K120986 · Focus Diagnostics, Inc. · OYP · Aug 16, 2012 · Microbiology
Device Facts
Record ID
K120986
Device Name
STRATIFY JCV DXSELECT
Applicant
Focus Diagnostics, Inc.
Product Code
OYP · Microbiology
Decision Date
Aug 16, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3336
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K120986 · Aug 16, 2012
STRATIFY JCV DXSELECT
Focus Diagnostics, Inc.
Archived clinical trial specimens (AFFIRM and STRATIFY-2 studies); Retrospective serum samples from confirmed PML patients
The device performance was assessed using archived samples from MS patients enrolled in clinical trials and retrospective samples from patients who developed PML. This data was used to demonstrate clinical agreement with a predicate assay and to support the device's role in PML risk stratification.
Pre-PML Sample Analysis; Retrospective analysis; Follow-up/Duration: Samples collected at least 6 months prior to clinical diagnosis
Confirmed PML patients; Sample Size: 31; Number of Sites: 1
Validated laboratory methodology
Positive agreement with validated assay
AFFIRM and STRATIFY-2 Archived Samples; Retrospective analysis of archived clinical trial samples
Multiple Sclerosis (MS) patients; Sample Size: 1330; Number of Sites: 3 (2 external, 1 internal)
Cleared Anti-JCV assay
Positive and Negative Percent Agreement
Indications for Use
The Focus Diagnostics’ STRATIFY JCV® DxSelect™ assay is intended for the qualitative detection of antibodies to JC virus in human serum or plasma. The assay is intended for use in conjunction with other clinical data, in multiple sclerosis patients receiving or considering natalizumab therapy, as an aid in risk stratification for progressive multifocal leukoencephalopathy development. The assay is for professional use only. The assay is not intended for donor screening. The performance of this assay has not been established for use in other immunocompromised patient populations or patients with different disease conditions or undergoing other treatments or in neonates and pediatric patient populations.
Device Story
ELISA-based assay for qualitative detection of anti-JCV antibodies in human serum or plasma. Uses 96-well microtiter plates pre-coated with JC virus-like particles (VLP). Patient samples incubated; JCV-specific antibodies bind to VLP. Peroxidase-conjugated anti-human antibodies added; enzyme substrate/chromogen produce color change quantified by spectrophotometric optical density (OD). Results reported as index values compared to cut-off calibrators. Indeterminate results undergo confirmation assay where soluble JCV VLP competes for antibodies. Used in clinical laboratories by professional staff. Output aids clinicians in assessing PML risk for MS patients on natalizumab therapy, alongside other risk factors (treatment duration, prior immunosuppressant use).
Clinical Evidence
Clinical performance evaluated using 31 pre-PML serum samples and 1,330 MS patient samples. Pre-PML samples showed 100% (31/31) positive agreement with validated methodology. Comparative agreement studies in natalizumab-treated patients showed 97.0% PPA and 90.6% NPA; in patients considering natalizumab, 98.5% PPA and 91.8% NPA. Seroprevalence in MS cohort was 55-59%.
Technological Characteristics
ELISA-based immunoassay. Uses JC virus-like particles (VLP) as antigen. Peroxidase-conjugated donkey anti-human JCV antibodies. 96-well microtiter plate format. Spectrophotometric detection (OD). Requires professional laboratory equipment. No specific ASTM material standards cited.
Indications for Use
Indicated for qualitative detection of anti-JCV antibodies in human serum or plasma for multiple sclerosis patients receiving or considering natalizumab therapy to aid in PML risk stratification. Not for donor screening, neonates, pediatric populations, or other immunocompromised patients.
Regulatory Classification
Identification
John Cunningham Virus serological reagents are devices that consist of antigens and antisera used in serological assays to identify antibodies to John Cunningham Virus in serum and plasma. The identification aids in the risk stratification for the development of progressive multifocal leukoencephalopathy in multiple sclerosis and Crohn's disease patients undergoing natalizumab therapy. These devices are for adjunctive use, in the context of other clinical risk factors for the development of progressive multifocal leukoencephalopathy.
Special Controls
The device is classified as Class II under regulation 21 CFR 866.3336 with special controls. The special control guidance document "Class II Special Controls Guidance Document: John Cunningham Virus Serological Reagents," will be available shortly.
*Classification.* Class II (special controls). The special control for this device is the FDA guideline document entitled “Class II Special Controls Guideline: John Cunningham Virus Serological Reagents.” For availability of the guideline document, see § 866.1(e).
