The Mahurkar™ Elite Acute Dual Lumen Catheter is intended to be used for short-term central venous access for hemodialysis, apheresis, and infusion. The Mahurkar™ Elite Acute Triple Lumen Catheter is intended to be used for short-term central venous access for hemodialysis, apheresis, infusion, central venous pressure monitoring, and pressure injection of contrast media. The maximum recommended infusion rate is 5 ml/sec for power injection of contrast media.
Device Story
Mahurkar™ Elite Acute Catheters are central venous catheters for short-term hemodialysis, apheresis, and infusion. Dual lumen version features two lumens with color-coded adapters; triple lumen version adds a third lumen for infusion and pressure monitoring. Catheters consist of a hub, extension tubes, and a shaft with side slots at the distal tip. Available in 12.0 Fr or 13.5 Fr (dual) and 12.5 Fr (triple) diameters, with lengths 13-30 cm. Used by clinicians in clinical settings for venous access. Triple lumen model supports pressure injection of contrast media. Devices are provided as single units or convenience kits.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench-top functional and performance testing, including cleaning/lock solution compatibility, dynamic flow, column strength, clamp functionality, shaft stiffness, tensile strength, and resistance to kink, leak, burst, catheter collapse, and fatigue. Triple lumen catheter specifically evaluated for central venous pressure monitoring and simulated contrast media injection. Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Central venous catheters; dual or triple lumen configurations. Materials evaluated for biocompatibility per ISO 10993. Features color-coded adapters, extension tubes, and multi-lumen shafts with distal side slots. Dimensions: 12.0-13.5 Fr diameters, 13-30 cm lengths. Sterilization method not specified. No software or electronic components.
Indications for Use
Indicated for short-term central venous access for hemodialysis, apheresis, and infusion in patients requiring such access. Triple lumen version also indicated for central venous pressure monitoring and power injection of contrast media (max 5 ml/sec).
Regulatory Classification
Identification
A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
Special Controls
*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Mahurkar™ High Pressure Triple Lumen Acute Dialysis Catheter (K102605)
Submission Summary (Full Text)
{0}------------------------------------------------
K120674 Page 1 of 2
APR - 4 2012
Image /page/0/Picture/2 description: The image shows the word "COVIDIEN" in bold, black letters. To the left of the word is a black square with a white rectangle in the middle. The white rectangle has two black bars on the top and bottom.
#### 510(K) SUMMARY
#### 1. 510(k) Owner:
Covidien 15 Hampshire Street Mansfield, MA 02048 Telephone: (508) 261 - 6596 (508) 261 - 6596 Fax:
Contact: Wing Ng Title: Manager, Regulatory Affairs Date Prepared: March 01, 2012
## 2. Device:
Trade Names: Mahurkar™ Elite Acute Dual Lumen Catheter Mahurkar™ Elite Acute Triple Lumen Catheter Common Name: Central Venous Catheter Classification Name: Non-Implanted Hemodialysis Catheter (MPB) Non-Implanted Triple Lumen Hemodialysis Catheter (NIE) Regulation Number: 21 CFR 876.5540 Product Code(s): MPB. NIE Classification: Class II
### 3. Predicate Devices:
| Proposed Device | Predicate Device |
|------------------------------------------------|---------------------------------------------------------------------------|
| Mahurkar™ Elite Acute<br>Dual Lumen Catheter | Mahurkar™ Q Plus 13.5 Fr Dual Lumen<br>Acute Dialysis Catheter (K030209) |
| Mahurkar™ Elite Acute<br>Triple Lumen Catheter | Mahurkar™ High Pressure Triple Lumen<br>Acute Dialysis Catheter (K102605) |
#### 4. Device Description:
The Mahurkar™ Elite Acute Dual Lumen Catheter features a two lumen design. The proximal end has color-coded adapters to indicate arterial and venous flow. The adapters are connected to extension tubes which are available in curved or straight configurations. The extension tubes are connected to the hub which is joined to a dual lumen shaft available in pre-curved and straight configurations. The shaft extends to side slots near the distal tip. The dual lumen catheter is available in 12.0 Fr or 13.5 Fr outer diameters and a variety of implant lengths ranging from 13 cm to 30 cm. It is offered as a single device or as convenience kits.
