K120490 · Nipro Medical Corporation · JKA · Dec 27, 2012 · Clinical Chemistry
Device Facts
Record ID
K120490
Device Name
QUICK SAMPLER HOLDER
Applicant
Nipro Medical Corporation
Product Code
JKA · Clinical Chemistry
Decision Date
Dec 27, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The Quick Sampler Holder is intended to be used as a part of vacuum blood collection equipment for the collection of blood samples for various types of blood tests
Device Story
The Quick Sampler Holder is a non-contacting blood specimen collection device used in blood banking. It consists of a double-ended 18-gauge needle attached to a blood tube holder. The proximal needle tip is blunt for insertion into bloodline y-ports to prevent accidental needlesticks; the distal end is sharp and covered by a rubber sleeve to prevent blood loss during vial replacement. The device is used in conjunction with vacuum blood collecting vessels and bloodlines. It is operated by healthcare professionals to facilitate the collection of blood samples for diagnostic testing.
Clinical Evidence
No clinical testing required or submitted. Evidence consists of bench testing, including integrity tests of tamper sealing, rubber sleeve functional tests, hub-needle functional tests, hub-holder functional tests, and chemical, microbiological, and dimensional characterization.
Technological Characteristics
18-gauge double-ended needle; blood tube holder; blunt proximal tip; rubber sleeve for distal needle protection. Materials and functional specifications are equivalent to the predicate. Non-clinical testing included integrity, functional, chemical, microbiological, and dimensional assessments.
Indications for Use
Indicated for use as part of vacuum blood collection equipment for the collection of blood samples for various types of blood tests.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Nipro Medical Corporation. The logo consists of a stylized, intertwined symbol on the left, followed by the word "NIPRO" in bold, sans-serif font. Below "NIPRO" is the text "MEDICAL CORPORATION" in a smaller, sans-serif font.
3150 NW 107th Avenue, Miami FL 33172 Tel: 305.599.7174 Fax: 305.592.4621
# DEC 2 7 2012
## 510(k) Summary - Quick Sampler Holder
807.92(a) (1)
Contact Person:
Jessica Oswald Regulatory Affairs Specialist
Date of summary preparation:
January 19, 2012
807.92(a) (2)
Trade Name: Quick Sampler Holder
Common Name: Blood specimen collection device
Classification Name: tubes, vials, systems, serum separators, blood collection Regulation Number: 862.1675
Regulatory Class: II
Panel: Clinical Chemistry ·
Product Code: JKA
#### 807.92(a) (3)
Legally marketed substantial equivalent device: ITL Samplok Luer Kit (K021941)
#### 807.92(a) (4)
Description of device:
The Quick Sampler Holder is a non-contacting patient device. It is intended to be used in conjunction with a vacuum blood collecting vessel and bloodlines with y-ports, for blood banking. The Quick Sampler Holder consists of a double ended 18 gauge needle attached to a blood tube holder. The proximal needle tip, which is inserted into the yport, is blunt to prevent accidental needlesticks; the distal end, inside the holder, is sharp and is covered by a rubber sleeve to prevent blood loss during vial replacement/sample collection.
#### 807.92(a) (5)
Indications for Use:
The Quick Sampler Holder is intended to be used as a part of vacuum blood collection equipment for the collection of blood samples for various types of blood tests
## 807.92(a) (6)
Comparison of technological characteristics:
The Quick Sampler Holder is identical to the predicate in terms of indications for use and mode of operation. It is substantially equivalent to the predicate in terms of functional, mechanical and material specifications. The only technological difference between the
{1}------------------------------------------------
two devices is the proximal tip. The quick Sampler holder has a blunt tip, whereas, the ITL Samplok has a luer tip.
807.92(b) (1)
Non-clinical tests submitted:
Performance characteristics: Integrity test of tamper sealing, Rubber sleeve functional tests, Hub -needle functional tests, Hub -Holder functional tests, Chemical and microbiological characteristics and Dimensional characteristics.
807.92(b) (2)
Clinical tests submitted:
No clinical testing required or submitted with this application.
#### 807.92(b) (3)
Conclusions drawn from non-clinical and clinical tests:
Based on the testing described above, we believe the Quick Sampler Holder to be substantially equivalent to ITL Samplok, in terms of safety and effectiveness for its · intended use.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, with three abstract figures or lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## December 27, 2012
Ms. Jessica Oswald-McLeod, ASQ CQA Regulatory Affairs Specialist Nipro Medical Corporation 3150 North West, 107th Avenue MIAMI FL 33172
Re: K120490
Trade/Device Name: Quick Sampler Holder Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: II Product Code: JKA Dated: September 11, 2012 Received: September 13, 2012
Dear Ms. Oswald-McLeod:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that I Dr Has Internations administered by other Federal agencies. You must or any I edolul statutes and regulaments, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the qualis) of bections (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regalations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
Indications for Use
349D
Device Name: Quick Sampler Holder
Indications for Use:
The Quick Sampler Holder is intended to be used as a part of vacuum blood collection equipment for the collection of blood samples for various types of blood tests
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
igitally signed by Sajjad H. Syed Saijad H. Syed DA, ou=People, cn=Sajjad H. Syed 0200300 100.1.1=200060174
(Division Sign-Off) (Division Sign-Off)
Division of Anesthestology, General Hospital Division Control, Dental Devices
K120490 510(k) Number;
Quick Sampler Holder
-4.1-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.