Predicate Devices
STRATIFY JCV™ Antibody ELISA
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K120986
B. Purpose for Submission:
New device
C. Measurand:
Anti- JC Virus (JCV) antibodies
D. Type of Test:
Enzyme Linked Immunosorbent Assay (ELISA)
E. Applicant:
Focus Diagnostics Inc.
F. Proprietary and Established Names:
STRATIFY JCV® DxSelect™
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3336: John Cunningham Virus serological reagents
2. Classification:
Class II Special Controls
3. Product code:
OYP
4. Panel:
Microbiology (83)
{2}
H. Intended Use:
1. Intended use(s):
The Focus Diagnostics’ STRATIFY JCV® DxSelect™ assay is intended for the qualitative detection of antibodies to JC virus in human serum or plasma. The assay is intended for use in conjunction with other clinical data, in multiple sclerosis patients receiving or considering natalizumab therapy, as an aid in risk stratification for progressive multifocal leukoencephalopathy development. The assay is for professional use only.
The assay is not intended for donor screening. The performance of this assay has not been established for use in other immunocompromised patient populations or patients with different disease conditions or undergoing other treatments or in neonates and pediatric patient populations.
2. Indication(s) for use:
The same as intended use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
None
I. Device Description:
The Focus Diagnostics’ STRATIFY JCV® DxSelect™ consists of two devices. The first is an initial detection ELISA and the second device is a confirmatory (inhibition) ELISA. Both tests utilize the same recombinant antigen which is used in two different formats as described in section L. The devices include the following reagents; Recombinant JC virus like particles (VLP), a JCV high and low positive controls, consisting of human sera that is positive for JCV antibodies, and JCV negative control consisting of human sera that is negative for JCV antibodies. The conjugate, substrate, wash buffers, and blocking buffers needed for the test are not supplied with the device and are listed together with other reagents and consumables in the device labeling.
J. Substantial Equivalence Information:
1. Predicate device name(s):
STRATIFY JCV™ Antibody ELISA
2. Predicate 510(k) number(s):
K112394
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | The assay is intended for the qualitative detection of antibodies to JC virus in human serum or plasma. | Same. |
| Measurand | Anti-JCV antibodies | Same. |
| Technology | Detection ELISA with a secondary Confirmation ELISA for specimens that have results in the indeterminate zone levels. | Same. |
| Matrix (sample type) | Serum and Plasma | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for use | The assay is intended for use in conjunction with other clinical data, in multiple sclerosis patients receiving or considering natalizumab therapy, as an aid in risk stratification for progressive multifocal leukoencephalopathy development. The assay is for professional use only. The assay is not intended for donor screening. The performance of this assay has not been established for use in other immunocompromised patient populations or in neonates and pediatric patient populations. | The assay is intended for use in conjunction with other clinical data, in multiple sclerosis and Crohn's disease patients receiving natalizumab therapy, as an aid in risk stratification for progressive multifocal leukoencephalopathy development. The assay is for professional use only and is to be performed only at Focus Diagnostics' Reference Laboratory. The assay is not intended for donor screening. The performance of this assay has not been established for use in other immunocompromised patient populations or in neonates and pediatrics patient populations |
| Assay Cut-off points | Index < 0.2, negative | nOD < 0.1, negative |
| | Index ≥ 0.4, positive | nOD ≥ 0.2, positive |
{4}
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2; Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline 2nd Edition (2004).
CLSI EP07-A2; Interference Testing in Clinical Chemistry; Approved Guideline 2nd Edition (2005).
CLSI H18-A4; Procedures for the Handling and Processing of Blood Specimens; Approved Guideline 4th Edition (2010).
L. Test Principle:
In the Focus Diagnostics’ STRATIFY JCV® DxSelect™ test, JC virus-like particles (VLP) are pre-coated onto 96-well microtiter plates. Diluted serum or plasma specimens and controls are incubated in the wells to allow JCV-specific antibodies present in the specimens to react with the JC VLP antigen. Nonspecific reactants are removed by washing. Peroxidase-conjugated anti-human antibodies are added to react with JCV-specific antibodies. Excess conjugate is removed by washing. Enzyme substrate and chromogen are added, and the color is allowed to develop. After adding the Stop Reagent, the resultant color change is quantified by a spectrophotometric reading of optical density (OD). Specimen OD readings are compared with Cut-Off Calibrator OD readings to determine results. Each specimen result is reported as an index value. A specimen with an index value that is greater than a specified upper cut-off is reported as positive for detectable JCV-specific antibodies, whereas a specimen with an index value less than the specified lower cut-off is reported as negative for detectable JCV-specific antibodies. A specimen with an index value that is equal to or between the upper and lower cut-off values is reported as indeterminate. An indeterminate result requires further evaluation in the confirmation (inhibition) assay.
