K120839 · Tangent Medical Technologies, Inc. · FOZ · Jul 20, 2012 · General Hospital
Device Facts
Record ID
K120839
Device Name
NOVACATH SECURE IV CATHETER SYSTEM-18G
Applicant
Tangent Medical Technologies, Inc.
Product Code
FOZ · General Hospital
Decision Date
Jul 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NovaCath™ Secure IV Catheter System is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature aids in the prevention of needlestick injuries. Upon catheter insertion, blood is contained within the device to aid in the prevention of blood exposure. The NovaCath™ may be utilized in any patient population with consideration given to vascular anatomy and appropriateness of procedure. The 18, 20 and 22 Gauge NovaCath™ devices are suitable for use with power injectors for up to a maximum of 300 psi.
Device Story
NovaCath™ Secure IV Catheter System is an over-the-needle peripheral intravascular catheter; features integrated extension tubing, secondary stabilization hub, female luer lock, clamp, and passive needle-shielding mechanism. Device functions as a closed system to contain blood during insertion and prevent exposure. Used by clinicians in clinical settings for vascular access; enables blood sampling, pressure monitoring, and fluid administration. Passive needle-shielding aids in preventing needlestick injuries. Stabilization platform and wings provide 360° catheter securement. 18G, 20G, and 22G versions support power injection up to 300 psi. Clinicians observe flashback via dual-flash feature to confirm vessel entry. Device benefits include reduced blood exposure risk and improved catheter stability.
Clinical Evidence
No prospective clinical trials were conducted. Evidence consists of bench testing and a simulated clinical use study performed in accordance with FDA Guidance for Medical Devices with Sharps Injury Prevention Features. Biocompatibility testing was conducted per ISO 10993 and FDA Blue Book Memorandum #G95-1. Performance testing confirmed compliance with ISO 10555-1 and ISO 10555-5 requirements.
Technological Characteristics
Radiopaque, flexible polyurethane catheter; integrated extension tubing with Y-configuration; passive needle-shielding mechanism; female luer lock. Stabilization via platform and wings. EO sterilized (ANSI/AAMI/ISO 11135-2). Materials meet ISO 10993 biocompatibility standards. Compatible with power injectors (18G, 20G, 22G) up to 300 psi. Shelf-life validated per ASTM F1980-07.
Indications for Use
Indicated for patients requiring short-term (less than 30 days) vascular access for blood sampling, blood pressure monitoring, or fluid administration. Suitable for any patient population based on vascular anatomy and procedure appropriateness. 18G, 20G, and 22G models are compatible with power injectors up to 300 psi.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Retractable Technologies' VanishPoint® IV Catheter (K051355)
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 20 2012
# TANGENT MEDICAL TECHNOLOGIES, INC. NOVACATH™ SECURE IV CATHETER SYSTEM
510(k) SUMMARY
## July 11, 2012
The following summary is provided pursuant to Section 513 (I) (3) (A) of the Federal Food Drug and Cosmetic Act:
Submitter Information: Tangent Medical Technologies, Inc. 8170 Jackson Road, STE A Ann Arbor, MI 48103
9/23/2005.
| Contact Information: | Kay Fuller, RAC<br>VP, Regulatory and Clinical Affairs<br>kay @tangentmedical.com<br>1+734-527-4060 / 1+734-846-7852 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | NovaCath™ Secure IV Catheter System |
| Proprietary Name: | NovaCath™Secure IV Catheter System |
| Common Name: | Intravascular Catheter (short-term) |
| Classification Name: | Catheter, intravascular, therapeutic, short-term less than 30 days |
| Classification Code: | FOZ |
| Regulation Number: | 21 CFR § 880.5200 |
| Predicate Device Equivalence: | The Tangent NovaCath™ Secure IV Catheter System is<br>substantially equivalent to the BD Nexiva™ Device cleared for<br>U.S. commercialization via K032843 on 2/27/200, the BD Insyte®<br>Catheter(s), cleared for US commercialization via K971339 on |
12/24/1997 and the Retractable Technologies' VanishPoint® IV Catheter , cleared for US commercialization via K051355, on
{1}------------------------------------------------
#### Device Description:
The NovaCath™ Secure IV Catheter System is comprised of an over-the-needle, peripheral intravascular catheter made from a radiopaque slender, flexible plastic, integrated extension tubing with a secondary stabilization hub system, a female luer lock and clamp, and a passive needle shielding mechanism. The design of the NovaCath™ System is considered a closed system since it protects users from blood exposure during the catheter insertion procedure. The needle is withdrawn through a septum that will seal after the needle has been withdrawn and the luer lock is closed, blood is contained within the NovaCath™ System during insertion. The clamp on the extension tubing is provided to eliminate blood exposure when the vent plug is replaced with an infusion set connection or other end cap.
