Pre-production clinical evaluations by dentists; Post-market surveillance/clinical practice data from European CE-marked sales
The sponsor utilized field evaluations and four months of European clinical use data to support the safety and performance profile of the device.
Clinical evaluations; Post-market surveillance; Field reports
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Pre-production clinical evaluations and European market surveillance; Observational/Field evaluation; Follow-up/Duration: Four months (European market); Study Period: Pre-submission
Patients receiving temporary prostheses in routine dental practice; Number of Sites: Multiple (dentist offices)
Not applicable for this study
Clinical performance and adverse event reporting
Indications for Use
Pulpdent Tuff-Temp 2.0 is a dual-cure, glass-filled, resin composite used by the dental professional to make a temporary prosthesis, such as a crown or bridge, to be used until a permanent restoration can be fabricated.
Device Story
Pulpdent Tuff-Temp 2.0 is a dual-cure, glass-filled resin composite material used by dental professionals in clinical settings to fabricate temporary dental prostheses (crowns/bridges). The material is supplied in two parts; it utilizes both chemical and photo-initiators to achieve polymerization. The device is applied by the clinician to create a temporary restoration that remains in the patient's mouth until a permanent restoration is fabricated. The material features improved esthetics, reduced oxygen-inhibited layer, and lower heat rise during setting compared to its predecessor. It provides a temporary functional and esthetic solution for patients during the interim period of restorative dental treatment.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing (flexural strength 84.0 MPa, compressive strength 205.0 MPa, heat rise 5.4°C) and pre-production clinical evaluations by dental consultants. The device has been marketed in Europe as a CE-marked device with no reported adverse events. Biocompatibility is supported by the established safety profile of the original Tuff-Temp formulation.
Technological Characteristics
Dual-cure, glass-filled resin composite. Composition: blended dimethacrylate resins, chemical and photo-initiators, barium glass, and submicron silica fillers. Form factor: two-part system. Setting mechanism: dual-cure (light-cure and self-cure).
Indications for Use
Indicated for use by dental professionals to fabricate temporary prostheses, including crowns and bridges, for patients requiring interim restorations until a permanent restoration is completed.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
{0}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
# 510(k) SUMMARY
March 13, 2012
Kenneth J. Berk 80 Oakland Street Watertown, MA 02472 USA Telephone: 617-926-6666 Fax: 617-926-6262 Email: ken@pulpdent.com
DEVICE: Pulpdent Tuff-Temp 2.0 Trade Name: Classification Name: Temporary Crown and Bridge Resin Class: Il FDA Product Code: 76 EBG, 21 CFR Part 872.3770
#### PREDICATE DEVICES:
Pulpdent Tuff-Temp Luxatemp, DMG
#### DESCRIPTION AND INTENDED USE:
Pulpdent Tuff-Temp 2.0 is a dual-cure, glass-filled, resin composite used by the dental professional to make a temporary prosthesis, such as a crown or bridge, to be used until a permanent restoration can be fabricated.
## COMPARISON WITH PREDICATE PRODUCTS:
Pulpdent Tuff-Temp 2.0 is substantially equivalent in design, composition, performance and intended use to the predicate products. The predicate products have been found substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Devices under CFR 872.3770. Tuff-Temp 2.0 is an enhanced version of the original Tuff-Temp with improved esthetics, reduced oxygen inhibited layer and lower heat rise on setting.
| Product | Description | Intended Use | Composition |
|-------------------------------|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pulpdent Tuff-Temp 2.0 | Dual-cure, glass-filled<br>composite material in<br>two parts | To make a temporary<br>prosthesis, such as a crown<br>or bridge, for use until a<br>permanent restoration can<br>be fabricated. | Blended dimethacrylate resins<br>Chemical and Photo-initiators<br>Barium glass and submicron silica fillers |
| Pulpdent Tuff-Temp<br>K081810 | Dual-cure, glass-filled<br>composite material in<br>two parts | To make a temporary<br>prosthesis, such as a crown<br>or bridge, for use until a<br>permanent restoration can<br>be fabricated. | Blended dimethacrylate resins<br>Chemical and Photo-initiators<br>Barium glass and submicron silica fillers |
| DMG Luxatemp Solar<br>K013674 | Dual cure, bis-acryl<br>composite material in<br>two parts | To make a temporary<br>prosthesis, such as a crown<br>or bridge, for use until a<br>permanent restoration can<br>be fabricated. | Bis-acryl resins (Urethane dimethacrylate, aromatic<br>dimethacyrlate, glycol methacrylate)<br>Chemical and Photo-initiators<br>Barium glass and submicron silica fillers |
{1}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
### SUMMARY OF PERFORMANCE TESTING - BENCH:
The following test results demonstrate that Tuff-Temp 2.0 performs as intended:
| Flexural strength | $84.0 \pm 4.0$ MPa | |
|------------------------------|---------------------------------|-------------------|
| Compressive strength | $205.0 \pm 18.0$ MPa | |
| Deflection at break | $2.3 \pm 0.3$ mm | |
| Heat rise | 5.4°C | |
| | <b>Self-cure</b> | <b>Light cure</b> |
| Initial setting time | 90 seconds | 20 seconds |
| Final self-cure setting time | 5 minutes from beginning of mix | Not applicable |
#### B10COMPATIBILITY:
Pre-production clinical evaluations of Tuff-Temp 2.0 by dentists in the field and by dentist consultants were positive. Tuff-Temp 2.0 has been sold in Europe as a CE marked medical device for the past four months. Its use in clinical practice has met both user and patient needs and there have been no reports of adverse events.
The original Tuff-Temp has been on the market for two years. Field Reports from sales people, dentists. and other customers have praised Tuff-Temp for its qualities. During that time there have been no reports of adverse events, i.e., allergic reactions, soft tissue irritation, etc. There is sufficient similarity between Tuff-Temp and Tuff-Temp 2.0 to expect that it will have equivalent biocompatibility.
#### CONCLUSION:
From the above comparisons, the bench testing and more than ten years of problem-free use of similar temporary crown and bridge resins in the dental profession, it can be concluded that Pulpdent Tuff-Temp 2.0 is substantially equivalent in design, composition, performance and intended use to the predicate products listed above.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol on the right. To the left of the symbol is a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the perimeter of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN - 8 2012
Mr. Kenneth J. Berk Director Pulpdent Corporation 80 Oakland Street Watertown, Massachusetts 02472
Re: K120784
Trade/Device Name: Pulpdent Tuff-Temp 2.0 Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown and Bridge Resin Regulatory Class: II Product Code: EBG Dated: March 13, 2012 Received: March 15, 2012
Dear Mr. Berk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Mr. Berk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh foc
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 6120784 Indications for Use
# 510(k) Number: 《 |Q()7&l)
Device Name: Pulpdent Tuff-Temp 2.0
## Indications For Use:
Pulpdent Tuff-Temp 2.0 is a dual-cure, glass-filled, resin composite used by the dental professional to make a temporary prosthesis, such as a crown or bridge, to be used until a permanent restoration can be fabricated.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR .
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runse
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
K1267 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.