In the confirmation assay, soluble JC VLP antigen will compete with plate bound JC VLP antigen for the JCV-specific antibodies present in the serum or plasma specimens. After washing away the unbound antibodies, peroxidase-conjugated anti-human antibodies are added and bind to any captured JCV-specific antibodies. Excess conjugate is removed by washing. Enzyme substrate and chromagen are added, and the color is allowed to develop. After adding the Stop Reagent, the resultant color change is quantified by a spectrophotometric reading of OD. The percent inhibition is calculated to confirm presence of JCV-specific antibodies in the specimen. Specimens with a percent inhibition value that is greater than the specified cut-off are reported as positive for detectable JCV-specific antibodies, whereas specimens with percent inhibition values less than or equal to the cut-off are reported as negative for detectable JCV-specific antibodies.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{5}
a. Precision/Reproducibility:
Reproducibility
The reproducibility of the assay was assessed using a sample panel consisting of four levels of sera, four levels of (EDTA) plasma samples, low and high positive controls. Three replicates of each panel member were tested on two runs a day for a total of five days at two external testing sites and one internal site. The four levels included a negative, low, indeterminate and moderate positive samples prepared in a serum and a plasma matrices. The indeterminate samples were tested simultaneously in the confirmation assay and the detection assay.
Table: Reproducibility
| JCV Gen-2 Validation Reproducibility Results | | | | | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Parameter | Sample Matrix | Sample Name | Qualitative Results | | | | | Quantitative Results | | | | | | | | | | |
| | | | ND | I | D | NV | Total | N | Mean | Variability Components | | | | | | | | |
| | | | | | | | | | | Between Sites | | Between Days | | Between Runs/Operators | | Within Assay (Repeatability) | | Total |
| | | | | | | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD |
| ID | | Cut-Off | | | | | | 120 | 1.131 | 0.145 | 12.8 | 0.067 | 5.9 | 0.088 | 7.8 | 0.037 | 3.2 | 0.186 |
| INDEX | Controls | Negative Control | 89 | | | 1 | 90 | 89 | 0.102 | 0.010 | 9.5 | 0.007 | 7.1 | 0.010 | 9.7 | 0.006 | 6.1 | 0.017 |
| | | Positive Control | | | 90 | | 90 | 90 | 1.202 | 0.008 | 0.7 | 0.010 | 0.8 | 0.022 | 1.8 | 0.036 | 3.0 | 0.044 |
| | Plasma | Indeterminate | | 90 | | | 90 | 90 | 0.270 | 0.010 | 3.7 | 0.009 | 3.4 | 0.011 | 4.2 | 0.014 | 5.1 | 0.023 |
| | | Low positive | | 39 | 51 | | 90 | 90 | 0.410 | 0.000 | 0.0 | 0.000 | 0.0 | 0.022 | 5.5 | 0.023 | 5.7 | 0.032 |
| | | Medium positive | | | 90 | | 90 | 90 | 0.858 | 0.041 | 4.7 | 0.025 | 3.0 | 0.018 | 2.1 | 0.059 | 6.9 | 0.078 |
| | | Negative | 89 | | | 1 | 90 | 89 | 0.129 | 0.002 | 1.8 | 0.007 | 5.5 | 0.013 | 9.8 | 0.008 | 5.8 | 0.017 |
| | Serum | Indeterminate | | 90 | | | 90 | 90 | 0.283 | 0.012 | 4.3 | 0.000 | 0.0 | 0.011 | 3.9 | 0.018 | 6.3 | 0.024 |
| | | Low positive | | 18 | 72 | | 90 | 90 | 0.422 | 0.004 | 0.9 | 0.004 | 1.1 | 0.018 | 4.2 | 0.013 | 3.1 | 0.023 |
| | | Medium positive | | | 90 | | 90 | 90 | 1.055 | 0.048 | 4.5 | 0.000 | 0.0 | 0.041 | 3.9 | 0.045 | 4.3 | 0.078 |
| | | Negative | 84 | 2 | | 4 | 90 | 86 | 0.109 | 0.024 | 22.3 | 0.000 | 0.0 | 0.018 | 16.5 | 0.013 | 11.7 | 0.033 |
{6}
6
| JCV Gen-2 Validation Reproducibility Results | | | | | | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Parameter | Sample Matrix | Sample Name | Qualitative Results | | | | | Quantitative Results | | | | | | | | | | | |
