K120679 · Fujifilm Medical Systems U.S.A, Inc. · LLZ · Jun 1, 2012 · Radiology
Device Facts
Record ID
K120679
Device Name
SYNAPSE 3D NODULE ANALYSIS
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
LLZ · Radiology
Decision Date
Jun 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Synapse 3D Nodule Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Nodule Analysis accepts DICOM compliant medical images acquired from CT and PT. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Addition to Synapse 3D Base Tools, Synapse 3D Nodule Analysis is intended to: use non-contrast and contrast enhanced computed tomography (CT) images, provide custom workflows and UI, and reporting functions for nodule analysis including boundary detection of nodules based on the location specified by the user, measurement of nodules, temporal comparison of nodule images, and fusion of a CT nodule image on a PT nodule image. use positron emission tomography (PT) images, provide custom workflows and UI, and reporting functions for nodule analysis including SUV measurement of nodules with sphere VOI based on the location specified by the user, temporal comparison of nodule images, and fusion of a PT nodule image on a CT nodule image.
Device Story
Synapse 3D Nodule Analysis is an optional software module for the Synapse 3D platform; operates on Windows server/client configurations. Inputs: DICOM-compliant CT and PT images. Functionality: provides custom workflows, UI, and reporting for nodule analysis; performs boundary detection (user-specified location), nodule measurement, temporal image comparison, and image fusion (CT/PT). For PET images, calculates standardized uptake value (SUV) using sphere volume of interest (VOI). Used by trained medical professionals in clinical settings to assist in treatment planning and response evaluation (RECIST/PERCIST). Output: visualized analysis, measurements, and reports; supports clinical decision-making by providing quantitative data and comparative views of nodule progression.
Clinical Evidence
Bench testing only. Verification and validation activities, including testing against Software Requirements Specifications and Traceability Matrix, were performed to establish performance, functionality, and reliability.
Technological Characteristics
Software-based medical image processing module; runs on commercial general-purpose Windows-compatible hardware. Connectivity: DICOM-compliant. Software class: Moderate concern per FDA guidance. Features: boundary detection, nodule measurement, temporal comparison, image fusion, and SUV calculation.
Indications for Use
Indicated for trained medical professionals to aid in reading, interpreting, reporting, and treatment planning using CT and PT medical images. Not for primary diagnostic interpretation of mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Advantage Windows Lung Analysis: GE Healthcare (GE Medical Systems) (K042694)
{1}------------------------------------------------
K120679
FUJIFILM Medical Systems U.S.A., Inc. Synapse 3D Nodule Analysis 510(k)
# 5. 510(k) Summary
# Date Prepared:
February 28, 2012
### Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902
Telephone: (301) 251-1092 Fax: (203) 602-3785 Contact: Jyh-Shyan Lin
### Device Trade Name:
Synapse 3D Nodule Analysis
### Device Common Name:
Medical Image Processing and Analysis Software
### Requlation Number:
21 CFR 892.2050
#### Device Classification:
Class II
### Device Classification Name :
Picture Archiving and Communications System (PACS)
### Panel:
Radiology
### Product Code:
LLZ
### Date Received:
TBD
### Decision Date :
TBD
### Decision :
TBD
# JUN - 1 2012
{2}------------------------------------------------
## FUJIFILM Medical Systems U.S.A. Synapse 3D Nodule Analysis 510(k)
### ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
· Advantage Windows Lung Analysis: GE Healthcare (GE Medical Systems) (K042694).
### Description of the Device
Synapse 3D Nodule Analysis is an optional software module that works with Synapse 3D Base Tools (V3.0) (K120361, cleared on April 6, 2012). Synapse 3D Nodule Analysis, Synapse 3D Base Tools (V3.0), as well as other optional software modules, all belong to the Synapse 3D product family.
Synapse 3D is the medical application software running on Windows server/client configuration installed on a commercial general-purpose Windows-compatible computer. It offers software tools which can be used by trained medical professionals to interpret medical images obtained from various medical devices, to create reports, or to develop treatment plans.
