AERODR STITCHING SYSTEM
K120752 · Konica Minolta Medical & Graphic, Inc. · KPR · Jun 8, 2012 · Radiology
Device Facts
| Record ID | K120752 |
| Device Name | AERODR STITCHING SYSTEM |
| Applicant | Konica Minolta Medical & Graphic, Inc. |
| Product Code | KPR · Radiology |
| Decision Date | Jun 8, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1680 |
| Device Class | Class 2 |
Intended Use
The AeroDR Stitching System is used with Konica Minolta AeroDR SYSTEM which is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures. This device is used for examinations of long areas of anatomy such as the leg and spine. This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.
Device Story
AeroDR Stitching System is an accessory for the Konica Minolta AeroDR SYSTEM; extends capability to capture long-length radiographic images (up to 1196mm x 349mm). System components: Stitching Unit, X-ray Auto Barrier Unit, Power Supply Unit. Operation: 14x17 inch AeroDR Detector moves vertically within the Stitching Unit; captures up to 3 sequential X-ray images; images transferred to Console CS-7. Console CS-7 performs image stitching to create a single long-length diagnostic image. Used in clinical/hospital radiology departments by trained personnel. Output allows clinicians to visualize and diagnose long anatomical structures (e.g., spine, leg) in a single view, facilitating clinical decision-making for orthopedic or alignment assessments.
Clinical Evidence
Bench testing only. Validation performed using sample images and non-clinical testing under simulated use conditions to confirm performance, safety, and effectiveness against predefined acceptance criteria.
Technological Characteristics
Accessory to AeroDR SYSTEM (FPD-based). Components: Stitching Unit, Auto Barrier, Power Supply. Vertical exposure range varies by SID (max 50" at 2.4m). Operation environment: 15-30C, 30-80% RH. Connectivity: Integrates with Console CS-7. Standards: IEC 60601-1 (Safety), IEC 60601-1-2 (EMC).
Indications for Use
Indicated for generating radiographic images of human anatomy, specifically long areas such as leg and spine, in general-purpose diagnostic procedures. Not for mammography, fluoroscopy, tomography, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- REGIUS MODEL 170 (K051998)
Reference Devices
- Konica Minolta AeroDR SYSTEM (K102349)
- Console CS3000 (K051523)
Related Devices
- K021829 — MODIFICATION TO KODAK DIRECTVIEW CR 800/900 SYSTEM · Eastman Kodak Company · Jul 2, 2002
- K122119 — DX-D FULL LEG FULL SPINE · Afga Healthcare N.V. · Sep 13, 2012
- K081836 — CARESTREAM DR LONG LENGTH IMAGING SYSTEM FOR DR 7500 SYSTEM · Carestream Health, Inc. · Jul 30, 2008
- K130567 — DR LONG LENGTH IMAGING SOFTWARE · Carestream Health, Inc. · Jun 11, 2013
- K042602 — REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM WITH MANUAL IMAGE PASTING · Ge Medical Systems, LLC · Jan 13, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K120752
### 510(k) Summary
as required by 807.92
JUN - 8 2012
#### 1. Company Identification
Konica Minolta Medical & Graphic, Inc. No.1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Establishment Registration Number: 3004485675
#### 2. Submitter's Name and Address
Shigeyuki Kojima Manager Regulations and Standards Section, Quality Assurance Center No. 1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Telephone: 81-42-589-8429 Fax: 81-42-589-8053
- 3. Date of Submission
March 9, 2012
- 4. Device Trade Name AeroDR Stitching System
- 5. Classification, Product Code Class II, 21CFR892.1680, 90 KPR
#### 6. Predicate Device
REGIUS MODEL 170. 510(k) number K051998
#### 7. Indications for Use
The AeroDR Stitching System is used with Konica Minolta AeroDR SYSTEM which is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures. This device is used for examinations of long areas of anatomy such as the leg and spine. This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.
#### 8. Device Description
The AeroDR Stitching System is used with 510(k) cleared Konica Minolta AeroDR SYSTEM (K102349) which is indicated for use in generating radiographic images of human anatomy.