| | | | ND | I | D | NV | Total | N | Mean | Variability Components | | | | | | | | | |
| | | | | | | | | | | Between Sites | | Between Days | | Between Runs/Operators | | Within Assay (Repeatability) | | Total | |
| | | | | | | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| %Inhibition | Plasma | Indeterminate | 11 | | 77 | 2 | 90 | 88 | 51.05 | 2.006 | 3.9 | 0.000 | 0.0 | 3.648 | 7.1 | 3.357 | 6.6 | 5.348 | 10.5 |
| | Serum | Indeterminate | 2 | | 85 | 3 | 90 | 87 | 62.59 | 3.889 | 6.2 | 2.438 | 3.9 | 0.000 | 0.0 | 5.550 | 8.9 | 7.202 | 11.5 |
ND = Not Detected, I = Indeterminate, D = Detected, NV = Invalid
## Reproducibility around the Assay cut-off
In order to demonstrate precision near the assay cut-off, two contrived samples, one serum and one (EDTA) plasma samples were prepared to be near the lower cut-point of the assay. Each sample was diluted twenty times and tested in the detection assay and the confirmation assay. As depicted in the following table the %CV was ≤ 2.6%.
Table: Reproducibility at the Limit of Detection
| Sample Matrix | Descriptive Statistics of Detection Assay (Index) | | | | | | Descriptive Statistics of Confirmation Assay (% Inhibition) | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | Min | Max | Mean | SD | %CV | N | Min | Max | Mean | SD | %CV |
| Plasma | 20 | 0.33 | 0.36 | 0.34 | 0.01 | 2.3 | 20 | 61.66 | 66.97 | 65.11 | 1.43 | 2.2 |
| Serum | 19* | 0.19 | 0.21 | 0.2 | 0.01 | 2.6 | 19 | 55.89 | 60.92 | 58 | 1.16 | 2 |
*One replicate was Invalid
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not Applicable
d. Detection limit:
{7}
Not Applicable
e. Analytical specificity:
Cross Reactivity
Cross reactivity was evaluated in a three part study conducted at an internal testing site.
I. The first study evaluated cross reactivity with commercially available human antibodies spiked into JCV negative serum and plasma as determined by the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay. The potentially cross reacting antibodies were spiked in at concentrations estimated to be approximately 2-4 times higher than the estimated detection limit of JCV antibody. Each antibody tested was not detected with the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay. There was no observed reactivity with the three potentially cross-reacting antibodies tested in this part of the study and the results are presented in the table below.
Table: Cross Reactivity - Part One – Spiked Antibodies
| Cross Reactant | Concentration | Agents Detected | |
| --- | --- | --- | --- |
| | | Serum | Plasma |
| Antibody to Escherichia coli | 0.4 μg/mL | 0/3 | 0/3 |
| Antibody to Mycobacterium tuberculosis | 0.4 μg/mL | 0/3 | 0/3 |
| Antibody to Pneumocystis jiroveci | Unquantified | 0/3 | 0/3 |
II. The second part of the study included a panel consisting of a least twenty left over specimens that previously tested positive for each of the potential cross reacting antibodies. Each member of the panel was evaluated using the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay along with appropriate controls. The seroprevalence of JCV in the panel was compared to the expected seroprevalence of JCV in the normal population. If a group demonstrates a higher than expected seroprevalence of JCV (set as $>70\%$ for the purposes of this study) it may be an indicator of potential cross reactivity. Four groups of patients (C. pneumoniae, HIV, CMV and HSV 1) exhibited a positivity rate that was slightly above that observed in previously reported studies. The results suggest that these groups demonstrate potential cross reactivity.
