K120932 · The Yoshida Dental Mfg. Co., Ltd. · GEX · Dec 7, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120932
Device Name
OPELASER PRO II AND THE OPELASER LITE II
Applicant
The Yoshida Dental Mfg. Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Opelaser Family is CO2 Medical Laser Systems (and the delivery system that are used to deliver laser energy) are indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology (including laparoscopy), neurosurgery, orthopedics (soft tissue), arthroscopy (knee), general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery.
Device Story
The OPELASER PRO II and Lite II are CO2 gas laser systems (Class 4) for soft tissue surgery. The device uses RF discharge excitation to generate a 10.6um wavelength laser. Energy is delivered via either a SiO2 hollow fiber or a 6-point articulated arm. The PRO model provides up to 7 watts, while the Lite model provides up to 5 watts. Operated by clinicians in surgical settings, the device allows for precise ablation, vaporization, excision, and coagulation of soft tissue. The healthcare provider directs the laser energy to the target tissue to achieve the desired surgical effect, facilitating various procedures across multiple medical specialties.
Clinical Evidence
Bench testing only. The device underwent extensive safety and bench testing, including software validation and risk analysis. Sterilization of laser tips was validated. The systems were certified by an NRTL to IEC/UL 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
CO2 gas laser (Class 4); RF discharge excitation oscillation; 10.6um wavelength; 7W (PRO) or 5W (Lite) max power; SiO2 hollow fiber or 6-point articulated arm delivery; IEC/UL 60601-1 and 60601-1-2 compliant.
Indications for Use
Indicated for patients requiring soft tissue ablation, vaporization, excision, incision, and coagulation across various surgical specialties including dermatology, plastic surgery, podiatry, ENT, gynecology, neurosurgery, orthopedics, general/thoracic surgery, dental/oral surgery, and genitourinary surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
#### 510(K) Summary, 510(k) K120932 Submitter: DEC 7 2012 THE YOSHIDA DENTAL MFG. CO., LTD 1-3-6 Kotobashi, Sumida-ku Tokyo, Japan ZIP 130-8516 TEL: +81-3-3631-2165 (International Business Div.) FAX: +81-3-3631-2685 (International Business Div.) Contact: Hidenori Watanabe, Regulatory Affairs, hi-watanabe@yoshida-net.co.jp Date Prepared: March 15, 2012
### 1. Identification of the Device:
Proprietary-Trade Name: OPELASER PRO II and OPELASER Lite II Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology. Product Code GEX Common/Usual Name: Powered laser surgical instrument
- Equivalent legally marketed devices: K031440, DEKA, Model Smart US 20D; K030147 2. Lumenis Model: UltraPulse SurgiTouch
- Description of the Device: These are CO2 Laser devices designed for surgical applications. The 3. maximum power output for the PRO model is 7 watts, and for the LITE model, 5 watts. Type of laser is a CO2 gas laser (class 4); Oscillation system: RF discharge excitation oscillation; Light guiding system: SiO2 hollow fiber or 6 point articulated arm depending on model; Oscillation wavelength: 10.6um.
- Indications for Use. The Opelaser Family is CO2 Medical Laser Systems (and the delivery system 4. that are used to deliver laser energy) are indicated for use in surgical applications requiring the ablation, vaporization, excision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology (including laparoscopy), neurosurgery, orthopedics (soft tissue), arthroscopy (knee), general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery.
- Safety and Effectiveness, comparison to predicate device. This device has the same indications 5. for use as the predicate device and employs nearly identical technology to accomplish the same task. Delivered energy levels and wavelengths are essesntially the same.
- Description of Testing, nonclinical tests submitted: The OPELASER systems have undergone 6. extensive safety and bench testing as well as software validation and risk analysis. The ability to sterilize the laser tips has been validated. The OPELASER systems have been tested and certified by the NRTL (Nationally Recognized Test Lab) Intertek. Certification to IEC/UL 60601-1 and IEC 60601-1-22 has been accomplished.
- Substantial Equivalence: A detailed comparison of specifications and technologies, as well as 7. comparisons of indications for use shows that the OPELASER device are substantially equivalent to the predicate devices. Safety standards applied are the same, as well.
- The conclusions drawn from the nonclinical and clinical tests demonstrate that the 8. Conclusion: device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph 2, above.
{1}------------------------------------------------
## Exhibit 6. Truthful and Accuracy Statement as required per 21CFR807.87(k).
I certify that, in my capacity as Executive Chief Operating Officer of the Laser division of The Yoshida Dental Mfg. Co., Ltd., I believe, to the best of my knowledge, that all data and information submitted in this premarket notification are truthful and accurate, and that no material fact has been omitted.
高橋三浄
(Signature)
Mioki Hirohashi Date: March 13, 2012
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure embracing a sphere, symbolizing the department's mission to protect the health of all Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem. The logo is presented in black and white.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service .
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-002
December 7, 2012
The Yoshida Dental Mfg. Co., Ltd. % Kamm and Associates Mr. Daniel Kamm 8870 Racello Court Naples, Florida 34114
Re: K120932
Trade/Device Name: Opelaser Pro II and Opelaser Lite II Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: November 29, 2012 Received: December 03, 2012
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have and have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 – Mr. Daniel Kamm
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Peter D. Rumm -Si 2012.12.07 17:09:12 -05'00'
> Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K120932
Device Name: QPELASER PRO II and OPELASER Lite II
Indications For Use:
The Opelaser Family is CO2 Medical Laser Systems (and the delivery system that are used to deliver laser energy) are indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology (including laparoscopy), neurosurgery, orthopedics (soft tissue), arthroscopy (knee), general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery.
Prescription Use _X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
AND/OR
Neil R Ogden
2012.12.07 16:46:19 -05'00'
(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.