← Product Code [DQA](/productcode/DQA) · K120984

# MASIMO RAS-125C ACOUSTIC RESPIRATION SENSOR (K120984)

_Masimo Corporation · DQA · Apr 26, 2013 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K120984

## Device Facts

- **Applicant:** Masimo Corporation
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Apr 26, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K120984 · Apr 26, 2013 | MASIMO RAS-125C ACOUSTIC RESPIRATION SENSOR | Masimo Corporation | Hospitalized post-surgical patient medical records/clinical monitoring data | Clinical validation of the acoustic respiratory rate monitoring sensor was performed using data from hospitalized post-surgical adult and pediatric patients to demonstrate accuracy compared to manual respiratory rate counts. | Post-surgical monitoring; Hospitalized patients; Clinical validation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical validation in hospitalized post-surgical subjects; Non-randomized clinical validation study | Hospitalized post-surgical adult and pediatric subjects; Sample Size: 26 adult subjects; 26 pediatric subjects; Number of Sites: 3 hospital centers in the United States (pediatric study) | Manual count of respiratory rate | Respiratory rate accuracy |

## Indications for Use

The Rainbow Acoustic Monitoring™ sensors and cables are indicated for continuous, noninvasive monitoring of respiratory rate (RRaTM). The RAS-125 is indicated for adult patients in hospital-type facilities, mobile and home environments. The RAS-125c is indicated for adult and pediatric patients in hospital-type facilities, mobile and home environments.

## Device Story

Masimo Acoustic Monitoring Sensor (RAS-125/RAS-125c) is a single-patient use piezoelectric sensor; attaches to patient's neck via adhesive strip. Device captures acoustic signals from upper airway; transmits data via cable to compatible Masimo pulse oximeter monitors (e.g., Radical 7, Rad 87). Monitor processes acoustic input to calculate and display continuous respiratory rate (RRa). Used in hospital, mobile, and home settings; operated by clinicians or patients/caregivers. Provides real-time respiratory rate data to assist healthcare providers in patient monitoring; enables continuous assessment of respiratory status; facilitates timely clinical intervention.

## Clinical Evidence

Clinical validation included three non-randomized studies: healthy adult volunteers (n=26), hospitalized post-surgical adults (n=26), and hospitalized post-surgical pediatric subjects (n=26). Primary endpoint was respiratory rate accuracy compared to manual counts. Validated range: 4-70 breaths/min (bench); clinical validation up to 30 breaths/min (adults >30 kg) and 50 breaths/min (pediatrics >10 kg, age ≥2 years). Accuracy reported as ± 1 breath per minute.

## Technological Characteristics

Piezoelectric sensor element; adhesive neck attachment. Single-patient use. Connectivity via cable to compatible Masimo pulse oximeter monitors. Complies with ISO 14971 (risk management), EN ISO 10993 (biocompatibility), ISO 9919 (safety/mechanical), IEC 60601-1 (electrical safety), and IEC 60601-1-2 (EMC).

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Masimo Rainbow SET® Rad-87 Pulse CO-Oximeter and Masimo Acoustic Monitoring Sensor (adult) ([K091241](/device/K091241.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary: K120984

# APR 2 6 2013

| Submitter:                | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>Phone: (949) 297-7000<br>FAX: (949) 297-7592     |
|---------------------------|---------------------------------------------------------------------------------------------------------|
| Contact:                  | Patricia Milbank<br>Vice President, Regulatory Affairs                                                  |
| Date Summary<br>Prepared: | April 23, 2013                                                                                          |
| Trade Name:               | Masimo Acoustic Monitoring Sensors (adult/pediatric)                                                    |
| Common Name:              | Pulse Oximeter                                                                                          |
| Classification Name:      | DQA - Oximeter (870.2700)<br>BZQ - Breathing frequency monitor (868.2375)                               |
| Predicate Device:         | K091241 – Masimo Rainbow SET® Rad-87 Pulse CO-Oximeter<br>and Masimo Acoustic Monitoring Sensor (adult) |

# Description of the Device

The Masimo Acoustic Monitoring Sensor is a single-patient use device consisting of a piezoelectric sensor element and an adhesive strip that is attached to the patient's neck for the purpose of noninvasive respiratory rate monitoring in adult and pediatric patients (> 10 kg). As with the predicate Masimo Acoustic Monitoring Sensor (model RAS-125 for adults only), the RAS-125c for pediatric patients is intended to be attached via cable to a pulse oximeter monitor incorporating the Masimo Rainbow® Acoustic Monitoring™ technology to provide a continuous display of the patient's acoustic respiratory rate. Suitable pulse oximeter monitors include the previously cleared Radical 7 and Rad 87 devices.

