RANGE SPINAL SYSTEM, UNIT ROD

K120899 · K2m, Inc. · MNI · May 25, 2012 · Orthopedic

Device Facts

Record IDK120899
Device NameRANGE SPINAL SYSTEM, UNIT ROD
ApplicantK2m, Inc.
Product CodeMNI · Orthopedic
Decision DateMay 25, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Range Spinal System is comprised of the DENALI, DENALI DEFORMITY, and MESA Spinal Systems and the ARI Anterior Vertebral Body Staples, all of which are cleared for the following indications: Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scollosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L5 to sacrum) with removal of the implants after the attainment of a solid fusion. Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

Device Story

Range Spinal System is a top-loading, posterior thoracic-lumbar spinal fixation system; components include pedicle screws, rods, locking set screws, and hooks. System functions as adjunct to fusion to provide immobilization and stabilization of spine. Used in clinical settings by surgeons; implanted during spinal surgery. Benefits include stabilization of spinal segments to facilitate fusion. Submission adds a unit rod to existing system.

Clinical Evidence

Bench testing only. Modified implants underwent static compression, static torsion, and dynamic compression testing in accordance with ASTM F1717. Results demonstrated substantial equivalence to predicate devices.

Technological Characteristics

Materials: Titanium Alloy and Cobalt Chrome per ASTM and ISO standards. System type: Top-loading, posterior thoracic-lumbar spinal fixation. Components: Pedicle screws, rods, locking set screws, hooks, and unit rod. Mechanical testing performed per ASTM F1717.

Indications for Use

Indicated for skeletally mature patients requiring posterior or anterolateral thoracolumbar stabilization as an adjunct to fusion. Conditions include trauma (fracture/dislocation), spinal stenosis, curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, failed previous fusion, and degenerative disc disease. Severe spondylolisthesis (grades 3-4) of L5-S1 specifically indicated for fusion with autogenous bone graft.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K120899 ## 510(k) Summary for the Range Spinal System: Unit Rod MAY 2 5 2012 This 510(k) summary for the Range Spinal System is provided as required per Section 513(i)(3) of the Food, Drug and Cosmetic Act. 1. Submitter : K2M, Inc. 751 Miller Drive SE, Suite F 1 Leesburg, VA 20175 Contact Person : Nancy Giezen K2M, Inc. Telephone: 703-777-3155 Date Prepared: 05/07/12 | 2. Tradename: | Range Spinal System | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Spinal Fixation System | | Classification Name: | Pedicle Screw Spinal System (21CFR 888.3070)<br>Spinal Interlaminal Fixation Orthosis (21CFR 888.3050)<br>Orthosis, Spondylolisthesis Spinal Fixation (21CFR 888.3070) | | Device Product Code: | MNI, KWP, MNH | | Regulatory Class: | Class II | # 3. Predicate or legally marketed devices which are substantially equivalent : - call of legally marketed 201000 129, K072914,K080792, K081301, K100851, K112037, K112920) . - Medtronic TSRH (K110676) . - Synthes USS, Click'X, VAS, Dual -Opening, Small Stature (K022949) . #### 4. Description of the device: The Range Spinal System is a top-loading, multiple component, posterior (thoracic-lumbar) spinal fixation system which consists of pedicle screws, rods, locking set screws, and hooks. Materials: The devices are manufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards. Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of the posterior thoracic and lumbar spine. ### 5. Purpose of Submission: The purpose of this submission is to add a unit rod to the system. #### 6. Intended Use: The Range Spinal System is comprised of the DENALI, DENALI DEFORMITY, and MESA Spinal The Rango Spinal Bycent is compressional Body Staples, all of which are cleared for the following indications: Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scollosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the Kyphosis, after of severe spondylolisthesis ( grades 3 and 4 ) of the LS-S i vertebra in skeletally mature patients of 1 2 t treating fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion. Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral the discoses (DD Non-cervical, noll-podler spinal thation are nees following indications: degenerative disc disease (DDD) Page 1 of 2 {1}------------------------------------------------ K120899 (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and (detailed over pair of anooglolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. - 7. Comparison of the technological characteristics of the device to predicate and legally marketed devices : The design features and sizing of the components were compared to predicate devices and the Range Spinal System was found to be substantially the same as these systems. - 8. Comparison of the performance data of the device to predicate and legally marketed devices : The modified implants were tested were tested in static compression, static torsion and dynamic compression in accordance with ASTM F1717 and compared with the original Range Spinal System components. The modified implants were determined to be substantially equivalent to the predicate devices. #### 9. Conclusion: - There are no significant differences between the Range Spinal System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use. Page 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY 2 5 2012 K2M, Inc. % Ms. Nancy Giezen 751 Miller Drive SE, Suite F1 Leesburg, Virginia 20175 Re: K120899 Trade/Device Name: Range Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI, KWP Dated: May 09 , 2012 Received: May 10, 2012 Dear Ms. Nancy Giezen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Ms. Giezen comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark A. Millenson Mark N. Melkerson · Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # K 120899 # Indications for Use 510(k) Number (if known): KI20899 Device Name: Range Spinal System Indications for Use: The Range Spinal System is comprised of the DENALI DEFORMITY, and MESA Spinal Systems and the ARI Anterior Vertebral Body Staples, all of which are cleared for the following indications: Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( LS to sacrum) with removal of the implants after the attainment of a solid fusion. Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion for the following indications: degenerative disc disease (DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ; spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. Prescription Use X (Part 21 CFR 801 Subpart D) ## AND/OR Over-the-counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS-LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) signature (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices K120899 510(k) Number_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...