Table: Cross Reactivity – Part Two – Sero-prevalence Comparison
| Sample Matrix | Cross Reactant | Total No. of Replicates or Samples | Screening Result Count (Based on Index) | Confirmation Result Count (Based on %Inhibition) | Final Interpretation | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | Count | | % | | | |
| | | | D | IND | ND | Not Tested | D | ND | D | ND | D | ND |
| Unknown | HIV | 22 | 16 | 2 | 4 | 20 | 1 | 1 | 17 | 5 | 77.3 | 22.7 |
| Serum | C. pneumoniae | 48 | 35 | 5 | 8 | 43 | 2 | 3 | 37 | 11 | 77.1 | 22.9 |
| | C. trachomatis | 20 | 11 | 2 | 7 | 18 | 1 | 1 | 12 | 8 | 60.0 | 40.0 |
{8}
| | CMV | 40 | 27 | 7 | 6 | 33 | 5 | 2 | 32 | 8 | 80.0 | 20.0 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Candida | 40 | 26 | 6 | 8 | 34 | 1 | 5 | 27 | 13 | 67.5 | 32.5 |
| | EBV | 40 | 23 | 9 | 8 | 31 | 2 | 7 | 25 | 15 | 62.5 | 37.5 |
| | HSV 1 | 48 | 32 | 6 | 10 | 42 | 2 | 4 | 34 | 14 | 70.8 | 29.2 |
| | HSV 2 | 20 | 8 | 5 | 7 | 15 | 1 | 4 | 9 | 11 | 45.0 | 55.0 |
| | HHV6 | 20 | 11 | 2 | 7 | 18 | | 2 | 11 | 9 | 55.0 | 45.0 |
| | Listeria | 17 | 8 | 3 | 6 | 14 | | 3 | 8 | 9 | 47.1 | 52.9 |
| | Mycoplasma | 20 | 10 | 4 | 6 | 16 | | 4 | 10 | 10 | 50.0 | 50.0 |
| | Treponema pallidum | 19 | 12 | 2 | 5 | 17 | 1 | 1 | 13 | 6 | 68.4 | 31.6 |
| | VZV | 20 | 12 | 4 | 4 | 16 | 1 | 3 | 13 | 7 | 65.0 | 35.0 |
| All | | 334 | 208 | 48 | 78 | 286 | 15 | 33 | 223 | 111 | 66.8 | 33.2 |
D= Detected, ND = Not Detected, IND = Indeterminate
III. The third part of the cross reactivity evaluation consisted of an evaluation of the potential cross reactivity with other polyoma viruses. This included a cross absorption study using BK virus like particles (VLP). A total of 40 clinical specimens that have previously tested positive for JCV antibodies in the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay were tested in a confirmation style assay using JC VLP and BK VLP. A control set consisted of samples spiked with JC VLP at the same concentration of VLP that is used in the confirmation assay. The test samples were with spiked with BK VLP at the same concentration. The tests samples would be considered to be cross reactive if the % change in signal between the unspiked sample and the sample spiked with BK VLP is > 45%. The control set demonstrated % change in OD that was consistent with expectations. The test samples demonstrated % change in OD values due to BK VLP that ranged from -15 to 27%. No samples exhibited >45% change in the OD signal when spiked with BK VLP, indicating that the assay does not cross react with BKV.
The potential cross reactivity with other human polyoma viruses was not evaluated. However, additional analysis of the structure of the VP1 protein of other polyoma viruses indicates that BKV and JCV are more closely related (79.4% similarity) than other polyoma virus such as Merkel Cell virus and WU virus and KI virus (30.8 to 50.8% similarity). Due to the differences in viral structure a similar experiment with other polyoma virus VLP was not conducted.
# Interferences
Potential interference due to endogenous substances was evaluated using a testing panel consisting of sera and plasma samples that exhibit JCV antibody values close to the assay cut-off. The samples are spiked with the potential interferent as compared to baseline of the same serum and plasma samples that did not contain the interferent.
{9}
The interferents were spiked at the highest possible endogenous level. For all of the potential interferents with the exception of $\gamma$ globulin; the observed differences in signal did not cause any changes in interpretation of the final result. A potentially interference is suspected if the $\%$ change from the baseline sample is $>20\%$ . Commercially available $\gamma$ globulin is produced using normal human serum containing IgG antibodies, due to the high seroprevalence of antibodies to JCV virus in the normal population this may explain the observed cross reactivity.
