device. Data collected included device performance,<br>procedural data, and clinical outcomes. Data was collected for<br>69 devices used in 69 clinical procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                   | The data provides objective evidence that the TRUE Dilatation<br>Catheter can be used safely and effectively in BAV procedures<br>compared to the predicate device. In addition, the clinical<br>experience confirms that the TRUE Dilatation Catheter can be<br>successfully used for pre-dilatation prior to transcatheter aortic<br>valve implantation, in comparison to the predicate. The success<br>of the TAVI procedure confirms that the TRUE Dilatation<br>Catheter as effective as the predicate in BAV procedures and<br>allows for successful placement of TAVI devices. The absence<br>of any adverse events (0/69) indicates that the device is safe as<br>compared to the predicate. In addition, there have been no<br>reports of any adverse events since the European commercial<br>launch representing more than 400 TRUE procedures to date.<br>The absence of device failures and device related adverse events<br>demonstrates the equivalency of the TRUE Dilatation Catheter<br>to the predicate NuCLEUS device. |
| Conclusion:                       | Loma Vista Medical considers the TRUE Dilatation Catheter to<br>be substantially equivalent to the predicate device listed above.<br>This conclusion is based upon the devices' similarities in<br>principles of operation, performance requirements, and<br>indications for use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

.

{1}------------------------------------------------

・

.

- - - -

.

.

| Test Performed                | TRUE Dilatation Acceptance Criteria                                                                                                                                              | TRUE Dilatation Results    |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Visual Inspection             | When examined with 2.5X magnification, the external surface of the effective length of the catheter shall appear free from extraneous material, and process and surface defects. | 61 of 61 catheters passed. |
| Balloon<br>Preparation Test   | Each catheter shall be prepped per the procedure without functional difficulties or anomalies.                                                                                   | 61 of 61 catheters passed. |
| Diameter and<br>Profile Test  | Balloon diameter shall not deviate from stated balloon diameter by more than 1mm when inflated to 6 atm.                                                                         | 61 of 61 catheters passed. |
| Balloon<br>Distensibility     | Inflated balloon diameter shall not increase more than 2% between nominal inflation pressure (3 atm) and maximum inflation pressure (6 atm).                                     | 61 of 61 catheters passed. |
| Repeated Balloon<br>Inflation | The balloon catheter shall not herniate or rupture after 10 inflation/deflation cycles to the maximum inflation pressure of 6 atm.                                               | 61 of 61 catheters passed. |

.

:

・

{2}------------------------------------------------

| Test Performed                            | TRUE Dilatation Acceptance Criteria                                                                                                                                                                                               | TRUE Dilatation Results                                                                                      |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Balloon Minimum<br>Burst Strength         | The results must show statistically that with<br>at least 95% confidence,<br>99.9% of the balloons will not burst at or<br>below the maximum recommended rated<br>burst pressure. All sizes must withstand a<br>minimum of 6 atm. | A total of 98 samples across all<br>device sizes tested and passed 95%<br>confidence with 99.9% reliability. |
| Balloon Inflation /<br>Deflation Test     | The balloon shall inflate in no more than 5<br>seconds and deflate in no more than 10<br>seconds.                                                                                                                                 | 60 of 60 catheters met acceptance<br>criteria.                                                               |
| Balloon<br>Deflatability Test             | There should be no interference with balloon<br>deflation.                                                                                                                                                                        | As evidenced by observed deflation,<br>60 of 60 samples passed.                                              |
| Tip Pull and<br>Torque Test               | Must withstand at least 10 turns without<br>breaking.                                                                                                                                                                             | 20 of 20 catheters passed. No<br>reported breaks. There is only 1<br>shaft size.                             |
| Bond Strength Test                        | All bonds must withstand at least 15N (3.4<br>lbs) pull strength.                                                                                                                                                                 | 29 of 29 samples passed.                                                                                     |
| Catheter Body<br>Maximum Pressure<br>Test | All samples must withstand 30<br>ATM (400 psi).                                                                                                                                                                                   | 20 of 20 catheters withstood 30<br>ATM.                                                                      |

:

.

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined serpents. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus. The seal is black and white.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

OCT 11 2012

Loma Vista Medical c/o Ms. Tiffini Diage Regulatory Project Manager, NAMSA 1307 South Mary Avenue, Suite 280 Sunnyvale, CA 94807

Re: K121083

Trade Name: TRUE Dilatation™ Balloon Valvuloplasty Catheter Regulation Number: 21 CFR 870.1255 Regulation Name: Balloon Aortic Valvuloplasty Catheter Regulatory Class: Class II (two) Product Code: OZT Dated: August 24, 2012 Received: August 28, 2012

Dear Ms. Diage:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

### Page 2 - Ms. Tiffini Diage

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

# Indications for Use

# 510(k) Number (if known): K121083

Device Name: TRUE Dilatation Balloon Valvuloplasty Catheter

Indications For Use:

The TRUE Dilatation Balloon Valvuloplasty Catheter is indicated for balloon aortic valvuloplasty.

Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

. 2

(Division Rign-Off) Division of Cordiovascular Devices 510(k) Mumber .. K12) 083

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/OZT/K121083](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/OZT/K121083)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