The key features in Synapse 3D Nodule Analysis include boundary detection of nodules, measurement of nodules, temporal comparison of nodule images, and fusion of CT and PT nodule images on the non-contrast and contrast enhanced CT or PT images.
Synapse 3D Nodule Analysis offers physicians the following clinical applications in addition to the features available from Synapse 3D Base Tools (V3.0) to analyze the image data retrieved from CT and PT devices.
- . Response Evaluation Criteria in Solid Tumors (RECIST) Tracker Use non-contrast and contrast enhanced computed tomography (CT) images, provide custom workflows, user interface (UI), and reporting functions for nodule analysis including boundary detection of nodules (based on the location specified by the user), measurement of nodules, temporal comparison of nodule images, and fusion of a CT nodule image on a PT nodule image.
- Positron Emission Response Criteria in Solid Tumors (PERCIST) Tracker . Use positron emission tomography (PT) images, provide custom workflows user interface (UI), and reporting functions for nodule analysis including standardized uptake value (SUV) measurement of nodules with sphere VOI (based on the location specified by the user), temporal comparison of nodule images, and fusion of a PT nodule image on a CT nodule image.
## Indication for Use
Synapse 3D Nodule Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Nodule Analysis accepts DICOM compliant medical images acquired from CT and PT.
{3}------------------------------------------------
FUJIFILM Medical Systems U.S.A., Inc Synapse 3D Nodule Analysis 510(k)
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Addition to Synapse 3D Base Tools, Synapse 3D Nodule Analysis is intended to:
- use non-contrast and contrast enhanced computed tomography (CT) images, provide A custom workflows and UI. and reporting functions for nodule analysis including boundary detection of nodules based on the location specified by the user, measurement of nodules, temporal comparison of nodule images, and fusion of a CT nodule image on a PT nodule image.
- 公 use positron emission tomography (PT) images, provide custom workflows and UI, and reporting functions for nodule analysis including SUV measurement of nodules with sphere VOI based on the location specified by the user, temporal comparison of nodule images, and fusion of a PT nodule image on a CT nodule image.
### Technological Characteristics
Synapse 3D Nodule Analysis introduces no new safety or efficacy issues other than those already indentified with the predicate devices. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices."
### Testing
Synapse 3D Nodule Analysis is tested successfully with reference to its Software Requirements Specification, as well as design verification and validation documents and Traceability Matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Synapse 3D Nodule Analysis software, which is found to be safe and effective and substantially equivalent to the currently-cleared predicate device.
### Conclusion
This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract shapes that resemble human figures or birds in flight.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Jyh-Shyan Lin Senior Manager, Regulatory, Quality and Clinical Affairs FUJIFILM Medical Systems, U.S.A., Inc. 419 West Avenue STAMFORD CT 06902
JUN - 1 2012
Re: K120679
Trade/Device Name: Synapse 3D Nodule Analysis Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 28, 2012 Received: March 6, 2012
### Dear Mr. Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device . referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice mourements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket with anony four to FDA finding of substantial equivalence of your device to a legally marketed nonication. The I Drivinant ssification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 Jul deen before the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Synapse 3D Nodule Analysis
Indications for Use:
Synapse 3D Nodule Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading. interpreting, reporting, and treatment planning. Synapse 3D Nodule Analysis accepts DICOM compliant medical images acquired from CT and PT.
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Addition to Synapse 3D Base Tools, Synapse 3D Nodule Analysis is intended to;
- > use non-contrast and contrast enhanced computed tomography (CT) images, provide custom workflows and Ul, and reporting functions for nodule analysis including boundary detection of nodules based on the location specified by the user, measurement of nodules, temporal comparison of nodule images, and fusion of a CT nodule image on a PT nodule image.
- use positron emission tomography (PT) images, provide custom workflows and UI, and A reporting functions for nodule analysis including SUV measurement of nodules with sphere VOI based on the location specified by the user, temporal comparison of nodule images, and fusion of a PT nodule image on a CT nodule image.
Prescription Use x AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
---
(Division Sign Off)
Division of Radiological Devices Office of In 510K
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.