#### Indications for Use

The Rainbow Acoustic Monitoring™ sensors and cables are indicated for continuous, noninvasive monitoring of respiratory rate (RRaTM).

The RAS-125 is indicated for adult patients in hospital-type facilities, mobile and home environments.

The RAS-125c is indicated for adult and pediatric patients in hospital-type facilities, mobile and home environments.

{1}------------------------------------------------

# Contraindications

The RAS-125 and RAS-125c sensors are contraindicated in patients who exhibit allergic reactions to foam rubber products and/or adhesive tape.

# Results from Clinical Studies

Accuracy data was obtained from a comparison of a manual count of respiratory rate in healthy adult subjects and in hospitalized pediatric subjects for whom post-operative monitoring of respiratory rate was indicated.

Respiration rate accuracy for the Rainbow Acoustic Monitoring sensor and instrument has been validated for the range of 4 to 70 breaths per minute in bench top testing. Clinical validation was performed with the RAS-125 and RAS-125c sensors and instrument for up to 30 breaths per minute in adult subjects >30 kg, and up to 50 breaths per minute in hospitalized pediatric subjects, age 2 years or older and >10kg.

|                                                            | RAS-125                       | RAS-125c                      |
|------------------------------------------------------------|-------------------------------|-------------------------------|
| <span style="vertical-align: middle;"> </span> Body Weight | Adult > 30 kg                 | Adult, Pediatric > 10 kg      |
| Application Site                                           | Neck                          | Neck                          |
| Breaths per Minute,<br>Accuracy Range*                     | 4 to 70 ± 1 breath per minute | 4 to 70 ± 1 breath per minute |

The accuracy of this device has not been validated for monitoring of respiratory rate in patients with specific medical conditions, e.g., asthma, COPD or cystic fibrosis. Accuracy of the sensor has not been validated in infants or neonates.

# Test Summary

The RAS-125 and RAS-125c sensors comply with the voluntary standards as detailed in Section 9 of this submission and the following quality assurance, verification and validation measures were included in the development process:

- Design control activities per 21 CFR 820 .
- Risk analysis and risk management activities per ISO 14971 .
- Software verification and validation testing per 21 CFR 820 and the applicable FDA . guidance documents (General Principles of Software Validation, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, and Off-The-Shelf Software Use in Medical Devices)
- Biocompatibility testing per EN ISO 10993 for patient skin contacting, prolonged . duration materials, including cytotoxicity, skin irritation and sensitization
- Environmental testing (operating temperature, humidity, fluid ingress), safety and . mechanical testing per ISO 9919
- Electromagnetic compatibility per IEC 60601-1-2 .
- Electrical safety testing per IEC 60601-1 �
- Two non-randomized clinical validation studies in healthy adult volunteers (n=26) and .

{2}------------------------------------------------

hospitalized post-surgical adult subjects (n=26)

- A non-randomized clinical validation study in hospitalized post-surgical pediatric . subjects (n=26), conducted at 3 hospital centers in the United States .
#### Conclusion

The information in this 510(k) submission demonstrates that the Masimo RAs-125 and RAS-125c Acoustic Respiration Sensors are substantially equivalent to the predicate device, with respect to safety, effectiveness, and performance, when used in adult and pediatric patients.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized eagle or bird-like figure with three curved lines representing its wings or feathers.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

### April 26, 2013

Ms. Patricia Milbank Vice President, Regulatory Affairs Masimo Corporation 40. Parker_ IRVINE CA 92618

Re: K120984

Trade/Device Name: Masimo Acoustic Respiration Sensor (Adult/Pediatric) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, BZA Dated: March 13, 2013 Received: March 14, 2013

Dear Ms. Milbank:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Milbank

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Digitally signed by Mary 5. Runner -5
ON: c=US, o=U.S. Government,
ou-HHS, ou=FDA, ou=People,
Susan Runner, DOSMA cn=Mary 5. Runner -S,
0.9.2342.19200300.100.1.1=13000879
50
Date: 2013.04.26 07:26:55 -04'00'

for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number: K120984

Masimo Acoustic Respiration Sensor (Adult/Pediatric) Device Name: ________

Indications for Use:

The Rainbow Acoustic Monitoring™ sensors and cables are indicated for continuous, noninvasive monitoring of respiratory rate (RRaTM).

The RAS-125 is indicated for adult patients in hospital-type facilities, mobile and home environments.

The RAS-125c is indicated for adult and pediatric patients in hospital-type facilities, mobile and home environments.

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

Chan O. Lee -S
2013.04.25 14:13:49
-04'00' For LS

thesiology, General Hospital on Control, Dental Devices

510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K120984](https://fda.innolitics.com/device/K120984)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