The NovaCath™ Secure IV Catheter System is inserted into the Indications for Use: patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature aids in the prevention of needlestick injuries. Upon catheter insertion, blood is contained within the device to aid in the prevention of blood exposure. The NovaCath™ System may be utilized in any patient population with consideration given to vascular anatomy and appropriateness of procedure. The 18, 20 and 22 gauge NovaCath™ System catheters are suitable for use with power injectors for a maximum of 300 psi.
> The technological similarities between the subject NovaCath™ Secure IV Catheter System and the noted predicate devices are substantially equivalent. There are no new questions raised regarding safety or efficacy of the subject NovaCath™ Secure IV Catheter System. Table 1, below provides a summary comparison of the technological characteristics.
Comparison of Technological Characteristics:
{2}------------------------------------------------
# Table 1: Comparison Between NovaCath™ Secure IV Catheter System and Predicates
| | NovaCath<br>Secure IV Catheter System | K032843<br>Nexiva | K971339<br>Insyte | K051355<br>VanishPoint |
|------------------------------------------|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|------------------------------------------------------------------------------|
| Comparison Criteria | | | | |
| Same Intended Use | Yes | Yes | Yes | Yes |
| Prescription Device (Rx Only) | Yes | Yes | Yes | Yes |
| Biocompatible Polyurethane Catheter | Yes | Yes | Yes | Yes |
| Biocompatible Materials of Fabrication | Yes | Yes | Yes | Yes |
| Closed System | Yes | Yes | No | No |
| Built-In Extension from Main Assembly | Yes, connecting tubing & reverse Y-configuration | Yes, connecting tubing & Y-configuration | No | No |
| Flashback Visualization | Yes, dual flash feature, allows for visualization at pre-priming & upon insertion | Yes, single, allows flash view upon insertion | Yes, single, allows flash view upon insertion | Yes, single, allows flash view upon insertion |
| Sharps Injury Prevention Feature | Yes, integrated sharps injury prevention device (passive) | Yes, integrated sharps injury prevention device | Yes integrated sharps injury prevention device | No, Automatic needle retraction occurs with depression of the needle housing |
| General Specifications | Yes, overall size is consistent w/ other IV catheter devices once extension tubing is added | Yes, overall size is slightly different than other IV catheter devices, due to Y-connector tubing & manual needle retraction | Yes, overall size is consistent w/ other IV catheter devices | Yes, overall size is consistent with other IV catheter devices |
| Intended anatomical location | Yes, peripheral vasculature | Yes, peripheral vasculature | Yes, peripheral vasculature | Yes, peripheral vein or artery |
| Distal End Configuration | Yes, Catheter & Needle consistent w/ ISO 10555-5 | Yes, Catheter & Needle consistent w/ ISO 10555-5 | Yes, Catheter & Needle consistent w/ ISO 10555-5 | No, Hub w/ female locking luer fitting per ISO 594 |
| Proximal End Configuration | Yes, female luer access port | Yes, female luer access port & open hub | Yes, Open hub | Yes, Tube w/ tapered tip & one outlet. No side ports<br>UNK |
| Flow Rate (ml/min) | Yes,<br>20G: 52.55 ml/min | Yes,<br>20G 45 ml/min | Yes,<br>20G: 65 ml/min | UNK |
| Mechanical | Needle retraction is manual | Needle retraction is manual | Stainless steel spring for needle retraction | Automatic needle retraction w/slight depression of the needle housing |
| Catheter Stabilization | Yes, primary & secondary stabilization (platform and wings) provide 360° stabilization | Yes, stabilization wings | Yes, available w/winged hub | No |
| EO Sterilization Method | Yes | Yes | Yes | Yes |
| Packaged Sterile, Single Use | Yes | Yes | Yes | Yes |
| Non-pyrogenic | Yes | Yes | Yes | Yes |
| Made without Latex Rubber? | Yes | Yes | Yes | Yes |
| Made without DEHP? | Yes | Yes | No | Yes |
| Multiple Gauge Sizes & Needle Lengths | Yes | Yes | Yes | No |
| For use w/ Power Injectors up to 300 psi | Yes<br>18G,20G,22G; | Yes | Yes | No |
{3}------------------------------------------------
# Summary of Device Evaluation:
The NovaCath™ Secure IV Catheter System is substantially equivalent in device description, intended use, function, principle of operation, and basic composition to the predicate devices noted herein.
