Retrospective clinical images were used to perform verification and validation testing of the CardIQ Xpress 2.0 software.
Retrospective clinical images; Software validation; Routine clinical care
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Non-significant risk reader study of care patient images; Retrospective analysis of clinical images
Patients undergoing routine clinical CT imaging
Not applicable for this study
Verification and validation of software performance
Indications for Use
CardIQ Xpress 2.0 is intended to provide an optimized non-invasive application to analyze cardiovascular anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images. CardIQ Xpress 2.0 is a CT, noninvasive, image analysis software package, which aids in diagnosing of cardiovascular disease to include, coronary artery disease, functional parameters of the heart, heart structures and follow-up for stent placement, bypasses and plaque imaging. CardIQ Xpress 2.0 offers unique tools such as automatic tracking, which will pre-process the CT data into multiple viewing ports to allow for an expedited read time improving workflow. With CardIQ Xpress 2.0, the user can color code the myocardial tissue to show hypo/hyperdense areas in the myocardial tissue of the heart. With the IVUS-like view the user can color code the HU units of the plaque to better visualize the difference between calcified and non-calcified plaque in the wall of the vessel and the lumen to determine the amount of atherosclerosis. The user can see the different valve planes along with a variety of new layouts to align the heart. The IVUS like view is created by applying GE's Volume Rendering on a cross-section perpendicular to the detected centerline. This view merely displays a cross section as in IVUS imaging and color codes like IVUS images. No new or additional diagnostic information is added. CardIQ Xpress 2.0 is for use on the Advantage Workstation (AW) platform, CT scanner, PAC or Centricity stations, which can be used in the analysis of 2D or 3D CT angiography images/data derived from DICOM 3.0 CT scans.
Device Story
CardIQ Xpress 2.0 with SnapShot Freeze is post-processing software for CT angiography images. It processes DICOM 3.0 CT data to visualize coronary vessels, heart structures, and myocardial tissue. Key features include automatic tracking for multi-view port display, myocardial tissue color-coding (hypo/hyperdense areas), and an IVUS-like view created via volume rendering on cross-sections perpendicular to detected centerlines. The software is used on GE Advantage Workstations, CT scanners, or PACS/Centricity stations by clinicians to assess atherosclerosis, plaque composition (calcified vs. non-calcified), and heart function (e.g., ejection fraction). SnapShot Freeze reduces coronary artery motion blurring. Output includes 2D/3D visualizations and text/graphic reports with templates. It aids clinical decision-making by streamlining workflow and providing anatomical/pathological insights for treatment planning.
Clinical Evidence
No clinical trials were conducted. Verification and validation testing utilized clinical images from a non-significant risk reader study of patient images to confirm performance.
Technological Characteristics
Software-based image analysis application. Operates on DICOM 3.0 CT data. Features include automatic tracking, volume rendering for IVUS-like views, and color-coding of Hounsfield Units (HU) for plaque and myocardial tissue analysis. Designed for integration with GE Advantage Workstation, CT scanners, and PACS/Centricity systems.
Indications for Use
Indicated for analysis of cardiovascular anatomy and pathology in patients undergoing CT angiography. Aids in diagnosis of coronary artery disease, assessment of heart structures, functional parameters, and follow-up for stent placement, bypasses, and plaque imaging.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular border. The letters and the border are black, creating a strong contrast. The logo is a well-known symbol associated with the General Electric company.
GE Healthcare 510(k) Premarket Notification Submission
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | March 21, 2012 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems, SCS (d.b.a. GE Healthcare) |
| | 283, rue de la Miniere BP34 |
| | 78530 Buc Cedex - France |
| | FDA Registration Number: 9611343 |
| Primary Contact<br>Person: | Stephen Slavens, RAC<br>Regulatory Affairs Director, MICT<br>GE Healthcare<br>Phone: (262) 548-4992<br>FAX: 262-548-3844<br>Stephen.Slavens@ge.com |
| Secondary Contact<br>Person: | Helen Peng, RAC<br>Regulatory Affairs Leader<br>GE Healthcare<br>Phone: (262) 548-5091<br>FAX: 262-548-3844<br>Hong.Peng@ge.com |
| Device:<br>Trade Name: | CardIQ Xpress 2.0 with SnapShot* Freeze Option |
| Common/Usual<br>Name: | Accessory to: System, x-ray, tomography, computed |
| Classification Names: | 21CFR 892.1750 |
| Product Code: | JAK |
| Predicate Device(s): | Predicate Device Name: CardIQ Xpress Version 2.0<br>Predicate 510k Number: K073138<br>Predicate Manufacturer: GE Medical Systems |
| Device Description: | The GE Medical Systems CardIQ Xpress 2.0 software is a<br>post processing software option for the Advantage<br>Workstation (AW) Platform, CT scanner, PACS or Centricity<br>systems. This product can be used in the analysis of CT<br>angiographic images to view the coronary vessels to<br>determine if the patient has normal coronary arteries,<br>arteriolosclerosis or severe stenosis, which needs to go on |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are solid black, creating a strong contrast against the white background.
## GE Healthcare 510(k) Premarket Notification Submission
for treatment. This software also can look at the heart structures to include valve imaging, heart motion and ejection fraction. CardIO Xpress 2.0 contains both graphic and text report capabilities with predefined templates for ease of use.