Table: Interference Summary - Signal Comparison to Baseline
| Substance Name | Substance Concentration | Plasma | | | Serum | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Average Index | | %Change from Baseline | Average Index | | %Change from Baseline |
| | | Baseline Sample | Interference Sample | | Baseline Sample | Interference Sample | |
| Albumin | 120 mg/mL | 0.35 | 0.38 | 8.6 | 0.43 | 0.37 | -14.0 |
| Ascorbic Acid | 0.03 mg/mL | 0.40 | 0.42 | 5.0 | 0.42 | 0.44 | 4.8 |
| Bilirubin | 0.2 mg/mL | 0.33 | 0.33 | 0.0 | 0.42 | 0.38 | -9.5 |
| Cholesterol | 5 mg/mL | 0.48 | 0.40 | -16.7 | 0.46 | 0.44 | -4.3 |
| Gamma Globulin | 60 mg/mL | 0.38 | 3.54 | 831.6 | 0.41 | 3.52 | 758.5 |
| Hemoglobin | 110 mg/mL | 0.36 | 0.41 | 13.5 | 0.44 | 0.46 | 5.0 |
| Hemoglobin | 165 mg/mL | | 0.38 | 6.1 | | 0.39 | -10.1 |
| Hemoglobin | 220 mg/mL | | 0.35 | -2.4 | | 0.37 | -16.0 |
| Triglycerides | 10 mg/mL | 0.36 | 0.35 | -2.8 | 0.40 | 0.37 | -7.5 |
Note: Three dilutions of Gamma Globulin stock solution exhibited Index values - 3.22, 3.57 and 3.40.
# f. Assay cut-off:
To characterize antibody responses against infectious agents in humans, it is critical to have reference sera from both infected and non-infected individuals. However, JCV infection is clinically asymptomatic, thereby making it difficult to generally distinguish JCV infected from non-infected individuals. Even though about $20 - 30\%$ of JCV infected individuals shed viral DNA in the urine, JCV DNA is often not detected in the blood or urine of infected individuals, even when the infection results in the development of PML. There is no evidence that testing for JCV DNA in blood or urine can identify all JCV infected individuals, however, those who shed virus in the urine are confirmed to be infected with JCV. Therefore, sera from viruric patients were used to establish the positive reference sera for the assay. The assay cut point was established from the distribution of the serological responses of samples collected from JCV viruric patients in the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay. The lower cut point for the detection assay was set at an index of 0.2 along with an upper cut point at 0.4, and inhibition cut point of the confirmation assay was set at $45\%$ . The assay has an estimated false negative rate of $2.2\%$ based on a negative JCV antibody results for 4 out of 184 serum samples collected from viruric patients in the STRATIFY-1 trial.
{10}
# 2. Comparison studies:
a. Method comparison with predicate device:
See clinical studies
b. Matrix comparison:
A panel of 109 paired sera and (EDTA) plasma samples were evaluated in the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay. Of the total 109 pairs one pair was eliminated from the regression analysis because the result of the sera sample was invalid, of the remaining 108 pairs 7 pairs were qualitatively detected with OD values that exceeded the range of the spectrophotometer. These pairs of samples are included in the qualitative result comparison below, but eliminated from the regression analysis since an index value could not be calculated. Passing-Bablok regression analysis of the pairs of sera and plasma specimens demonstrates a slope of 1 with a $95\%$ CI of (0.9928 to 1.0167)) and an intercept of 0 with a $95\%$ CI of (-0.0068 to 0.0016). Analysis of the qualitative results yielded Positive Percent Agreement = $96.7\%$ (58/60), $95\%$ CI: (88.6 to $99.1\%$ ), and Negative Percent Agreement = $97.9\%$ (47/48), $95\%$ CI: (89.1 to $99.6\%$ )
Table: Sample Matrix Comparison - Sera vs. Plasma Qualitative Results
| Final Plasma Result | Final Serum Result (Count) | | | All |
| --- | --- | --- | --- | --- |
| | ND | D | Invalid | |
| ND | 47 | 2 | 1 | 50 |
| D | 1 | 58 | | 59 |
| All | 48 | 60 | 1 | 109 |
{11}

Figure: Passing-Bablok Regression Analysis (Sera vs. Plasma)
# 3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable): There is an increased risk of PML in natalizumab treated multiple sclerosis (MS) and Crohn's disease (CD) patients. Three risk factors for PML have been identified in MS and CD patient populations undergoing natalizumab treatment: natalizumab treatment duration, prior immunosuppressant use, and the presence of antibodies to JCV. Utilizing these three risk factors, sub-groups of patients can be identified with both higher and lower risk for PML. Please consult the current, locally available, prescribing or supplemental physician information for natalizumab for detailed information on the known risks associated with JCV serological status and the development of PML in natalizumab-treated patients. Patients with all three known risk factors have the highest risk for the development
{12}
of PML. The risk stratification for PML in natalizumab treated patients who are positive for the presence of JCV antibodies is shown in the following table.