The AeroDR Stitching System is an accessory of stationary X-ray system which
1
{1}------------------------------------------------
extends the capability of the AeroDR SYSTEM to allow the capture of long length images with an image area up to 1196mm x 349mm. It consists of AeroDR Stitching Unit, AeroDR Stitching X-ray Auto Barrier Unit and Power Supply Unit.
The AeroDR Detecter (K102349) loaded in the AeroDR Stitching Unit takes up to 3 images and transfer them to the Console CS-7 (K102349). Combining the transferred images in the CS-7 enables diagnosis of long images.
## 9. Risk Analysis (Summary of Design Control Activities)
The Risk Analysis for the AeroDR Stitching System was conducted on the basis of ISO14971, "Medical devices - Application of risk management to medical devices". As a result of risk control measures, the risk associated with all of the identified hazards was reduced to an acceptable level or ALARP (as low as reasonably practicable).
## 10. Compliance of Recognized Consensus Standard
In terms of Indications for use of this AeroDR Stitching System, it has been tested and · shown to meet the requirements of IEC 60601-1 and IEC 60601-1-2 as follows. Safety standard: IEC60601-1 Ed.2 (1988) + A1 (1991)+A2(1995) Electromagnetic Compatibility: IEC60601-1-2 Ed.3(2007)
## 11. Substantial Equivalence to Predicate Device
The predicate device is our REGIUS MODEL 170, 510(k) number K051998. Both AeroDR Stitching System and the predicate are used for examinations of long areas of anatomy such as the leg and spine.
Comparison of the principal characteristics of these devices, please refer to the Substantial Equivalence Comparison Table as follows.
| | | Medical Device Applied for Clearar |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant, etc. | Company : KonicaMinoIta Medical | Company: KonicaMinoIta Medical &Graphic. |
| | &Graphic, Inc. | Inc. |
| | Product Name : REGIUS MODEL 170 | Product Name: AeroDR Stitching System |
| | CR cassette:RC-110T/RC-110R | |
| | K Number : K051198 | |
| Configuration | The REGIUS MODEL 170 is an X-ray | The AeroDR Stitching System is used with |
| | image reader which uses a stimulable | 510(k) cleared Konica Minolta AeroDR |
| | phosphor plate (Plate) as X-ray detector | SYSTEM (K102349). |
| | installed in a separate cassette. | The AeroDR Stitching System extends the |
| | The REGIUS MODEL 170 consists of<br>REGIUS MODEL 170 image reader and CR<br>Cassette / Plate.<br>It reads the image recorded on the Plate<br>and transfers the image data to an<br>externally connected device such as<br>Console CS3000 (K051523).<br>Using the CR cassette with multiple (three)<br>stimulable phosphor plates aligned along<br>the long side, latent images are produced,<br>and the visual image is generated by<br>scanning with a separate reader device.<br>Combining the transferred images in the<br>CS3000 enables diagnosis of long images. | capability of the AeroDR SYSTEM to allow<br>the capture of long length images.<br>AeroDR Stitching System consists of<br>AeroDR Stitching Unit, AeroDR Stitching<br>X-ray Auto Barrier Unit and Power Supply<br>Unit.<br>The AeroDR Detecter (K102349) loaded in<br>the AeroDR Stitching Unit takes up to three<br>images and transfers them to the Console<br>CS-7 (K102349). Combining the transferred<br>images in the CS-7 enables diagnosis of long<br>images. |
| Principle of<br>Operation | Using the CR cassette with multiple<br>stimulable phosphor plates aligned along<br>the long side, position the cassette at the<br>back of the patient and make a single X-ray<br>exposure over the necessary area. Reload<br>the exposed image into the other cassettes,<br>and feed these cassettes into the reader<br>device in order so that the image is<br>independently scanned and multiple<br>images are stitched and displayed on the<br>console by the image processing. | Position the AeroDR Stitching unit at the back<br>of the patient. The AeroDR Stitching System<br>produces long length images of subjects by<br>moving a 14 x 17 inch AeroDR Detector up<br>and down inside the AeroDR Stitching Unit<br>and takes up to 3 images. Transfer these<br>multiple images to the Console CS-7<br>(K102349) and stitch multiple images by<br>image processing on the console. |