The NovaCath™ Secure IV Catheter System has been designed and tested to meet the requirements of relevant FDA consensus standards, voluntary standards and FDA regulations and guidance documents applicable to the subject and predicate devices. Results of the non-clinical testing support the conclusion of substantial equivalence to the NovaCath™ Secure IV Catheter System to the predicate devices noted herein.
#### Performance Testing:
The NovaCath™ Secure IV Catheter System has been designed and successfully tested to meet the applicable requirements outlined in ISO 10555-1 Sterile, single use intravascular catheters - Part 1 General requirements and ISO 10555-5 Sterile, single use intravascular catheters - Part 5 Over-needle perioheral catheters.
#### Biocompatibility Testing:
The NovaCath™ Secure IV Catheter System is classified as an external communicating, prolonged (24hrs to less than 30 days) circulating blood contact device. The device was successfully tested per the FDA Blue Book Memorandum - #G95-1 Table 1 Initial Evaluation Test for Consideration and relevant ISO 10993 related biocompatibility standards to establish that the NovaCath™ Secure IV Catheter System meets the appropriate biocompatibility testing requirements.
{4}------------------------------------------------
Sterilization and Shelf-life Testing:
The NovaCath™ System will be released to market with a Sterility Assurance Level of 10-6, per the requirements set forth in ANSI/AAMI/ISO 11135-2. The NovaCath™ System is intended for single use only and not intended for reuse or re-sterilization by the user. The maximum levels of residues of ethylene oxide and ethylene chlorohydrin will not exceed the limits presented in ISO 10993-7. Shelf-life and expiry dating meet the requirements of ASTM F1980-07, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, which is also supported by real-time shelf -life data.
#### Clinical Data:
A NovaCath™ Secure IV Catheter System simulated clinical use study was successfully conducted in accordance with: FDA Guidance for Medical Devices with Sharps Injury Prevention Features.
No prospective clinical trials were conducted in support of this Traditional 510(k).
Conclusions:
Based on the information contained herein, we conclude the NovaCath™ Secure IV Catheter System is substantially equivalent to the noted legally marketed Predicate devices and is safe and effective for its intended use.
pwie Sats
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Kay Fuller, RAC Vice President, Regulatory & Clinical Affairs Tangent Medical Technologies, Incorporated 8170 Jackson Road, Suite A Ann Arbor, Michigan 48103
JUL 20 2012
Re: K120839
Trade/Device Name: NovaCath™ Secure IV Catheter System Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: July 3, 2012 Received: July 5, 2012
Dear Ms. Fuller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Ms. Fuller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 1 1 5809 htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Antony D. arar
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K120839
Device Name: NovaCath™ Secure IV Catheter System
### Indications for Use:
The NovaCath™ Secure IV Catheter System is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature aids in the prevention of needlestick injuries. Upon catheter insertion, blood is contained within the device to aid in the prevention of blood exposure. The NovaCath™ may be utilized in any patient population with consideration given to vascular anatomy and appropriateness of procedure. The 18, 20 and 22 Gauge NovaCath™ devices are suitable for use with power injectors for up to a maximum of 300 psi.
Over-The-Counter Use Prescription Use × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Atricia Bustos for David Chapman
Ivision of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _K 120039 Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.