Table: Estimated Incidence of PML Stratified by Risk Factor
| Natalizumab Exposure† | JCV Antibody Positive * | |
| --- | --- | --- |
| | No Prior Immunosuppressant Use | Prior Immunosuppressant Use |
| 1-24 months | <1/1,000 | 2/1,000 |
| 25-48 months | 4/1,000 | 11/1,000 |
Notes: Based on postmarketing PML data and Natalizumab use data as of September 1, 2011.
†Data beyond 4 years of treatment are limited.
*Risk in anti-JCV antibody positive patients was estimated based on the assumptions that 18% of Natalizumab-treated MS patients have a history of prior immunosuppressant treatment and that 55% of Natalizumab-treated MS patients are anti-JCV antibody positive.
# Clinical Performance (Comparative Agreement)
Because PML is an infrequent event in natalizumab-treated patients, data collected from both clinical trial and post-marketing reports of confirmed cases of PML were used to assess the clinical performance of the STRATIFY JCV® DxSelect™ assay for PML risk stratification. A clinical plan was developed for collection of serum samples obtained from natalizumab-treated patients prior to the onset of PML for JCV antibody testing. A total of 31 available serum samples from confirmed PML patients collected at least 6 months prior to clinical diagnosis of PML were tested for JCV antibody status using the STRATIFY JCV® DxSelect™ assay. In addition to the pre-PML samples, 1330 Samples from MS patients were tested by the STRATIFY JCV® DxSelect™ assay. A total of 707 patients receiving treatment tested positive for antibodies to JCV using the test, and the positivity rate was estimated to be 58.7% (414/707) with a 95% CI of 54.9% to 62.1%. An assay is statistically informative if the percentage of positive results in patients with the disease of interest is higher than the percentage of positive results in the population at risk. The 100% JCV antibody positivity demonstrated in the 31 natalizumab-treated PML patients prior to PML diagnosis was significantly different than the 58.7% JCV antibody positivity in the MS population, and represents an approximately 2-fold increased risk of PML compared to the PML incidence in the overall natalizumab-treated population.
In prior clinical studies for natalizumab the risk of developing PML was estimated using statistical modeling. The relative risk for patients who have received natalizumab for at least 18 months is shown in the table below. Risk was calculated based on statistical modeling with an assumption that there is one hypothetical PML case with a negative test result (38 PML cases: 37 positive and 1 hypothetical negative) and an assumption that the study has 13,227 patients.
{13}
Table: Estimated incidence of PML by JCV serological status
| | | | Number with PML | Number without PML | Total number patients treated |
| --- | --- | --- | --- | --- | --- |
| JCV Serological Status | Positive | 37 | 7,229 | 7,266 | |
| | | Negative | 1* | 5,960 | 5,961 |
| Total | | | 38 | 13,189 | 13,227 |
| | | | | | |
| Risk of PML (per 1,000) treated with ≥ 18 months for Positive result | | | 5.09
95% CI: 3.70 to 7.01 | | |
| Risk of PML (per 1,000) treated with ≥ 18 months for Negative result | | | 0.17
95% CI: 0.03 to 0.95 | | |
| Relative risk | | | 30.4
95% CI: 5.3 to 437.4 | | |
*For the negative result, a hypothetical case was assumed in order to allow for calculation of the point estimate.
The studies demonstrated that the positivity rate for JCV antibodies is not dependant upon prior IS use or the duration of natalizumab treatment. The results of the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay can be used along with other established PML risk factors, of prior IS use and natalizumab treatment duration, to stratify an individuals risk for PML, please refer to Table 1 and to the prescribing information for additional risk estimates.
# Performance with pre-PML samples.
The STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay was compared to an FDA cleared assay (STRATIFY $\mathrm{JCV}^{\text{TM}}$ Antibody ELISA) using serum samples obtained from PML patients at least 6 months prior to PML diagnosis. Because PML is an infrequent event in natalizumab-treated patients, data collected from both clinical trial and post-marketing reports of confirmed cases of PML were used to assess the clinical performance of the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay for PML risk stratification. Thirty-one available serum samples from confirmed PML patients collected at least 6 months prior to clinical diagnosis of PML were tested at one internal testing site for JCV antibody status using the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay and the validated laboratory methodology. The sample set demonstrated $100\%$ (31/31) positive agreement (95% CI: $89.0\%$ to $100\%$ ) with the validated assay.