| Specifications | REGIUS MODEL 170<br>CR cassette:RC-110T<br>1) Vertical Exposure range<br>14×42inch (14×14×3plates)<br>11×24inch (11×12×2plates)<br>10×36inch (10×12×3plates)<br>14×51inch (14×17×3plates)<br>2) Sampling Pitch : 87.5/175μm<br>3)Operation Environment<br>Temperature : 15~30 C<br>Humidity : 30~80%RH | AeroDR Detector (K102349) : 14×17inch<br>FPD<br>Pixel Size: 175μ<br>AeroDR Stitching System (Device)<br>1) Vertical Exposure range<br>SID=2.4m : Max. 50"(3shoots)<br>SID=2.0m : Max. 41"(2shoots)<br>SID=1.5m : Max. 31"(2shoots)<br>Width 14inch only<br>2)Operation Environment<br>Temperature : 15~30 C<br>Humidity : 30~80%RH |
| Indications for<br>Use | The Direct Digitizer, REGIUS MODEL 170<br>is an X-ray image reader which uses a<br>stimulable phosphor plate (Plate) as X-ray<br>detector installed in a separate cassette. It | The AeroDR Stitching System is used with<br>Konica Minolta AeroDR SYSTEM which is<br>indicated for use in generating radiographic<br>images of human anatomy. It is intended to |
| reads the image recorded on the Plate and<br>transfers the image data to an externally<br>connected device such as a host computer,<br>an order input device, an image display<br>device, a printer, an image data filing<br>device, and other image reproduction<br>devices.<br>REGIUS MODEL 170 is used to obtain<br>image data of long areas of anatomy such<br>as the whole spine or the whole lower leg.<br>REGIUS MODEL 170 is also used to obtain<br>image data to verify the position for a<br>radiotherapy location.<br>It is designed intended to use in a clinic, a<br>radiology department in a hospital and in<br>other medical facilities. It is not intended for<br>use with digital mammography system. | replace radiographic film/screen systems in<br>general-purpose diagnostic procedures.<br>This device is used for examinations of long<br>areas of anatomy such as the leg and spine.<br>This device is not indicated for use in<br>mammography, fluoroscopy, tomography<br>and angiography applications. | |
## AeroDR Stitching System Substantial Equivalence Comparison Table
{2}------------------------------------------------
:
{3}------------------------------------------------
## 12. Performance-Testing
Performance testing was conducted to verify the design output met the design input requirements and to validate AeroDR Stitching SYSTEM conformed to the defined user needs and intended uses upon the quality of the device software. Through validation results of sample images and non-clinical testing under simulated use conditions, safe, effectiveness and performances are confirmed the achievement of predefined acceptance criteria, as well as substantially equivalence to the predicate device, besides product safety and electromagnetic compatibility.
## 13. Conclusion
Comprehensively, we judged that the AeroDR Stitching System has the same technological characteristics as the predicate devices. This 510(k) has demonstrated substantial equivalence as the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Konica Minolta Medical & Graphic, Inc. % Mr. Russel Munves Official Correspondent Storch, Amini & Munves, P.C. 140 East 45th Street, 25th Floor Two Grand General Tower NEW YORK NY 10017
Re: K120752
Trade/Device Name: AeroDR Stitching System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II . Product Code: KPR Dated: May 22, 2012 Received: May 23, 2012
# JUN - 8 2012
#### Dear Mr. Munves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known)
: KI20752
Device Name
AeroDR Stitching System
Indications for Use:
The AeroDR Stitching System is used with Konica Minolta AeroDR SYSTEM which is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures. This device is used for examinations of long areas of anatomy such as the leg and spine. This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.
Prescription Use_ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Arshad D'Silva
Division Sign-off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K120752