CardIO Xpress 2.0 with SnapShot* Freeze is additionally designed to reduce the coronary artery motion blurring in a CT image.
Indications for Use:
CardIO Xpress 2.0 is intended to provide an optimized noninvasive application to analyze cardiovascular anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images. CardIQ Xpress 2.0 is a CT, noninvasive, image analysis software package, which aids in diagnosina of cardiovascular disease to include, coronary artery disease, functional parameters of the heart, heart structures and follow-up for stent placement, bypasses and plaque imaging.
CardIQ Xpress 2.0 offers unique tools such as automatic tracking, which will pre-process the CT data into multiple viewing ports to allow for an expedited read time improving workflow. With CardIQ Xpress 2.0, the user can color code the myocardial tissue to show hypo/hyperdense areas in the myocardial tissue of the heart. With the IVUS-like view the user can color code the HU units of the plaque to better visualize the difference between calcified and non-calcified plaque in the wall of the vessel and the lumen to determine the amount of atherosclerosis. The user can see the different valve planes along with a variety of new lavouts to alian the heart. The IVUS like view is created by applying GE's Volume Rendering on a crosssection perpendicular to the detected centerline. This view merely displays a cross section as in IVUS imaging and color codes like IVUS images. No new or additional diagnostic information is added.
CardIQ Xpress 2.0 is for use on the Advantage Workstation (AW) platform, CT scanner, PAC or Centricity stations, which can be used in the analysis of 2D or 3D CT angiography
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "k120910_S001". The text is in a sans-serif font and is black. The background is white. The text appears to be a file name or identifier.
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The letters are stylized with curved lines, and the circle has decorative elements around its perimeter.
# GE Healthcare 510(k) Premarket Notification Submission
# images/data derived from DICOM 3.0 CT scans.
Technology: The proposed medical device, CardIQ Xpress 2.0 with Snapshot Freeze Option, employs the same fundamental scientific technology as its predicate device CardIQ Xpress 2.0.
Determination of Substantial
Equivalence:
### Summary of Non-Clinical Tests:
CardIQ Xpress 2.0 with SnapShot Freeze Option complies with voluntary standards as discussed in Section 9, and 11 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- . Requirements Reviews
- Design Reviews .
- . Testing on unit level (Module verification)
- Integration testing (System verification) .
- Performance testing (Verification) .
- . Safety testing (Verification)
- . Simulated use testing (Validation)
### Summary of Clinical Tests:
As referenced in Section 20, the CT acquired clinical images used for the completion of verification and validation testing for CardIQ Xpress 2.0 with Snapshot Freeze Option were obtained from a non-significant risk reader study of care patient images.
- GE Healthcare considers the CardIQ Xpress 2.0 with Conclusion: Snapshot Freeze Option application to be as safe, as effective, and performance is substantially equivalent to the predicate device.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping bodies, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 8 2012
GE Healthcare c/o Mr. Stephen Slavens Regulatory Affairs Director, MICT 3000 N Grandview Blvd W 1140 Waukesha, WI 53188
#### Re: K120910
Trade Name: CardIO Xpress 2.0 with Snapshot Freeze Option Regulation Number: 21.CFR892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II (two) Product Codes: JAK Dated: June 17, 2012 Received: June 22, 2012
#### Dear Mr. Slavens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 –Mr. Slavens
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part -807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram DA Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular frame. The letters and the frame are in black, set against a white background. The design is simple and recognizable, representing the brand identity of General Electric.
### k120910 S001
# GE Healthcare 510(k) Premarket Notification Submission
#### 510(k) Number (if known):
Device Name: CardIO Xpress 2.0 with SnapShot* Freeze Option
#### Indications for Use:
CardlQ Xpress 2.0 is intended to provide an optimized non-invasive application to andyze cardiovascular anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images. CardlQ Xpress 2.0 is a CT, noninvasive, image analysis software package, which aids in diagnosing of cardiovascular disease to include, coronary artery disease, functional parameters of the heart, heart structures and follow-up for stent placement, bypasses and plaque imaging.
CardIQ Xpress 2.0 offers unique tools such as automatic tracking, which will pre-process the CT data into multiple viewing ports to allow for an expedited read time improving workflow. With CardIQ Xpress 2.0, the user can color code the myocardial tissue to show hyperdense areas in the myocardial tissue of the heart. With the IVUS-like view the user can color code the HU units of the plaque to better visualize the difference between calcified and noncalcified plaque in the wall of the vessel and the lumen to determine the amount of atherosclerosis. The user can see the different valve planes along with a variety of new layouts to alian the heart. The IVUS like view is created by applying GE's Volume Rendering on a cross-section perpendicular to the detected centerline. This view merely displays a cross section as in IVUS imaging and color codes like IVUS images. No new or additional diagnostic informotion is added.
CardIQ Xpress 2.0 is for use on the Advantage Workstation (AW) platform, CT scanner, PAC or Centricity stations, which can be used in the analysis of 2D or 3D CT angiography images/data derived from DICOM 3.0 CT scans.
*ShapShot is a trademark of General Electric Company
Prescription Use x . AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subport C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510k)
Page 1 of 1
.
Signature
11:37 32:005
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.