{1}------------------------------------------------
K120674 Pgs 2 of 2
The Mahurkar™ Elite Acute Triple Lumen Catheter features a three lumen design. The proximal end has color-coded adapters to indicate arterial flow, venous flow, and medial infusion. The adapters are connected to extension tubes which are available in curved or straight configurations. The extension tubes are connected to the hub which is joined to a triple lumen shaft that extends to side slots near the distal tip. The triple lumen catheter is available in 12.5 Fr outer diameter and a variety of implant lengths ranging from 13 cm to 30 cm. It is offered as a single device or as convenience kits.
## 5. Intended Use:
The Mahurkar™ Elite Acute Dual Lumen Catheter is intended to be used for short-term central venous access for hemodialysis, apheresis, and infusion.
The Mahurkar™ Elite Acute Triple Lumen Catheter is intended to be used for short-term central venous access for hemodialysis, apheresis, infusion, central venous pressure monitoring, and pressure injection of contrast media. The maximum recommended infusion rate is 5 ml/sec for power injection of contrast media.
#### 6. Technological Characteristics:
The proposed devices have the same technological characteristics as compared to their respective predicate devices.
### 7. Performance Data:
Bench top functional and performance testing was completed to support substantial equivalence between the proposed device and the current device. The test regimen evaluated the devices for cleaning agent compatibility, lock solution compatibility, dynamic flow, column strength, clamp functionality, shaft stiffness, tensile strength at various locations, and resistance to kink, leak, burst, catheter collapse, and fatigue. Additionally, the triple lumen catheter has been evaluated for central venous pressure monitoring performance and simulated injection of contrast media. The results of the performance testing show that the proposed devices continue to meet the relevant product specifications.
Biocompatibility testing per ISO 10993: Biological Evaluation of Medical Devices was completed to support biocompatibility of the proposed device. The results of the biocompatibility testing show that the proposed devices continue to be biocompatible for its intended use.
The results of functional, performance, and biocompatibility testing support the determination of substantial equivalence.
#### 8. Conclusion:
- Based on non-clinical testing results, Covidien has demonstrated that the proposed Mahurkar™ Elite Acute Catheters are substantially equivalent to its respective predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Wing Ng Manager, Regulatory Affairs Covidien 15 Hampshire Street MANSFIELD MA 02048
APR - 4 2012
Re: K120674
Trade/Device Name: Mahurkar™ Elite Acute Dual Lumen Catheter Mahurkar™ Elite Acute Triple Lumen Catheter Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: MPB, NIE Dated: March 1, 2012 Received: March 5, 2012
Dear Mr.Ng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including (f.e.; anninonea) and rainer. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I Rase be advised that i Dr. I be rice complies with other requirements of the Act or any I Dri has made a acteriminens administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807);
{3}------------------------------------------------
## Page 2 - Mr. Wing Ng
labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains Lidocaine and ChloraPrep, which are subject to regulation as drugs.
Our substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:
Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) I mo letter will and my are FDA finding of substantial equivalence of your device to a legally premaince nodicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference Compilance at (210) 210 01197 Patt 807.97). You may obtain other general information on your to prematice notification (21 Orethe Division of Small Manufacturers, International and ICsponsibilities and its toll free number (800) 638-2041 or (240) 276-3150, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Benjamin K. Teeter
Benjam R. Fisher, Ph.D. Difector Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K120674
## Section 4 Indications for Use Statement
510(k) Number (if known):
K120674
Device Name: Mahurkar™ Elite Acute Dual Lumen Catheter Mahurkar™ Elite Acute Triple Lumen Catheter
Indications for Use:
The Mahurkar™ Elite Acute Dual Lumen Catheter is indicated for short-term central venous access for hemodialysis, apheresis, and infusion.
The Mahurkar™ Elite Acute Triple Lumen Catheter is indicated for short-term central venous access for hemodialysis, apheresis, infusion, central venous pressure monitoring, and pressure injection of contrast media. The maximum recommended infusion rate is 5 ml/sec for power injection of contrast media.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Estaluation (ODE) | |
|------------------------------------------------------------------------------------------|--|
| | |
| (Division Sign-Off)<br>Division of Reproductive, Gastro-Renal, and<br>Upplogical Devices | |
| 510(k) Number = | |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.