The Polaris Dial-a-Date Contact Lens Case is intended for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.
Device Story
Polaris Dial-a-Date is a manual contact lens storage case. Device consists of a molded dual-compartment base and two screw-down lids. Each lid features a manual rotating dial marked with numbers (0-3 and 0-9) to allow users to set a reminder for lens replacement. Lenses are immersed in chemical disinfection solution within the compartments. Device is for OTC use by contact lens wearers. The outer lids and dials do not contact the lens care solution. The device provides a physical mechanism for tracking lens replacement schedules, aiding patient compliance with recommended replacement intervals.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (skin sensitization and ocular irritation), and ISO 10993-11 (systemic toxicity) showed no adverse effects. Leakage testing conducted on 192 units (filled to 2/3 capacity, inverted for 15 minutes, repeated 3 times) demonstrated zero leakage.
Technological Characteristics
Materials: Polyolefin (>95%) and mixture (<5%). Dimensions: 69.50mm x 33.50mm x 18.91mm. Volume: 4.4 ml per side. Principle: Manual storage container with integrated mechanical dial for replacement tracking. Non-powered. Standalone device.
Indications for Use
Indicated for storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection. Not for use with heat disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
K120904
Image /page/0/Picture/1 description: The image shows the word "RELIANCE" in bold, sans-serif font. To the left of the word is a logo that appears to be an eagle head inside of a square with a check mark. The logo and the word are both in black. The background is white.
NOV 26 2012 1
# 510(k) Summary
(per 21 CFR 807.92)
# Submission Correspondent
Emergo Group, Inc. www.emergogroup.com/
## Address
611 West Fifth Street Third Floor Austin, TX 78701 USA
# Phone
(512) 327-9997 Fax (512) 327-9998
# Contact
Caroline Tontini to projectmanagement@emergogroup.com
### Submission Sponsor
Reliance Design & Manufacture Corp. No.19, Lane 166, Yanhe St., Yongkang Dist. Tainan City 71082, Taiwan R.O.C
FDA Establishment Registration #: 3008388282
# Date Prepared
February 10th, 2012
# Trade Name
Polaris Dial-a-Date Contact Lens Case
# Classification Name
Soft (hydrophilic) contact lens care products
# Regulation Number 886.5928
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "RELIANCE" in bold, black letters. To the left of the word is a black square with a white bird-like figure inside. The font of the word is sans-serif and the letters are closely spaced together.
Product Code LRX
Classification Panel Ophthalmic Devices
Device Class Class 2
## Predicate Devices
Polaris Contact Lens Case (K093377) from Reliance Design & Manufacture Corp.
#### Intended Use
The Polaris Dial-a-Date Contact Lens Case is intended for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.
#### Device Description
The Polaris Dial-a-Date contact lens case is a device for the storage of soft (hydrophilic) and rigid gas permeable contact lenses. Contact lenses can be fully immersed into the chamber and accommodates all lenses currently being sold in the market. The contact lens case consists of a molded dual compartment bottom base and two screwdown lids. The bottom of each compartment is marked with L (left) and R (right). The lids are also labeled with L (left) and R (right), and include a dial with molded-in numbers from 0 to 3 and 0 to 9, respectively. The outer lids are colored to make the molded-in numbers of the dials visible at the 'dialed position'. The dials can be incrementally rotated to let the user set reminders (i.e. day of the month to replace the contact lenses).
The outer lids and number dials do not come into contact with the lens care solution.
Image /page/1/Picture/12 description: The image shows an exploded view of a contact lens case and its components. The case itself is a double-sided container, with two circular depressions for holding the lenses. Above each side of the case, there is a lens and a corresponding cap, with dashed lines indicating the alignment and assembly order of the components. The image illustrates how the lens and cap fit together into the case.
Figure 5A Exploded View of the lens case illustrating the coloring of the Outer Lid
{2}------------------------------------------------
K120904
Image /page/2/Picture/1 description: The image shows the word "RELIANCE" in bold, black letters. To the left of the word is a logo that features a white bird head inside of a black square. A black wing extends from the square. The logo and the word are aligned horizontally.
### Comparison of Technological Characteristics
The following table compares the Polaris Dial-a-Date to the POLARIS (201) with respect to intended use, technological characteristics and principles of operation.