# Performance with archived clinical specimens
Two groups of prospectively collected and archived clinical samples obtained from the STRATIFY-2 and the AFFIRM clinical studies were used to assess the performance of the STRATIFY $\mathrm{JCV}_{\text{®}}$ DxSelect™ assay compared to the validated laboratory methodology used in the STRATIFY-2 and AFFIRM clinical studies. One group consisted of patients who were receiving natalizumab, the other group of patients had not received natalizumab therapy (and were considering receiving natalizumab). The samples were blinded and randomly distributed to two external testing sites and one internal testing site. The data was analyzed for each group separately. Results for the two individual groups are presented in the tables below. The Positive and Negative Percent Agreement are shown in the table below.
{14}
Table: Agreement for Clinical Samples – Patients Receiving Natalizumab
| STRATIFY JCV® DxSelect™ | Predicate Device | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| Positive | 385 | 29 | 414 |
| Negative | 12 | 281 | 293 |
| Total | 397 | 310 | 707 |
| Positive Percent Agreement (PPA) | 97.0% (385/397)
95% CI: 94.8 to 98.3% | | |
| Negative Percent Agreement (NPA) | 90.6% (281/310)
95% CI: 86.9 to 93.4% | | |
14
{15}
15
Table: Agreement for Clinical Samples - Patients Considering Natalizumab
| STRATIFY JCV® DxSelect™ | Predicate device | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| Positive | 326 | 24 | 350 |
| Negative | 5 | 268 | 273 |
| Total | 331 | 292 | 623 |
| Positive % Agreement (PPA) | 98.5% (326/331)
95% CI: 96.5 to 99.4% | | |
| Negative % Agreement (NPA) | 91.8% (268/292)
95% CI: 88.1 to 94.4% | | |
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
The STRATIFY JCV® DxSelect™ assay been used to evaluate JCV antibody positivity rate in serum and plasma samples from a geographically diverse cohort of 1330 MS patients. The cohort was comprised from MS patients from clinical trials including a completed Phase 3 clinical study of natalizumab in MS patients (AFFIRM C-1801), an ongoing study to evaluate seroprevalence in the MS population (STRATIFY-2 [101JC402]). The clinical characteristics for the MS patients within each study are shown in the following table. The age and gender distribution of the MS cohort tested with the STRATIFY JCV® DxSelect™ are similar to the age and gender distribution of MS patients treated with natalizumab in the post-marketing setting. JCV antibody positivity rate in the MS cohort was 55-59%. JCV antibody positivity rate was shown to increase with age and was lower in women compared to men, consistent with what has been reported in the literature in healthy adults using similar assay methodologies.
Seroprevalence data was evaluated for each study. The observed seroprevalence using the STRATIFY JCV® DxSelect™ assay was 59% (467/792) in the STRATIFY-2 study and 55% (296/538) in the AFFIRM Study. These seroprevalence values are consistent with the seroprevalence values observed during the clinical studies, when the relationship between JCV serological status and PML development were described. Additionally the STRATIFY JCV® DxSelect™ assay demonstrated 100% concordance with 31 pre-PML samples.
{16}
Table: Demographic Data and JCV Antibody Prevalence for MS Patients
| | AFFIRM (N=538) | STRATIFY-2 (N=792) |
| --- | --- | --- |
| Age (years) | | |
| • Range | 18-50 | 19-78 |
| • Mean | 35.8 | 46.4 |
| • Median | 36 | 46 |
| Gender (%)* | | |
| • Male | 173/538 (32.2%) | 202/792 (25.5%) |
| • Female | 365/538 (67.8%) | 590/792 (74.5%) |
| Geography | North America and EU/Rest of World | US |
| JCV Antibody Positivity Rate (95% CI) | | |
| • % JCV Antibody Positive | 297/538 (55.2%) (51.0 to 59.4) | 467/792 (59.0%) (55.5 to 62.3) |
| • % JCV Antibody Negative | 241/538 (44.8%) (40.6 to 49.0) | 325/792 (41.0%) (37.7 to 44.5) |
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.