| Table 5A – Comparison of Characteristics | | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Reliance Design & Manufacture Corp. | Reliance Design & Manufacture Corp. |
| Product Name | POLARIS (201) Contact Lens Case | Polaris Dial-a-Date (202)<br>Contact Lens Case |
| 510(k) Number | K093377 | Pending |
| Classification<br>Advisory<br>Committee | Ophthalmic | Ophthalmic |
| Product Code | LRX | LRX |
| Regulation # | 886.5928 | 886.5928 |
| Class | 2 | 2 |
| Intended Use | For storage of soft (hydrophilic), rigid<br>gas permeable, and hard contact<br>lenses.<br>Use for storage during chemical<br>disinfection only. Not for use during<br>heat disinfection. | For storage of soft (hydrophilic), rigid<br>gas permeable, and/or hard contact<br>lenses.<br>Use for storage during chemical<br>disinfection only. Not for use during heat<br>disinfection. |
| Composition (%) | - Polyolefin > 95%<br>- Mixture < 5% | - Polyolefin > 95%<br>- Mixture < 5% |
| Size | Length: 67.97mm<br>Width: 31.97mm<br>Height: 21.50mm | Length: 69.50mm<br>Width: 33.50mm<br>Height: 18.91 |
| Volume | 3.0 ml each side | 4.4 ml each side |
| Number of Outer<br>Lid Colors | Four (4) | Four (4) |
## Table 5A - Comparison of Characteristics
## Summary of Differences
The Polaris Dial-a-Date Contact Lens Case differs from its predicate device POLARIS (201) in some aspects that however do not affect its safety or performance. The differences are as follows:
- a) the Polaris Dial-a-Date (202) has a moveable part on the top lid to set a reminder while the predicate device POLARIS (201) has a plain top
- b) the dimensions length, height and width differ slightly
- c) the compartment volume is larger: 4.4 ml vs. 3.0 ml
## Summary of Non-Clinical Data Submitted
The following testing has been performed to support substantial equivalence:
- · Biocompatibility Testing according to the applicable standards, as described in Section 15 Biocompatibility
- Leakage Testing, as described in Section 18 Performance Testing Bench Studies
{3}------------------------------------------------
K120904
Image /page/3/Picture/1 description: The image shows the word "RELIANCE" in bold, black letters. To the left of the word is a black square with a white bird-like figure inside. The bird figure appears to be flying or diving. The logo is simple and modern.
#### Biocompatibility Testing
The Polaris Dial-a-Date contact lens case has been evaluated for biocompatibility in accordance with Part 10993 of the ISO Standard tests, which include:
- . ISO 10993-5:1999, Biological evaluation of medical devices -- Part 5: Tests for In Vitro Cytotoxicity.
The test article did not induce cytotoxic effects and did not inhibit cell proliferation after being exposed for 24 to 48 hours in L929 colorectal carcinoma cells
- ISO 10993-10:2006, Biological evaluation of medical devices Part 10: Tests for . irritation and delayed-type hypersensitivity (Skin Sensitization). Under the conditions of this study, the test article extracted by Sodium Chloride 0.9% Inj. showed no evidence of causing delayed dermal contact sensitization in guinea pigs.
- . ISO 10993-10:2006, Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity (Ocular Irritation). All grades of ocular irritation calculated by maximum average score were zero. Based on the conditions and results of the study, the test article extract administered to NZW rabbits by ocular instillation did not cause any observable irritation response in vivo. All data generated from this study will provide as safety reference for human exposure.
- . ISO 10993-11:2006, Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity.
Under the conditions of this study, there was no mortality or evidence of systemic toxicity from the test article extract injected into ICR mice. All data generated from this study will provide as safety criteria for human exposure.
- . ISO 10993-12:2007, Biological evaluation of medical devices -- Part 12: Sample preparation and reference materials. (Biocompatibility).
#### Leakage Testing
Production units of the Polaris Dial-a-Date contact lens case have undergone leakage testing
The Polaris Dial-a-Date contact lens case was tested for leakage following a written test protocol. 192 combinations of different tops and bottoms were filled to 2/3 with liquid. Each set was turned upside down for 15 minutes and the tests were repeated 3 times. None of the tested lens cases showed any leakage and all of the Leakage Tests passed successfully. The results were documented
The Polaris Dial-a-Date Contact Lens Case complies with the voluntary standards as detailed in Section 9 Declarations of Conformity and Summary Reports of this submission.
{4}------------------------------------------------
K120904
Image /page/4/Picture/1 description: The image shows the word "RELIANCE" in bold, black letters. To the left of the word is a black graphic that appears to be a stylized bird or eagle. The text is the main focus of the image, and the graphic serves as a visual element associated with the brand or company name.
# Safety and Effectiveness
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the differences between the Polaris Diala-Date Contact Lens Case and the predicate device do not raise any questions regarding its safety and effectiveness. The Polaris Dial-a-Date Contact Lens Case, as designed and manufactured, therefore is determined to be substantially equivalent to the referenced predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a circular logo for the Department of Health & Human Services - USA. The logo features the department's name in a circular arrangement around the perimeter. In the center is a stylized symbol, possibly representing a bird or a flame, composed of three curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
November 26, 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
NOV 2 6 2012
Reliance Design and Manufacture Corp. % Ms. Caroline Tontini International Project Manager Emergo Group, Incorporated 611 West 5th Street, Third Floor Austin, TX 78701
K120904 Re:
> Trade/Device Name: Polaris Dial-a-Date Contact Lens Case Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LRX Dated: October 22, 2012 Received: October 31, 2012
Dear Ms. Tontini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
**Kesia Y. Alexander**
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K120904
Device Name: Polaris "Dial-A-Date" Contact Lens Case
Indications for Use:
Polaris 'Dial-A-Date' is a contact lens case for storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses. Use for storage during chemical disinfection only.
Not for use during heat disinfection.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
MARC ROBBOY
sion Sign-O Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.